COAMATIC FACTOR VIII

K981038 · Instrumentation Laboratory CO · GGP · Apr 16, 1998 · Hematology

Device Facts

Record IDK981038
Device NameCOAMATIC FACTOR VIII
ApplicantInstrumentation Laboratory CO
Product CodeGGP · Hematology
Decision DateApr 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7290
Device ClassClass 2

Indications for Use

Coamatic® Factor VIII is an in vitro diagnostic test for the quantitative determination of Factor VIII activity in human citrated plasma using manual, microplate and automated methods. The measurement of Factor VIII is used as an aid in identifying factor VIII deficiency or monitoring patients on replacement therapy as well as for potency estimation of FVIII concentrates.

Device Story

Coamatic® Factor VIII is an in vitro diagnostic reagent kit for measuring Factor VIII activity in human citrated plasma. It is used in clinical laboratory settings by trained personnel. The device utilizes manual, microplate, or automated methods to quantify Factor VIII levels. By measuring this activity, the test assists clinicians in diagnosing Factor VIII deficiency, monitoring patients undergoing replacement therapy, and assessing the potency of Factor VIII concentrates. The output provides quantitative activity levels, which inform clinical decisions regarding patient management and therapeutic dosing.

Clinical Evidence

Method comparison study of 69 plasma samples comparing Coamatic® Factor VIII on a Cobas Mira S to the predicate Coatest® Factor VIII (manual method) yielded a correlation (r) of 0.99. Precision testing (within-run) showed CVs of 1.87% at 1.0 IU/mL, 2.57% at 0.21 IU/mL, and 2.76% at 0.025 IU/mL.

Technological Characteristics

In vitro diagnostic test for quantitative Factor VIII activity. Compatible with manual, microplate, and automated laboratory instrumentation. Uses the same test principle as the predicate device.

Indications for Use

Indicated for quantitative determination of Factor VIII activity in human citrated plasma to aid in identifying Factor VIII deficiency, monitoring patients on replacement therapy, and estimating potency of FVIII concentrates.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## APR 1 6 1998 # Section 3 Coamatic® Factor VIII - 510(k) SUMMARY (Summary of Safety and Effectiveness) #### Submitted by: Carol Marble Senior Regulatory Affairs Specialist Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02173 Phone: (781) 861-4467 (781) 861-4464 Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: ### Contact Persons: Carol Marble Phone: (781) 861-4467 #### Summary Prepared: March 18, 1998 #### Name of the device: Coamatic® Factor VIII ### Classification name(s): 864.7290 Factor Deficiency Test Class II Test, Qualitative and Quantitative Factor Deficiency 81GGP #### Identification of predicate device(s): Coatest® Factor VIII K833892 ## Description of the device/intended use(s): Coamatic® Factor VIII is an in vitro diagnostic test for the quantitative determination of Factor VIII activity in human citrated plasma using manual, microplate and automated methods. The measurement of Factor VIII is used as an aid in the indentification of factor VIII deficiency or to monitor patients on replacement therapy as well as for potency estimation of FVIII concentrates. # Statement of How the Technological Characteristics of the Device Compare to the Predicate Device: Coamatic® Factor VIII uses the same general test principle as the predicate Coatest® Factor VIII and is substantially equivalent in performance, intended use, and safety and effectiveness. ## Summary of Performance Data: In a method comparison study evaluating 69 plasma samples, the correlation (r) of the new Coamatic® Factor VIII on a Cobas Mira S as compared to the predicate Coatest® Factor VIII using a manual test tube method was 0.99. Within run precision assessed over multiple runs gave a CV of 1.87% (at a Factor VIII concentration of 1.0 IU/mL), 2.57% (at a Factor VIII concentration of 0.21 IU/mL) and 2.76% (at a Factor VIII concentration of 0.025 IU/mL). Coamatic® Factor VIII 510(k) {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread. The eagle is facing to the right. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 APR | 6 1398 Carol Marble . Senior Requlatory Affairs Specialist Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, Massachusetts 02173-3190 K981038 Re : Coamatic® Factor VIII Requlatory Class: II Product Code: GGP Dated: March 18, 1998 Received: March 19, 1998 Dear Ms. Marble: We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the qeneral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਕਿ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, -as setforth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification™ (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Litman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): __ Device Name: Coamatic® Factor VIII ## Indications for Use: Coamatic® Factor VIII is an in vitro diagnostic test for the quantitative determination of Factor VIII activity in human citrated plasma using manual, microplate and automated methods. The measurement of Factor VIII is used as an aid in identifying factor VIII deficiency or monitoring patients on replacement therapy as well as for potency estimation of FVIII concentrates. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number. 981038 Prescription Use (Per 21 CFR 801.019) OR Over-The-Counter Use _ Section 2 Coamatic® Factor VIII 510(k)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...