MODIFICATION TO MEDAMICUS FLOWGUARD PEELABLE INTRODUCER

K040150 · Medamicus, Inc. · DYB · Feb 18, 2004 · Cardiovascular

Device Facts

Record IDK040150
Device NameMODIFICATION TO MEDAMICUS FLOWGUARD PEELABLE INTRODUCER
ApplicantMedamicus, Inc.
Product CodeDYB · Cardiovascular
Decision DateFeb 18, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The MedAmics FlowGuard™ Peelable Introducer is indicated for use in the percutaneous insertion of pacing leads or catheters in the venous system.

Device Story

FlowGuard™ Peelable Introducer is a small-diameter tubular device with integrated proximal handles; facilitates atraumatic implantation of catheters and pacemaker leads into venous system. Features 'tear-away' design allowing introducer removal while leaving inserted catheter or lead in place. Available as individual sterile units, bulk non-sterile units, or convenience kits (including needle, syringe, guidewire). Used by clinicians in clinical settings. Device provides mechanical access path for vascular procedures.

Clinical Evidence

Bench testing only. Design verification performed via risk analysis comparing new handle material to predicate. Accelerated aging shelf-life testing conducted to verify handle integrity.

Technological Characteristics

Tubular sheath with integrated proximal handles; peelable 'tear-away' design. Material change to sheath handle. Sterile and non-sterile configurations. No electronic or software components.

Indications for Use

Indicated for percutaneous insertion of pacing leads or catheters into the venous system.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K040150 #### V. 510(k) Summary #### Submitter MedAmicus, Incorporated 15301 Highway 55 West Minneapolis, MN 55447 Main: 763-15301 763-559-0148 Fax: email: www.medamicus.com #### Date Prepared January 22, 2004 ### Trade Name MedAmicus FlowGuard™ Peclable Introducer # Common Name Catheter Introducer #### Predicate Device MedAmicus FlowGuard TM Peelable Introducer 510(k) K030905. #### Device Description The McdAmicus FlowGuard™ Peelable Introducer is a small diameter tubular shaped device with integrated proximal handles. The FlowGuard 100 Peelable Introducer is designed to provide a relatively atraumatic method for implanting catheters and pacemaker leads into the venous system. The FlowGuard™ Peelable Introducer has a "tear-away" feature common to the predicate devices. This feature allows the user to remove the introducer without removing the inserted catheter or pacing lead. The MedAmicus FlowGuard" Peelable Introducer is packaged in three configurations: 1) a convenience kit containing a MedAmicus FlowGuard11M Peelable Introducer, a thin-wall needle, a disposable syringe, and a flexible guidewire, packaged in a tray and a sealed pouch. 2) individually pouched, sterile MedAmicus FlowGuard™ Peelable Introducer and 3) bulk, nonsterile MedAmicus FlowGuard™ Peclable Introducers. 6 {1}------------------------------------------------ Confidential # Intended Use There are no changes to the Intended Use of the device from the currently approved device. The MedAmics FlowGuard™ Peelable Introducer is indicated for use in the percutaneous insertion of pacing leads or catheters in the venous system. ## Technological Characteristics The material revision to the sheath handle material is technologically equivalent to the currently approved device. The method of use is not revised as a result of the material change. # Summary of Studies Design Verification of the device has been completed to verify the changes to the device. These validations were completed based on a Risk Analysis per internal procedures that are compliant with European Standard EN1441: 1998 Medical Devices -- Risk Analysis. This Risk Analysis was completed by comparing the risks associated with the new material sheath handle compared to the currently approved device. Shelf Life testing, using accelerated aging, has been completed to verify that the new material will yield consistent handle integrity. The validations are summarized in Section X of this submission. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three wavy lines or a caduceus without the snake, representing health and human services. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 1 8 2004 MedAmicus Incorporated c/o Ms. Kary Haskell Quality Assurance and Regulatory Affairs Manager 15301 Hwy, 55 West Minneapolis, MN 55447 Re: K040150 MedAmicus FlowGuard Peelable Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II (two) Product Code: DYB Dated: January 22, 2004 Received: January 23, 2004 Dear Ms. Haskell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Kary Haskell Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Name R. Holmes ()_Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known):__K040150 Device Name: __ MedAmicus FlowGuard™ Peelable Introducer Indications For Use: The MedAmicus FlowGuard Peelable Introducer in indicated for indications i or ooc. This modifines pacing leads or catheters in the venous system. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Dung R. Voelker Page 1 of ____________________________________________________________________________________________________________________________________________________________________ /ision Sign-Off) ്. Jision of Cardiovascular Devices 510(k) Number ŁO4 D150 _______________________________________________________________________________________________________________________________________________________
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...