NON STERILE, POWDER FREE NITRILE EXAMINATION GLOVE, BLUE, POLYMER COATING WITH CHEMOTHERAPY CLAIM

K080520 · Siam Sempermed Corp., Ltd. · LZA · Apr 30, 2008 · General Hospital

Device Facts

Record IDK080520
Device NameNON STERILE, POWDER FREE NITRILE EXAMINATION GLOVE, BLUE, POLYMER COATING WITH CHEMOTHERAPY CLAIM
ApplicantSiam Sempermed Corp., Ltd.
Product CodeLZA · General Hospital
Decision DateApr 30, 2008
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination betweeen patient and examiner.

Device Story

Non-sterile, powder-free blue nitrile examination glove with polymer coating; designed for use by healthcare professionals during patient examinations to provide a barrier against contamination. Device features chemotherapy drug resistance. Used in clinical settings; worn on examiner's hands. Provides protection for both patient and examiner by preventing cross-contamination. Performance verified against ASTM D6319-00a and ASTM D6124-2006 standards.

Clinical Evidence

Bench testing only. No clinical data. Performance verified via ASTM D6319-00a (dimensions, physical properties, pinholes) and ASTM D6124-2006 (powder residue). Biocompatibility testing (primary skin irritation and guinea pig sensitization) passed.

Technological Characteristics

Nitrile material; powder-free; blue color; polymer coating. Standards: ASTM D6319-00a (dimensions, physical properties, pinholes), ASTM D6124-2006 (powder residue). Biocompatibility: Primary skin irritation and guinea pig sensitization testing passed.

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Suitable for use with chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K080520 APR 3 0 2008 # 510 (k) Summary As Required by 21 section 807.92 (c) - 1. Submitter Name: Siam Sempermed Corp., Ltd - 2. Address: 110 Kanjanavanit Road. Pathong Hatyai Songkhla.Thailand 90230 - (+66) 74 291 648 3. Phone: - 4. Fax: (+66) 74 291 650 - Contract Person: Mrs. Sureerat Choosri (Product Manager) న్. - 6. Date summary prepared: December 30, 2005 7. Official Correspondent: Sempermed USA Inc. | 8. Address: | 13900 49th Street North<br>Clearwater, USA, FL 33762 | |-------------|------------------------------------------------------| | 9. Phone: | 727 787 7250 | | 10. Fax: | 727 787 7558 | - 11. Contact person: Mr. William E. Harris 12. Device Trade or Proprietary Name: Non sterile, Powder free Nitrile Examination Glove, Blue, Polymer coating with chemotherapy claim. 13. Device Common or usual name: Examination glove 14. Device Classification Name: Nitrile Patient Examination Glove (Powder free, Blue color) - 15. Description of the Device: Non sterile, Powder free Nitrile Examination Glove, Blue ,Polymer coating with Chemotherapy claim . #### 16. Intended use of the device: This device is a disposable device intended for medical purpose that is worn on the examiner 's hand to prevent contamination between patient and examiner. #### 17. Summary of The Technological Characteristics of The devices: Non sterile, Powder free Nitrile Examination Glove, Blue ,Polymer coating with Chemotherapy claim are summarized with the following technological characteristics compared to ASTM or equivalent standards. | CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | |-----------------------|---------------------------------------|---------------------| | Dimensions | ASTM D 6319-00a (Re-approved<br>2005) | Meets | | Physical Properties | ASTM D 6319-00a (Re-approved<br>2005) | Meets | | Freedom from pinholes | ASTM D 6319-00a (Re-approved<br>2005) | Meets | | Powder Free Residue | ASTM D 6124-2006 | $\u2264$ 2 mg/glove | | Biocompatibility | Primary Skin Irritation in Rabbits | Passes | | | Guinea Pig Sensitization | Passes | #### 18. Substantial Equivalents Based on Assessment of Non-Clinical Performance Data 31 {2}------------------------------------------------ The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above. #### 19. Conclusion : It can be concluded that the Non sterile, Powder free Nitrile Examination Glove, Blue , Polymer coating with chemotherapy claim will perform according to the glove performance standards referenced in section 17 above and meet ASTM standards, and FDA requirements. Consequently, this device is substantially equivalent to currently marketed devices. This device is safe and effective as the predicate device Siam Sempermed Non sterile Powder free Nitile Examination Glove Blue with Polymer coaling. Indeed, it is equivalent. 1This is better expressed in the tabulated comparison as below. Technical comparison of specific elements is attached in the main submission. | FDA file reference number | 510k number | | |-----------------------------------------------------|---------------------------------------------------------------------------------------------------|--| | Attachments inside notification<br>submission file | REFER TO APPENDIX 1 | | | TECHNOLOGICAL<br>CHARACTERISTICS | Comparison result<br>REFER TO ADDITIONAL TECHNICAL<br>COMPARATIVE TABLE WITHIN 510K<br>SUBMISSION | | | Indications for use | Identical | | | Target population | Identical | | | Design | Identical | | | Materials | Identical | | | Performance | Identical | | | Sterility | Identical (Not applicable) | | | Biocompatibility | Identical | | | Mechanical safety | Identical | | | Chemical safety | Identical | | | Anatomical sites | Identical | | | Human factors | Identical | | | Energy used and/or delivered | Identical (Not applicable) | | | Compatibility with environment and<br>other devices | Identical | | | Where used | Identical | | | Standards met | Identical | | | Electrical safety | Identical (not applicable) | | | Thermal safety | Identical (not applicable) | | | Radiation safety | Identical (not applicable) | | ? ? {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the bird-like figure. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 3 0 2008 Siam Sempermed Corporation, Limited C/O Mr. William E. Harris President and Chief Executive Officer Sempermed USA Incorporated 13900 49th Street North Clearwater, Florida 33762 Re: K080520 Trade/Device Name: Non-Sterile, Powder-Free Nitrile Examination Glove, Blue with Polymer Coating, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: April 17, 2008 Received: April 23, 2008 Dear Mr. Harris: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register. {4}------------------------------------------------ #### Page 2 - Mr. Harris Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Sudite Y. Michaudmd. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use ## 510(k) Number (if known): K0805 20 Non-sterile, Powder-Free Nitrile Examination Glove, Blue with Polymer Device Name: Coating, Tested for Use with Chemotherapy Drugs Indications For Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination betweeen patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR × Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Shila A. Murphy, MD (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page 1 of 1 510(k) Number: K081520 12
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...