RENACLEAN SH DIALYZER CLEANING SYSTEM
Device Facts
| Record ID | K033505 |
|---|---|
| Device Name | RENACLEAN SH DIALYZER CLEANING SYSTEM |
| Applicant | Minntech Corp. |
| Product Code | LIF · Gastroenterology, Urology |
| Decision Date | Apr 27, 2004 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5820 |
| Device Class | Class 2 |
Indications for Use
Minntech Corporation's Renaclean SH Dialyzer Cleaning System is used to facilitate the cleaning of blood and other debris from multiple-use hollow fiber dialyzers, prior to being reprocessed on a Renatron Dialyzer Reprocessing System.
Device Story
Renaclean SH Dialyzer Cleaning System is a stand-alone countertop device used to pre-clean multiple-use hollow fiber dialyzers. It removes blood and debris before the dialyzer undergoes full reprocessing on a Renatron system. The device uses ANSI/AAMI quality water and a pre-diluted sodium hypochlorite solution. An operator controls the system via a membrane switch front panel, selecting from three cycles: High Flux dialyzer clean, Low Flux dialyzer clean, or System Sanitize. By automating the pre-cleaning phase, the device prepares dialyzers for subsequent testing and sterilization, supporting the maintenance of multiple-use dialyzers in clinical settings.
Clinical Evidence
Bench testing only. Functional testing was performed to demonstrate safety and performance equivalence to the predicate device. Material compatibility testing was conducted for all system components exposed to the cleaning chemicals.
Technological Characteristics
Stand-alone countertop system. Utilizes ANSI/AAMI quality water and pre-diluted sodium hypochlorite solution. Features a membrane switch front panel for operator control. Cycles include High Flux clean, Low Flux clean, and System Sanitize. Materials tested for chemical compatibility.
Indications for Use
Indicated for cleaning blood and debris from multiple-use hollow fiber dialyzers prior to reprocessing in a Renatron Dialyzer Reprocessing System.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
- Renatron II Dialyzer Reprocessing System (K904210)