K080921 · Chest M.I., Inc. · BZG · Jan 21, 2009 · Anesthesiology
Device Facts
Record ID
K080921
Device Name
SPIRO-MASTER PC-10 SPIROMETRY SYSTEM
Applicant
Chest M.I., Inc.
Product Code
BZG · Anesthesiology
Decision Date
Jan 21, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Spiro-Master PC-10 Spirometry System is intended to acquire, view, store and print measures and waveforms of the exhaled breath of a person and evaluate pulmonary function These measures are used in the diagnosis and monitoring of lung diseases and interventions for the treatment of certain lung diseases The spirometer should only be used with patients able to understand the instructions for performing the test
Device Story
Handheld spirometer; connects to PC via USB. Input: patient exhaled breath through disposable mouthpiece/bacterial filter into pneumotach sensor. Sensor measures differential pressure across mesh ring; pressure transducer converts to electrical signal. PC software (Windows-based) performs A/D conversion; calculates flow/volume parameters in real-time using calibration polynomial algorithm. Output: flow-volume loops, volume-time graphs, and pulmonary function indices (FVC, FEV1, etc.) displayed on PC monitor; reports printed. Used in hospitals, clinics, and industrial settings by qualified clinicians. Assists clinicians in diagnosing/monitoring lung disease and evaluating pulmonary health status.
Clinical Evidence
Bench testing only. Device performance validated against ATS 1994 and 2005 standards. Calibration verified using 3-liter syringe. No clinical study data presented.
Technological Characteristics
Handheld pneumotach spirometer; plastic handle with pressure transducer, A/D converter, USB microcontroller. Powered by PC USB port (5V DC). Disposable mouthpiece/bacterial filter. Connectivity: USB 1.1/2.0 to PC. Software: Windows-based application. Standards: ATS 1994/2005, IEC 60601-1, IEC 60601-2.
Indications for Use
Indicated for use with people of all ages, excluding infants and neonates, to acquire, view, store, and print measures and waveforms of exhaled breath to evaluate, monitor, or describe pulmonary function, disease effects, risk, pre/post-surgical status, therapeutic interventions, pollutant exposure, drug toxicity, rehabilitation, insurance/legal evaluations, and epidemiological surveys.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
Predicate Devices
IQTeQ Spirometer 2001 (k020102)
Brentwood IQMarkTM Digital Spirometer (k002499)
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(k) Summary
# a. Owner/Company name, address CHEST MI , INC
3-6-10 Hongo, Bunkyo-ku, Tokyo, 113-0033, Japan
Mr Norıo Suzukı Quality Assurance Specialist Phone 011-81-3-3812-7271 Fax 011-81-3-3812-7496 Email bck04184@nifty com
#### b. Contact/Application Correspondent
Fumıakı Kanaı, Ph D President and CEO MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan
011-81-3-3818-8577 Phone Fax 011-81-3-3818-8577 Email fkanaı@micinternational com
#### c. Date prepared
January 14, 2008
#### d. Name of device
| Trade Name | Spiro-Master PC-10 Spirometry System |
|---------------------------|--------------------------------------|
| Common Name | Diagnostic spirometer |
| Classification Name | Diagnostic spirometer |
| Classification Regulation | 21 CFR 868 1840 |
#### e. Predicate devices
The Spiro-Master PC-10 Sprometry System is substantially equivalent to the following legally marketed devices
| 510(k) | k020102 |
|--------------|---------------------------------------|
| Trade name | IQTeQ Spirometer 2001 |
| Product code | BZG |
| 510(k) | k002499 |
| Trade name | Brentwood IQMarkTM Digital Spirometer |
| Product code | BZG |
SPIRO-MASTER PC-10 SPIROMETRY SYSTEM (K080921)
JAN 2 1 2009
{1}------------------------------------------------
### SPIRO-MASTER PC-10 SPIROMETRY SYSTEM PREMARKET NOTIFICATION 510(k)
## f. Description of the device
The "Spiro-Master PC-10 Spirometry System" is a handheld spirometer that connects to a Personal Computer (PC) via a Universal Serial Bus (USB 11 and 20) cable Windows 10-based software on the PC runs the diagnostic spirometer application
The device consists of a plastic handle which houses the amplified pressure transducer, analog to digital converter. USB microcontroller and connection via a USB cable The electronics in the handle are powered by the 5 volts DC of the PC USB port The top of the handle has a connection for the flow sensor, which connects to the mouthpiece
Image /page/1/Picture/5 description: The image shows a laptop computer next to a piece of equipment. The equipment has a cylindrical component at the top, which is connected to a rectangular base by an arm. A cable is connected to the base of the equipment. The laptop is open and the screen is visible.
Spiro-Master PC-10 Spirometry System Figure 1
A Pneumotach sensor is incorporated into the Spiro-Master PC-10 Spirometry System to detect arrilow The flow sensor consists of two flow tubes that connect to each other around a sensor mesh ring which provides flow resistance A differential pressure across the mesh is generated depending on the inhalation/exhalation strength of the person being tested Inside the flow tube, the front and back of the mesh is connected to a pressure sensor through pressure outlet ports and tubes This differential pressure transducer measures the pressure difference across the flow sensor (mesh ring) as the air-stream passes through it The resulting pressure change is converted to a signal proportional to the airflow rate
Image /page/1/Figure/8 description: The image shows a diagram of a flow sensor. The air flows through a pipe, and a flow sensor mesh ring is placed in the middle of the pipe. The pressure difference (ΔP) is measured, and the data is sent to a processor, which outputs measurement values.
Figure 2 Block diagram of air flow detection circuit
{2}------------------------------------------------
The patient places their mouth on the mouthpiece, which is connected to a bacterial filter (KoKo Moe White Filter) between the mouthpiece and the sensor The mouthpiece is a single-use, disposable component.
Image /page/2/Figure/3 description: The image shows a diagram of a device with several components. The components are labeled as "Mouthpiece", "Bacterial filter", and "Sensor". An arrow shows the direction of flow from the mouthpiece to the bacterial filter and then to the sensor. A note says, "Check to be sure that the sensor is connected in the correct direction".
Figure 3 Patient connection components
The patient breathes as instructed through the mouthpiece The pressure transducer in the handle continuously transmits the differential pressure in the flow tube to the analog to digital converter The digitized 12 bit data is sent to the PC via the USB cable at 12 bit samples per second The Spiro-Master PC-10 Spirometry System application software (which requires Windows 98 or later), calculates the flow and integrates the flow to calculate volume The flows are calculated from the 12 bit data stream by applying a calıbration polynomial algorithm All flow and volume data are calculated in real tıme for the flow-volume measurements which includes Peak Flow, Forced Vital Capacity, Forced Expiratory Flows Forced Inspiratory Vital Capacity as well as FEF and FIVC
The spirometer technician/operator uses the software on the PC to perform the tests The software includes a number of screens and 'buttons' that assust to logically perform the tests in the correct order The tests can be saved to the PC hard drive, printed (report format) and recalled for later review There are various parameters that are displayed to assust the technician in the quality control of the test procedure per the ATS 1994 update and the ATS 2005 standards The system calıbration procedure requires sıx (6) varyıng strokes from a three (3) liter syringe
{3}------------------------------------------------
Image /page/3/Picture/35 description: The image shows a simple line drawing of a cube. The cube is drawn in a way that emphasizes its three-dimensional form. The lines are thin and black, and the background is white.
# g. Intended Use, Indications for Use, and Environment
#### Intended Use
The Spiro-Master PC-10 Spirometry System is intended to acquire, view, store and print measures and waveforms of the exhaled breath of a person and evaluate pulmonary function These measures are used in the diagnosis and monitoring of lung diseases and interventions for the treatment of certain lung diseases The spirometer should only be used with pattents able to understand the instructions for performing the test
#### Indication for Use
The Spiro-Master PC-10 Spirometry System is indicated for use with people of all ages, of either gender, excluding infants and neonates, to acquire, view, store and print measures and waveforms of the exhaled breath of a person and evaluate, access, describe, measure, or monitor
- Symptoms, signs, or abnormal laboratory tests 1
- Effects of disease on pulmonary function 2
- 3 Individuals at risk for pulmonary disease
- 4 Preoperative risk
- 5 Post-surgical prognosis
- ર્ભ Pre-treatment health status
- 7 Therapeutic interventions
- 8 The course of disease affecting lung function
- 9 Persons exposed to pollutants
- Adverse reactions to drugs with known pulmonary toxicity 10
- Rehabilitation programs 11
- Risks as part of an insurance evaluation 12
- Individuals for legal reasons 13
- 14 Epidemiological surveys
- Derivation of reference equations । રે
#### Environment of Use
Places where a qualified cluncian desires to take lung function measurement such as in hospitals, clinics, physicians' offices, laboratories and industrial health screenings
#### h. Statement of substantial equivalence
The characteristics of the Spiro-Master PC-10 Spirometry System are similar to those of the predicates described in Item e above The similarities are
- . same intended use
- . same operating principle
- same spirometry parameters �
- same algorithms for spirometry parameters calculation .
- same physical aspect (shape, size and weight) 0
There are no differences regarding safety or effectiveness Refer to Table 1 "Companson Table" on the following two pages for a complete assessment
{4}------------------------------------------------
SPIRO-MASTER PC-10 SPIROMETRY SYSTEM
PREMARKET NOTIFICATION 510(k)
m
| | Table 1 Comparison Table | | | |
|---|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Characteristic | IQTEQ Spirometer 2001 | Predicate Devices<br>Brentwood IQmark™ Digital Spirometer | Spiro-Master PC-10 Spirometry System |
| 1 | PC Based | Yes | Yes | Yes |
| 2 | Physical<br>configuration | The IQTEQ Spirometer2001 is a handheld<br>spirometer that connects to a Personal<br>Computer (PC) via a Universal Serial Bus<br>(USB) port cable connection Windows<br>based software on the PC runs the<br>diagnostic spirometer application | Disposable Pneumotach Mouthpiece with<br>handle (containing electronics) that<br>connects directly to PC serial port via<br>cable system software disk for PC<br>installation | Spiro-Master PC-10 is a handheld<br>spirometer that connects to a Personal<br>Computer (PC) via a Universal Serial<br>Bus (USB) port cable connection<br>Windows based software on the PC<br>runs the diagnostic spirometer<br>application |
| 3 | Minimum PC<br>requirements | Pentium II with USB port, 128mb RAM,<br>Windows 98 Second Edition, CD-ROM<br>drive, mouse, and 800X600 screen<br>resolution | Pentium 100 MHz Processor, 16MB RAM<br>(Windows 95&98), 32MB RAM<br>(Windows NT), 64MB RAM (Windows<br>2000 Professional), 10MB HD space,<br>Windows compatible printer | Windows 98, 2000, XP, Vista<br>Keyboard, mouse, and a CD-Rom<br>drive<br>VGA display accommodating<br>1024x786, 1280x1024, or higher<br>resolution<br>At least one USB port |
| 4 | Power source | Power derived from USB port | Disposable Pneumotach Mouthpiece with<br>handle two series AAA batteries (1 5v)<br>inside handle | Power derived from USB port |
| 5 | ATS Spirometry<br>performance<br>recommendations | Complies (1994 Update) | Complies (1994 Update) | Complies (1994 Update)<br>Complies (2005) |
| 6 | Cross<br>contamination<br>control | Disposable mouth piece & bacterial filter | Disposable mouth piece & bacterial filter | Disposable mouth piece & bacterial<br>filter |
| 7 | Flow detection<br>principle | Differential pressure Laminar flow<br>resistance element | Disposable bi-directional pressure<br>differential measuring pneumotach<br>mouthpieces for expiratory/inspiratory<br>testing | Reusable pressure differential<br>measuring pneumotach |
| 8 | Flowmeter<br>calibration method | Injection of known fixed volume from<br>calibrated syringe | Injection of known fixed volume from<br>calibrated syringe | Injection of known fixed volume from<br>calibrated syringe |
| 9 | Display & printer<br>used | PC monitor screen (LCD/CRT) and PC<br>printer | PC monitor screen (CRT), alphanumerics<br>and graphics | PC monitor screen (LCD/CRT) and PC<br>printer |
CONFIDENTIAL
{5}------------------------------------------------
| Table 1 Comparison Table (continued) | | | | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| | | Predicate Devices | | | | |
| Characteristic | IQTEQ Spirometer 2001 | Brentwood IQmark Digital Spirometer | Spiro-Master PC-10 | | | |
| 10 Graphic output | Flow-Volume loop,<br>Volume/Time graph,<br>predicted curve | Flow-Volume loop, Volume-Time curve,<br>predicted curve, pre & post bronchodilator<br>comparison | Flow-Volume loop, Volume/Time<br>graph, predicted curve, pre & post<br>bronchodilator comparison,<br>bronchodilator-challenge | | | |
| 11 Tests performed | FVC, FIVC, PRE/POST Bronchodilator,<br>Flow/Volume Loop, Volume/Time graph,<br>SVC | FVC, F-V loop, MVV, VC(SVC), IVC,<br>respiratory pattern, PRE/POST<br>comparisons, bronchodilator-challenge | FVC, MVV, MV, VC(SVC),<br>respiratory pattern, pre/post<br>comparisons, broncho-challenge | | | |
| 12 Indices calculated<br>(bold text indicates<br>that an index used<br>by Brentwood<br>spirometer is also<br>used by a listed<br>predicate) | FVC, FEV1, FEV3, FEV6, FEV1/FVC,<br>FEV1/FEV6, PEFR, FEF25%, FEF50%,<br>FEF75%, FEF25-75, FEF75-85, FIVC, PIFR,<br>FIV0.5, FIV1, FIV3, FIV1/FIVC, FIF50%,<br>FET100%, SVC Extrapolated V%, and<br>V(L), Peak risk time | FVC, FEV0.5, FEV10, FEV30,<br>FEV1/FVC; FEF25-75, FEF75-85, FEF25,<br>FEF50, FEF75, FEF200-1200, PEF, FIVC,<br>FIF50, FEF50/FIF50, PIF, MVV, VC, VT,<br>ERV, RR, tE, Vext, FIV0.5, FIV0.5/FIV0.5,<br>MVT, IRV | SVC (VC), TV, FVC, FEV0.5, FEV10,<br>FEV30, FEV60, FEV1/FVC,<br>FEV1/SVC, MMEF, PEF, FEF25,<br>FEF50, FEF75, FEF25-75, FEF200-1200,<br>FET, FIVC, FIV0.5, FIV10, PIF, FIF50,<br>FEF50/FIF50, FIF50/FEF50, MVV, RR,<br>MV | | | |
| 13 Predictive models<br>used (bold text<br>indicates<br>predicated model<br>used by Brentwood<br>spirometer also<br>used by predicate) | Crapo, ECCS, Chernisk, Morris,<br>Knudoson1983, Hsu, Schoenberg,<br>NHANES III 1999 (Hankınson), Plogar<br>79 | Adult Kundson 76 & 83, Crapo(ITS),<br>ECCS 93<br>Pediatric Plogar 71 and Knudson 76 &<br>83u | Crapo-Hsu, ECCS, ITS, Knudson,<br>Morrıs, Polgar, NHANES III 1999<br>(Hankınson) | | | |
SPIRO-MASTER PC-10 SPIROMETRY SYSTEM
(K080921)
ATTACHMENT 4
4-6 OF 4-6 OF 4-7
CONFIDENTIAI
SPIRO-MASTER PC-10 SPIROMETRY SYSTEM
PREMARKET NOTIFICATION 510(k)
{6}------------------------------------------------
#### SPIRO-MASTER PC-10 SPIROMETRY SYSTEM PREMARKET NOTIFICATION 510(k)
As with the predicate devices, the Spiro-Master PC-10 Spirometry System has been tested for compliance to the ATS1994 (Update), IEC60601-1 and IEC60601-2 standards In addition the Spiro-Master PC-10 Spirometry System also complies with the ATS2005 standard
Therefore CHEST MI, INC believes that the Spiro-Master PC-10 Spirometry System is substantially equivalent to the predicate devices and does not raise any new questions regarding safety or effectiveness
#### Conclusion i.
Based on the above discussion and enclosed sections regarding substantial equivalence to predicate devices. CHEST MI. INC concludes that the Spiro-Master PC-10 Spirometry System is substantially equivalent to the IQTeQ Spirometer 2001 (k020102) and the Brentwood IQMark" Digital Spirometer (k002499) and does not raise any new questions regarding safety or effectiveness
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized caduceus, a symbol of medicine and health. The caduceus consists of a staff with two snakes coiled around it, and a pair of wings at the top.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Chest M I , Incorporated C/O Fumiakı Kanaı, Ph D President and Chief Executive Officer MIC International 4-1-17 Hongo Bunkyo-ku Tokyo JAPAN 113-0033
JAN 2 1 2009
Re K080921
> Trade/Device Name Spiro-Master PC-10 Spirometry System Regulation Number 21 CFR 868 1840 Regulation Name Diagnostic Spirometer Regulatory Class II Product Code BZG Dated January 14, 2009 Received January 16, 2009
Dear Dr Kanaı
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affectung your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addition, FDA may publish further announcements concerning your device in the Federal Register
{8}------------------------------------------------
Page 2 - Dr Kanaı
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requrements, including, but not hmited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801), good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474 For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http //www fda gov/cdrh/industry/support/index html
Sincerely yours,
Anthony V. Arintor for
Ginette Michaud, M D Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (If known) K080921
Device Name Spiro-Master PC-10 Spirometry System
#### Intended Use
The Spiro-Master PC-10 Spirometry System is intended to acquire, view, store and print measures and waveforms of the exhaled breath of a person and evaluate pulmonary function These measures are used in the dragnosis and monitoring of lung diseases and interventions for the treatment of certain lung diseases The spirometer should only be used with patients able to understand the instructions for performing the test
#### Indication for Use
The Spiro-Master PC-10 Spirometry System is indicated for use with people of all ages, of either gender, excluding infants and neonates, to acquire, view, store and print measures and waveforms of the exhaled breath of a person and evaluate, access, describe, measure, or monitor
- Symptoms, signs, or abnormal laboratory tests 1
- 2 Effects of disease on pulmonary function
- 3 Individuals at risk for pulmonary disease
- 4 Preoperative risk
- 5 Post-surgical prognosis
- Pre-treatment health status 6
- 7 Therapeutic interventions
- 8 The course of disease affecting lung function
- 9 Persons exposed to pollutants
- 10 Adverse reactions to drugs with known pulmonary toxicity
- Rehabilitation programs l I
- 12 Risks as part of an insurance evaluation
- 13 Individuals for legal reasons
- 14 Epidemiological surveys
- ા ર Derivation of reference equations
#### Environment of Use
Places where a qualified clinician desires to take lung function measurement such as in hospitals, clinics, physicians' offices, laboratories and industrial health screenings
Image /page/9/Picture/24 description: The image shows the words "Prescription Use" followed by the text "(Per 21 CFR 801 Subpart D)". There is a large "X" marked over the words "Prescription Use". The text is in black font and is on a white background.
AND/OR
Over-the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
A
(P
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number K080921
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.