EBI DFS OPTIROM ELBOW FIXATOR

K021312 · Ebi, L.P. · KTT · May 23, 2002 · Orthopedic

Device Facts

Record IDK021312
Device NameEBI DFS OPTIROM ELBOW FIXATOR
ApplicantEbi, L.P.
Product CodeKTT · Orthopedic
Decision DateMay 23, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The EBI XFIX DFS OptiROM Elbow Fixator is intended for use in upper extremity treatment of bone and soft tissue conditions and other bone conditions amenable to treatment by use of the external treatment modality. Possible applications include: Fracture dislocation with ligamentos instability. Comminuted intra-articular fractures. Post traumatic reconstruction for joint stiffness.

Device Story

External fixation system for upper extremity bone/soft tissue conditions. Bone screws inserted through skin/tissue into bone; fixator frame attached to screw shanks. Modification adds distractor component for gradual movement during rehabilitation. Used in clinical settings by physicians. Provides mechanical stabilization and controlled joint movement to assist in fracture healing and joint stiffness reconstruction. Benefits patient by enabling external treatment of complex fractures and post-traumatic conditions.

Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and intended use comparisons to the predicate device.

Technological Characteristics

External fixation system consisting of bone screws and a fixator frame. Materials include Aluminum Alloy, Stainless Steel, Pyromet® 718, Carbon Fiber, and Titanium Alloy. Features a distractor component for gradual movement. Provided non-sterile. Mechanical fixation principle.

Indications for Use

Indicated for patients requiring upper extremity treatment for bone and soft tissue conditions, including fracture dislocation with ligamentous instability, comminuted intra-articular fractures, and post-traumatic reconstruction for joint stiffness.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 021312 page 1 of 2 ## Special 510(k) Device Modification: EBI® XFIX® DFS® OptiROM® Elbow Fixator #### 510(k) Summary This 510(k) Summary for the EBI® XFIX® DFS® OptiROM® Elbow Fixator is provided as required per Section 513(1)(3) of the Federal Food, Drug and Cosmetic Act. Contact Person: Frederic Testa, RAC 1. Sponsor: Telephone: (973) 299-9300, ext.2208 EBI, L.P. 100 Interpace Parkway Parsippany, NJ 07054 Date Prepared: April 24, 2002 | 2. Proprietary Name: | EBI® XFIX® DFS® OptiROM® Elbow Fixator | |----------------------|--------------------------------------------------------------------------------------------------| | Common Name: | External Fixation Device | | Classification Name: | Single Multiple Component Metallic Bone Fixation<br>Appliances and Accessories, 21 CFR 888.3030. | #### 3. Predicate or Legally Marketed Device: - EBI® XFIX® DFS® OptiROM® Elbow Fixator (K992515, K981483, ● K0202403) #### 4. Description of Device: The System consists of external fixation components and implantable bone screws. The EBI XFIX DFS OptiROM Elbow Fixator is utilized in the following manner: bone screws are inserted through the patient's skin and soft tissue and into the bone. The fixator frame of the EBI XFIX DFS OptiROM Elbow Fixator is attached to the shanks of the bone screws. This submission is for the addition of distractor component, which allows for gradual movement used in rehabilitation. The intended use and fundamental scientific technology have not changed from the previously cleared submission. #### 5. Intended Use: The EBI XFIX DFS OptiROM Elbow Fixator is intended for use in upper extremity treatment of bone and soft tissue conditions and other bone conditions amenable to treatment by use of the external treatment modality. Possible applications include: {1}------------------------------------------------ 021312 page 2sf2 Fracture dislocation with ligamentos instability. Comminuted intra-articular fractures. Post traumatic reconstruction for joint stiffness. #### 6. Materials: The components of the System may be manufactured from materials such as Aluminum Alloy, Stainless Steel, Pyromet® 718, Carbon Fiber, and Titanium Alloy. ### 7. Comparison of the technological characteristics of the device to predicate devices: The modified EBI XFIX DFS OptiROM Elbow Fixator is substantially equivalent to the following predicate device: EBI XFIX DFS OptiROM Elbow Fixator (K981483, K992515, K020403) - The additional EBI XFIX DFS OptiROM Elbow Fixator components . are fabricated from the same materials as the components of the currently marketed EBI XFIX DFS OptiROM Elbow Fixator. - The modified EBI XFIX DFS OptiROM Elbow Fixator and the currently . marketed EBI XFIX DFS OptiROM Elbow Fixator are both intended for use in upper extremity treatment of bone and soft tissue conditions and other bone conditions amenable to treatment by use of the external treatment modality. Possible applications include: Fracture dislocation with ligamentos instability. Comminuted intra-articular fractures. Post traumatic reconstruction for joint stiffness. - The bone screw clamps of the modified EBI XFIX DFS OptiROM Elbow . Fixator, like the bone screw.clamps currently marketed in the EBI XFIX DFS OptiROM Elbow Fixator, are designed for fixation of the bar and bone screw. - The modified EBI XFIX DFS OptiROM Elbow Fixator, like the currently . marketed EBI XFIX DFS OptiROM Elbow Fixator, is provided nonsterile. There are no significant differences between the modified EBI XFIX DFS OptiROM Elbow Fixator and the currently marketed EBI XFIX DFS OptiROM Elbow Fixator. It is substantially equivalent* to the predicate device with regard to intended use, materials, and function. <sup>*</sup> Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate to whether the product can be lawfully market approval or reclassification and is not intended to be interpreted as an admission or any other type of evident infringement litigation. [Establishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)) {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # AY 2 3 2007 Mr. Frederic Testa, RAC Regulatory Affairs Specialist EBI. L.P. 100 Interpace Parkway Parsippany, NJ 07054 Re: K021312 K021912 Trade/Device Name: EBI® XFIX® DFS® OptiROM® Elbow Fixator Regulation Number: 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: April 24, 2002 Received: April 25, 2002 Dear Mr. Testa: We have reviewed your Section 510(k) premarket notification of intent to market the device wo nave reviewed your your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050 {3}------------------------------------------------ Page 2 -- Mr. Frederic Testa, RAC This letter will allow you to begin marketing your device as described in your Section 510(k) I mis leter will anow you to ought and ing of substantial equivalence of your device to a legally premits to the month. The results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you desire specific acres 809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 Of It Far are 659. Additionally, for questions on the promotion and advertising of Compination at (301) 597-1699 of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). regulation onlined, "Thisocrange jesponsibilities under the Act may be obtained from the Other general information of the mational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark McMillan Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # STATEMENT OF INDICATIONS FOR USE 510(k) Number (if known): KON / 310 Device Name: EBI® XFIX® DFS® OptiROM® Elbow Fixator Indications For Use: The EBI® XFIX® DFS® OptiROM® Elbow Fixator is intended for use in upper extremity treatment of bone and soft tissue conditions and other bone conditions amenable to treatment by use of the external treatment modality. Possible applications include: - 1. Fracture dislocation with ligamentos instability. - 2. Comminuted intra-articular fractures. - 3. Post traumatic reconstruction for joint stiffness. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) Marl N. Mllkerson Division Sign-Off) (Division of General, Restorative and Neurological Devices 510(k) Number K021312
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...