K093268 · Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · MHX · May 11, 2010 · Cardiovascular
Device Facts
Record ID
K093268
Device Name
INTELLIVUE PATIENT MONITOR
Applicant
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN
Product Code
MHX · Cardiovascular
Decision Date
May 11, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Philips MP2, X2, MP5, MP20, MP30, MP40, MP50, MP70, MP80, and MP90 IntelliVue Patient Monitors are intended for monitoring and recording of and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP2, X2, MP5, MP20, MP30, MP40, and MP50 are additionally intended for use in transport situations within hospital environments. The MP5 is also intended for use during patient transport outside of a hospital environment. The monitors are not intended for home use. They are intended for use by health care professionals.
Device Story
IntelliVue Patient Monitors (MP2, X2, MP5, MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90) function as multi-parameter physiological monitoring systems. Devices accept inputs from various sensors (ECG, blood pressure, SpO2, CO2, temperature, etc.); process signals to calculate vital signs and detect arrhythmias; and generate real-time visual/audible alarms. Operated by healthcare professionals in hospital settings (including transport). Output displayed on integrated screens for clinical decision-making. Software release G.08 introduces enhancements to the respiration algorithm. Benefits include continuous patient surveillance and timely clinical intervention.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level and regression testing, as well as hazard analysis testing. Pass/Fail criteria were based on specifications cleared for the predicate devices. Results demonstrated substantial equivalence in performance, functionality, and reliability.
Technological Characteristics
Multi-parameter patient monitoring system. Components include ECG, blood pressure, SpO2, CO2, and temperature sensors. Connectivity via network/communication systems. Software release G.08. Class II medical device.
Indications for Use
Indicated for use by health care professionals for monitoring physiological parameters in adult, pediatric, and neonatal patients in hospital environments, including transport. Not for home use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
## K093268
## 510(k) Summary
. .
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
- 1. The submitter of this premarket notification is:
Andreas Suchi Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-1291 e-mail: andreas.suchi@philips.com
Fax: ++49 7031 463-2442
This summary was prepared on May 10, 2010.
- 2. The names of the devices are the Philips MP2, X2, MP5, MP20, MP30, MP40, MP50, MP60, MP70, MP80, and MP90 IntelliVue Patient Monitors Classification names are as follows:
| Device Panel | Classification | ProCode | Description |
|---------------------------|----------------|---------|--------------------------------------------------------------------------|
| Cardiovascular<br>Devices | §870.1025, II | DSI | Detector and alarm, arrhythmia |
| | §870.1025, II | MLD | Monitor, ST Segment with Alarm |
| | §870.1025, II | MHX | Monitor, Physiological, Patient (with<br>arrhythmia detection or alarms) |
| | §870.1100, II | DSJ | Alarm, Blood Pressure |
| | §870.1110, II | DSK | Computer, Blood Pressure |
| | §870.1130, II | DXN | System, Measurement, Blood-Pressure,<br>Non-Invasive |
| | §870.1435, II | DXG | Computer, Diagnostic, Pre-Programmed,<br>Single-Function |
| | §870.1915, II | KRB | Probe, Thermodilution |
| | §870.2060, II | DRQ | Amplifier and Signal Conditioner,<br>Transducer Signal |
| | §870.2300, II | DRT | Monitor, Cardiac (incl. Cardiotachometer<br>& Rate Alarm) |
| | §870.2340, II | DPS | Electrocardiograph |
| | §870.2340, II | MLC | Monitor, ST Segment |
| | §870.2350, II | DRW | Electrocardiograph, Lead Switching<br>Adapter |
| | §870.2370, II | KRC | Tester, Electrode, Surface,<br>Electrocardiograph |
| | §870.2600, I | DRJ | System, Signal Isolation |
| | §870.2700, II | DQA | Oximeter |
| | §870.2770, II | DSB | Plethysmograph, Impedance |
| | §870.2800, II | DSH | Recorder, Magnetic tape, Medical |
| | §870.2810, I | DSF | Recorder, Paper Chart |
| | §870.2850, II | DRS | Extravascular Blood Pressure Transducer |
| | §870.2900, I | DSA | Cable, Transducer and Electrode, incl.<br>Patient Connector |
| | - | MSX | System, Network and Communication,<br>Physiological Monitors |
{1}------------------------------------------------
| Device Panel | Classification | ProCode | Description |
|----------------------------------------------|----------------|---------|----------------------------------------------------------------------------|
| | §870.2910, II | DRG | Transmitters and Receivers, Physiological<br>Signal, Radiofrequency |
| Anesthesiology<br>Devices | §868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide, Gaseous-<br>Phase |
| | §868.1500, II | CBQ | Analyzer, Gas, Enflurane, Gaseous-Phase<br>(Anesthetic Concentration) |
| | §868.1500, II | NHO | Analyzer, Gas, Desflurane, Gaseous-Phase<br>(Anesthetic Concentration) |
| | §868.1500, II | NHP | Analyzer, Gas, Sevoflurane, Gaseous-<br>Phase (Anesthetic Concentration) |
| | §868.1500, II | NHQ | Analyzer, Gas, Isoflurane, Gaseous-Phase<br>(Anesthetic Concentration) |
| | §868.1620, II | CBS | Analyzer, Gas, Halothane, Gaseous-Phase<br>(Anesthetic Concentration) |
| | §868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide, Gaseous-<br>Phase (Anesthetic Concentration) |
| | §868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-Phase |
| | §868.1880, II | BZC | Data calculator Pulmonary-function |
| | §868.2375, II | BZQ | Monitor, Breathing Frequency |
| | §868.2480, II | LKD | Monitor, Carbon Dioxide, Cutaneous |
| | §868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for Infant<br>not under Gas Anesthesia |
| General Hospital and<br>Personal Use Devices | §880.2910, II | FLL | Thermometer, Electronic, Clinical |
| Neurological Devices | §882.1400, II | GWR | Electroencephalograph |
| | §882.1420, II | GWS | Analyzer, Spectrum,<br>Electroencephalogram Signal |
- 3. The modified devices are substantially equivalent to previously cleared Philips IntelliVue Patient Monitors marketed pursuant to K021778, K030038, K032858, K033444, K033513, K040304, K040357, K041235, K042845, K050762, K051106, K052801, K053522, K060221, K060541, K061052, K061610, K062283, K063315, K062392, K063725, K71426, K72020.
- 4. The modification is the introduction of software release G.08 for the entire IntelliVue Patient Monitors family, models MP2, X2, MP5, MP20, MP30, MP30, MP60, MP70, MP80, and MP90 including enhancements in the Respiration algorithm.
- 5. The modified devices have the same intended use as the legally marketed predicate devices. The Philips MP2, X2, MP5, MP20, MP30, MP40, MP50, MP70, MP80, and MP90 IntelliVue Patient Monitors are intended for monitoring and recording of and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP2, X2, MP5, MP20, MP30, MP40, and MP50 are additionally intended for use in transport situations within hospital environments. The MP5 is also intended for use during patient transport outside of a hospital environment. The monitors are not intended for home use. They are intended for use by health care professionals.
- 6. The modified devices have the same technological characteristics as the legally marketed predicate device.
{2}------------------------------------------------
- 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified devices with respect to the predicate. Testing involved system level and regression tests as well as testing from the hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitors meet all reliability requirements and performance claims.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo includes the department's emblem, which features a stylized caduceus-like symbol with three figures representing people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Philips Medizin Systeme Boeblingen GmbH c/o Mr. Andreas Suchi Senior Regulatory Affairs Engineer Hewlett-Packard Str. 2 D-71034 Boeblingen Germany
MAY 1 1 2010
Re: K093268
> Trade/Device Name: Philips Intellivue Patient Monitor Models MP2, X2, MP5, MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 Regulatory Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II (Two) Product Code: MHX Dated: May 4, 2010 Received: May 5, 2010
Dear Mr. Suchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Mr. Andreas Suchi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
K093268 510(k) Number (if known):
Philips MP2, X2, MP5, MP20, MP30, MP40, MP50, MP60, Device Name: MP70, MP80, and MP90 IntelliVue Patient Monitors, Software Revision G.08.
Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
| Prescription Use (Part 21 CFR 801 Subpart D) | yes |
|----------------------------------------------|--------|
| | AND/OR |
| Over-The-Counter Use (21 CFR 807 Subpart C) | No |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number 5/11/10
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.