The Vantage software has been developed to aid clinicians in the assessment and quantification of pathologies derived from PET and PET-CT brain scans. The software is deployed via Advantage Windows Workstation and PET-CT scanner console and is organized as a series of workflows steps, which are specific to use with particular drug and disease combination. The software aids in the assessment of human brain scans enabling automated analysis through quantification of the comparison of local peak activity values at standardized anatomical locations with the corresponding reference normal peak activity in age stratified control subjects. The resulting quantification is presented to the user through 3D stereotactic surface projection maps of the brain. The package allows the user to generate information regarding relative changes in PET-FDG metabolic activity between a subjects images and age stratified controls, which may be resulting from brain function alterations by neurodegenerative processes. CT fusion to PET offers visualization of structural abnormalities, which may result from brain injury, trauma, disease or dysfunction, such as subdural heratom tumor, stroke (or cerebrovascular disease) etc. Such tools may aid the physician in the rule out process.
Device Story
Vantage PET Neuro Software; post-processing tool for PET/CT brain scans. Inputs: patient PET/CT images. Operation: realigns PET volume to standard atlas; compares local peak activity values at standardized anatomical locations against age-stratified normal control database. Output: 3D stereotactic surface projection maps; Z-score quantification of metabolic activity deviations. Used in clinical settings on Advantage Windows Workstation or scanner console by physicians/technicians. Assists clinicians in visualizing structural abnormalities and metabolic changes; aids in rule-out processes for neurodegenerative disease, trauma, or stroke. Benefits: provides objective quantitative assessment of brain function alterations.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
Software-based post-processing tool for PET/CT imaging. Operates on Advantage Windows Workstation and PET-CT scanner consoles. Uses 3D stereotactic surface projection mapping. Quantitative analysis based on comparison of local peak activity values against age-stratified normal control databases.
Indications for Use
Indicated for clinicians to aid in assessment and quantification of pathologies in human brain PET/CT scans. Used for patients with suspected neurodegenerative processes, brain injury, trauma, or cerebrovascular disease. No specific age or gender contraindications stated.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
K062393
# 510(k) Summary of Safety and Effectiveness (in accordance to 21
SEP - 1 200
CFR 807.87(h))
Device Name Proprietary Device Name: Date prepared:
GE Vantage PET Neuro Software May 26, 2006
| Establishment Name and Registration Number of Submitter | |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name: | GE Healthcare (Formerly GE Medical Systems) |
| Registration Number: | 2126677 |
| Corresponding Official: | D. Duersteler<br>GE Healthcare<br>P.O. Box 414<br>Milwaukee, WI 53201<br>Phone: 262-312-7029<br>FAX: 262-312-7144<br>Email: david.duersteler@med.ge.com |
| Device Classification | |
| Classification Code: | 90 KPS |
| Panel Identification: | Radiology |
| Classification Name: | Emission Computed Tomography Syst<br>(Per 21CFR 892.1200) |
Common Name: Classification Class: Reason for 510(k) Submission:
em (Per 21CFR 892.1200) PET/CT Imaging Software Class II Product New device
#### Device Description
The Vantage package includes automated analysis and review of PET and PET-CT brain scans. It has been developed to aid in assessment of human brain scans hrough quantification of the comparison of local peak activity values at standardized anatomical locations with the corresponding reference normal peak activity in age stratified normal database (or asymptomatic control subjects as defined). It rage a basic review of any PET brain scan, realignment of the PET volume to a standard attas and comparison to brain scans derived from control subjects or prior PET and/or MR scans. Results are displayed in user-friendly graphical format as Z-score, where Zscore is the scale of the deviation from the normal mean, relative to corresponding standard deviation.
# Identification of Legally Marketed Equivalent Devices
| NeuroQ | Syntermed, Inc. | K04102 |
|---------|------------------------|---------|
| Scenium | Mirada Solutions, Ltd. | K042863 |
General Electric Company
Image /page/0/Picture/14 description: The image shows a black and white logo. The logo appears to be a stylized design, possibly representing a company or brand. The design is circular and contains intricate patterns and shapes within it. The image is somewhat pixelated, suggesting it may be a low-resolution or scanned image.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the word "Gammailhouse". The text is in a simple, sans-serif font. The image is a simple black and white picture.
Image /page/1/Picture/1 description: The image shows a logo with the letters 'GE' in a stylized font inside a circle. To the left of the circle are the letters 'P' and 'M' stacked vertically. To the right of the circle is the number '53201'. The logo appears to be a company or brand identifier, with the number possibly representing a product code or identifier.
510(k) Summary of Safety and Effectiveness May 26, 2006 Page 2
### Comparison with Predicate Devices
Vantage as compared with the above predicate devices provides a clinically focused software solution for brain analysis with PET and PET/CT imaging. Similar to Scenium and Neuro Q, Vantage™ is a post processing tool that assists in regional assessment of human brain scans by providing comparison between a patient brain scan to a brain scan derived from FDG-PET studies of a defined group of control subjects, where the asymptomatic control group is as defined and age stratified by a neurologist with medical expertise. The algorithms used in Vantage™ to calculate quantitative measures compares local peak activity values instead of average peak activity values used by predicate devices; at standardized anatomical locations with the corresponding reference normal peak activity in age stratified control subjects.
#### Indications for Use of Device
The GE Vantage PET Neuro Software has been developed to aid clinicians in the assessment and quantification of pathologies derived from PET and PET-CT brain scans. The software is deployed via Advantage Windows Workstation and PET-CT scanner console and is organized as a series of workflows steps, which are specific to use with particular drug and disease combination.
The software aids in the assessment of human brain scans enabling automated analysis through quantification of the comparison of local peak activity values at standardized anatomical locations with the corresponding reference normal peak activity in age stratified control subjects. The resulting quantification is presented to the user through 3D stereotactic surface projection maps of the brain. The package allows the user to generate information regarding relative changes in PET-FDG metabolic activity between a subjects images and age stratified controls, which may be resulting from brain function alterations by neurodegenerative processes.
CT fusion to PET offers visualization of structural abnormalities, which may result from brain injury, trauma, disorder, disease or dysfunction, such as subdural hematoma tumor, stroke (or cerebrovascular disease) etc. Such tools may aid the physician in the rule out process.
### Conclusion
In the opinion of General Electric Medical Systems, the GE Vantage PET Neuro Software is substantially the same in design, materials, energy sources, and technology, does not introduce new safety concerns, performs as well as currently marketed devices, and is therefore substantially equivalent in terms of safety and effectiveness to the currently marketed Syntermed NeuroQ (K041022) and Miraga Solutions Scenium (K042863) products.
{2}------------------------------------------------
General Electric Company
P.O. Box 414
Milwaukee, WI 53201
:
、
・
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing and tail feathers. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## SEP - 1 2006
GE Healthcare % Mr. Tamas Borsai Responsible Third Party Official TÜV Rheinland of North America 12 Commerce Road NEWTON CT 06470
Re: K062393
Trade/Device Name: Vantage Regulation Number: 21 CFR 8892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: August 11, 2006 Received: August 16, 2006
#### Dear Mr. Borsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a black and white circular logo or seal. The text "1906 - 2006" is at the top of the logo. The letters "FDA" are in a bold, stylized font in the center of the logo. Below the letters, the word "Centennial" is written in a cursive font. Three stars are below the word "Centennial".
Promoting Public Hoalth
{4}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing proctice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logal|| marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE
#### 510(k) Number (if known): _ Ob23393 Device Name: Vantage Indications for Use:
The Vantage software has been developed to aid clinicians in the assessment and quantification of pathologies derived from PET and PET-CT brain scans. The software is deployed via Advantage Windows Workstation and PET-CT scanner console and is organized as a series of workflows steps, which are specific to use with particular drug and disease combination.
The software aids in the assessment of human brain scans enabling automated analysis through quantification of the comparison of local peak activity values at standardized anatomical locations with the corresponding reference normal peak activity in age stratified control subjects. The resulting quantification is presented to the user through 3D stereotactic surface projection maps of the brain. The package allows the user to generate information regarding relative changes in PET-FDG metabolic activity between a subjects images and age stratified controls, which may be resulting from brain function alterations by neurodegenerative processes.
CT fusion to PET offers visualization of structural abnormalities, which may result from brain injury, trauma, disease or dysfunction, such as subdural heratom tumor, stroke (or cerebrovascular disease) etc. Such tools may aid the physician in the rule out process.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801-109)
OR Over-The-Counter Use_
David E. Lyon
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices K062393 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.