K052864 · Tyrx Pharma, Inc. · FTL · Dec 22, 2005 · General, Plastic Surgery
Device Facts
Record ID
K052864
Device Name
TYRX SURGICAL MESH, MODEL SMPC-0501
Applicant
Tyrx Pharma, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Dec 22, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TyRx Surgical Mesh is intended for the repair of hernias and other abdominal fascial deficiencies requiring the addition of a reinforcing or bridging material to obtain the desired surgical result. Examples of applications where TyRx Surgical Mesh may be used include, but are not limited to: inguinal, femoral, umbilical, abdominal, incisional and intramuscular hernias and muscle flap reinforcement.
Device Story
TyRx Surgical Mesh is a sterile, dual-component prosthesis for soft tissue reconstruction. It consists of a non-resorbable polypropylene mesh coated with a bioresorbable polyarylate polymer. The coating, comprising approximately 10% of the device weight, provides stiffness to facilitate intraoperative handling. Once implanted, the coating resorbs within approximately 90 days, leaving a permanent, lightweight polypropylene mesh. The device is used by surgeons in clinical settings for hernia repair and fascial reinforcement. It provides mechanical support to the surgical site, promoting tissue ingrowth as demonstrated in animal studies. The benefit is a combination of ease of handling during placement and long-term tissue reinforcement.
Clinical Evidence
No clinical data. Evidence consists of non-clinical laboratory testing performed in accordance with FDA guidance for surgical mesh and animal testing. Animal studies demonstrated satisfactory tissue ingrowth compared to control devices via histopathology.
Technological Characteristics
Dual-component sterile prosthesis. Materials: knitted polypropylene filaments and bioresorbable polyarylate coating. Coating represents ~10% of total device weight. Resorption time: ~90 days. Form factor: surgical mesh. Intended for permanent implantation.
Indications for Use
Indicated for patients requiring repair of hernias and abdominal fascial deficiencies, including inguinal, femoral, umbilical, abdominal, incisional, and intramuscular hernias, and muscle flap reinforcement.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{1}------------------------------------------------
TyRx Pharma Inc.
TyRx Surgical Mesh - Premarket Notification [510(k)] Submission
K052864 Page
November 9, 2005 1/2
# SECTION 2. 510(K) SUMMARY
510(k) Summary (as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)3A of the Safe Medical Devices Act of 1990, TyRx Pharma, Inc. is providing a summary of the safety and effectiveness information available for TyRx Surgical Mesh, as well as the substantial equivalence decision making process used for TyRx Surgical Mesh.
Sponsor/Applicant Name and Address: TyRx Pharma, Inc. 1 Deer Park Drive, Suite G Monmouth Junction, NJ 08852
Sponsor Contact Information: Mason Diamond, DDS Vice President, Clinical and Regulatory Affairs Phone: 732-246-8676 Fax: 732-246-8677 E-mail: mason@tyrxpharma.com
Date of Preparaton of 510(k) Summary: October 6, 2005
New Device Trade/Proprietary Name: TyRx Surgical Mesh
Device Common/Classification Name: Polymeric Surgical Mesh
Predicate Devices Name and 510(k) Numbers: VYPRO Mesh VICRYL-PROLENE partially absorbable synthetic surgical mesh (K002672) PROLENE polypropylene mesh nonabsorbable synthetic mesh (K962530) GORE POLYPROPYLENE HERNIA Mesh (K043081)
Device Description:
TyRx Surgical Mesh is dual component (resorbable and non-resorbable), sterile prosthesis designed for the reconstruction of soft tissue deficiencies. TyRx Surgical Mesh is constructed of a non-resorbable mesh comprised of knitted filaments of polypropylene and a bioresorbable polyarylate coating on the mesh. The resorbable coating represents approximately 10% of the total weight of the device. The purpose of the resorbable coating is to provide additional stiffness to the mesh in order to facilitate interoperative handling during placement. Once placed, the polymer resorbs in approximately 90 days leaving a lighter permanent mesh incorporated into the tissue. In
{2}------------------------------------------------
TyRx Pharma Inc. TyRx Surgical Mesh - Premarket Notification [510(k)] Submission
K052864 Page
November 9, 2005 2/2
addition, animal testing demonstrated that TyRx Surgical Mesh would achieve satisfactory tissue ingrowth compared to commercial surgical mesh.
#### Intended Use:
TyRx Surgical Mesh is intended for the repair of hernias and other abdominal fascial deficiencies requiring the addition of a reinforcing or bridging material to obtain the desired surgical result. Examples of applications where TyRx Surgical Mesh may be used include, but are not limited to: inguinal, femoral, umbilical, abdominal, incisional and intramuscular hernias and muscle flap reinforcement.
#### Performance Data:
Non-clinical laboratory testing was performed in accordance with the FDA guidance document "Guidance for the Preparation of a Premarket Notification for a Surgical Mesh" demonstrating that the device is comparable to standard surgical mesh devices that are indicated for hernia repair and other abdominal fascial or muscular deficiencies requiring the addition of a reinforcing or bridging material to obtain the desired surgical result. The TyRx Surgical Mesh and the polyarylate polymer have been demonstrated to be biocompatible. In addition, animal testing demonstrated that TyRx Surgical Mesh will achieve satisfactory tissue ingrowth compared to control as evidenced by histopathology.
### Conclusions:
Based on the 510(k) summaries (21 CFR 807) and the information provided herein, we conclude that the new device is substantially equivalent to the Predicate Device under the Federal Food Drug and Cosmetic Act.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the caduceus symbol.
DEC 2 1 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Mason W. Diamond, D.D.S. Vice President, Clinical & Regulatory Affairs TyRx Pharma, Inc. 1 Deer Park Drive, Suite G Monmouth Junction, New Jersey 08852
Re: K052864
Trade/Device Name: TyRx Surgical Mesh, Model SMPC-0501 Regulatory Number: 21 CFR 878.3300 Regulatory Name: Implantable clip Regulatory Class: II Product Code: FTL Dated: December 2, 2005 Received: December 6, 2005
Dear Dr. Diamond:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 – Dr. Diamond
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis icher will anow you to ough marketing of substantial equivalence of your device to a legally premits to the are month in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at (240) 276-0115. Also, please note the regulation entitled, Contact the Office of Compullier as (21 the Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general miorination on Jour respumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Barbara Richards
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
November 9, 2005
TyRx Pharma Inc. TyRx Surgical Mesh - Premarket Notification [510(k)] Submission
## SECTION 1. INDICATION FOR USE STATEMENT
Indications for Use
510(k) Number (if known): K052864
Device Name: _________________ TyRx Surgical Mesh
Indications for Use:
Is intended for the repair of hernias and other abdominal fascial deficioncies requiring the addition of a reinforcing or bridging material to obtain the desired surgical result. . Examples of applications where TyRx Surgical Mesh may be used include, but are not limited to: inguinal, femoral, umbilical, abdominal, incisional and intramuscular hernias and muscle flap reinforcement.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Anbare Bnchmp for Xxm
Division Sign Off
Page 1 of \
Division of General. Restorative. and Neurological Devices
510(k) Number K052864
CONFIDENTIAL - PROPRIETARY INFORMATION
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.