THE PHILIPS MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90 INTELLIVUE PATIENT MONITORS

K062283 · Philips Medical Systems North America, Inc. · MHX · Sep 20, 2006 · Cardiovascular

Device Facts

Record IDK062283
Device NameTHE PHILIPS MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90 INTELLIVUE PATIENT MONITORS
ApplicantPhilips Medical Systems North America, Inc.
Product CodeMHX · Cardiovascular
Decision DateSep 20, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Indications for Use

Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alparming of multiple physiological parameters of adults, pediatrics and neonates in healthcare facilities. The MP20, MP30, MP40 and MP50. are additionally intended for use in transport situations within healthcare facilities.

Device Story

IntelliVue patient monitors (MP20-MP90) collect multi-parameter physiological data (ECG, blood pressure, gas analysis, etc.) from patients in healthcare settings; operated by clinicians. Device processes signals to provide real-time monitoring, recording, and alarm functions. Release E.01 software update maintains core functionality of predicate monitors. Output displayed on-screen for clinician review to support clinical decision-making; enables continuous patient surveillance and timely intervention. Benefits include improved patient safety through automated monitoring and alarm notification.

Clinical Evidence

Bench testing only. Verification and validation activities included system-level tests, performance tests, and safety testing derived from hazard analysis. Pass/Fail criteria were based on specifications cleared for the predicate devices. Results demonstrate substantial equivalence.

Technological Characteristics

Multi-parameter patient monitoring system. Includes ECG, blood pressure, gas analysis, and neurological signal analysis capabilities. Connectivity via network/communication systems. Software-based signal processing and alarm management. Modular form factor (MP20-MP90).

Indications for Use

Indicated for monitoring, recording, and alarming physiological parameters in adult, pediatric, and neonatal patients in healthcare facilities. MP20-MP50 models also indicated for intra-facility transport. ST segment monitoring restricted to adults. Transcutaneous gas measurement (tcpO2/tcpCO2) restricted to neonates.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510 (k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92. 1. The submitter of this premarket notification is: Andreas Suchi Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany . Tel: ++49 7031 463-1291 Fax: ++49 7031 463-2442 e-mail: andreas.suchi@philips.com This summary was prepared on August 4, 2006. - 2. The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors. Classification names are as follows: | Device Panel | Classification | ProCode | Description | |------------------------------------------|----------------|---------|-----------------------------------------------------------------------------| | Circulatory<br>System Devices<br>(12625) | \$870.1025, II | DSI | Detector and alarm, arrhythmia | | | \$870.1025, II | MLD | Monitor, ST Segment with Alarm | | | \$870.1025, II | MHX | Monitor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms) | | | \$870.1100, II | DSJ | Alarm, Blood Pressure | | | \$870.1110, II | DSK | Computer, Blood Pressure | | | \$870.1130, II | DXN | System, Measurement, Blood-<br>Pressure, Non-Invasive | | | \$870.1435, II | DXG | Computer, Diagnostic, Pre-<br>Programmed, Single-Function | | | \$870.1915, II | KRB | Probe, Thermodilution | | | \$870.2060, II | DRQ | Amplifier and Signal<br>Conditioner, Transducer Signal | | | \$870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm) | | | \$870.2300, II | MSX | System, Network and<br>Communication, Physiological<br>Monitors | | | \$870.2340, II | DPS | Electrocardiograph | | | \$870.2340, II | MLC | Monitor, ST Segment | | | \$870.2350, II | DRW | Electrocardiograph, Lead<br>Switching Adapter | | | \$870.2370, II | KRC | Tester, Electrode, Surface,<br>Electrocardiograph | | | \$870.2450, II | DXJ | Display, Cathode-Ray Tube,<br>Medical | | | \$870.2600, I | DRJ | System, Signal Isolation | | | \$870.2700, II | DQA | Oximeter | | | \$870.2770, II | DSB | Plethysmograph, Impedance | | | \$870.2800, II | DSH | Recorder, Magnetic tape,<br>Medical | | | \$870.2810, I | DSF | Recorder, Paper Chart | | | \$870.2850, II | DRS | Extravascular Blood Pressure<br>Transducer | SEP 2 0 2006 {1}------------------------------------------------ | \$870.2900, I | DSA | Cable, Transducer and Electrode, incl. Patient Connector | | |------------------------------------------------------|-------------------------|------------------------------------------------------------------|------------------------------------------------------------------------| | \$870.2910, II | DRG | Transmitters and Receivers, Physiological Signal, Radiofrequency | | | Anesthesiology<br>and Respiratory<br>Therapy (12624) | \$868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide, Gaseous-Phase | | | \$868.1500, II | CBQ | Analyzer, Gas, Enflurane, Gaseous-Phase (Anesthetic Concentration) | | | \$868.1500, II | NHO | Analyzer, Gas, Desflurane, Gaseous-Phase (Anesthetic Concentration) | | | \$868.1500, II | NHP | Analyzer, Gas, Sevoflurane, Gaseous-Phase (Anesthetic Concentration) | | | \$868.1500, II | NHQ | Analyzer, Gas, Isoflurane, Gaseous-Phase (Anesthetic Concentration) | | | \$868.1620, II | CBS | Analyzer, Gas, Halothane, Gaseous-Phase (Anesthetic Concentration) | | | \$868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide, Gaseous-Phase (Anesthetic Concentration) | | | \$868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-Phase | | | \$868.2375, II | BZQ | Monitor, Breathing Frequency | | | \$868.2480, II | LKD | Monitor, Carbon Dioxide, Cutaneous | | | \$868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for Infant not under Gas Anesthesia | | General Hospital<br>and Personal Use<br>(12520) | \$880.2910, II | FLL | Thermometer, Electronic, Clinical | | | Neurological<br>(12513) | \$882.1400, II | GWR | | \$882.1420, I | | GWS | Analyzer, Spectrum, Electroencephalogram Signal | - 3. The modified devices are substantially equivalent to previously cleared Philips devices marketed pursuant to, K021778, K030038, K032858, K033513, K040304, K041235, K042845, K050141, 050762, K051106, K052801, K053522, K060221, K060541 and K061052 - 4. The modification is the introduction of Release E.01 software for the IntelliVue patient monitor devices. - 5. The modified devices have the same intended use as the legally marketed predicate devices. They are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare facilities and during transport within healthcare facilities. - 6. The modified devices have the same technological characteristics as the legally marketed predicate devices. {2}------------------------------------------------ - 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wing-like shapes, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 0 2006 Philips Medizin Systeme Boblingen GmbH c/o Ms. Linda Stone Philips Medical Systems 3000 Minuteman Road Andover. MA 01810 Re: K062283 Trade Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors, Release E.01 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-segment measurement and alarm Regulatory Class: Class II (two) Product Code: MHX Dated: August 21, 2006 Received: August 23, 2006 Dear Ms. Stone: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2 - Ms. Linda Stone Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantials squisvalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bram D. Zuckerman, M.D. Director Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors, Release E.01 Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alparming of multiple physiological parameters of adults, pediatrics and neonates in healthcare facilities. The MP20, MP30, MP40 and MP50. are additionally intended for use in transport situations within healthcare facilities. ST Segment monitoring is restricted to adult patients only. The transcutaneous gas measurement (tcp02 / tcpC02) is restricted to neonatal patients only. Prescription Use (Part 21 CFR 801 Subpart D) yes___________________________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use No (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Bhumanna
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...