K002185 · Medison America, Inc. · IYN · Aug 1, 2000 · Radiology
Device Facts
Record ID
K002185
Device Name
SONOACE SA 9900 DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
Medison America, Inc.
Product Code
IYN · Radiology
Decision Date
Aug 1, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Fetal OB/GYN . Abdominal . . Small Organs (breast, thyroid, testicle) Pediatric . Neonatal Cephalic . Trans-Vaginal . . Trans-Rectal Peripheral Vascular . Cardiac . Typical examinations performed using the system are: . General abdominal and pelvic studies including organ surveys, assessment, and retroperitoneal cavity studies. Study of small parts and superficial structures including breasts, shoulders, . thyroid, and the abdominal wall. . Pediatric scans of organs, superficial, and bony structures. Monitoring procedures for infertility studies (other than in vitro fertilization). . First, second and third trimester pregnancy studies. . Neonatal head studies. . Podiatry scans of superficial structures including muscles, tendons and bones. . General cardiac studies in adults. . Prostate, bladder and rectum visualization.
Device Story
Cart-mounted diagnostic ultrasound system; uses digital beamforming technology; supports multiple transducers (1.0–20.0 MHz). Inputs: ultrasound signals from piezoelectric transducers. Processing: digital beamforming; software-based parameter controls (transmit focusing, filtering, image enhancement, dynamic window curves). Outputs: B-mode, M-mode, Tissue Harmonic Imaging (THI), Color-flow, Continuous (CW) Doppler, Pulsed (PW) Doppler, Power Doppler, and 3D imaging. Used in clinical settings by healthcare professionals for diagnostic imaging and biopsy guidance. Features include cine loop, image filing, and on-screen biopsy guidelines. Benefits: high-resolution imaging and measurement capabilities for diverse clinical applications.
Clinical Evidence
Bench testing only. Biocompatibility testing performed on patient contact materials. Acoustic output measurements provided to demonstrate compliance with FDA Track 3 limits.
Technological Characteristics
Digital beamforming ultrasound system. Transducers use piezoelectric material. Operating modes: B, M, THI, Color/Power/PW/CW Doppler. Frequencies: 1.0–20.0 MHz. Biocompatible patient contact materials. Acoustic output: ISPTA 720 mW/cm2, MI 1.9. Software-based parameter control.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including fetal, abdominal, pediatric, small organ (breast, thyroid, testicle), neonatal/adult cephalic, cardiac, trans-rectal, trans-vaginal, peripheral vascular, and musculoskeletal applications. Includes biopsy guidance and infertility monitoring.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
510(k) Premarket Notification
# AUG - 1 2000
2185
SA - 9900 Ultrasound System
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
- Submitter's name, address, telephone number, contact person: 1) Regulatory Management Services 16303 Panoramic Way San Leandro, CA 94578-1116 Gary J. Allsebrook Regulatory Affairs Consultant (510) 276-2648 Telephone: (510) 276-3559 Fax: regman1@home.com Email June 19, 2000 Prepared
- Name of the device, including the trade or proprietary name if 2) applicable, the common or usual name, and the classification name, if known:
Common/Usual Name:
Diagnostic Ultrasound System and Accessories
## Proprietary Name:
SA - 9900 Diagnostic Ultrasound System and Transducers.
Classification Names:
| | FR Number | Product Code |
|---------------------------------------|-----------|--------------|
| Ultrasound Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
{2}------------------------------------------------
#### Identification of the predicate or legally marketed device: 3)
Medison America, Inc believes that SA-9900 Ultrasound system is substantially equivalent to the currently marketed Medison/Kretztechnik Combison 530D (K940942 & K992155), SonoAce SA-8800 (K974269) and SonoAce 6000C/Ultramark 400C (K990970 & K992470) .
#### 4) Device Description:
### SonoAce SA-9900:
The SonoAce SA-9900 scanner is a multiple-mode, multiple-application ultrasound imaging system. The cart-mounted console contains an ultrasound generator/receiver offering a full complement of conventional operating modes, software-based parameter controls, and recording. The selection of eight transducers to be offered with the system permits a wide range of clinical applications. The various transducers adapt the system for the specific imaging tasks.
Eight different models of transducers are available. In addition to the initial operational settings for each transducer preprogrammed in the system, usercustomized parameter settings for each transducer may be inserted by the operator and stored for recall as needed via the system control panel. Customization includes transmit focusing, filtering, image enhancement processing, dynamic window curve selection. Controls are also provided to select display format (single and various combinations), to activate zoom features, and to utilize the cine loop function. Patient contact materials have been tested for biocompatibility in accordance to their intended use and are used for each individual transducer.
The SonoAce SA-9900 uses digital beamforming technology. The SonoAce SA-9900 supports a variety of Linear and Convex probes for wide variety of applications. It is an ultrasound scanner, which provides high resolution, high
{3}------------------------------------------------
penetration performance, and various measurement functions. Probes are supported in frequencies from 1.0 MHz to 20.0 MHz. The SonoAce SA-9900 provides high quality images and various measuring functions. Biopsy guidelines are provided on screen to assist in the collection of tissue samples, using biopsy guide adapters offered as an optional accessory. Operating Modes of the SonoAce SA-9900 are B-Mode, M-Mode, Tissue Harmonic Imaging (THI), Color-flow Doppler, Continuous (CW) Doppler, Pulsed (PW) Doppler and Power Doppler. Management of patient history is possible by image-filing function. High-resolution images are provided by utilizing a technology called digital dynamic receive focusing. The same clinical uses were cleared for the predicate device(s), Medison/Kretztechnik Combison 530D (K940942 & K992155), SonoAce SA-8800 (K974269) and SonoAce 6000C/Ultramark 400C (K990970 & K992470) .
#### 5) Intended Use:
- Fetal OB/GYN .
- Abdominal .
- . Small Organs (breast, thyroid, testicle)
- Pediatric .
- Neonatal Cephalic .
- Trans-Vaginal .
- . Trans-Rectal
- Peripheral Vascular .
- Cardiac .
Typical examinations performed using the system are:
- . General abdominal and pelvic studies including organ surveys, assessment, and retroperitoneal cavity studies.
- Study of small parts and superficial structures including breasts, shoulders, . thyroid, and the abdominal wall.
{4}------------------------------------------------
- . Pediatric scans of organs, superficial, and bony structures.
- Monitoring procedures for infertility studies (other than in vitro fertilization).
- . First, second and third trimester pregnancy studies.
- . Neonatal head studies.
- . Podiatry scans of superficial structures including muscles, tendons and bones.
- . General cardiac studies in adults.
- . Prostate, bladder and rectum visualization.
#### 6) Technological Characteristics:
This device operates identical to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as a 2D and M-mode images. Scanhead patient contact materials are biocompatible.
The device's acoustic output limits are:
All Applications:
| ISPTAd | 720 mW/cm2 (maximum) | |
|-------------|----------------------|-----------|
| TIS/TIB/TIC | 0.0 - 5.0 | (Range) |
| MI | 1.9 | (Maximum) |
The limits are the same as predicate Track 3 devices.
{5}------------------------------------------------
Image /page/4/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of a bird or eagle with its wings spread.
#### 1 2000 AUG
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medison America, Inc. c/o Carole Stamp TUV Product Service 1775 Old Highway 8, NW, Suite 104 New Brighton, MN 55112-1891
Re: K002185
Sonoace SA - 9900 Diagnostic Ultrasound System Regulatory Class: II 21CFR 892.1550/Procode: 90 IYN 21CFR 892.1560/Procode: 90 IYO 21CFR 892.1570/Procode: 90 ITX Dated: July 18, 2000 Received: July 19, 2000
Dear Ms. Stamp:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranion.
This determination of substantial equivalence applies to the following transducers intended for use with the Sonoace SA - 9900 Diagnostic Ultrasound System, as described in your premarket notification:
#### Transducer Model Number
EC4-9ES, Curved Linear Array 6.5 MHz/10R/140D L5-121R, Linear Array 7.5 MHz/40mm/192 elements C2-51R, Curved Linear Array 3.5 MHz/40R/85D P2-5AC, Phased Array 3.5 MHz P3-7AC, Phased Array 4.5 MHz 2.0CW, Static CW 2.0 MHz S-VAW3-5, Volume Curved Array 3.5 MHz S-VAW4-7, Volume Curved Array 4.5 MHz S-VDW5-8, Volume Curved Array 6.5 MHz
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
#### Page -2- Ms. Carole Stamp
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally; for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Paul M. Gammell, Ph. D. at (301) 594-1212.
Sincerely yours,
Daniel A. Seymore
Daniel G. Schultz, M. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{7}------------------------------------------------
### Section 4.3 INDICATION FOR USE Ultrasound Device Indications Statement
510(k) Number:
Device Name: SonoAce 9900 Ultrasound System
Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | |
| Fetal (See Note 3) | N | N | N | N | | N | N | Note 1 | Note 2, 5, 6 |
| Abdominal | N | N | N | N | | N | N | Note 1 | Note 2, 5, 6 |
| Intra-Operative<br>(See Note 7) | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | |
| Pediatric | N | N | N | N | N | N | N | Note 1 | Note 2, 5, 6 |
| Small Organ<br>(See Note 4) | N | N | N | N | | N | N | Note 1 | Note 2, 5 |
| Neonatal Cephalic | N | N | N | N | N | N | N | Note 1 | Note 5, 6 |
| Adult Cephalic | N | N | N | N | N | N | N | Note 1 | Note 5, 6 |
| Cardiac | N | N | N | N | N | N | N | Note 1 | Note 5, 6 |
| Transesophageal | | | | | | | | | |
| Trans-Rectal | N | N | N | N | | N | N | Note 1 | Note 2, 5 |
| Trans-Vaginal | N | N | N | N | | N | N | Note 1 | Note 2, 5 |
| Trans-Urethral | | | | | | | | | |
| Intra-Vascular | | | | | | | | | |
| Peripheral -Vascular | N | N | N | N | | N | N | Note 1 | Note 5 |
| Laparoscopic | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | N | N | N | N | | N | N | Note 1 | Note 2, 5 |
| Muscular-Skeletal<br>Superficial | N | N | N | N | | N | N | Note 1 | Note 2, 5 |
| Others(Specify) | | | | | | | | | |
Mode of Operation (*includes simultaneous B-mode)
N = new indication; P = previously cleared in K990970, SA 6000C/ UM 400C system; E = added under Appendix E Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
510(k) Number
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: 3D Imaging
Note 6: Harmonic Imaging
Note 7: Abdominal organs and peripheral vessel
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Segner
(Division Sign-Off)
Division of Reproductive, Abdominal, EN and Radiological Devices
{8}------------------------------------------------
| 510(k) Number: | |
|----------------------|------------------------------------------------------------------------------------|
| Device Name: | SonoAce 9900 Ultrasound System |
| Transducer: | EC4-9ES, Curved Linear Array 6.5MHz/10R/140D |
| Indications for Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |
Mode of Operation (*includes simultaneous B-mode)
| | | | Color | Power | Combined | Other |
|--|--|--|-------|-------|----------|-------|
|--|--|--|-------|-------|----------|-------|
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) | |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------|--------------------|--|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-Operative<br>(See Note 7) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(See Note 4) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | N | N | N | | N | N | Note 1 | Note 2, 5 | |
| Trans-Vaginal | | N | N | N | | N | N | Note 1 | Note 2, 5 | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others(Specify) | | | | | | | | | | |
N = new indication; P = previously cleared in K990970, SA 6000C/ UM 400C system; E = added under Appendix E Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler (N)
Note 2: Includes imaging for guidance of biopsy (N)
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: 3D Imaging (N)
Note 6: Harmonic Imaging
Note 7: Abdominal organs and peripheral vessel
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Daniel le. Seaman
(Division Sign-Off) Division of Reproductive, Abdominal, El and Radiological D
510(k) Number
{9}------------------------------------------------
#### 510(k) Number:
SonoAce 9900 Ultrasound System Device Name:
L5-12IR. Linear Array 7.5MHz/40mm/192elements Transducer:
Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) | |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------|--------------------|--|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-Operative<br>(See Note 7) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | Note 1 | Note 2, 5 | |
| Small Organ<br>(See Note 5) | | N | N | N | | N | N | Note 1 | Note 2, 5 | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Peripheral -Vascular | | N | N | N | | N | N | Note 1 | Note 5 | |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | N | N | N | | N | N | Note 1 | Note 2, 5 | |
| Muscular-Skeletal<br>Superficial | | N | N | N | | N | N | Note 1 | Note 2, 5 | |
| Others(Specify) | | | | | | | | | | |
Mode of Operation (*includes simultaneous B-mode)
N = new indication; P = previously cleared in K990970, SA 6000C/ UM 400C system; E = added under Appendix E Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler (N)
Note 2: Includes imaging for guidance of biopsy (N)
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: 3D Imaging (N)
Note 6: Tissue Harmonic Imaging
Note 7: Abdominal organs and peripheral vessel
Concurrence of CDRH, Office of Device Evaluation (ODE)
✓ Prescription Use (Per 21 CFR 801.109)
David A. Hegmann
(Division Sign-Off) Division of Reproductive, Abdominal, E and Radiological De
510(k) Number K002185
{10}------------------------------------------------
510(k) Number: SonoAce 9900 Ultrasound System Device Name: C2-5IR, Curved Linear Array 3.5MHz/40R/85D Transducer: Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | |
| Fetal (See Note 3) | | N | N | N | | N | N | Note 1 | Note 2. 5. 6 |
| Abdominal | | N | N | N | | N | N | Note 1 | Note 2, 5, 6 |
| Intra-Operative<br>(See Note 7) | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | Note 1 | Note 2. 5. 6 |
| Small Organ<br>(See Note 4) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Cardiac | | | | | | | | | |
| Transesophageal | | | | | | | | | |
| Trans-Rectal | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | |
| Trans-Urethral | | | | | | | | | |
| Intra-Vascular | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | |
| Others(Specify) | | | | | | | | | |
Mode of Operation (*includes simultaneous B-mode)
N = new indication; P = previously cleared in K990970, SA 6000C/ UM 400C system; E = added under Appendix E Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler (N)
Note 2: Includes imaging for guidance of biopsy (N)
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: 3D Imaging (N)
Note 6: Tissue Harmonic Imaging (N)
Note 7: Abdominal organs and peripheral vessel
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
Torid A. Hegson
(Division Sign-Off) (Division Sign-On)
Division of Reproductive, Abdominal, E and Radiological
{11}------------------------------------------------
510(k) Number:
SonoAce 9900 Ultrasound System Device Name:
Transducer: P2-5AC, Phased Array 3.5MHz
Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation (*includes simultaneous B-mode)
| Clinical Application | A | B | M * | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|-----|------|-----|-------------------|----------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | |
| Fetal (See Note 3) | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-Operative<br>(See Note 7) | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | Note 1 | Note 5. 6 |
| Small Organ<br>(See Note 4) | | | | | | | | | |
| Neonatal Cephalic | | N | N | N | N | N | N | Note 1 | Note 5, 6 |
| Adult Cephalic | | | | | | | | | |
| Cardiac | | N | N | N | N | N | N | Note 1 | Note 5, 6 |
| Transesophageal | | | | | | | | | |
| Trans-Rectal | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | |
| Trans-Urethral | | | | | | | | | |
| Intra-Vascular | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | |
N = new indication; P = previously cleared in K990970, SA 6000C/ UM 400C system; E = added under Appendix E Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (N) Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: 3D Imaging (N)
Note 6: Tissue Harmonic Imaging (N)
Note 7: Abdominal organs and peripheral vessel
**Concurrence of CDRH, Office of Device Evaluation (ODE)**
Prescription Use (Per 21 CFR 801.109)
Daniel G. Legum
(Division Sign-Off) (Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K002183
{12}------------------------------------------------
| 510(k) Number: | |
|----------------|--|
|----------------|--|
SonoAce 9900 Ultrasound System Device Name:
Transducer: P3-7AC, Phased Array 4.5MHz
Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation (*includes simultaneous B-mode)
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) | |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------|--------------------|--|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-Operative<br>(See Note 7) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | Note 1 | Note 5, 6 | |
| Small Organ<br>(See Note 4) | | | | | | | | | | |
| Neonatal Cephalic | | N | N | N | N | N | N | Note 1 | Note 5, 6 | |
| Adult Cephalic | | N | N | N | N | N | N | Note 1 | Note 5, 6 | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others(Specify) | | | | | | | | | | |
N = new indication; P = previously cleared in K990970, SA 6000C/ UM 400C system; E = added under Appendix E Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (N) Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
510(k) Number
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: 3D Imaging (N)
Note 6: Tissue Harmonic Imaging (N)
Note 7: Abdominal organs and peripheral vessel
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Bet 21 CFR 801.109)
David C. Hanson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
Basic Information
{13}------------------------------------------------
| 510(k) Number: | |
|----------------------|------------------------------------------------------------------------|
| Device Name: | SonoAce 9900 Ultrasound System |
| Transducer: | 2.0CW, Static CW 2.0MHz |
| Indications for Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body |
as follows:
**Mode of Operation** (*includes simultaneous B-mode)
| | | | | Color | Power | Combined | Other |
|--|--|--|--|-------|-------|----------|-------|
|--|--|--|--|-------|-------|----------|-------|
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | |
| Fetal (See Note 3) | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-Operative<br>(See Note 7) | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | |
| Pediatric | | | | | N | | | | |
| Small Organ<br>(See Note 4) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Cardiac | | | | | N | | | | |
| Transesophageal | | | | | | | | | |
| Trans-Rectal | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | |
| Trans-Urethral | | | | | | | | | |
| Intra-Vascular | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
N = new indication; P = previously cleared in K990970, SA 6000C/ UM 400C system; E = added under Appendix E Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
510(k) Number
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: 3D Imaging
Note 6: Tissue Harmonic Imaging
Note 7: Abdominal organs and peripheral vessel
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Prescription Use (Per 21 CFR 801.109)
David A. Lynn
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
{14}------------------------------------------------
#### 510(k) Number:
SonoAce 9900 Ultrasound System Device Name:
S-VAW3-5, Volume Curved Array 3.5MHz Transducer:
Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation (*includes simultaneous B-mode)
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | |
| Fetal (See Note 3) | E | E | E | E | | E | E | Note 1 | Note 2, 5, 6 |
| Abdominal | E | E | E | E | | E | E | Note 1 | Note 2. 5, 6 |
| Intra-Operative<br>(See Note 7) | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | |
| Pediatric | E | E | E | E | | E | E | Note 1 | Note 2, 5, 6 |
| Small Organ<br>(See Note 4) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Cardiac | | | | | | | | | |
| Transesophageal | | | | | | | | | |
| Trans-Rectal | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | |
| Trans-Urethral | | | | | | | | | |
| Intra-Vascular | | | | | | | | | |
| Peripheral - Vascular | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | |
| Others(Specify) | | | | | | | | | |
N = new indication; P = previously cleared in K940942, Voluson C530D system; P = previously cleared in K974813, Voluson C530D Power Doppler mode; P2 = previously cleared in K993517, Voluson C530D Power Doppler mode; E = added under Appendix E
Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler (E)
Note 2: Includes imaging for guidance of biopsy (E)
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: 3D Imaging (E)
Note 6: Tissue Harmonic Imaging (P2)
Note 7: Abdominal organs and peripheral vessel
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David h. leppon
(Division Sign-Off) Division of Reproductive, Abdominal, El and Radiological Devices
510(k) Number
Basic Information
{15}------------------------------------------------
#### 510(k) Number:
SonoAce 9900 Ultrasound System Device Name:
S-VAW4-7, Volume Curved Array 4.5MHz Transducer:
Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation (*includes simultaneous B-mode)
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | |
| Fetal (See Note 3) | | P | P | P | | P | P | Note 1 | Note 2, 5, 6 |
| Abdominal | | P | P | P | | P | P | Note 1 | Note 2, 5, 6 |
| Intra-Operative<br>(See Note 7) | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | |
| Pediatric | | P | P | P | | P | N | Note 1 | Note 2, 5, 6 |
| Small Organ<br>(See Note 4) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Cardiac | | | | | | | | | |
| Transesophageal | | | | | | | | | |
| Trans-Rectal | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | |
| Trans-Urethral | | | | | | | | | |
| Intra-Vascular | | | | | | | | | |
| Peripheral - Vascular | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | |
| Others(Specify) | | | | | | | | | |
N = new indication; P = previously cleared in K992155, C530D Voluson system; P = previously cleared in K974813, Voluson C530D Power Doppler mode; P2 = previously cleared in K993517, Voluson C530D Power Doppler mode; E = added under Appendix E
Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler (P)
Note 2: Includes imaging for guidance of biopsy (P)
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: 3D Imaging (P)
Note 6: Tissue Harmonic Imaging (P2)
Note 7: Abdominal organs and peripheral vessel
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 2PCFR 801.109)
David A. Lymm
(Division Sign-Off)
production of Reproductive, Abdominal, EN
Division of Reproductive, Abdominal, EN and Radiological Devices
510(k) Number
{16}------------------------------------------------
| Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body<br>as follows:<br>Mode of Operation (*includes simultaneous B-mode) | | | | | | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---|-----|------|-----|-------------------|----------------------------|------------------------|--------------------|
| Clinical Application | A | B | M * | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Combined<br>(Specify)* | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal (See Note 3) | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-Operative<br>(See Note 7) | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | |
| Pediatric | | | | | | | | | |
| Small Organ<br>(See Note 4) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Cardiac | | | | | | | | | |
| Transesophageal | | | | | | | | | |
| Trans-Rectal | E | E | E | E | | E | E | Note 1 | Note 2, 5 |
| Trans-Vaginal | E | E | E | E | | E | E | Note 1 | Note 2, 5 |
| Trans-Urethral | | | | | | | | | |
| Intra-Vascular | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | |
| Muscular-Skeletal | | | | | | | | | |
N = new indication; P = previously cleared in K940942, Voluson C530D system; P = previously cleared in K974813, Voluson C530D Power Doppler mode; E = added under Appendix E
Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler (E)
Note 2: Includes imaging for guidance of biopsy (E)
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: 3D Imaging (E)
Note 5. 3D imaging (L)
Note 6. Tissue Handling
Note 6: Tissue Harmonic Imaging
Note 7: Abdominal organs and peripheral vessel
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR, 801.109)
Prescription Use (Per 21 CFR 801.109)
David A. Lipton
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT and Radiological Device
510(k) Number K002185
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.