K042576 · Instrumentation Laboratory CO · GGP · Dec 2, 2004 · Hematology
Device Facts
Record ID
K042576
Device Name
COATEST SP FVIII
Applicant
Instrumentation Laboratory CO
Product Code
GGP · Hematology
Decision Date
Dec 2, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7290
Device Class
Class 2
Indications for Use
Coatest SP FVIII is intended for the photometric determination of factor VIII activity in citrated plasma, such as when identifying factor VIII deficiency or monitoring patients on replacement therapy, as well as for potency estimation of FVIII concentrates. For in vitro diagnostic use.
Device Story
Coatest SP FVIII is an in vitro diagnostic reagent kit for photometric determination of factor VIII activity in citrated plasma. Used in clinical laboratories to identify factor VIII deficiency, monitor patients on replacement therapy, and estimate potency of FVIII concentrates. Operates via photometric measurement of plasma samples; results used by clinicians to assess coagulation status and guide therapeutic decisions. Provides quantitative activity levels to support diagnosis and treatment management.
Clinical Evidence
Bench testing only. Method comparison studies (n=181 vs. manual reference; n=90 vs. automated reference) demonstrated high correlation (r=0.9873 and r=0.9919 respectively). Precision assessed over 80 runs; total CV% ranged from 5.3% to 7.1% across normal and abnormal controls.
Technological Characteristics
Photometric chromogenic assay. Reagents: bovine factors IXa/X, synthetic phospholipids, chromogenic substrate S-2765, thrombin inhibitor I-2581, calcium chloride, Tris buffer with 10% BSA and antimicrobial. Form factor: lyophilized reagents for reconstitution. Connectivity: manual or automated on ACL 9000 series instruments. Measurement: absorbance at 405 nm. Stability: 12-month shelf life at 2-8°C.
Indications for Use
Indicated for the photometric determination of factor VIII activity in citrated plasma for patients requiring identification of factor VIII deficiency or monitoring of replacement therapy, and for potency estimation of FVIII concentrates.
{0}
Page 1 of 12
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K042576
B. Purpose for Submission:
To obtain clearance for the Coatest SP FVIII assay, a photometric assay for the determination of factor VIII activity.
C. Measurand:
Factor VIII
D. Type of Test:
Photometric qualitative and quantitative factor determination
E. Applicant:
Instrumentation Laboratory Company
F. Proprietary and Established Names:
Coatest SP FVIII
G. Regulatory Information:
1. Regulation section:
864.7290
2. Classification:
II
3. Product code:
GGP
4. Panel:
{1}
Page 2 of 12
81 Hematology
## H. Intended Use:
1. Intended use(s):
Coatest SP FVIII is intended for the photometric determination of factor VIII activity in citrated plasma.
2. Indication(s) for use:
Coatest SP FVIII is intended for the photometric determination of factor VIII activity in citrated plasma, such as when identifying factor VIII deficiency or monitoring patients on replacement therapy, as well as for potency estimation of FVIII concentrates. For in vitro diagnostic use.
3. Special conditions for use statement(s):
Not applicable
4. Special instrument requirements:
Not applicable
## I. Device Description:
Coatest SP FVIII is a modified version of Coatest Factor VIII (K833892) reformulated to European Pharmacopoeia Standards. Coatest SP FVIII is a photometric assay containing a chromogenic substrate, S-2765, with EDTA added as a preservative, lyophilized bovine factors IXa and X with bovine albumin added as a stabilizing agent. The device also contains calcium chloride, Tris buffer stock solution containing sodium chloride, bovine serum albumin with added antimicrobial in addition to a mixture of highly purified synthetic phospholipids.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Coatest Factor VIII
2. Predicate 510(k) number(s):
K833892
3. Comparison with predicate:
{2}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Same as predicate and identify factor VIII deficiency, or monitoring patients on replacement therapy as well as for potency estimation for FVIII concentrates | Photometric determination of FVIII activity in citrated plasma |
| Reagent: Factor IXa + X & CaCl2 | Same as predicate | Bovine factors IXa & X with bovine albumin added as stabilizer |
| Storage | Same as predicate | 2-8°C until expiration |
| Linearity | Same as predicate | 0-150% |
| Detection Limit | Same as predicate | 1% factor VIII |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Reagent: Chromogenic substrate | S-2765 + I-2581 with mannitol and EDTA as a preservative | S-2222 + I2581 with mannitol |
| Buffer Stock Solution | Increased BSA concentration from 2% to 10% and added antimicrobial (Ciprofiaxin) in buffer reagent | Tris buffer containing NaCl and BSA |
| Phospholipid | Synthetic phospholipids | Porcine brain emulsion |
# K. Standard/Guidance Document Referenced (if applicable):
# L. Test Principle:
In the presence of calcium and phospholipids, factor X is activated to factor Xa by factor IXa. This generation is greatly stimulated by factor VIII, which may be considered as a cofactor in this reaction. By using optimal amounts of $\mathrm{Ca^{2+}}$ and phospholipids and an excess of factors IXa and X, the rate of activation of factor X is solely dependent on the amount of factor VIII. Factor Xa hydrolyses the chromogenic substrate S-2765 thus liberating the chromophoric group, pNA. The
{3}
Page 4 of 12
color is then read photometrically at 405 nm. The generated factor Xa and thus the intensity of color are proportional to the factor VIII activity in the sample. Hydrolysis of S-2765 by thrombin formed is prevented by the addition of the synthetic thrombin inhibitor, I-2581, together with the substrate.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
**Manual Method:** A precision study was performed with the manual method using HemosIL Normal Control (K021023) for the normal range and HemosIL High Abnormal Control (K021024) diluted with factor diluent (saline) for the low range. Each control level was run in duplicate twice a day over twenty days (n=80).
Statistics calculated according to NCCLS Document EP5-T2
**Acceptance Criteria:**
| | Within Run %CV | Total %CV |
| --- | --- | --- |
| HemosIL Normal Control | < 6% | < 9% |
| HemosIL High Abnormal Control | < 6% | < 9% |
| Control | n | Mean % FVIII | Within run % CV | Between run % CV | Total % CV |
| --- | --- | --- | --- | --- | --- |
| Abnormal | 80 | 14.4 | 4.3 | 3.7 | 5.6 |
| Normal | 80 | 83 | 3.4 | 3.8 | 5.3 |
**Instrument Application:** An additional precision study was performed on an ACL 9000 using HemosIL Normal Control (K021023) for the normal range and HemosIL High Abnormal Control (K021024) diluted with factor diluent (saline) for the low range. Each control level was run in duplicate twice a day over twenty days (n=80).
Statistics calculated according to NCCLS Document EP5-T2
**Acceptance Criteria:**
| | Within Run %CV | Total %CV |
| --- | --- | --- |
| HemosIL Normal Control | < 6% | < 9% |
| HemosIL High Abnormal Control | < 6% | < 9% |
{4}
Page 5 of 12
| Control | n | Mean % FVIII | Within run % CV | Between run % CV | Total % CV |
| --- | --- | --- | --- | --- | --- |
| Abnormal | 80 | 17.3 | 5.7 | 1.0 | 6.3 |
| Normal | 80 | 102 | 4.7 | 3.7 | 7.1 |
b. Linearity/assay reportable range:
Manual Method: A linearity study was performed using a high FVIII sample (150%) diluted in factor VIII deficient plasma to prepare seven concentrations. Each level was tested in quadruplicate with the manual method.
Acceptance criteria: $\mathrm{R}^2\geq 0.99$
A graph of the results with the manual method show linearity throughout the claimed range in the product insert of $0 - 150\%$ factor VIII $(\mathrm{R}^2 >0.9911)$
Instrument Application: An additional linearity study was performed using a high FVIII sample (150%) diluted in factor VIII deficient plasma to prepare six concentrations. Each level was tested in quadruplicate on an ACL 9000.
Acceptance criteria: $\mathrm{R}^2\geq 0.99$
A graph of the results from the ACL 9000 show linearity throughout the claimed range in the instrument application sheet of $0 - 150\%$ factor VIII $(\mathrm{R}^2 > 0.9984)$ .
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Value assignments for the controls and calibrator were determined in multiple runs using specific lots of reagents and against a House Standard, which is traceable to the $4^{\text{th}}$ International Standard for FVIII/vWF (NIBSC Code: 97/586).
d. Detection limit:
Manual Method: Detection limit testing was performed using factor VIII deficient plasma run in replicates of five with the manual method using both the normal and low ranges. The mean value $(n = 5)$ plus three standard deviations were calculated.
NOTE: The low range results support the insert claim of a $1\%$ detection limit given that FVIII deficient plasma runs at $0.0\%$ in the low range.
Mean % FVIII = 0.04 $\mathrm{SD} = 0.0$ Mean + 3SD = 0.04
{5}
Page 6 of 12
Instrument Application: Additional detection limit testing was performed using factor VIII deficient plasma run in replicates of five on the ACL 9000 using both the normal and low ranges. The mean value (n=5) plus three standard deviations were calculated.
NOTE: The low range results support the insert claim of a 1% detection limit given that FVIII deficient plasma runs at 0.0% in the low range.
Mean % FVIII = -0.5
SD = 0.1
Mean + 3SD = -0.2
e. Analytical specificity:
Manual Method: Interference testing was performed using the manual method by spiking levels of each interferent into two different factor VIII plasma sample levels and comparing the results against the unspiked sample results. The two factor VIII levels were prepared using: 1) fresh frozen pooled plasma (FFP) for the normal level and 2) FFP diluted with factor VIII deficient plasma for the 30% level. All sample were tested in triplicate (n=3) with a single lot of Coatest SP FVIII reagents.
Acceptance criteria: Recovery of ± 10% of the unspiked sample result
The data support the interference claims in the Coatest SP FVIII product insert for no significant interference by:
- Triglycerides up to 700 mg/dl
- Bilirubin up to 20 mg/dl
- Hemoglobin up to 100 mg/dl
- Unfractionated heparin up to 1.0 IU/ml
The Coatest SP FVIII product insert includes the limitation that hemolyzed samples in the low range should not be analyzed.
Instrument Application: Additional interference testing was performed on an ACL 9000 by spiking levels of each interferent into two different factor VIII plasma sample levels and comparing the results against the unspiked sample results. The two factor VIII levels were prepared using: 1) HemosIL Normal Control (K021023) for the normal level and 2) HemosIL Normal Control (K021023) diluted with factor VIII deficient plasma for the 30% level. All samples were tested in triplicate (n=3) with a single lot of Coatest SP FVIII reagents.
Acceptance criteria: Recovery of ± 10% of the unspiked sample result
The data support the interference claims in the ACL 8000/9000/10000 instrument application sheet for no significant interference by:
{6}
Page 7 of 12
- Triglycerides up to 900 mg/dl
- Bilirubin up to 20 mg/dl
- Hemoglobin up to 50 mg/dl
- Unfractionated heparin up to 1.0 IU/ml
The ACL instrument application sheet (low range) for Coatest SP FVIII includes the limitation that hemolyzed samples should not be analyzed.
**Lupus anticoagulant**: The product insert for the predicate device, Coatest Factor VIII (K833892), states that due to the high dilutions used, there is no underestimation of FVIII activity in samples containing lupus anticoagulant. To verify this claim for the new Coatest SP FVIII, the results were compared from testing 10 lupus anticoagulant samples in duplicate with the new Coatest SP FVIII versus the predicate Coatest Factor VIII using the manual method. The two tests gave statistically equivalent results. The mean %FVIII of the predicate device was 104 with a SD of 23.2 and the mean %FVIII if Coatest SP was 105 with a SD of 21.6.
**System Sensitivity**:
**Manual Method**: System sensitivity for the low and normal ranges was calculated with the manual method as the absorbance change for 1% of factor VIII activity using the slope of the standard curve.
$\Delta A_{405}$ per 1% of FVIII activity: Low Range = 0.043 & Normal Range = 0.010.
**Instrument Application**: System sensitivity for the low and normal ranges was calculated on the ACL 9000 as the absorbance change for 1% of factor VIII activity using the slope of the standard curve.
$\Delta A_{405}$ per 1% of FVIII activity: Low Range = 0.024 & Normal Range = 0.005.
f. Assay cut-off:
Not applicable
2. **Comparison studies**:
a. Method comparison with predicate device:
**In-House Study**-
**Manual Method**: An in-house method comparison study was performed at Instrumentation Laboratory’s facility in Orangeburg, New York, to compare the performance of the new Coatest SP FVIII versus the predicate Coatest Factor VIII (K833892) using the manual method for both tests.
{7}
Page 8 of 12
The study used 181 citrated plasma samples (106 normal and 75 abnormal), each run in duplicate.
Samples reporting outside the tests' ranges were diluted and reanalyzed according to manufacturers' instructions. No artificially prepared samples were used in the study. The samples ranged in value from 1.2 to 588% factor VIII.
The clinical breakdown of the abnormal patient samples (obtained from CliniSys Associates in Atlanta, Georgia) is as follows:
| Abnormal Sample Type | Quantity |
| --- | --- |
| Low Factor VIII | 17 |
| vWF Disease | 10 |
| Heparin Therapy | 10 |
| Oral Anticoagulant Therapy | 10 |
| Lupus Anticoagulant | 10 |
| Liver Disease | 9 |
| High Factor VIII | 9 |
Acceptance criteria: Slope: 0.90-1.10 r: > 0.95
The results of running Coatest SP FVIII versus the predicate, Coatest Factor VIII, show a correlation of 0.987 and a slope of 1.09, indicating that the performance of the factor assays is statistically similar.
Instrument Application: An additional in-house method comparison study was performed at Instrumentation Laboratory's facility in Milan, Italy, to compare the performance of the new Coatest SP FVIII on an ACL 9000 versus Coamatic Factor VIII (K981038) on an ACL 10000. The study used 90 citrated plasma samples (41 normal and 49 abnormal), each run in duplicate.
Samples reporting outside the tests' ranges were diluted and reanalyzed according to manufacturers' instructions. The samples in the low range were obtained by dilution with Factor Diluent (saline). No spiked samples were used in the study. The samples ranged in value from 1.6 to 217% factor VIII.
The clinical breakdown of the abnormal patient samples (obtained from Sacco Hospital in Milan, Italy) is as follows:
{8}
| Abnormal Sample Type | Quantity |
| --- | --- |
| Low Factor VIII | 28 |
| Oral Anticoagulant Therapy | 10 |
| Liver Disease | 5 |
| High Factor VIII | 6 |
Acceptance criteria:
Slope: 0.90-1.10
$\mathrm{r}: > 0.95$
The results of running Coatest SP FVIII versus the predicate, Coatest Factor VIII, show a correlation of 0.992 and a slope of 1.00, indicating that the performance of the two tests is statistically similar.
# Field Site Study-
At A. Bianchi Bonomi, Haemofilia Centre (hemophilia center in Milan, Italy), a method comparison study of the new Coatest SP FVIII performed on an ACL 9000 versus Coamatic FVIII (K981038) performed on an ELECTRA 1600C was conducted using 336 citrated plasma samples (152 normal and 184 abnormal), each run in duplicate.
Samples reporting outside the tests' ranges were diluted and reanalyzed according to manufacturers' instructions. Of the 336 patient samples, 4 were artificially prepared by spiking normal patient samples to heparin concentrations of 0.1, 0.2, 0.4, and $1.9\mathrm{IU / ml}$ . The samples ranged in value from 1.0 to $373\%$ factor VIII.
The clinical breakdown of the abnormal patient samples is as follows:
| Abnormal Sample Type | Quantity |
| --- | --- |
| Hemophilia | 57 |
| vWF Disease | 33 |
| Oral Anticoagulant Therapy | 31 |
| Lupus Anticoagulant | 22 |
| Heparin Therapy | 10 |
| Liver Disease | 13 |
| DVT | 10 |
| Normal Sample Spiked with Heparin | 4 |
| Factor VII Deficient | 2 |
| Factor IX Deficient | 2 |
Acceptance criteria:
Slope: 0.90-1.10
$\mathrm{r}: > 0.95$
The results of running Coatest SP FVIII on an ACL 9000 versus Coamatic FVIII (K981098) performed on an ELECTRA show a correlation of 0.990 and
{9}
Page 10 of 12
a slope of 0.98, indicating that the performance of the two tests is statistically similar.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
An in-house study of 181 citrated plasmas (106 normal and 75 abnormal) were performed by manual method. The samples ranged in value from 1.2% to 588% factor VIII.
An additional in-house study of 90 citrated plasmas (41 normal and 49 abnormal) were performed on an ACL 9000. The samples ranged in value from 1.6% to 217% factor VIII.
A field-site study of 336 citrated plasmas (152 normal and 184 abnormal), each run in duplicate were performed on an ACL 9000. The samples ranged in value from 1.0% to 373% factor VIII.
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Stability:
Reconstituted/open vial stability testing at 2-8° C was performed to support the following product insert claims:
- S-2765 + I-2581: Reconstituted substrate is stable 3 months at 2-8° C
- Cacl2: Opened vial is stable 3 months at 2-8° C
- Buffer, stock solution: Opened vial is stable 3 months at 2-8° C
- Phospholipid: Opened vial is stable for 3 months at 2-8° C
- Factor Reagent (IXa + X): Aliquotted for -20° C for 3 months
For three different lots of Coatest SP FVIII reagents, vials of each component (Buffer, Phospholipids, Calcium, Factor Reagent and Substrate) were pooled, tested at time zero and stored at 2-8° C (-20° C for Factor IXa + X) for the duration of testing.
{10}
Page 11 of 12
HemosIL Normal Control and HemosIL High Abnormal Control were tested in triplicate using the manual test at the following time intervals (days): 0, 7, 14, 21, 28, 42, 56, 70, 90, and 120. The data indicate that the reagents are stable for the times specified in the product insert.
Working Factor Reagent Stability:
The Working Factor Reagent is a mixture of Phospholipid Reagent and Factor Reagent (phospholipids + factor IXa + factor X reagent). Stability testing of the Working Factor Reagent was performed to support the new claim added to the product insert for 12 hours on ice.
Using three different lots of Coatest SP FVIII reagents, the Working Factor Reagent was prepared per the insert instructions and placed on ice for the duration of testing. At each time interval (hours: 0, 2, 4, 8, 12, 24, 48), HemosIL Normal Control and HemosIL High Abnormal Control were tested in triplicate using the manual test. The data indicate that the Working Factor Reagent is stable on ice for a minimum of 12 hours.
Shelf-life Stability:
A shelf-life stability study is ongoing at 2-8° C using three different lots of Coatest SP FVIII reagents. At each time interval (Day 0, 3 months, 6, 9, and 12 months), HemosIL Normal Control and HemosIL High Abnormal Control were tested in duplicate using the manual method. The results to date support a shelf-life stability of 12 months.
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
121 citrated plasmas samples from healthy donors were tested in duplicate with Coatest SP FVIII using the manual method. The 95% reference intervals were calculated as recommended by NCCLS Document C28-A.
152 citrated plasmas samples from healthy donors were tested in duplicate with Coatest SP FVIII on the ACL 9000. The 95% reference intervals were calculated as recommended by NCCLS Document C28-A.
{11}
Page 12 of 12
| Mean | Manual Method | Instrument Application |
| --- | --- | --- |
| | 87.3 | 102.2 |
| SD | 19.3 | 23.4 |
| 95% Reference Interval | 48.6 to 126.0 | 55.4 to 148.9 |
**N. Proposed Labeling:**
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
**O. Conclusion:**
1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.