N-DEX FREE NIGHTHAWK BLACK NITRILE MEDICAL EXAMINATION GLOVE, POWDER -FREE, MODEL# 7700PFT

K082125 · Best Glove, Inc. · LZA · Nov 21, 2008 · General Hospital

Device Facts

Record IDK082125
Device NameN-DEX FREE NIGHTHAWK BLACK NITRILE MEDICAL EXAMINATION GLOVE, POWDER -FREE, MODEL# 7700PFT
ApplicantBest Glove, Inc.
Product CodeLZA · General Hospital
Decision DateNov 21, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The N-DEX® NightHawk® Black Nitrile Powder-Free Medical Examination Glove is a disposable device intended for medical purposes; that is worn on the examiner's hand, to prevent contamination between the patient and examiner (21 CFR 880.6250).

Device Story

N-DEX® NightHawk® is a disposable, non-sterile, powder-free black nitrile medical examination glove. Worn by clinicians on the hands during patient examinations to serve as a protective barrier, preventing cross-contamination between patient and examiner. Device is used in clinical settings. Output is the physical barrier provided by the glove material. Benefits include protection against contamination and suitability for environments where powder is undesirable.

Clinical Evidence

Bench testing only. Device demonstrated compliance with ASTM D 6319-00a, ASTM D 6124-06, and ASTM D 5151-06. Biocompatibility testing included ISO 10993-10 skin irritation and sensitization studies (passing) and ISO 10993-5 cytotoxicity study (passing).

Technological Characteristics

Nitrile rubber material. Powder-free. Non-sterile. Conforms to ASTM D 6319 (physical dimensions/properties), ASTM D 6124 (residual powder), and ASTM D 5151 (water tightness).

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hands to prevent contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K0+2/25 510(k) Submission For Black Nitrile Powder-Free Medical Exam Glove # Summary Of Safety And Effectiveness ### A. Information l. 2. | Submitter's | | | |----------------------------|-----------------------------------------------------------|-------------| | Name: | Best Glove, Inc. | NOV 21 2008 | | Address: | 579 Edison Street<br>Menlo, GA 30731-6335 | | | Telephone Number: | 706-862-6712 | | | Contact person: | Neil Dow | | | Name of Device | | | | Trade or Proprietary name: | Nitrile Powder-Free Medical Examination<br>Glove (Black). | | | Common or Usual name: | Non-sterile Powder-Free Patient Examination<br>glove. | | | Classification Name: | Patient Examination Glove (80LZA, 21 CFR<br>880.6250) | | - 3. Predicate Device: Nitrile Powder-Free medical Examination Glove (green) Submission Number: K012899 - 4. Description Of Device The N-DEX® NightHawk® Black Nitrile Powder-Free Medical Examination Glove is a disposable device intended for medical purposes; that is worn on the examiner's hand, to prevent contamination between the patient and examiner. - 5. Statement of intended use, including descriptions of the disease or conditions that the device will address: This is a disposable device intended for medical purposes that is worn on the examiners hands to prevent contamination between the patient and the examiner. Powder-Free examination gloves are suitable in situations where powder is not desirable. - 6. Explanation of similarities or differences to predicate device The proposed device is identical to the predicate device except for the following: The proposed device has been rendered black instead of green. ### B. If SE decision is based on performance: {1}------------------------------------------------ ## 1. Non-Clinical Tests | Specification | Predicate Device | Proposed Device | |----------------------------------|-------------------------------------------------------------|-------------------------------------------------------------| | | Nitrile Powder-Free<br>Medical Examination<br>Glove (green) | Nitrile Powder-Free<br>Medical Examination<br>Glove (black) | | Performance standards (conforms) | ASTM D 6319-00a<br>ASTM D 6124-01 | ASTM D 6319-00a<br>ASTM D 6124-06 | | Water tightness (conforms) | ASTM D 5151-99 | ASTM D 5151-06 | #### 2. Clinical Tests (Animal Studies) | Biocompatibility | Predicate Device | Proposed Device | |-----------------------------------------------------|------------------|-----------------| | ISO Skin Irritation Study (ISO 10993-10) | Passes | Passes | | ISO Closed Patch Sensitization Study (ISO 10993-10) | Passes | Passes | | Cytotoxicity Study (ISO 10993-5) | Not tested | Passes | ### REPORTS OF SAFETY OR EFFECTIVENSS DATA OBTAINED (With specific reference to adverse effects and complications) ### See Section J: Biocompatibility Testing. ## CONCLUSIONS DRAWN FROM NON CLINICAL AND CLINICAL TESTS THAT DEMONSTRATE SAFETY AND EFFECTIVENESS, AND PERFORMANCE EQUAL TO OR BETTER THAN THE PREDICATE PRODUCT The Nitrile Powder-Free Medical Examination Glove (black) has been carefully compared to a legally marketed device in the 510(k). The data summaries indicate that the proposed product meets or exceeds accepted scores for the predicate product in both physical and nonclinical tests and satisfies the requirements for a safe and effective powderfree medical glove. {2}------------------------------------------------ Pursuant to 21 C.F.R. 807.87 (k), I, Neil Dow, Regulatory Affairs and Quality Assurance Manager, certify that to the best of my knowledge and belief, and based upon the data and information submitted to me in the course of my responsibilities as Regulatory Affairs and Quality Assurance Manager for Best Glove, Inc, and in reliance thereupon, the data and information submitted in this premarket notification are truthful and accurate and that no facts material to a review of the substantial equivalence of this device have been knowingly omitted from this submission. Hii Dow Neil Dow Regulatory Affuirs and Quality Assurance Manager 24 Sep 2008 --- **Date** Section K {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Neil Dow Regulatory Affairs & Quality Assurance Manager Best Glove, Incorporated 579 Edison Street Menlo, Georgia 30731-6335 United States NOV 2 1 2008 Re: K082125 Trade/Device Name: N-DEX® NightHawk® Black Nitrile Powder-Free Medical Examination Glove, Non-Sterile Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: I.Z.A Dated: September 30, 2008 Received: October 20, 2008 Dear Mr. Dow: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration, If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Dow Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice rcquirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA linding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Conter for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Clas Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Indication for Use 510(k) Number (if known): Device Name: N-DEX® NightHawk® Black Nitrile Powder-Free Medical Examination Glove, Non-Sterile Indications For Use: The N-DEX® NightHawk® Black Nitrile Powder-Free Medical Examination Glove is a disposable device intended for medical purposes; that is worn on the examiner's hand, to prevent contamination between the patient and examiner (21 CFR 880.6250). Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use x (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office Of device Evaluation (ODE) Shiela A Murphy, D Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices 510(k) Number: Section D Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...