Exogenesis Hernia Mesh

K191545 · Exogenesis Corporation · FTL · Sep 26, 2019 · General, Plastic Surgery

Device Facts

Record IDK191545
Device NameExogenesis Hernia Mesh
ApplicantExogenesis Corporation
Product CodeFTL · General, Plastic Surgery
Decision DateSep 26, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Exogenesis Hernia Mesh is indicated for the repar of abdominal wall hernias and abdominal wall deficiencies that require the addition of reinforcing material to obtain the desired surgical result. The Exogenesis Hernia Mesh is indicated for the repair of abdominal wall hernia defects, including inguinal (direct & indirect). The Exogenesis Hernia Mesh is not indicated for transvaginal pelvic organ prolapse repair. The Exogenesis Hernia Mesh is intended for single patient one-time use only.

Device Story

Sterile, non-resorbable surgical mesh; composed of knitted polypropylene filaments; provides mechanical support for soft tissue reconstruction; used for abdominal wall hernia repair. Device undergoes proprietary Accelerated Neutral Atom Beam (ANAB) surface treatment to increase microscopic filament surface area. Implanted by surgeons in clinical settings to reinforce tissue defects; promotes tissue ingrowth while minimizing inflammatory response. Output is physical reinforcement of fascial deficiencies; provides long-term tissue support. Benefits include reduced implant mass and effective hernia repair.

Clinical Evidence

Bench and animal testing only. Bench testing compared physical/mechanical properties (thickness, pore size, tear resistance, tensile strength, suture pull-out, burst strength) of processed vs. untreated mesh and against predicate. Animal study (90 and 180 days) assessed localized tissue interaction and burst strength; results showed processed and untreated meshes were comparable and exceeded native abdominal wall burst strength.

Technological Characteristics

Material: Polypropylene. Construction: Warp-knitted filaments. Surface treatment: Accelerated Neutral Atom Beam (ANAB) processing. Form factor: Flexible mesh. Sterilization: Sterile. Connectivity: None (implantable).

Indications for Use

Indicated for patients requiring repair of abdominal wall hernias and deficiencies, including direct and indirect inguinal hernias. Not indicated for transvaginal pelvic organ prolapse repair.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal for the Department of Health & Human Services - USA. To the right of the seal, there is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue. September 26, 2019 Exogenesis Corporation % Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, Massachusetts 02180 Re: K191545 Trade/Device Name: Exogenesis Hernia Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: June 10, 2019 Received: June 11, 2019 Dear Ms. O'Connell: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K191545 Device Name Exogenesis Hernia Mesh Indications for Use (Describe) The Exogenesis Hernia Mesh is indicated for the repar of abdominal wall hernias and abdominal wall deficiencies that require the addition of reinforcing material to obtain the desired surgical result. The Exogenesis Hernia Mesh is indicated for the repair of abdominal wall hernia defects, including inguinal (direct & indirect). The Exogenesis Hernia Mesh is not indicated for transvaginal pelvic organ prolapse repair. The Exogenesis Hernia Mesh is intended for single patient one-time use only. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size:10pt"><svg height="15" width="15"><rect fill="none" height="15" stroke="black" width="15"></rect></svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:10pt"><svg height="15" width="15"><rect fill="none" height="15" stroke="black" width="15"></rect></svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ a Exogenesis Hernia Mesh Premarket Notification: Traditional 510(k) ## 510(k) Summary ### Exogenesis Hernia Mesh | Submitter: | Exogenesis Corporation<br>20 Fortune Drive<br>Billerica, MA 01821<br>978-439-0120 | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Maureen O'Connell<br>O'Connell Regulatory Consultants, Inc.<br>44 Oak Street<br>Stoneham, MA 02180<br>Phone: 978-207-1245 | | Date Prepared: | September 18, 2019 | | Trade Name: | Exogenesis Hernia Mesh | | Classification Name: | Surgical Mesh | | Regulation Number: | 21 CFR 878.3300 | | Product Code: | FTL | | Predicate Device: | Ethicon, Inc. Prolene Polypropylene Mesh Non-Absorbable<br>Synthetic Surgical Mesh – K180829 | | Device Description: | The Exogenesis Hernia Mesh is a sterile prosthesis designed to<br>provide mechanical support for reconstruction of soft tissue<br>deficiencies. The device is intended for the repair of hernias and<br>other fascial deficiencies requiring the addition of reinforcing or<br>bridging material to obtain the desired surgical result. The<br>Exogenesis Hernia Mesh is composed of knitted filaments of<br>extruded polypropylene, and knitted to provide elasticity in both<br>directions. This construction permits the mesh to be cut into any<br>desired shape or size without unraveling. The Exogenesis Hernia<br>Mesh is treated with a proprietary technology to modify the surface<br>of the mesh to increase surface area of the filaments on a<br>microscopic level. | {4}------------------------------------------------ Indications for Use: The Exogenesis Hernia Mesh is indicated for the repair of abdominal wall hernias and abdominal wall deficiencies that require the addition of reinforcing material to obtain the desired surgical result. The Exogenesis Hernia Mesh is indicated for the repair of abdominal wall hernia defects, including inguinal (direct & indirect). The Exogenesis Hernia Mesh is not indicated for transvaginal pelvic organ prolapse repair. > The Exogenesis Hernia Mesh is intended for single patient one-time use only. ### Summary of Technological Characteristics: The Exogenesis Hernia Mesh is constructed of polypropylene fibers warp knitted together to form the mesh. The knitting process creates a device with large pores and minimum density and thickness similar to the light weight meshes currently on the market. The result is an implant which allows tissue ingrowth that reinforces the tissue defect, while minimizing the inflammatory response and fibrous encapsulation related to implant mass. The mesh possesses the mechanical and physical properties necessary for long term tissue support. The Exogenesis Hernia Mesh is surface treated using a proprietary process to increase surface area of the filaments on a microscopic level. # Substantial Equivalence Discussion: Exogenesis Corporation believes that the Exogenesis Hernia Mesh described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to a legally marketed predicate device that is a Class II medical device. The following table presents the Exogenesis Hernia Mesh compared with the predicate device which is the Ethicon, Inc. Prolene Polypropylene Mesh Non-Absorbable Synthetic Surgical Mesh cleared in K180829. {5}------------------------------------------------ # | Characteristic | Exogenesis Hernia Mesh | Prolene Polypropylene Mesh Non-<br>Absorbable Synthetic Surgical Mesh | |----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Exogenesis Corporation | Ethicon Inc. | | 510(k) Number | TBD | K180829 | | Class | II | II | | Device Classification<br>Name | Mesh, Surgical, Polymeric | Mesh, Surgical, Polymeric | | Regulation Number | 878.3300 | 878.3300 | | Product Code | FTL | FTL | | Use | Prescription | Prescription | | Indications for Use | ... is indicated for the repair of abdominal wall<br>hernias and abdominal wall deficiencies that<br>require the addition of reinforcing material to<br>obtain the desired surgical result. The<br>Exogenesis Hernia Mesh is indicated for the<br>repair of abdominal wall hernia defects,<br>including inguinal (direct & indirect). The<br>Exogenesis Hernia Mesh is not indicated for<br>transvaginal pelvic organ prolapse repair.<br><br>...is intended for single patient one-time use<br>only. | ...is indicated for the repair of abdominal<br>wall hernias and abdominal wall<br>deficiencies that require the addition of a<br>reinforcing material to obtain the desired<br>surgical result. The PROLENE™ Hernia<br>System is indicated for the repair of<br>abdominal wall hernia defects, including<br>inguinal (direct & indirect). | | Device Composition | Polypropylene (non-resorbable) | Polypropylene (non-resorbable) | | Construction | Knitted filaments of extruded polypropylene | Knitted filaments of extruded<br>polypropylene | | Bench Testing for<br>Hernia Mesh | Pass | Pass | | Animal Data | Yes | No | | Biocompatibility | Biocompatible | Biocompatible | | Sterile | Yes | Yes | | Characteristic | Exogenesis Hernia Mesh | Prolene Polypropylene Mesh Non-Absorbable Synthetic Surgical Mesh | | Single Use | Yes | Yes | ## Substantial Equivalence Comparison {6}------------------------------------------------ The intended use of both the Exogenesis Hernia Mesh and the Prolene predicate device is as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue. The Exogenesis Hernia Mesh and the Prolene predicate device are both mesh devices composed of polypropylene which is non-resorbable. Both devices are made from knitted filaments of extruded polypropylene. Both devices are provided sterile and are for single use only. The Exogenesis Hernia Mesh has an additional processing step compared to the Prolene, which is Accelerated Neutral Atom Beam (ANAB) processing. This step does not add any new materials of construction and does not change the chemical composition of the product ### Performance Testing Biocompatibility testing has been performed on the final version of the Exogenesis Hernia Mesh which supports the substantial equivalence of the material including the processing. Performance testing is provided which compares the physical characteristics of the Exogenesis Hernia Mesh in both its untreated state and with processing as well as compared to the predicate device. There were no significant differences between the Exogenesis untreated Hernia Mesh and the Exogenesis processed Hernia Mesh in physical or mechanical parameters. These results confirm that the processing has no measurable impact on the structure and function of the polypropylene mesh and supports substantial equivalence. In addition, there were no significant differences between the Exogenesis Hernia Mesh and the Prolene predicate device with respect to mesh thickness and single-fiber thickness. However, the pore size for the Exogenesis Hernia Mesh was greater than that of the predicate device. In performance testing, there were no differences between the Exogenesis Hernia Mesh and the Prolene predicate device in tear resistance. Given that Prolene is a stiffer mesh, it demonstrated higher tensile, suture pull-out and burst strength than the Exogenesis Hernia Mesh. However, published results for the Prolite Ultra showed that the Exogenesis mesh was higher than the ProLite Ultra. In addition, all meshes reported higher burst strength than that of abdominal muscle. The results of this testing confirm that the Exogenesis Hernia Mesh is substantially equivalent to the predicate device. Additionally, animal testing was performed which assessed the localized tissue interaction and the mechanical properties of the untreated Exogenesis Hernia Mesh compared to the processed Exogenesis Hernia Mesh. Results at Day 90 and Day 180 showed that both versions of the Exogenesis Hernia Mesh were comparable in burst strength, and both were much greater in burst strength as compared to the native abdominal wall. Therefore, biocompatibility testing and bench and animal testing support substantial equivalence. {7}------------------------------------------------ Therefore, in summary, the new device has the same intended use and similar technological characteristics to the predicate device and is therefore, substantially equivalent. Performance data support substantial equivalence of the different technological characteristics.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%