PHILIPS MP60, MP70 AND MP90 INTELLIVUE PATIENT MONITORS WITH PORTAL TECHNOLOGY AND M3185 OPT C41 WIRELESS BEDSIDE ADAPTR

K030038 · Philips Medical Systems, Inc. · MHX · Mar 12, 2003 · Cardiovascular

Device Facts

Record IDK030038
Device NamePHILIPS MP60, MP70 AND MP90 INTELLIVUE PATIENT MONITORS WITH PORTAL TECHNOLOGY AND M3185 OPT C41 WIRELESS BEDSIDE ADAPTR
ApplicantPhilips Medical Systems, Inc.
Product CodeMHX · Cardiovascular
Decision DateMar 12, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Intended Use

Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments.

Device Story

Philips IntelliVue MP60, MP70, and MP90 are multi-parameter patient monitors designed for hospital use. Devices collect physiological data (ECG, blood pressure, SpO2, gas analysis, EEG, etc.) via various sensors and probes. The system processes these signals to provide real-time monitoring, recording, and alarming. The subject device introduces an optional externally mounted wireless network connection (M3185 opt. C41 Wireless Bedside Adapter) to facilitate data transmission. Operated by healthcare professionals, the monitors display patient status, enabling clinical decision-making through continuous surveillance of vital signs. The wireless capability allows for flexible bedside monitoring within the hospital network.

Clinical Evidence

Bench testing only. Verification testing activities were conducted to establish performance and reliability characteristics, including functional level tests and safety testing based on risk analysis.

Technological Characteristics

Multi-parameter patient monitoring system. Includes ECG, blood pressure, gas analysis, EEG, and pulse oximetry modules. Features an optional externally mounted wireless LAN adapter (M3185 opt. C41). Connectivity via wireless network. System performs signal conditioning, amplification, and alarm processing. Designed for hospital bedside use.

Indications for Use

Indicated for monitoring, recording, and alarming of physiological parameters in adult, pediatric, and neonatal patients in hospital environments by healthcare professionals.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## MAR 1 2 2003 ## 510 (k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92. 1. The submitter of this premarket notification is: David Osborn Philips Medical Systems 3000 Minuteman Road Andover, MA 01810-1099 Tel: 978 659 3178 Fax: 978 685 5624 e-mail: d.g.osborn@ieee.org This summary was prepared on December 31, 2002. - 2. The names of the devices are the Philips MP60, MP70, and MP90 IntelliVue Patient Monitor with Portal Technology and Wireless LAN. Classification names are as follows: | Device Panel | Classification | ProCode | Description | |------------------------------------------|-----------------|---------|-----------------------------------------------------------------------------| | Circulatory<br>System Devices<br>(12625) | \$870.1025, III | DSI | Detector and alarm, arrhythmia | | | \$870.1025, III | MLD | Monitor, ST Segment with Alarm | | | \$870.1025, III | MHX | Monotor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms) | | | \$870.1100, II | DSJ | Alarm, Blood Pressure | | | \$870.1110, II | DSK | Computer, Blood Pressure | | | \$870.1130, II | DXN | System, Measurement, Blood-<br>Pressure, Non-Invasive | | | \$870.1435, II | DXG | Computer, Diagnostic, Pre-<br>Programmed, Single-Function | | | \$870.1915, II | KRB | Probe, Thermodilution | | | \$870.2060, II | DRQ | Amplifier and Signal<br>Conditioner, Transducer Signal | | | \$870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm) | | | \$870.2340, II | DPS | Electrocardiograph | | | \$870.2340, II | MLC | Monitor, ST Segment | | | \$870.2350, II | DRW | Electrocardiograph, Lead<br>Switching Adapter | | | \$870.2370, II | KRC | Tester, Electrode, Surface,<br>Electrocardiograph | | | \$870.2450, II | DXJ | Display, Cathode-Ray Tube,<br>Medical | | | \$870.2600, I | DRJ | System, Signal Isolation | | | \$870.2700, II | DQA | Oximeter | | | \$870.2770, II | DSB | Plethysmograph, Impedance | | | \$870.2800, II | DSH | Recorder, Magnetic tape,<br>Medical | | | \$870.2810, I | DSF | Recorder, Paper Chart | | | \$870.2850, II | DRS | Extravascular Blood Pressure<br>Transducer | | | \$870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector | | | - | MSX | System, Network and<br>Communication, Physiological<br>Monitors | {1}------------------------------------------------ | Anesthesiology<br>and Respiratory<br>Therapy (12624) | \$868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase | |------------------------------------------------------|----------------|-----|------------------------------------------------------------------------------| | | \$868.1500, II | CBQ | Analyzer, Gas, Enflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) | | | \$868.1500, II | NHO | Analyzer, Gas, Desflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) | | | \$868.1500, II | NHP | Analyzer, Gas, Sevoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) | | | \$868.1500, II | NHQ | Analyzer, Gas, Isoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) | | | \$868.1620, II | CBS | Analyzer, Gas, Halothane,<br>Gaseous-Phase (Anesthetic<br>Concentration) | | | \$868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,<br>Gaseous-Phase (Anesthetic<br>Concentration) | | | \$868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-<br>Phase | | | \$868.2375, II | BZQ | Monitor, Breathing Frequency | | | \$868.2480, II | LKD | Monitor, Carbon Dioxide,<br>Cutaneous | | | \$868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia | | General Hospital<br>and Personal Use<br>(12520) | \$880.2910, II | FLL | Thermometer, Electronic,<br>Clinical | | Neurological<br>(12513) | \$882.1400, II | GWR | Electroencephalograph | | | \$882.1420, I | GWS | Analyzer, Spectrum,<br>Electroencephalogram Signal | - 3. The new devices are substantially equivalent to previously cleared Philips devices marketed pursuant to K001664, K021778, and ко23871. - The modification is primarily a hardware based change that offers, as 4 . an option, the addition of a an externally mounted wireless network connection to the Philips Medical System MP60, MP70 and MP90 IntelliVue patient monitor devices. - 5. The new devices have the same intended use as the legally marketed predicate devices. When used in the hospital environment, they are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates. - 6. The new devices have the same technological characteristics as the leqally marketed predicate devices. - 7. Verification testing activities were conducted to establish the performance and reliability characteristics of the new device. Testing involved functional level tests and safety testing from the risk analysis. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three swooping lines forming its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 2 2003 Phillips Medical Systems, Inc. c/o Mr. David Osborn Ouality Program Manager 3000 Minuteman Road Andover, MA 01810-1099 ## Re: K030038 Trade/Device Name: IntelliVue Models MP60, MP70 and MP90 Patient Monitors with Portal Technology and M3185 opt. C41 Wireless Bedside Adapter Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm Regulatory Class: Class III (three) Product Code: MHX Dated: February 24, 2003 Received: February 25, 2003 ## Dear Mr. Osborn: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 – Mr. David Osborn Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use Page of KO30038 510(k) Number (if known): Device Name: The Philips IntelliVue MP60, MP70, and MP90 Patient Monitors with Portal Technology and M3185 opt C41 Wireless Bedside Adapter. Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) (Optional Format 1-2-96) OR Over-The-Counter VCelkatil (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_ 9
Innolitics
510(k) Summary
Decision Summary
Classification Order