SONOACE 6000C DIAGNOSTIC ULTRASOUND SYSTEM AND TRANSDUCERS, ALSO CALLED ULTRASOUND SYSTEM AND TRANSDUCERS
Applicant
Medison America, Inc.
Product Code
IYN · Radiology
Decision Date
Apr 7, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
SonoAce 6000C/UM 400C intended uses as defined FDA guidance documents are: Fetal (includes infertility monitoring of follicle development) Abdominal Intra-operative (abdominal organs and peripheral vessel, neurological) Pediatric Small Organ Neonatal Cephalic Adult Cephalic Cardiac Trans-Rectal Trans-Vaginal Peripheral-Vascular Muscular-Skeletal (conventional, superficial)
Device Story
Mobile, software-controlled diagnostic ultrasound system. Acquires ultrasound data via piezoelectric transducers (2.0-9.0 MHz) to display B-mode, M-mode, Color-Flow, Pulsed Doppler, Power Doppler, and 3D images. Digital beamforming and digital dynamic receive focusing enhance image quality. Operator uses control panel to adjust transmit focusing, filtering, and image enhancement. System calculates anatomical measurements (distances, areas, volumes) and obstetric parameters (BPD, HC, AC, FL, etc.). Cine loop stores up to 64 sequential images; image-filing manages patient history. DICOM 3.0 output supports remote viewing. Used by healthcare professionals in clinical settings for diagnostic imaging and biopsy guidance. Real-time acoustic output (mechanical/thermal indices) displayed for safety. Benefits include high-resolution visualization of internal structures and blood flow to assist clinical diagnosis.
Clinical Evidence
Bench testing only. No clinical data provided. System performance validated through acoustic output measurements (SPTA 720 mW/cm2, MI 1.9) and compliance with international electromechanical safety standards (IEC 601-1, EN 60601-1-2).
Technological Characteristics
Digital beamforming ultrasound system. Transducers use piezoelectric materials. Supports linear and convex probes (2.0-9.0 MHz). Connectivity via DICOM 3.0. Software-controlled image processing (filtering, enhancement, dynamic windowing). Acoustic output: SPTA 720 mW/cm2, MI 1.9. Safety standards: EN 60601-1, UL 2601-1, C22.2 No. 601.1, CEMEC 1157:1992, EN 60601-1-2.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including fetal, abdominal, intra-operative, pediatric, small organ, neonatal/adult cephalic, cardiac, trans-rectal, trans-vaginal, peripheral-vascular, and muscular-skeletal applications. Includes biopsy guidance and 3D imaging. Contraindications: Not for in vitro fertilization or percutaneous umbilical blood sampling.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
4/7/99
510(k) Premarket Notification
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
Submitter's name, address, telephone number, contact person: 1) Medison America, Inc. 6616 Owens Drive Pleasanton, CA 94588 Bob Leiker Vice President, Regulatory and Quality Telephone: (510) 463-1830
Prepared February 25, 1999
Name of the device, including the trade or proprietary name if applicable, the common or 2) usual name, and the classification name, if known:
> Common/Usual Name: Diagnostic Ultrasound System and Accessories
#### Proprietary Name:
SonoAce 6000C Diagnostic Ultrasound System and Transducers. Also called Ultramark® 400C Diagnostic Ultrasound System and Transducers.
| Classification Names: | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasound Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
#### 3) Identification of the predicate or legally marketed device:
Medison America, Inc. believes that SA 6000C/UM 400C Ultrasound System is substantially equivalent to the currently marketed SA 8800/HDI® 1500 and the SA 6000/UM 400 systems (K974269 and K981510).
{2}------------------------------------------------
#### Device Description: 4)
The SA 6000C/UM 400C System is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B-mode. M-mode, Color-Flow Doppler, Pulsed (PW) Doppler, Power Doppler, 3D or in a combination of these modes. M-mode uses the sweep display method which has its images flow from the left to the right on the monitor The SA 6000C/UM 400C also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The system has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
Eight different models of transducers are available and any two may be connected at the same time. In addition to the initial operational settings for each transducer preprogrammed in the system, user-customized parameter settings for each transducer may be inserted by the operator and stored for recall as needed via the system control panel. Customization includes transmit focusing, filtering, image enhancement processing, dynamic window curve selection. Controls are also provided to select display format (single and various combinations), to activate zoom features, and to utilize the cine loop function. Patient contact materials are the same as those previously cleared in the predicate device, SA 8800/HDI 1500 and SA 6000C/UM 400C ultrasound systems and transducers (K974267 and K981510).
The SA 6000C/UM 400C uses digital beamforming technology, and supports a variety of Linear and Convex probes for a wide variety of applications. It is an ultrasound scanner, which provides high resolution, high penetration performance, and various measurement functions. Probes are supported in frequencies from 2.0 MHz to 9.0 MHz.
The system can be used to measure distances and calculate areas, circumferences and volumes, as well as calculate the date of delivery by using BPD (biparietal diameter), OFD (occipito-frontal diameter), HC (head circumference), AC (abdominal circumference), AD (abdominal diameter), FL (femur length), CRL (crown rump length), APTD (anteroposterior trunk diameter), TTD (transverse trunk diameter), GS (gestational sac), LMP (last menstural period.), Cardiac Analysis and Vascular Analysis.
Biopsy guidelines are provided on screen to assist in the collection of tissue samples, using biopsy guide adapters offered as an optional accessory. The SA 6000C/UM 400C supports the Cine function (capable of storing up to 64 sequential images) and real-time zoom function to the region-of-interest. The system provides the ability to perform remote viewing of images, without
{3}------------------------------------------------
compression, via a Dicom 3.0 compatible output. Management of patient history is possible by image-filing function. High-resolution images are provided by utilizing a technology called digital dynamic receive focusing.
The SA 6000C/UM 400C has been designed to meet the following electromechanical safety standards:
- EN 60601-1 (IEC 601-1,) European Norm, Medical Electrical Equipment (
- UL 2601-1, Underwriters Laboratories Standards, Medical Electrical Equipment (
- C22.2 No. 601.1, Canadian Standards Association, Medical Electrical Equipment (
- CEMEC 1157:1992, International Electrotechnical Commission, Requirements for the ( declaration of the acoustic output of medical diagnostic ultrasonic equipment
- EN 60601-1-2 (IEC 601-1-2,) European Norm, Collateral Standard: Electromagnetic ( Compatibility
- Compliant with the European Medical Device Directive Certificate issued by TUV. (
#### Intended Use: ર)
SonoAce 6000C/UM 400C intended uses as defined FDA guidance documents are:
- Fetal (includes infertility monitoring of follicle development) (
- ( Abdominal
- ( Intra-operative (abdominal organs and peripheral vessel, neurological)
- ( Pediatric
- Small Organ (
- Neonatal Cephalic (
- Adult Cephalic (
- { Cardiac
- ( Trans-Rectal
- Trans-Vaginal (
- ( Peripheral-Vascular
- ( Muscular-Skeletal (conventional, superficial)
{4}------------------------------------------------
Typical examinations performed using the system are:
- ( General abdominal and pelvic studies including organ surveys, assessment, and retro-peritoneal cavity studies.
- Study of small parts including breasts, shoulders, thyroid, and the abdominal wall. く
- Pediatric scans of organs and bony structures. (
- く Peripheral vascular applications including carotid arteries, legs, arms, feet, and penile artery.
- Monitoring procedures for infertility studies (other than in vitro fertilization). (
- First, second and third trimester pregnancy studies. ・(
- Prostate, prostate biopsy guidance, and rectal wall studies. (
- Neonatal head studies. (
- Transcranial studies of middle cerebral arteries, internal carotid artery, and vertebral arteries. (
- ( Cardiac studies in adults and children.
- Biopsy guidance for tissue or fluid sampling. (
- Conventional podiatry scans. (
- ( Intraoperative application including soft tissue structures.
#### Technological Characteristics: 6)
This device operates identical to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as a 2D and M-mode, Spectral Doppler, Color Doppler, Power Doppler, or 3D images. Transducer patient contact materials are biocompatible.
The device's acoustic output limits are:
All Applications:
| | (Maximum Range) |
|--------|-----------------|
| I SPTA | 720 mW/cm2 |
| MI | 1.9 |
The limits are the same as predicate Track 3 devices.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its head turned to the left. The eagle is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
7 1999 APR
Medison America, Inc. c/o Carol Stamp TUV Product Services 1775 Old Highway 8 New Brighton, MN 55112-1891
K990970 Re: SonoAce 6000C/Ultramark 400C Requlatory Class: II/21 CFR 892.1550, 21 CFR 892.1560 and 21 CFR 892.1570 Product Code: 90/IYN, 90 IYO and 90 ITX Dated: March 22, 1999 Received: March 23, 1999
Dear Ms. Stamp:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the followinq transducers intended for use with the SonoAce 6000C/Ultramark 400C, as described in your premarket notification:
Transducer Model Number
HLS-9 Linear Array 7.5 MHz/40 mm L5-9/60mm Linear Array 7.5 MHz/60mm C2-4/30R Curved Linear Array 3.0MHz/30R/60deg MC2-5 Curved Linear Array 3.5MHz/40R/89deg HC3-6 Curved Linear Array 3.5MHz/60R/60deg C4-7 Curved Linear Array 6.5MHz/10R/10R/140deg C4-9 Curved Linear Array 5.0MHz/40R/60deg EC 4-9/10R Curved Linear Array 6.5MHz/10R/140deg (Endocavity)
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in
{6}------------------------------------------------
Page 2 - Carol Stamps
requlatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the provision of the investigational device exemptions (IDE) regulations, 21 Therefore, your product labeling must be consistent with FDA's CFR, Part 812. position on this use.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Other general information on your responsibilities under the Act may 807.97). be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
{7}------------------------------------------------
Page 3 - Carol Stamps
If you have any questions regarding the content of this letter, please contact II you have all quosebood at (301) 594-1212.
Sincerely yours,
Daniel h. Syym
CAPT Daniel G. Schultz, M.D.
Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{8}------------------------------------------------
# Section 4.3 INDICATION FOR USE Ultrasound Device Indications Statement
510(k) Number:
SonoAce 6000C / Ultramark 400C Ultrasound System Device Name:
Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body
as follows:
Mode of Operation (*includes simultaneous B-mode)
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Abdominal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Intra-Operative<br>(See Note 5) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Intra-Operative<br>Neurological | | N | N | N | | N | N | | Note 1 | Note 6 |
| Pediatric | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Small Organ<br>(See Note 4) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Neonatal Cephalic | | P | P | N | | N | N | | Note 1 | Note 6 |
| Adult Cephalic | | N | N | N | | N | N | | Note 1 | Note 6 |
| Cardiac | | P | P | N | | N | N | | Note 1 | Note 6 |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Trans-Vaginal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | P | P | N | | N | N | | Note 1 | Note 6 |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Muscular-Skeletal<br>Superficial | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Others(Specify) | | | | | | | | | | |
N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes:
Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: Abdominal organs and peripheral vessel
Note 6: 3D Imaging
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
.
David he. Sezman
(Division Sign-Off)
Division of Reproductive, Abdominal and Radiological
510(k) Number K990970
{9}------------------------------------------------
| 510(k) Number: | |
|----------------------|------------------------------------------------------------------------------------|
| Device Name: | SonoAce 6000C / Ultramark 400C Ultrasound System |
| Transducer: | HL5-9, 7.5MHz/40mm, Linear Array Probe |
| Indications for Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |
| | Mode of Operation (*includes simultaneous B-mode) | |
|--|---------------------------------------------------|--|
|--|---------------------------------------------------|--|
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-Operative<br>(See Note 5) | | N | N | N | . | N | N | | Note 1 | Notes 2, 6 |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Small Organ<br>(See Note 4) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Neonatal Cephalic | | P | P | N | | N | N | | Note 1 | Note 6 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | P | P | N | | N | N | | Note 1 | Note 6 |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Muscular-Skeletal<br>Superficial | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Others(Specify) | | | | | | | | | | |
N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes:
Note 1: PWD/Color Doppler, PWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: Abdominal organs and peripheral vessel
Note 6: 3D Imaging
Concurrence of CDRH, Office of Device Evaluation (ODE)
L Prescription Use (Per 21 CFR 801.109)
Daniel h. Sesom
(Division Sign-Off) Reproductive, Abdominal, E and Radiological
510(k) Number K990970
{10}------------------------------------------------
510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: L5-9/60mm, 7.5MHz/60mm, Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | N | N | N | N | | N | N | | Note 1 | Note 6 |
| Abdominal | N | N | N | N | | N | N | | Note 1 | Note 6 |
| Intra-Operative<br>(See Note 5) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | N | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Small Organ<br>(See Note 4) | P | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Neonatal Cephalic | P | P | P | N | | N | N | | Note 1 | Note 6 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | P | P | P | N | | N | N | | Note 1 | Note 6 |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | N | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Muscular-Skeletal<br>Superficial | N | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Others(Specify) | | | | | | | | | | |
Mode of Operation (*includes simultaneous B-mode)
N = new indication: P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes:
Note 1: PWD/Color Doppler, PWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: Abdominal organs and peripheral vessel
Note 6: 3D Imaging
**Concurrence of CDRH, Office of Device Evaluation (ODE)**
Prescription Use (Per 21 CFR 801.109)
David h. Segner
(Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological D
510(k) Number K990970
{11}------------------------------------------------
510(k) Number: Device Name: SonoAce 6000C / Ultramark 400C Ultrasound System C2-4/30R, 3.0MHz/30R/60D, Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation (*includes simultaneous B-mode) |
|--|---------------------------------------------------|
|--|---------------------------------------------------|
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | P | P | N | | N | N | | Note 1 | Note 6 |
| Abdominal | | P | P | N | | N | N | | Note 1 | Note 6 |
| Intra-Operative | | | | | | | | | | |
| (See Note 5) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | N | N | N | | N | N | | Note 1 | Note 6 |
| Pediatric | | | | | | | | | | |
| Small Organ | | | | | | | | | | |
| (See Note 4) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | N | N | N | | N | N | | Note 1 | Note 6 |
| Cardiac | | P | P | N | | N | N | | Note 1 | Note 6 |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others(Specify) | | | | | | | | | | |
N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes:
Note 1: PWD/Color Doppler, PWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: Abdominal organs and peripheral vessel
Note 6: 3D Imaging
Concurrence of CDRH, Office of Device Evaluation (ODE)
[ Prescription Use (Per 21 CFR 801.109)
David C. Stagner
(Division Sign-Off)
Division of Reproductive, Abdominal, El and Radiological Devi
510(k) Number K990970
{12}------------------------------------------------
510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: MC2-5, 3.5MHz/40R/89D, Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | P | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Abdominal | P | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Intra-Operative<br>(See Note 6) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | P | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Small Organ<br>(See Note 5) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others(Specify) | | | | | | | | | | |
#### Mode of Operation (*includes simultaneous B-mode)
N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes:
Note 1: PWD/Color Doppler, PWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: Abdominal organs and peripheral vessel
Note 6: 3D Imaging
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
Daniel L. Segmen
productive, Abdominal, and Rac
{13}------------------------------------------------
510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: HC3-6, 3.5MHz/60R/60D, Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Abdominal | | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Intra-Operative<br>(See Note 5) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Small Organ<br>(See Note 4) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others(Specify) | | | | | | | | | | |
Mode of Operation (*includes simultaneous B-mode)
N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes:
Note 1: PWD/Color Doppler, PWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: Abdominal organs and peripheral vessel
Note 6: 3D Imaging
Concurrence of CDRH, Office of Device Evaluation (ODE)
| / Prescription Use (Per 21 CFR 801.109)
David A. Syum
Reproducti and Radiological
{14}------------------------------------------------
510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: C4-7, 5.0MHz/40R/60D, Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Mode of Operation (*includes simultaneous B-mode)
| Clinical Application | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Abdominal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Intra-Operative<br>(See Note 5) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Small Organ<br>(See Note 4) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Muscular-Skeletal<br>Superficial | | P | P | | | | | | | |
| Others(Specify) | | | | | | | | | | |
N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes:
Note 1: PWD/Color Doppler, PWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: Abdominal organs and peripheral vessel
Note 6: 3D Imaging
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
David le. Seyton
Division Sign-Off) Division of Reproductive, Abdominal, E. and Radiological De
510(k) Number K9902
{15}------------------------------------------------
| 510(k) Number: | |
|----------------------|------------------------------------------------------------------------------------|
| Device Name: | SonoAce 6000C / Ultramark 400C Ultrasound System |
| Transducer: | C4-9/10R, 6.5MHz/10R/140D, Curved Linear Array Probe |
| Indications for Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |
| | Mode of Operation (*includes simultaneous B-mode) |
The fulles A B M* PWD* CWD* Color . Power Color Combined Clinical Application
| Clinical Application | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Abdominal | | | | | | | | | | |
| Intra-Operative<br>(See Note 5) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(See Note 4) | | P | P | N | | N | N | | Note 1 | Note 6 |
| Neonatal Cephalic | | P | P | N | | N | N | | Note 1 | Note 6 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Trans-Vaginal | | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | P | P | N | | N | N | | Note 1 | Note 6 |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others(Specify) | | | | | | | | | | |
N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: Abdominal organs and peripheral vessel
Note 6: 3D Imaging
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
David R. Segerson
---
(Division Sign Off)
(Division Sign-Off) Division of Reproductive, Abdominal, E. and Radiological Devices
Radiological Devices
510(k) Number K990970
{16}------------------------------------------------
510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: EC 4-9/10R, 6.5MHz/10R/140D (Endocavity), Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation (*includes simultaneous B-mode)
| Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal (See Note 3) | | N | N | N | | N | N | | Note 1 | Notes 2, 6 |
| Abdominal | | | | | | | | | | |
| Intra-Operative<br>(See Note 5) | | | | | | | | | | |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(See Note 4) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Trans-Vaginal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others(Specify) | | | | | | | | | | |
N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes
Note 1: PWD/Color Doppler, PWD/Power Doppler
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 5: Abdominal organs and peripheral vessel
Note 6: 3D Imaging
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Toril A. Sogn
Reproductive, Abdominal, olnoical D and Rad 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.