Fortrex 0.035 OTW PTA Balloon Catheter

K142654 · Covidien, LLC · LIT · Oct 28, 2014 · Cardiovascular

Device Facts

Record IDK142654
Device NameFortrex 0.035 OTW PTA Balloon Catheter
ApplicantCovidien, LLC
Product CodeLIT · Cardiovascular
Decision DateOct 28, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The Fortrex 0.035" OTW PTA Balloon Catheter is intended to dilate stenoses in the iliac, femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

Device Story

Over-the-wire (OTW) dual-lumen PTA balloon catheter; used by physicians in clinical settings for peripheral vascular interventions. Device features distally mounted semi-compliant inflatable balloon and atraumatic tapered tip. Operates by tracking over 0.035" guidewire to target stenosis; balloon inflated via contrast/saline mixture through dedicated inflation lumen. Radiopaque markers facilitate precise positioning. Device dilates stenotic vessels or post-dilates stents to restore blood flow. Benefits include improved vessel patency and treatment of dialysis access obstructions. Design modifications from predicate include higher burst pressure and enhanced joint strength.

Clinical Evidence

Bench testing only. Performance testing included balloon dimensional verification, rated burst pressure, burst strength, fatigue, compliance, inflation/deflation time, wire movement, radiopacity, tracking, insertion/pull-back/reinsertion forces, torque, tensile strength, and kink resistance. Biocompatibility testing performed per ISO 10993-1 for short-term blood contact (<24 hours), including cytotoxicity, sensitization, reactivity, systemic toxicity, hemolysis, pyrogenicity, complement activation, thromboresistance, and mutagenicity. All tests met acceptance criteria.

Technological Characteristics

OTW dual-lumen catheter; 0.035" guidewire compatibility. Balloon diameters 4.0-12.0 mm; lengths 20-100 mm; shaft lengths 40-135 cm. Features radiopaque marker bands. Biocompatibility per ISO 10993-1. Mechanical design optimized for high burst pressure and joint integrity.

Indications for Use

Indicated for patients requiring dilation of stenoses in iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae, and stent post-dilatation in peripheral vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three profiles of human faces incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 29, 2014 Covidien llc. c/o Mark Job Regulatory Technical Services LLC 1394 25th Street NW Buffalo, MN 55313 Re: K142654 Trade/Device Name: Fortrex 0.035" OTW PTA Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: LIT Dated: September 16, 2014 Received: September 18, 2014 Dear Mr. Job: This letter corrects our substantially equivalent letter of October 28, 2014. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {1}------------------------------------------------ CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K142654 Device Name Fortrex™ 0.035" OTW PTA Balloon Catheter #### Indications for Use (Describe) The Fortrex 0.035" OTW PTA Balloon Catheter is intended to dilate stenoses in the iliac, femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the word "COVIDIEN" in a bold, sans-serif font. To the left of the word is a square logo. The logo is composed of smaller squares in varying shades of blue, creating a pattern within the larger square. # 510(k) Summary # Fortrex™ 0.035" OTW PTA Balloon Catheter This 510(k) summary is being submitted in accordance with the requirements of 21 C.F.R § 807.92. #### 1. Submitter Information | Applicant | Covidien llc<br>4600 Nathan Lane N<br>Plymouth, MN 55441-2651<br>Tel: 763-398-7000<br>Fax: 763-591-3248 | |----------------|---------------------------------------------------------------------------------------------------------| | Contact Person | Rupali Gupta<br>Senior Regulatory Affairs Specialist | | Date Prepared | August 25, 2014 | #### 2. Subject Device | Device Trade Name | Fortrex TM 0.035" OTW PTA Balloon Catheter | |----------------------|------------------------------------------------------------------------------------| | Device Common Name | PTA Dilatation Catheter | | Classification Name | Catheter, Angioplasty, Peripheral, Transluminal (21 CFR§870.1250, Product Code LIT | | Classification Panel | Cardiovascular | ### 3. Predicate Devices | Device Trade Name | EverCross TM .035” OTW PTA Dilatation Catheter (Primary Predicate) | |-----------------------|--------------------------------------------------------------------| | 510(k) Number | K103322, K110319 | | 510(k) Clearance Date | December 06, 2010, April 14, 2011 | | Device Trade Name | MustangTM Balloon Dilatation Catheter | | 510(k) Number | K103751 | | 510(k) Clearance Date | March 22, 2011 | ### 4. Device Description The Fortrex™ 0.035" PTA Balloon Catheter (Fortrex Catheter) is an over the wire (OTW) 0.035"dual lumen catheter with a distally mounted semi-compliant inflatable balloon and an atraumatic tapered tip. The manifold includes a lumen marked "THRU". This is the central lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire with a maximum diameter of 0.035". The lumen marked "BALLOON" is the {4}------------------------------------------------ ## Fortrex™ 0.035" OTW PTA Balloon Catheter 510(k) Summary Image /page/4/Picture/1 description: The image contains the word "COVIDIEN" in a sans-serif font. To the left of the word is a blue square with a lighter blue cross shape in the center. The word "COVIDIEN" is in a dark blue color. The image appears to be a logo or branding element. balloon inflation lumen, which is used to inflate and deflate the balloon with a mixture of contrast medium and saline solution. The balloon has two radiopague markers for positioning the balloon relative to the stenosis. The radiopaque marker bands indicate the dilating or working section of the balloon. The Fortrex Catheter is available with balloon diameters 4.0 mm to 12.00 mm, balloon lengths 20 mm, 40 mm, 80 mm, and 100 mm and with shaft lengths of 40 cm, 80 cm, and 135 cm. ## 5. Indications for Use The Fortrex™ 0.035" OTW PTA Balloon Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature. #### Comparison of Technological Characteristics 6. The proposed Fortrex Catheter uses very similar design and materials as the primary predicate EverCross Catheter. Modifications have been made to the proposed Fortrex device design and materials to improve performance of the balloon through higher burst pressure and improved joint strength and integrity. The Fortrex Catheter shares the following similarities to the predicate devices: - Similar fundamental scientific technology - . Same operating principle - . Similar balloon lengths - . Similar rated burst pressure Additionally, the indications for use, device materials, and manufacturing site and methods are similar between the proposed Fortrex Catheter and the predicate EverCross Catheter. ### 7. Performance Testing Summary To demonstrate substantial equivalence of the proposed Fortrex Catheter to the predicate EverCross device, bench testing and biocompatibility testing was performed. Results from these testing provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. Using internal Risk Analysis procedures, the following performance tests were performed: | Balloon Dimensional Verification | Device Tracking | |-----------------------------------|-----------------------------| | Balloon Rated Burst Pressure | Insertion Force | | Minimum Balloon Burst Strength | Balloon Pull-back Force | | Balloon Fatigue | Reinsertion Force | | Balloon Compliance | Torque Strength | | Catheter Dimensional Verification | Tensile Strength | | Inflation / Deflation Time | Kink | | Wire Movement | Marker Band to Balloon Body | | Radiopacity | | {5}------------------------------------------------ ## Fortrex™ 0.035'' OTW PTA Balloon Catheter 510(k) Summary Image /page/5/Picture/1 description: The image contains the word "COVIDIEN" in a bold, dark blue font. To the left of the word is a square logo. The logo is dark blue with a light blue vertical stripe and a white square in the center. The Fortrex Catheter was tested for biocompatibility per ISO 10993-1 for short duration contact with blood (<24 hours). The testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, complement activation, thromboresistance, partial thromboplastin time, platelet/leukocyte counts, mutagenicity, lymphoma and micronucleus assay. The Fortrex catheter met all acceptance criteria for the performance testing. The results from these tests demonstrate that the technological characteristics and performance criteria of the Fortrex Catheter are comparable to the predicate EverCross Catheter and that the Fortrex Catheter performs in a manner equivalent to the predicate EverCross device currently on the market. ### 8. Conclusions Based on the intended use, technological characteristics, safety and performance testing included in this submission, Covidien considers the proposed Fortrex Catheter to be substantially equivalent to the currently marketed EverCross™ .035" OTW PTA Dilatation Catheter (K103322, K110319) and Mustang™ Balloon Dilatation Catheter (K103751).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...