The intended use of the Vitalograph Model 6000 Alpha is the simple assessment of respiratory function through the measurement of dynamic lung volumes i.e. spirometry. The device measures patient respiratory parameters including FVC, FEV1, FEV6, PEF, MVV and VC. The device is designed to be operated by medical professionals trained in respiratory and lung function testing on adults and pediatrics, 5 years and older, in a variety of professional healthcare environments, e.g. primary care, hospitals and occupational health centers.
Device Story
Desktop spirometer for professional healthcare environments; operated by trained medical professionals. Inputs: patient respiratory effort via Fleisch Pneumotachograph flowhead. Device transforms flow signals into lung volume parameters (FVC, FEV1, FEV6, PEF, MVV, VC) using flow integration sampling at 100Hz. Outputs: real-time display on touchscreen, thermal printout, and digital data export via USB, Ethernet, or Wi-Fi. Data can be exported to Spirotrac software. Assists clinicians in diagnosing and monitoring respiratory function. Benefits: provides standardized, accurate lung function measurements for clinical decision-making.
Clinical Evidence
No clinical or animal testing performed. Evidence consists of bench testing, including performance verification per ATS/ERS (2019), ISO 23747:2015, and ISO 26782:2009. Biocompatibility evaluated per ISO 10993-5, -10, -18 and ISO 18562-2, -3. Electrical safety and EMC testing conducted per AAMI ANSI ES 60601-1 and IEC 60601-1-2.
Technological Characteristics
Desktop spirometer; Fleisch Pneumotachograph flowhead; 7.2V NiMH battery; 4-inch thermal printer; USB, Ethernet, Wi-Fi connectivity. Complies with AAMI ANSI ES 60601-1, IEC 60601-1-2, ATS/ERS (2019), ISO 23747, ISO 26782. Biocompatibility per ISO 10993 and ISO 18562. Software level of concern: moderate.
Indications for Use
Indicated for assessment of respiratory function in adults and pediatrics (5 years and older) in professional healthcare settings (primary care, hospitals, occupational health).
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 26, 2022
Vitalograph Ireland Ltd. % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704
Re: K212938
Trade/Device Name: Vitalograph Model 6000 Alpha Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: December 21, 2021 Received: December 22, 2021
### Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K212938
Device Name
Vitalograph Model 6000 Alpha Indications for Use (Describe)
The intended use of the Vitalograph Model 6000 Alpha is the simple assessment of respiratory function through the measurement of dynamic lung volumes i.e. spirometry. The device measures patient respiratory parameters including FVC, FEV1, FEV6, PEF, MVV and VC. The device is designed to be operated by medical professionals trained in respiratory and lung function testing on adults and pediatrics, 5 years and older, in a variety of professional healthcare environments, e.g. primary care, hospitals and occupational health centers.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
EF PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
### Page 1 of 5
| Date Prepared: | 25-January-2022 |
|----------------|-----------------|
|----------------|-----------------|
I. Submitter
Vitalograph Ireland Ltd.
Gort Road Business Park
Ennis Co Clare V95 HFT4 Ireland
Tel - +353-65-6864100
| Submitter Contact: | Tony O'Hanlon<br>Regulatory Affairs / Quality Assurance Manager |
|---------------------------|-----------------------------------------------------------------|
| Submission Correspondent: | Paul Dryden ProMedic, LLC |
II. Device
| Proprietary or Trade Name: | Vitalograph Model 6000 Alpha |
|----------------------------|-----------------------------------------|
| Common/Usual Name: | Diagnostic spirometer |
| Classification Name: | Diagnostic spirometer (21 CFR 868.1840) |
| Regulatory Class: | II |
| Product Code: | BZG |
III. Predicate Device:
K200550 - Vitalograph Model 2120 In2itive eDiary
| Proprietary or Trade Name: | Vitalograph Model 2120 In2itive eDiary |
|----------------------------|-----------------------------------------|
| Common/Usual Name: | Diagnostic spirometer |
| Classification Name: | Diagnostic spirometer (21 CFR 868.1840) |
| Regulatory Class: | II |
| Product Code: | BZG |
#### IV. Device Description:
The Vitalograph Alpha Model 6000 is a desktop spirometer which measures the following lung function parameters FVC, FEV1, FEV6, PEF, MVV and VC in professional healthcare environments, e.g., primary care, hospitals and occupational health centers.
It is externally powered from a Class II, IEC 60601-1 compliant medical power supply. It contains a rechargeable battery powered from the external supply. The device also contains an integral 4 inch thermal printer.
The device has a USB port for connection to other devices and an SD card slot for backup of stored data. The device also has wired ethernet and Wi-Fi for connection to a hospital network.
Its primary functions and technology are:
- Spirometry measurements using single breath and multiple-breath testing techniques, the ● display and recording of measured lung volumes and flow rates (including FVC, FEV1, FEV6, PEF, MVV and VC) are identical to the predicate device
- Record subject data
- Storage of data and test results on unit for later printing or export to Spirotrac software which was cleared under 510(k) K201562
- . The Flowhead utilizes a Fleisch Pneumotachograph. The operating principle is identical to the predicate K200550
- User Interface navigation via touch screen display ●
{5}------------------------------------------------
#### V. Indications for Use:
The intended use of the Vitalograph Model 6000 Alpha is the simple assessment of respiratory function through the measurement of dynamic lung volumes i.e. spirometry. The device measures patient respiratory parameters including FVC, FEV1, FEV6, PEF, MVV and VC. The device is designed to be operated by medical professionals trained in respiratory and lung function testing on adults and pediatrics, 5 years and older, in a variety of professional healthcare environments, e.g. primary care, hospitals and occupational health centers.
Environments of use: Professional healthcare environments, e.g., primary care, hospitals and occupational health centers.
#### VI. Comparison of Technological Characteristics and Performance with the Predicate
Table 1 is a comparison - Subject Device vs. the Predicate, K200550.
{6}------------------------------------------------
## Page 3 of 5
| | Predicate | Subject Device | Comparison |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| | Model 2120 In2itive eDiary | Model 6000 Alpha | |
| K# | K200550 | K212938 | - |
| Product Code | BZG | BZG | Identical |
| CFR | 868.1840 | 868.1840 | Identical |
| Classification | Spirometer, diagnostic | Spirometer, diagnostic | Identical |
| Indications for Use | The Vitalograph Model 2120 In2itive eDiary<br>device is a battery-operated spirometer which<br>measures three basic patient respiratory<br>parameters (FVC, MVV and VC). The<br>Vitalograph Model 2120 In2itive eDiary is a<br>handheld spirometer designed for lung function<br>testing in a variety of environments such as<br>hospital wards, health centers and private homes.<br>The Vitalograph Model 2120 In2itive eDiary can<br>be configured as a stand-alone spirometer or<br>connected to a printer. | The intended use of the Vitalograph Model 6000<br>Alpha is the simple assessment of respiratory<br>function through the measurement of dynamic lung<br>volumes i.e. spirometry. The device measures<br>patient respiratory parameters including FVC,<br>FEV1, FEV6, PEF, MVV and VC. The device is<br>designed to be operated by medical professionalstrained in respiratory and lung function testing on<br>adults and pediatrics, 5 years and older, in a variety<br>of professional healthcare environments, e.g.<br>primary care, hospitals and occupational health<br>centers. | Equivalent |
| Fundamental<br>scientific<br>technology | Fleisch Pneumotachograph type Flowhead is<br>connected to a transducer, with related signal-<br>processing circuitry. | Fleisch Pneumotachograph type Flowhead is<br>connected to a transducer, with related signal-<br>processing circuitry. | Similar |
| Parameters<br>measured | FVC, FEV1, FEV6, PEF, MVV and VC | FVC, FEV1, FEV6, PEF, MVV and VC | Identical parameters |
| Hand-held | Yes | Not a desktop | Same functionality in a different<br>form |
| Patient Interface | Flowhead cone | Flowhead cone | Similar |
| Cleaning | Unit surface cleaning and 70% alcohol | Unit surface cleaning and 70% alcohol | Similar |
| Patient use | Unit surface cleaning and 70% alcohol | Unit surface cleaning and 70% alcohol | Similar |
## Table 1 - Comparison of Subject and Predicate Devices
{7}------------------------------------------------
## Page 4 of 5
| Specifications | Predicate<br>Model 2120 In2itive eDiary | Subject Device<br>Model 6000 Alpha | Comparison |
|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Back pressure | Less than 0.1kPa/L/second @ 14L/s | Less than 0.1kPa/L/second @ 14L/s | Similar |
| Volume detection | Flow integration sampling @ 100Hz | Flow integration sampling @ 100Hz | Similar |
| Maximum<br>displayed volume | 10L | 10L | Similar |
| Volume accuracy | $\pm$ 3% or 0.05L ( $\pm$ 2.5% for device and $\pm$ 0.5% for syringe) | $\pm$ 2.5% | Similar |
| Flow Accuracy<br>when operated in<br>operating<br>temperature range<br>conditions | Flow $\pm$ 10% or 0.3 L/s<br>Max. flow rate $\pm$ 16 L/s<br>Min. flow rate $\pm$ 0.02 L/s | Flow Accuracy $\pm$ 10%<br>Max. flow rate $\pm$ 16 L/s<br>Min. flow rate $\pm$ 0.02 L/s | Similar |
| Operating<br>temperature range | 10 $-$ 40 $\degree$ C | 10 $-$ 40 $\degree$ C | Similar |
| Performance<br>standards | ATS/ERS (2005)<br>ISO 23747<br>ISO 26782 | ATS/ERS (2019)<br>ISO 23747<br>ISO 26782 | Similar, the subject device<br>complies with the latest version of<br>ATS/ERS |
| Electrical Safety<br>and EMC | ES 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 | ES 60601-1<br>IEC 60601-1-2 | Similar but subject device is not<br>intended for home use |
| Communications | USB | USB, Ethernet, WiFi | Similar, additional connectivity in<br>subject device |
| Device weight | 0.230 kg | 1.5 kg | Not critical too equivalence |
| Dimensions | 160 x 100 x 45mm | 204mm (length) x 253mm (width) | Not critical too equivalence |
| Interface | Touchscreen LCD | Color touchscreen | Similar |
| Power / Energy<br>Source | 5V/3.7V Li-ion rechargeable battery | 7.2V, 2.2Ahr NiMH | Similar |
| Biocompatibility | Surface Contact, Skin / Mucosa, Limited<br>Duration | Externally Communicating, Tissue and Surface<br>Contact, Skin / Mucosa, Limited Duration | Similar |
{8}------------------------------------------------
### Page 5 of 5
#### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility -
The patient contacting materials were evaluated using: ISO 10993-5(2009) – Cytotoxicity, ISO 10993-10 (2010) - Sensitization and Irritation, ISO 10993-18 (2020) - Chemical Characterization, ISO 18562-2 (2017) - Particulate Materials, ISO 18562-3(2017) - VOC, with a toxicological risk assessment. The material were found to be acceptable for their intended use.
### Electrical Safetv and EMC
Electrical safety and EMC testing were conducted on the subject device. The system complies with AAMI ANSI ES 60601-1: 2005 + A1: 2012 and IEC 60601-1-2:2014 for EMC.
### Software Verification and Validation Testing
Software verification and validation testing were conducted. The software for this device was considered as a "moderate" level of concern.
### Mechanical, Animal, and Clinical Testing
Cleaning validation, aging. There was no animal or clinical testing was performed.
### Bench Testing
Performance testing per ATS/ERS (2019) for peak flow and timed forced expired volume per ISO 23747:2015- Peak Flow and ISO 26782:2009 - Timed Forced Expired Volume was performed.
#### Conclusion IX.
### Discussion of Differences -
The identified differences do not raise new or different concerns of safety or effectiveness relative to the predicate.
### Substantial Equivalence Conclusion
The performance testing has demonstrated that the subject devise met the applicable standard performance requirements. Through performance testing, design and features, and non-clinical testing, the proposed device and predicate are found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.