VASCUTRAK 2, MODELS: V14 AND V18

K082343 · Ymed, Inc. · PNO · Sep 11, 2008 · Cardiovascular

Device Facts

Record IDK082343
Device NameVASCUTRAK 2, MODELS: V14 AND V18
ApplicantYmed, Inc.
Product CodePNO · Cardiovascular
Decision DateSep 11, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The VascuTraK™ 2 PTA Dilatation Catheter is intended for dilatation of stenoses in the peripheral arteries including the illiac, femoral, poplited, illio-femoral, infrapopliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Device Story

VascuTraK™ 2 PTA Dilatation Catheter; sterile, single-use percutaneous transluminal angioplasty device. Features flexible shaft with through-lumen and semi-compliant balloon. Balloon inflated via central lumen connected to proximal luer fitting. Used by physicians in clinical settings for peripheral artery stenosis dilatation and arteriovenous dialysis fistula treatment. Modification from predicate (K073025) introduces larger balloon diameters (5.0–7.0 mm) and lengths (up to 300 mm). Device tracks over 0.014" or 0.018" guidewires. Clinical benefit involves mechanical restoration of vessel patency. No new materials or sterilization changes; performance verified via tensile strength, balloon fatigue, compliance, and burst testing.

Clinical Evidence

Bench testing only. No clinical data presented. Performance verified through tensile strength, balloon fatigue, compliance, and burst testing. Biocompatibility previously established per ISO 10993-1.

Technological Characteristics

Sterile, single-use PTA catheter. Flexible shaft with single through-lumen and semi-compliant balloon. Materials identical to predicate. Sterilization validated to SAL 10^-6. Compatible with 0.014"/0.018" guidewires. Requires minimum sheath/guide catheter I.D. up to 0.120". Mechanical operation via manual inflation of balloon.

Indications for Use

Indicated for patients requiring dilatation of stenoses in peripheral arteries (iliac, femoral, popliteal, ilio-femoral, infrapopliteal, renal) or treatment of obstructive lesions in native/synthetic arteriovenous dialysis fistulae. Contraindicated for use in coronary arteries.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a simple, clean design. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 15, 2016 YMed, Inc. c/o Mr. Thomas P. Schroeder 9951B Business Park Avenue San Diego, CA 92131 Re: K082343 Trade/Device Name: Vascutrak™ 2 PTA Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PNO Dated: August 13, 2008 Received: August 15, 2008 Dear Mr. Schroeder: This letter corrects our substantially equivalent letter of September 11, 2008. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {2}------------------------------------------------ ## Page 2 - Mr. Thomas P. Schroeder device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Misti L. Malone -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ YMed, Inc. Special 510(k) Device Modification August 12, 2008 ・・ . #### INDICATIONS FOR USE STATEMENT 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: VascuTraK™ 2 PTA Dilatation Catheter Indications for use: The VascuTraK™ 2 PTA Dilatation Catheter is intended for dilatation of stenoses in the peripheral arteries including the illiac, femoral, poplited, illio-femoral, infrapopliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Prescription Use: × (Part 21 CFR 801 Subpart D) AND/OR Over-the-counter Use: (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Duna R. Vahner (Division Sim. Off) Division of Carolovascular Devices 510(k) Number_Ko8234 3 Page 17 of 55 {4}------------------------------------------------ K082343 page 1 of 2 # SEP 1 1 2008 YMed, Inc. Special 510(k) Device Modification August 12, 2008 ### 510 (k) SUMMARY Applicant and Manufacturer YMed, Inc. 9951B Business Park Avenue San Diego. California 92131 Phone: (858) 549-1337 Fax: (858) 549-1717 | Contact Person: | Thomas Schroeder, Director, RA/QA | |----------------------|-------------------------------------------------------------------| | Prepared: | August 12, 2008 | | Common Name: | PTA Catheter | | Classification Name: | Catheter, Angioplasty, Peripheral, Transluminal; Product Code LIT | | Proprietary name: | VascuTraK™ 2 PTA Dilatation Catheter | #### Unmodified Device The VascuTraK™ 2 Catheler is a sterile single use device that consists of a flexible shaft with a single through lumen terminating in a semi-compliant balloon. The catheter has a 24 cm lumen oroximal to the balloon that accepts a 0.014" guidewire and an 18 mm lumen distal to the balloon that accepts a 0.014" or 0.018" guidewire. The distal balloon is inflated via a central lumen terminated in a luer fitting at the proximal end. The unmodified device utilizes balloons with diameters up to 4.0 mm and lengths up to 120 mm. #### Modified Device Description The YMed VascuTraK™ 2 PTA Dilatation Catheter is a modification to the VascuTraK™ 2 PTA Dilatation Catheter cleared under K073025. The device is fabricated of the same materials as the unmodified device and has the same intended use. The distal balloon sizes for the modified device include diameters from 5.0 - 7.0 mm and lengths up to 300mm. {5}------------------------------------------------ KC82343 page 302 YMed. Inc. Special 510(k) Device Modification August 12. 2008 #### Indications for Use The VascuTraK™ 2 PTA Dilatation Catheter is intended for dilatation of stenoses in the peripheral arteries including the illac, femoral, popliteal, ilio-femoral, infrapopliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. #### Technological Characteristics Comparison The VascuTraK™ 2 PTA Dilatation Catheter is identical in design and construction to the currently marketed VascuTraK™ 2 PTA Dilatation Catheter. This version of the catheter features larger balloons for use in the larger peripheral arteries. #### Performance and Safety No new materials were used for this modification. The biological safety of the device was previously demonstrated through biocompatibility studies of patient contact materials in accordance with the requirements outlined in ISO 10993-1. The modification does not affect the ability to sterilize the device; thus, the prior sterilization testing is valid. The device conforms to a Sterility Assurance Level (SAL) of 10° The modified device will require a minimum sheath or guide catheter I.D of up to 0.120" depending upon balloon size. The revised labeling reflects the applicable sheath size for the larger balloon sizes. The modified device was tested using the same test procedures that were used for the unmodified device: these tests included tensile strength, balloon fatigue, compliance and burst. All test results were satisfactory. The supplied instructions for use provide the user with the applicable warnings and cautions during use. The device is contraindicated for the coronary arteries. There are no new safety or effectiveness issues related to this device. The modified device is considered safe for its intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...