IMALUX OCT PROBE GUIDE

K041077 · Imalux Corporation · KKX · Jun 25, 2004 · General, Plastic Surgery

Device Facts

Record IDK041077
Device NameIMALUX OCT PROBE GUIDE
ApplicantImalux Corporation
Product CodeKKX · General, Plastic Surgery
Decision DateJun 25, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4370
Device ClassClass 2

Indications for Use

The Imalux OCT (Optical Coherence Tomography) Probe Sheath is a sterile, single-use, protective sheath for use as an accessory with the Imalux OCT Imaging System. The Probe Sheath is intended to serve as a microbial barrier between the probe and the patient's tissue, helping to prevent the transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the probe.

Device Story

Imalux OCT Probe Sheath is a sterile, single-use protective accessory for the Imalux OCT Imaging System. Device consists of rigid tubing, a flexible end-cap membrane window, and a plastic handle with a lock nut. The OCT probe is inserted through the handle and tubing until the distal tip is flush with the membrane window. The pliable membrane allows optical imaging without requiring additional coupling gel. Used in clinical settings to prevent transfer of microorganisms, body fluids, and particulates between patient and healthcare worker during probe reuse. The sheath acts as a physical microbial barrier; it does not alter the imaging system's function. Benefits include enabling safe reuse of the OCT probe while maintaining sterility and patient safety.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per AAMI/ANSI/ISO 10993 (cytotoxicity, systemic toxicity, intracutaneous reactivity, sensitization). Sterilization validated for ethylene oxide to SAL 10^-6. Microbial barrier testing confirmed prevention of transmission of a 25-27 nm challenge virus. Mechanical verification confirmed tensile strength and imaging performance integrity.

Technological Characteristics

Rigid tubing with flexible end-cap membrane and plastic handle. Materials evaluated for biocompatibility per ISO 10993. Sterilization via 100% ethylene oxide. Packaging: Tyvek/Mylar pouches. No electronic or software components; purely mechanical protective accessory.

Indications for Use

Indicated for use as a microbial barrier between the Imalux OCT Imaging System probe and patient tissue (intact skin, mucosal membranes, or breached surfaces) to prevent cross-contamination during probe reuse. Applicable for surface and tissue-communicating contact of limited duration (<24 hours).

Regulatory Classification

Identification

A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.

Special Controls

*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ### JUN 2 5 2004 510(k) SUMMARY This summary is being submitted in accordance with 21 CFR 807.92. ### Submitter's name, address, telephone number, initial importer, contact A. person | Submitter's Name: | Imalux Corporation | |-------------------|--------------------------------------------------| | Address: | 1771 East 30th Street | | Address: | Cleveland, OH 44114 | | Official Contact: | Stephanie A. S. Harrington | | Title: | Vice President, Regulatory & Clinical<br>Affairs | | Telephone: | 216 502-0755 | | Fax: | 216 622-0723 | | E-mail: | harrington@imalux.com | #### Device Name, Common Name B. #### Common/Usual Name 1. Probe Sheath #### Device Name 2. Imalux OCT Probe Sheath #### Classification Name 3. | Name | Classification<br>Regulation | Product Code | Class | |-----------------------------------------------------------|------------------------------|--------------|-------| | System, Imaging, Optical<br>Coherence Tomography<br>(OCT) | 892.1560 | NQQ | II | #### C. Identification of the predicate or legally marketed device | Device Name | 510(k) Number | |----------------------------------------------------------------------------------|---------------| | CIVCO Medical Instruments Synthetic Polyisoprene<br>Ultrasound Transducer Cover | K013721 | | Vision-Sciences Slide-On™ EndoSheath® System for<br>use with Flexible ENT Scopes | K990354 | {2}------------------------------------------------ #### Device Description D. #### Summary 1. The Imalux OCT (Optical Coherence Tomography) Probe Sheaths (Probe Sheaths) are sterile, single-use, protective sheaths for use with the Imalux OCT Imaging System. The sheaths are provided to fit the Imalux OCT Imaging System probe geometry and to help to prevent the transfer of microorganisms, body fluids, and material to the patient and healthcare worker during reuse of the Probe. The Probe Sheath does not impair the imaging performance of the Imalux OCT Imaging System. Adequate coupling between the Probe Sheath and the probe's distal end can be achieved without the use of an additional medium, such as gel, due to the pliability of the membrane window material. #### Design 2. The Probe Sheath is constructed of rigid tubing with a flexible end cap membrane window at one end and with a plastic handle at the other end. The probe is inserted into the tubing at the handle and fed through the tubing until the probe's distal tip is flush with the Probe Sheath's membrane window. The handle includes a lock nut to secure the probe in place. The Probe Sheath is provided in various dimensional configurations necessary to accommodate the probe as well as to assist the user in probe positioning. Probe Sheaths are provided as sterile, single-use sheaths, individually packaged in Tyvek/Mylar pouches. #### Materials 3. The Probe Sheath is constructed of commonly used medical device materials and has been evaluated for biocompatibility. This evaluation has demonstrated that the Probe Sheath is safe for its intended use. The intended use and indications for use place the Probe Sheath in the device body contact categories as follows: - a. surface devices, intact skin / mucosal membranes / breached surfaces, limited contact duration (<24 hours) - b. external communicating devices, tissue communicating, limited contact duration (<24 hours) {3}------------------------------------------------ #### Intended Üse E. The Imalux OCT (Optical Coherence Tomography) Probe Sheath is a sterile, single-use, protective sheath for use as an accessory with the Imalux OCT Imaging System. The Probe Sheath is intended to serve as a microbial barrier between the probe and the patient's tissue, helping to prevent the transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the probe. The Imalux OCT Probe Sheath is substantially equivalent in safety and performance to the CIVCO Synthetic Polyisoprene Ultrasound Transducer Cover and to the Vision-Sciences EndoSheaths (Section C). #### Safety & Performance F. To assure the safety of the device, the Probe Sheath was tested for biocompatibility per AAMI/ANSI/ISO 10993 for cytotoxicity, acute systemic toxicity, intracuteanous reactivity, sensitization, and ethylene oxide sterilization residuals. Testing was performed in accordance with ISO 10993-Part 1 Biological Evaluation of Medical Devices, FDA Blue Book Memorandum #G95-1, and FDA-Good Laboratory Practices 21 CFR Part 58 (GLP). These tests demonstrated that the Probe Sheath device is non-toxic, non-irritating, non-sensitizing, and can be safely sterilized with ethylene oxide. The Probe Sheath was verified for imaging, mechanical, and packaging performance. The Probe Sheath's materials and design have been shown to not impair the optical imaging performance of the Imalux OCT Imaging System. Mechanical verification of the Sheath, such as tensile strength, demonstrated that the Probe Sheath's distal end will withstand normal use. Microbial barrier testing demonstrated that the Probe Sheath was carable of preventing the transmission of a challenge virus with a diameter of 25-27 nanometers (nm), smaller than most clinically claimeter (1 22 to a detection limit of less than 5.5 x 107 milliliters (mL) of challenge virus suspension. The Probe Sheath, individually packaged in its Tyvek/Mylar package, was successfully sterilized with 100% ethylene oxide. The sterilization was seconding broad and verified to provide a sterility assurance level (SAL) of 10 . In addition, the Tyvek/Mylar package was verified for strength and integrity. {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image is a black and white circular seal. The seal contains the logo for the U.S. Department of Health and Human Services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the perimeter of the circle. Inside the circle is an abstract image of an eagle. Public Health Service JUN 2 5 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Stephanie A. S. Harrington Vice President, Regulatory and Clinical Affairs Imalux Corporation 1771 East 30th Street Cleveland, Ohio 44114 Re: K041077 Trade/Device Name: Imalux OCT Probe Sheath Regulation Number: 878.4370, 892.1560 Regulation Name: Surgical Drape and Drape Accessories, Ultrasonic Pulsed Echo Imaging System Regulatory Class: II Product Code: KKX, NQQ Dated: May 28, 2004 Received: June 1, 2004 Dear Ms. Harrington: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 -- Ms. Harrington Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. Chris Lins, Ph.D. Chiu Lin, Ph.D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## INDICATIONS FOR USE K041077 510(k) Number (if known): Device Name: Imalux OCT Probe Sheath Indications For Use: The Imalux OCT (Optical Coherence Tomography) Probe Sheath is a sterile, single-use, protective sheath that is used as an accessory with the Imalux OCT Imaging System. The Imalux OCT Probe Sheath is intended to serve as a microbial barrier between the probe and the patient's tissue, helping to prevent the transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the probe. AND/OR Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Kein Muly (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number: K
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...