HALLU PLATES

K021626 · New Deal, S.A. · HRS · Aug 12, 2002 · Orthopedic

Device Facts

Record IDK021626
Device NameHALLU PLATES
ApplicantNew Deal, S.A.
Product CodeHRS · Orthopedic
Decision DateAug 12, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HALLU® -PLATES are intended to be implanted for fixation of fractures, osteotomies or arthrodesis of the first metatarso-phalangeal joint. Examples include: -Hallux rigidus -Severe hallux valgus (IM angle > 20° and HV angle > 40°) -Deformity from rheumatoid arthritis -Failed previous surgical procedure -Traumatic arthritis -Neuromuscular instability.

Device Story

HALLU®-PLATES are low-profile titanium bone plates designed for first metatarso-phalangeal (MTP) joint arthrodesis; pre-bent for 10° valgus and 10° dorsiflexion. System includes two designs (HALLU®-C and HALLU®-S) in six sizes for anatomical fit. Implants are secured to bone using titanium SNAP-OFF® screws (2.7 mm and 3.0 mm diameters). Bone contact surface is sand-blasted to enhance stability. Used by orthopedic surgeons in clinical/surgical settings to stabilize bone during healing or fusion. Provides rigid internal fixation to correct deformities or treat fractures, potentially improving patient mobility and reducing pain associated with MTP joint instability or arthritis.

Clinical Evidence

Bench testing only. Testing included determination of bending strength and stiffness per French Standard ISO 9585 and ASTM F-382-99, and fatigue strength testing to ensure minimal risk of plate rupture.

Technological Characteristics

Titanium alloy (TiAl6V4) bone plates; low-profile; pre-bent for 10° valgus/10° dorsiflexion; sand-blasted bone contact surface. Fixation via 2.7 mm and 3.0 mm titanium SNAP-OFF® screws. Complies with ASTM F-382-99 and ISO 9585 for bending strength/stiffness.

Indications for Use

Indicated for patients requiring fixation of fractures, osteotomies, or arthrodesis of the first metatarso-phalangeal joint, including those with hallux rigidus, severe hallux valgus, rheumatoid arthritis deformity, failed prior surgery, traumatic arthritis, or neuromuscular instability.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ **AUG 1 2 2002** K021626 p113 #### SUMMARY OF SAFETY AND EFFECTIVENESS 3. - SPONSOR IDENTIFICATION: A. NewDeal SA Parc d'Activitiés Garigliano Rue de la Convention 38 200 VIENNE FRANCE Telephone: (33) 4 74 78 15 15 Fax: (33) 4 74 78 15 16 #### B. ESTABLISHMENT REGISTRATION NUMBER: 9615741 ## OFFICIAL CONTACT PERSON C. Norman F. Estrin, Ph.D., RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac, MD 20854 Tel: (301) 279 -2899 Fax: (301) 294-0126 #### DATE OF PREPARATION OF THIS SUMMARY: April 11th, 2002 D. #### HALLU®PLATES PROPRIETARY (TRADE) NAME: E. - Metatarsal-phalangeal arthrodesis plate F. COMMON NAME: #### CLASSIFICATION NAME AND REFERENCE G. Plate, Fixation, Bone (21 CFR, Section 888. 3030) - H. PROPOSED REGULATORY CLASS: Class II - DEVICE PRODUCT CODE: I. 87 HRS - 87 OR Orthopedic J. PANEL CODE: - K. DESCRIPTION OF DEVICE: The HALLU® -PLATES are available in two different designs : - the HALLU®-C PLATE - - the HALLU®-S PLATE - The HALLU®-C PLATE and the HALLU® -S PLATE are low profile Titanium plates dedicated to first metatarso-phalangeal arthrodesis. Those implants are pre-bent for optimal anatomical adaptation (10° valgus and 10°dorsiflexion). Their fixation is provided by Titanium SNAP-OFF® screws available in a two diameters: 3.0 and 2.7 mm. They exist in different colors for size identification. The bone contact surface is sand blasted in order to maximize plate stability. The range of HALLU -C PLATE and 01 2005 {1}------------------------------------------------ K021624 p²/3 HALLUS S PLATE include 6 sizes ( 31eft/3right) for optimal anatomic fit. Moreover, the HALLU®-S PLATE has an anatomical design providing optimal bone coverage. - The HALLU® -PLATES are intended to be implanted for L. INTENDED USE: fixation of fractures, osteotomies or arthrodesis of the first metatarso-phalangeal joint. ### INDICATIONS M. The HALLU®-PLATES are intended to be implanted for FOR USE: fixation of fractures, osteotomies or arthrodesis of the first metatarso-phalangeal joint. Examples include: -Hallux rigidus -Severe hallux valgus (IM angle > 20° and HV angle > 40°) -Deformity from rheumatoid arthritis -Failed previous surgical procedure -Traumatic arthritis -Neuromuscular instability. #### N. PREDICATE DEVICE: The HALLU - PLATES are substantially equivalent to the Howmedica Luhr plate (K935448) and Howmedica Profyle® Hand and Small Fragment System (K961497) , the Acumed Congruent bone plate system (K012655) and the Synthes modular foot system (K001941). #### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS: 0. The HALLU® - PLATES, the Howmedica Luhr and Howmedica Profyle plates, the Acumed plate and Synthes modular foot system are intended for internal bone fixation for bone fractures or reconstruction. Luhr® HALLU®- PLATES, the Howmedica The and Howmedica Profyle® plates, the Acumed plate and Synthes modular foot system are all indicated for bone fractures or reconstruction in the foot. The HALLU® - PLATES, the Howmedica Profyle " plates and the Acumed plate are in Titanium alloy TIAI6V4 whereas the Synthes modular foot system is manufactured in Stainless steel 316 L and the Howmedica Luhr in Vitallium alloy . {2}------------------------------------------------ All those systems are fixed with screws except for the Synthes modular foot system attached to bone via buttress pins. KOZIG 26 Test 1 : Determination of bending strength and stiffness of SUMMARY OF STUDIES: bone plates was compared with requirement of the French Standard ISO 9585 and ASTM standard F-382-99 and found to have a bending stiffness and strength in compliance with the selected standards. > Test 2 : Determination of fatigue strength of bones plates were conducted and have shown the risk of rupture of the plate is minimal. > > () . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling three stylized human profiles or birds in flight, stacked on top of each other. ## Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NewDeal SA c/o Norman F. Estrin, PhD, RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac, Maryland 20854 AUG 1 2 2002 Re: K021626 Trade/Device Name: HALLU®-Plates System Regulation Number: 21 CFR §888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: May 14, 2002 Received: May 17, 2002 Dear Dr. Estrin; We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Norman F. Estrin, PhD, RAC This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark H-Milliman Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ # Page 1 of 1 # 510(k) Number (if known): 《ՀՀՀՆՆՀՀՀ HALLU®-PLATES SYSTEM Device Name: Indications for Use: > The HALLU® -PLATES are intended to be implanted for fixation of fractures, osteotomies or arthrodesis of the first metatarso-phalangeal joint. Examples include: > > -Hallux rigidus -Severe hallux valgus (IM angle > 20° and HV angle > 40°) -Deformity from rheumatoid arthritis -Failed previous surgical procedure -Traumatic arthritis -Neuromuscular instability. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-the-Counter Use<br>(Optional Format 1-2-96) | |------------------------------------------|----|--------------------------------------------------| |------------------------------------------|----|--------------------------------------------------| for (Division Sign-Off) Division of General, Restorative and Neurological Devices | 510(k) Number | K021626 | |---------------|---------| |---------------|---------| 004E
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...