K020102 · Iqteq Development · BZG · Mar 21, 2003 · Anesthesiology
Device Facts
Record ID
K020102
Device Name
IQTEQ SPIROMETER
Applicant
Iqteq Development
Product Code
BZG · Anesthesiology
Decision Date
Mar 21, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The IQTeQ Spirometer is intended for use as a prescription-use-only clinical diagnostic spirometer for pulmonary function evaluation and data management. The IQTeQ Spirometer is for use in hospitals, clinics and physician/clinician offices by individuals that have received minimal instruction or training in the administration of spirometry tests. The IQTeQ Spirometer operates with an IBM or compatible computer using a USB port (Universal serial bus) connection and the PC's installed IQTeQ Spirometer Windows™ compatible software. Minimum PC and operating system requirements are specified in the IQTeQ Spirometer Manual. The IOTeO Spirometer is indicated for use with male/female adult patients and male/female pediatric patients to evaluate, access, describe, measure, or monitor. - Symptoms, signs, or abnormal laboratory tests. - Effects of disease on pulmonary function. - Individuals at risk for pulmonary disease. - 4. Preoperative risk - 5. Post-surgical prognosis - 6. Pre-treatment health status - 7. Therapeutic interventions - The course of disease affecting lung function 8. - Persons exposed to pollutants 9. - 10. Adverse reactions to drugs with known pulmonary toxicity - 11. Rehabilitation programs - 12. Risks as part of an insurance evaluation - 13. Individuals for legal reasons - 14. Epidemiological surveys - 15. Derivation of reference equations
Device Story
Handheld diagnostic spirometer; connects to PC via USB. Hardware: plastic handle, amplified pressure transducer, analog-to-digital converter, USB microcontroller; electronics sealed in epoxy (Eli-Cast FR 453). Flow tube contains inert ceramic (CORDIERITE) laminar flow resistance element. Patient breathes through bacterial filter into flow tube; pressure transducer transmits differential pressure to PC. Software calculates flow and volume in real-time using calibration polynomial algorithm. Used in hospitals, clinics, physician offices, and homes (prescribed) by trained operators. Output: flow-volume measurements (Peak Flow, FVC, Forced Expiratory Flows, FIVC, MVV, SVC) displayed on PC. Assists clinicians in evaluating lung disease, therapy course, and pulmonary function. Benefits: real-time data acquisition, standardized pulmonary testing, and digital record management.
Clinical Evidence
Bench testing only. Device performance validated against ATS 1994 standards for spirometry. Electrical safety and emission compliance verified via EN60601-1 and IEC60601-1.
Indicated for male/female adult and pediatric patients to monitor, measure, and evaluate pulmonary function parameters, including forced flow/volume, forced flows, slow vital capacity, forced inspiratory volumes/flows, and MVV, for clinical diagnosis, disease progression, and therapeutic monitoring.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Section 2: h)
510(k) Summary
IOTeQ Spirometer 2001
020102
- 1. Applicant:
Choice Decisions 86 (PTY) LTD trading as IQTeQ Development The Penthouse. Jaga House, 61 Wale Street Cape Town 8001
Contact person: Ray Wright, Director, Research and Development
Phone: 27 21 422 2222 Fax: 2721 422 2223 Cell: 27 (0) 82 293 1051
Date prepared: January 6, 2002
- 2. Device name: Proprietary name: IQTeQ Spirometer 2001 Common/Usual name: Diagnostic Spirometer Classification Name: Diagnostic Spirometer Classification Panel: Anesthesiology Classification Code: 73BZG Classification: Class II, according to 21 CFR, 868.1840
- 3. Predicate Devices/ Substantial Equivalence: The IQTeQ Spirometer is substantially equivalent to:
- 3.1: Serial Flow or also known as the PCFlow+ K900673 Manufactured Spirometrics Medical Equipment Company
- 3.2 Brentwood IQMark tm Digital Spirometer K002499 Manufactured by Brentwood Medical Technology Corporation
AN
Page 1 A
{1}------------------------------------------------
4. Device Description:
The IQTeQ Spirometer2001 is a handheld spirometer that connects to a Personal Computer (PC) via a Universal Serial Bus (USB) port cable connection. Windows based software on the PC runs the diagnostic spirometer application.
The device consists of a plastic handle which houses the amplified pressure transducer, analog to digital conversion, USB microcontroller and connection to the USB cable. The electronic component is screened and connects to the USB cable screen. The entire electronic component is sealed in epoxy (Eli-Cast FR 453) and permanently fixed into the handle with epoxy. The electronics in the handle is powered by the 0 and 5volt power supply, which is provided with all USB ports. The top of the handle has a circular tube for connection to the flow tube, which is inserted into the handle tube.
The flow tube, which includes a laminar flow air resistance element, is inserted into the handle tube. Porting of the flow tube and the handle tube allows air pressure to connect to the pressure transducer in the handle. Two 'O' rings maintain an airtight seal between the tubes. The flow tube includes an inert ceramic (CORDIERITE) laminar flow air resistance element.
The patient is connected by mouth to a bacterial filter, which is fitted to the flow tube. The patient breathes as instructed through the flow tube. The pressure transducer in the handle continuously transmits the differential pressure in the flow tube to the analog to digital converter. The digitized 12 bit data is sent to the PC via the USB port at 335, 12bit samples per second. The IQTeQ application software which requires Windows 98 SE or later calculates the flow and integrates flow to calculate volume. The flows are calculated from the 12 bit data by applying the calibration polynomial algorithm. Therefor all flow and volume data is calculated in real time for the flow-volume measurements which includes: Peak Flow, Forced Vital Capacity, Forced Expiratory Flows Forced Inspiratory Vital Capacity as well as MVV and SVC.
M
Page 2A
{2}------------------------------------------------
The spirometer technician/operator uses the software on the PC to conduct the tests. The software includes a number of screens and 'buttons' that help to logically perform the tests in the correct order. The tests can be saved to the hard drive, printed (reports) and recalled. There are various parameters that are displayed to assist the technician in quality control of the test procedure as is requested by the ATS 1994 update document.
The system has a calibration procedure that requires 6 varying strokes from a 3 liter syringe.
5. The IOTeO Spirometer is indicated for use by Adult/Pediatric, Male/Female, to monitor, measure and evaluate pulmonary function parameters the in hospitals, clinics, physicians offices, industrial health screening and homes (as prescribed by a physician) of the following:
Forced Flow/Volume measurements, Forced Flows, Slow Vital capacity, Forced Inspiratory Volumes and Flows, MVV.
To evaluate the presence, extent, progression, physiologic disturbance and course of therapy of lung disease.
6. Comparison to predicates:
The IQTeO Spirometer is similar to the predicates in #3. The only significant difference is the method of communication to the PC. The predicate devices both connected by serial port. The IQTeO Spirometer connects via the newly adopted Universal Serial Bus (USB) PC standard, which supercedes the serial port. The USB has a number of advantages over the serial port. The USB port supplies a 5 volts power supply. Devices that connect to the PC via USB are 'Plug n Play', devices which communicate at higher speeds and more reliably than the serial port. PC's do not come equipped with serial ports anymore and only USB ports are fitted. Apart from small software design layout differences and other minor differences in characteristics The IQTeQ Spirometer is substantially equivalent to the predicate devices and does not raise any new questions regarding safety or effectiveness. As with the predicate devices, the IQTeQ Spirometer has been tested for compliance to the ATS1994 standards. The IQTeQ Spirometer also complies with the EN60601-1, IEC60601-1 electrical safety and emmision standards.
t
Page 3A
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol on the right side. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 1 2003
Mr. Raymond E. Wright Director Research & Development IOTeO Development The Penthouse Jaga House, 61 Wale Street Cape Town, Western Cape 8001 SOUTH AFRICA
Re: K020102
Trade/Device Name: IOTeO Spirometer Regulation Number: 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG, BTY, BZC, BZM Dated: January 26, 2003 Received: February 7, 2003
Dear Mr. Wright
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Wright
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646 Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Susan Runner, DDS, MA
Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
## Device Name: IQTeQ Spirometer.
## 510(k) number: k020102
## Indications for use:
The IQTeQ Spirometer is intended for use as a prescription-use-only clinical diagnostic spirometer for pulmonary function evaluation and data management.
The IQTeQ Spirometer is for use in hospitals, clinics and physician/clinician offices by individuals that have received minimal instruction or training in the administration of spirometry tests. The IQTeQ Spirometer operates with an IBM or compatible computer using a USB port (Universal serial bus) connection and the PC's installed IQTeQ Spirometer Windows™ compatible software. Minimum PC and operating system requirements are specified in the IQTeQ Spirometer Manual.
The IOTeO Spirometer is indicated for use with male/female adult patients and male/female pediatric patients to evaluate, access, describe, measure, or monitor.
- Symptoms, signs, or abnormal laboratory tests 】.
- Effects of disease on pulmonary function 2.
- Individuals at risk for pulmonary disease 3.
- 4. Preoperative risk
- 5. Post-surgical prognosis
- 6. Pre-treatment health status
- 7. Therapeutic interventions
- The course of disease affecting lung function 8.
- Persons exposed to pollutants 9.
- 10. Adverse reactions to drugs with known pulmonary toxicity
- 11. Rehabilitation programs
- 12. Risks as part of an insurance evaluation
- 13. Individuals for legal reasons
- 14. Epidemiological surveys
- 15. Derivation of reference equations
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B. Huth
Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number. K020102
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter USE
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.