TEMPORARY TRANSVENOUS PACEMAKER PLACEMENT ASSIST DEVICE

K032613 · Peter M. Rothenberg, M.D. · LDF · Nov 20, 2003 · Cardiovascular

Device Facts

Record IDK032613
Device NameTEMPORARY TRANSVENOUS PACEMAKER PLACEMENT ASSIST DEVICE
ApplicantPeter M. Rothenberg, M.D.
Product CodeLDF · Cardiovascular
Decision DateNov 20, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3680
Device ClassClass 2

Indications for Use

This device is designed to connect an intravascular electrode to a monitor through a standard EKG patient lead for the purpose of displaying an intravascular signal in real time. Only the black pinjack is "active". The red pinjack is electrically blind and designed to isolate the proximal electrode from inadvertent stimulation.

Device Story

Device acts as an electrical interface between an intravascular electrode and a standard bedside EKG monitor. Input: electrical signal from proximal pin of an intravascular electrode. Transformation: signal passes through a 1000 ohm resistor to a standard EKG eyelet, allowing the signal to travel retrograde along a standard EKG cable. Output: real-time intravascular signal displayed on a bedside monitor. Used in clinical settings (e.g., during pacemaker placement) by physicians to guide catheter positioning without fluoroscopy. Benefits include reduced reliance on expensive or cumbersome fluoroscopic procedures. Device is a simple passive electrical connector.

Clinical Evidence

No clinical data provided; device is a passive electrical accessory.

Technological Characteristics

Passive electrical connector consisting of a pin jack connected in series via a 1000 ohm resistor to a standard EKG eyelet, encased in a housing. Provides electrical isolation for the proximal electrode. No energy source, software, or complex electronics.

Indications for Use

Indicated for patients requiring transvenous pacemaker or intravenous catheter placement where EKG guidance is used to facilitate positioning, as an alternative to fluoroscopy.

Regulatory Classification

Identification

Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.

Submission Summary (Full Text)

{0}------------------------------------------------ Copy of file in entirety sent electronically KO32613 PETER M. ROTHENBERG, M.D., M.A. 657 CAMINO DE LOS MARES SUITE 137 SAN CLEMENTE, CA 92673 PHONE: (949) 489-9039 FAX: (949) 489-8136 EMAIL: PMRMDINC@SBCGLOBAL.NET ## 510(k) SUMMARY OF DEVICE TRADE NAME: PACER ASSIST DEVICE (FORMAL NAME TO BE COPYRIGHTED ON RECEIPT OF 510(K) CLEARANCE) COMMON NAME: ACCESSORY TO EKG CABLE CLASSIFICATION NAME: ACCESSORY TO EKG CABLE (CLASSIFICATION TO BE DETERMINED) ## CONTACT PERSON: PETER M. ROTHENBERG, M.D., M.A. As noted in my 510 (k) application, this device is of such simplicity, there are no others directly comparable. Features have been incorporated into those far more technologically advanced. In essence, the subject device provides an electrically safe connection between the proximal pin of an intravascular electrode and a patient EKG lead , the signal ultimately traveling retrograde along a standard EKG cable to a bedside monitor. It would be used to facilitate placement of a transvenous pacemaker or appropriately designed intravenous catheter with EKG guidance rather than the more cumbersome (and expensive) fluoroscopic approach. The device entirely consists of a pin jack (to accept the proximal pin of the electrode) connected in series via a 1000 ohm resistor to a standard EKG eyelet all encased in an appropriate housing. The snap-fit connection between the EKG lead and device eyelet completes the connection to the monitor. ## DATE PREPARED: 10/27/03 {1}------------------------------------------------ Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 2 0 2003 Peter M. Rothenberg, M.D., M.A. 657 Camino De Los Mares Suite 137 San Clemente, CA 92673 Re: K032613 Trade Name: Transvenous Pacemaker Placement Assist Device Regulation Number: 21 CFR 870.3680 and 870.2050 Regulation Name: Accessory to cardiovascular permanent or temporary pacemaker electrode; and biopotential amplifier and signal conditioner Regulatory Class: Class II (two) Product Code: LDF and DRR Dated: October 30, 2003 Received: November 4, 2003 Dear Dr. Rothenberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Peter M. Rothenberg, M.D., M.A. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Dini Huschur for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## INDICATIONS FOR USE This device is designed to connect an intravascular electrode to a monitor through a standard EKG patient lead for the purpose of displaying an intravascular signal in real time. Only the black pinjack is "active". The red pinjack is electrically blind and designed to isolate the proximal electrode from inadvertent stimulation. Dona Seemle X Prescription device 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...