K992470 · Advanced Technology Laboratories, Inc. · IYO · Aug 23, 1999 · Radiology
Device Facts
Record ID
K992470
Device Name
ULTRAMARK 400C/SONOACE 6000 ULTRASOUND SYSTEM
Applicant
Advanced Technology Laboratories, Inc.
Product Code
IYO · Radiology
Decision Date
Aug 23, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal (includes infertility monitoring of follicle development) - Abdominal - Intra-operative (abdominal organs and peripheral vessel, neurological) - Pediatric Small Organ - Neonatal Cephalic . Adult Cephalic - Cardiac - Trans-Rectal - Trans-Vaginal - Peripheral-Vascular - Muscular-Skeletal (conventional, superficial) -
Device Story
Mobile, software-controlled diagnostic ultrasound system; acquires ultrasound data via piezoelectric transducers (2.0-9.0 MHz). Transforms reflected sound waves into B-mode, M-mode, Color-Flow/Power Doppler, Pulsed Doppler, 3D, and Tissue Harmonic Imaging (THI) outputs. Operated by healthcare professionals in clinical settings. Features digital beamforming, digital dynamic receive focusing, and cine loop (64 images). Provides anatomical measurements (distances, areas, volumes) and obstetric/cardiac/vascular analysis packages. Displays real-time mechanical and thermal indices. Supports biopsy guidance via optional adapters. Output viewed on monitor; aids clinical diagnosis and tissue sampling. DICOM 3.0 compatible for remote viewing.
Clinical Evidence
Bench testing only. Acoustic output measurements provided to demonstrate compliance with FDA Track 3 limits. No clinical data presented.
Technological Characteristics
Piezoelectric transducer technology; digital beamforming; digital dynamic receive focusing. Supports 2.0-9.0 MHz probes. Connectivity: DICOM 3.0. Safety standards: UL 2601-1, CSA C22.2 No. 601.1, EN 60601-1-2. Acoustic output: MI 1.9, ISPTA 720 mW/cm2. Software-controlled.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, intra-operative, pediatric, small organ, neonatal/adult cephalic, cardiac, trans-rectal, trans-vaginal, peripheral-vascular, and musculoskeletal applications. Includes biopsy guidance and infertility monitoring. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
UM 400C/ SA 6000C Ultrasound System
K992470
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
1) Submitter's name, address, telephone number, contact person: Advanced Technology Laboratories, Inc. P.O. Box 3003 Bothell, WA 98031-3003 Vice President, Worldwide Quality & Regulatory Affairs Telephone: (425) 487-7602
Prepared: August 23, 1999
Name of the device, including the trade or proprietary name if applicable, the common or 2) usual name, and the classification name, if known:
> Common/Usual Name: Diagnostic Ultrasound System and Accessories
Proprietary Name:
Ultramark® 400C Diagnostic Ultrasound System and Transducers.
Also called: SonoAce 6000C Diagnostic Ultrasound System and Transducers.
| Classification Names: | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasound Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
#### Identification of the predicate or legally marketed device: 3)
Advanced Technology Laboratories, Inc. believes that the UM 400C/ SA 6000C Ultrasound System is substantially equivalent to the currently marketed SA 8800/HDI® 1500 system (K974269).
{2}------------------------------------------------
510(k) Premarket Notification Tissue Harmonic Imaging
## UM 400C/ SA 6000C Ultrasound System
#### Device Description: 4)
The UM 400C/SA 6000C system is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B-mode, M-mode, Color-Flow Doppler, Pulsed (PW) Doppler, Power Doppler, 3D, Tissue Harmonic Imaging or in a combination of these modes. M-mode uses the sweep display method which has its images flow from the left to the right on the monitor The UM 400C/SA 6000C also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The system has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
Nine different models of transducers are available and any two may be connected at the same time. In addition to the initial operational settings for each transducer preprogrammed in the system, user-customized parameter settings for each transducer may be inserted by the operator and stored for recall as needed via the system control panel. Customization includes transmit focusing, filtering, image enhancement processing, dynamic window curve selection. Controls are also provided to select display format (single and various combinations), to activate zoom features, and to utilize the cine loop function.
The UM 400C/SA 6000C system uses digital beamforming technology, and supports a variety of Linear and Convex probes for a wide variety of applications. It is a diagnostic ultrasound scanner, which provides high resolution, high penetration performance, and various measurement functions. Probes are supported in frequencies from 2.0 MHz to 9.0 MHz.
The system can be used to measure distances and calculate areas, circumferences and volumes, as well as calculate the date of delivery by using BPD (biparietal diameter), OFD (occipitofrontal diameter), HC (head circumference), AC (abdominal circumference), AD (abdominal diameter), FL (femur length), CRL (crown rump length), APTD (anteroposterior trunk diameter), TTD (transverse trunk diameter), GS (gestational sac), LMP (last menstural period.), Cardiac Analysis and Vascular Analysis.
Biopsy guidelines are provided on screen to assist in the collection of tissue samples, using biopsy guide adapters offered as an optional accessory. The UM 400C/SA 6000C supports the Cine function (capable of storing up to 64 sequential images) and real-time zoom function to the region-of-interest. The system provides the ability to perform remote viewing of images, without compression, via a Dicom 3.0 compatible output. Management of patient history is possible by
{3}------------------------------------------------
image-filing function. High-resolution images are provided by utilizing a technology called
digital dynamic receive focusing.
The UM 400C/SA 6000C has been designed to meet the following electromechanical safety standards:
- -UL 2601-1, Underwriters Laboratories Standards, Medical Electrical Equipment
- -C22.2 No. 601.1, Canadian Standards Association, Medical Electrical Equipment
- declaration of the acoustic output of medical diagnostic ultrasonic equipment
- EN 60601-1-2 (IEC 601-1-2,) European Norm, Collateral Standard: Electromagnetic -Compatibility
#### Intended Use: 5)
Ultramark 400C/ SonoAce 6000C intended uses as defined FDA guidance documents are:
- Fetal (includes infertility monitoring of follicle development) -
- Abdominal -
- Intra-operative (abdominal organs and peripheral vessel, neurological) -
- -Pediatric
- Small Organ -
- Neonatal Cephalic .
- Adult Cephalic -
- Cardiac -
- Trans-Rectal -
- Trans-Vaginal -
- Peripheral-Vascular -
- Muscular-Skeletal (conventional, superficial) -
Typical examinations performed using the system are:
- General abdominal and pelvic studies including organ surveys, assessment, and retro-peritoneal cavity studies.
- Study of small parts including breasts, shoulders, thyroid, and the abdominal wall. .
- Pediatric scans of organs and bony structures. -
- -Peripheral vascular applications including carotid arteries, legs, arms, feet, and penile artery.
- -Monitoring procedures for infertility studies (other than in vitro fertilization).
- -First, second and third trimester pregnancy studies.
- Prostate, prostate biopsy guidance, and rectal wall studies. -
- -Neonatal head studies.
- Transcranial studies of middle cerebral arteries, internal carotid artery, and vertebral arteries. -
- -Cardiac studies in adults and children.
- Biopsy guidance for tissue or fluid sampling. -
- Conventional podiatry scans. -
- Intraoperative application including soft tissue structures. .
{4}------------------------------------------------
K992470 p. 34,4
UM 400C/ SA 6000C Ultrasound System
#### 6) Technological Characteristics:
This device operates identical to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as a 2D and M-mode, Spectral Doppler, Color Doppler, Power Doppler, 3D, or Tissue Harmonic Imaging images. Transducer patient contact materials are biocompatible.
The device's acoustic output limits are:
All Applications:
| TIS/TIB/TIC | 0.0 - 5.0 | (Range) |
|-------------|------------|-----------|
| ISPTA | 720 mW/cm2 | (Maximum) |
| MI | 1.9 | (Maximum) |
The limits are the same as predicate Track 3 devices.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines above it, representing the department's mission to protect and promote the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is simple and recognizable, conveying the department's authority and purpose.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# AUG 7 3 1999
Mr. Terrence J. Sweeney
Vice President, Worldwide Quality and Regulatory Affairs
ATL Ultrasound
Post Office Box 3003
· Bothel, Washington 98041-3303
K992470 Re:
> Trade Name: Ultramark® 400C/Sonoace 6000C Diagnostic Ultrasound System with Tissue Harmonic Imaging (THI) Regulatory Class: II 21 CFR §892.1560/Procode: 90 IYO Dated: July 23, 1999 Received: July 26, 1999
Dear Mr. Sweeney:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducer intended for use with the Ultramark® 400C/Sonoace 6000C Diagnostic Ultrasound System with Tissue Harmonic Imaging (THI), as described in your premarket notification:
## Transducer Model Number
## C3-7 Curved Linear Array 4.5 MHz/60R/60D/128 elements
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing
{6}------------------------------------------------
### Page 2 - Mr. Terrence J. Sweeney
Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address: "http://www.fda.gov/cdrh/dsmamain.html".
{7}------------------------------------------------
# Page 3 - Mr. Terrence J. Sweeney
l :
If you have any questions regarding the content of this letter, please contact Paul M. Gammell, Ph.D. at (301) 594-1212.
Sincerely yours,
CAPT Daniel G. Schultz, M.D.
Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{8}------------------------------------------------
## 4.3 INDICATIONS FOR USE
### STIC ULTRASOUND INDICATIONS STATEMENT DIA
510(k) Number: Ultramark® 400C/SonoAce 6000C Ultrasound System
System: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|------------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | P | P | P | P | P | Note 1 | Notes 2, 5, 6 |
| | Abdominal | P | P | P | P | P | Note 1 | Notes 2, 5, 6 |
| | Intra-operative (Abdominal,<br>vascular) | P | P | P | P | P | Note 1 | Notes 2, 5 |
| | Intra-operative (Neuro.) | P | P | P | | P | Note 1 | Note 5 |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | Note 1 | Notes 2, 5, 6 |
| | Small Organ (See Note 4) | P | P | P | P | P | Note 1 | Notes 2, 5, 6 |
| | Neonatal Cephalic | P | P | P | | P | Note 1 | Note 5 |
| | Adult Cephalic | P | P | P | | P | Note 1 | Note 5 |
| | Trans-rectal | P | P | P | P | P | Note 1 | Notes 2, 5 |
| | Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2, 5 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Notes 2, 5 |
| | Musculo-skel. (Superfic.) | P | P | P | | P | Note 1 | Notes 2, 5 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | Note 1 | Note 5 |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 5 |
| | Trans-esophageal (card) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Note 1 | Note 5 |
| | Other (spec.) | | | | | | | |
N=new indication; P= previously cleared by FDA in K981510 & K990970; E= added under Appendix E Additional Comments:
Color Doppler includes Color Amplitude Doppler (P)
Note 1: PWD/Color Doppler, PWD/Power Doppler (P)
Note 2: Includes imaging for guidance of biopsy (P)
Note 3: Includes infertility monitoring of follicle development (P)
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P)
Note 5: 3D Imaging (P)
Note 6: Tissue Harmonic Imaging (N)
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
Vinil h. Kym
Division Sign-Off) Division of Reproductive, Abdominal, E and Radiological I 510(k) Number
{9}------------------------------------------------
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
:
K992470
510(k) Number:
System:
Ultramark® 400C/ SonoAce 6000C Ultrasound System
C3-7 Curved Linear Array 4.5 MHz/60R/60D/128 elements Scanhead:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|------------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | N | N | N | | N | Note 1 | Notes 2, 5, |
| | Abdominal | N | N | N | | N | Note 1 | Notes 2, 5, |
| | Intra-operative (Abdominal,<br>vascular) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Notes 2, 5, |
| | Small Organ (See Note 4) | N | N | N | | N | Note 1 | Notes 2, 5, |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Color Doppler includes Color Amplitude Doppler
Note 1: PWD/Color Doppler, PWD/Power Doppler (N)
Note 2: Includes imaging for guidance of biopsy (N)
Note 3: Includes infertility monitoring of follicle development (N)
Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (N)
Note 5: 3D Imaging (N)
Note 6: Tissue Harmonic Imaging (N)
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
Chid A. Segmen
(Division Sign-Off)
Division of Reproductive, Abdominal, EN and Radiological Devi
510(k) Number K992470
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.