ENPATH MEDICAL VIASEAL VALVED PEELABLE INTRODUCER, MODEL 10866-XXX

K063182 · Enpath Medical, Inc. · DYB · Dec 21, 2006 · Cardiovascular

Device Facts

Record IDK063182
Device NameENPATH MEDICAL VIASEAL VALVED PEELABLE INTRODUCER, MODEL 10866-XXX
ApplicantEnpath Medical, Inc.
Product CodeDYB · Cardiovascular
Decision DateDec 21, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The Enpath ViaSeal Valved Peelable Introducer is intended for use in the percutaneous insertion of pacing leads or catheters in the venous system.

Device Story

ViaSeal Valved Peelable Introducer is a tubular device with proximal handles; used for atraumatic implantation of pacing leads or catheters into venous system. Device features integrated hemostatic valve to seal against venous pressures; includes 'peel-away' mechanism allowing introducer removal while leaving inserted lead/catheter in place. Supplied in sterile convenience kit containing introducer, needle, syringe, and guidewire. Used by clinicians in clinical settings. Device provides access to venous system; hemostatic valve minimizes blood loss during procedure; peelable design facilitates lead/catheter placement.

Clinical Evidence

No clinical evaluations conducted. Bench testing only, including dimensional, functional, and hemostasis testing to verify performance against specifications.

Technological Characteristics

Tubular sheath and dilator (7F to 10.5F); integrated proximal handles; hemostatic valve; peel-away mechanism. Sterile, single-use, ETO sterilized. Packaged in PETG tray with Tyvek lid and Tyvek/LDPE pouch.

Indications for Use

Indicated for percutaneous insertion of pacing leads or catheters into the venous system.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Enpath Medical, Inc. ViaSeal Valved Peelable Introducer Traditional 510(k) Submission # 510(k) Summary Submitter Enpath Medical Inc. 15301 Highway 55 West Minneapolis, MN 55447 Main: 763-559-2613 763-559-0148 Fax: email: www.enpathmed.com ## Contact Person James Jenkins Regulatory Associate 952-653-2412 # Date Prepared December 14, 2006 Trade Name Enpath Medical ViaSeal Valved Peelable Introducer Common Name Catheter Introducer ## Classification Name Catheter Introducer (21 CFR 870.1340, Product Code DYB) ## Predicate Device Enpath Medical FlowGuard Valved Peelable Introducer; K040150. Thomas Medical Products Inc., SafeSheath MSP Introducer Kit with Integral Hemostasis Valve; K003731 ## Device Description The Enpath ViaSeal Valved Peelable Introducer is a small diameter tubular shaped device with integrated proximal handles. The ViaSeal Valved Peelable Introducer is designed to provide a relatively atraumatic method for implanting catheters and pacemaker leads into the venous system while providing hemostatic sealing to venous pressures. The ViaSeal Valved Peelable Introducer has a "peel-away" feature common to the predicate device. This feature allows the user to remove the introducer without removing the inserted catheter or pacing lead. The ViaSeal Valved Peelable Introducer is packaged in a sterile 5-pack convenience kit containing a ViaSeal Valved Peelable Introducer, a thin-wall 510(k) Summary K063182 page 1 of 2 レビしな | 2006 {1}------------------------------------------------ Enpath Medical, Inc. ViaSeal Valved Peelable Introducer Traditional 510(k) Submission 510(k) Summary K063182 page 2 of 2 needle, a disposable syringe, and a flexible guidewire. The kit is packaged in a PETG tray with a Tyvek lid, and a sealed Tyvek/LDPE pouch. Introducer sizes of the sheath and dilator range from 7 French to 10.5 French. The materials and construction are the same for all French sizes. Enpath ViaSeal Valved Peelable Introducer kits will be packaged and ETO sterilized for one time use and sealed in a Tyvek/LDPE pouch. ## Intended Use The Enpath ViaSeal Valved Peelable Introducer is intended for use in the percutaneous insertion of pacing leads or catheters in the venous system. #### Comparison of Technological Characteristics All technological characteristics of the Enpath ViaSeal Valved Peelable Introducer are substantially equivalent to the predicate device; FlowGuard Valved Peelable Introducer (K040150) including indications for use, operation, manufacturing process, biocompatibility, sterilization method, and packaging. The Enpath ViaSeal Valved Peelable Introducer is also substantially equivalent to the Thomas Medical Products Inc. SafeSheath Introducer Kit with Integral Hemostasis Valve, K003731, with respect to indications for use, sterilization method, materials, packaging, hemostatic sealing to venous pressures, French sizes, and available lengths (13 and 25 cm). ## Summary of Studies The performance testing for this device included testing to verify that the device functions in a safe and effective manner. The performance testing included the device specifications, dimensional and functional testing, and hemostatsis of the device to venous pressures. Test results verify that the device performs per specification requirements and is equivalent to the predicate device without creating additional risk to the patient or user. No clinical evaluations for this submission have been conducted. #### Substantial Equivalence The ViaSeal Valved Peelable Introducer is as safe, as effective, and performs as well as, and is substantially equivalent to the original "cleared" FlowGuard Valved Peelable Introducer, K040150. The ViaSeal Valved Peelable Introducer is also substantially equivalent to the Thomas Medical Products Inc. SafeSheath Introducer Kit with Integral Hemostasis Valve (K003731) with respect to intended use (the introduction of various pacing leads and catheters), device classification (Class II, 21 CFR 870.1340), product code (DYB), and hemostatic sealing to venous pressures. (Note: This document uses the term "substantial equivalence" as intended in 21 CFR 807.88, and not as defined in Title 35 of the U.S. Code.) {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract eagle design with three lines forming the body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 2 1 2006 Enpath Medical Inc. James Jenkins, Regulatory Associate 15301 Highway 55 West Minneapolis MN 55447 Re: K063182 Trade/Device Name: Enpath Medical ViaSeal Valved Peelable Introducer Regulation Number: 21 CFR Sec. 870.1340 Regulation Name: Catheter, Introducer Regulatory Class: Class II Product Code: DYB Dated: December 14, 2006 Received: December 15, 2006 Dear Mr. Jenkins We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ # Page 2 - Mr. James Jenkins or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, B/Simmons for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Enpath Medical, Inc. ViaSeal Valved Peelable Introducer Traditional 510(k) Submission #### 2.0 Indications for Use 510(k) Number (if known): Device Name: ViaSeal Valved Peelable Introducer Indications for Use: The Enpath ViaSeal Valved Peelable Introducer is indicated for use in the percutaneous insertion of pacing leads or catheters in the venous system. Prescription Use な (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) K043182 (Division Sign Off) Division of Cardiovascular Devices 510(k) Number Blimonume