K023869 · American Bio Medica Corp. · DJR · Apr 30, 2003 · Clinical Toxicology
Device Facts
Record ID
K023869
Device Name
'RAPIDTEC'-5M-MULTIPLE DIP TEST
Applicant
American Bio Medica Corp.
Product Code
DJR · Clinical Toxicology
Decision Date
Apr 30, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3620
Device Class
Class 2
Indications for Use
'RapidTec'-5M-Multiple Drug Test is used for the qualitative detection of the following abused substances in human urine: methamphetamines, benzoyl ecgonine, phencyclidine, opiates and cannabinoids. This immunoassay is a simplified screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e. gas chromatography/mass spectrometry (GC/MS.)
Device Story
RapidTec-5M-Multiple Dip Test is a lateral flow, one-step competitive immunoassay for qualitative detection of drugs of abuse in human urine. Device consists of a membrane strip with immobilized drug conjugates and dried colloidal gold-antibody complexes. In absence of drug, gold-antibody complex binds to immobilized conjugates, forming visible test lines. In presence of drug, drug competes for antibody sites, preventing line formation; absence of line indicates positive result. A control line confirms proper flow. Used in professional settings for preliminary screening; results require confirmation via GC/MS. Provides rapid, visual qualitative assessment to guide clinical decision-making regarding further testing.
Clinical Evidence
Bench testing only. Reproducibility evaluated using control urines with concentrations above and below stated cut-offs. All concentrations verified by GC/MS. Samples tested four times, twice daily, for five days. Results confirmed reproducibility of performance.
Technological Characteristics
Lateral flow immunoassay; membrane strip with immobilized drug conjugates and colloidal gold-antibody complex. Detects methamphetamines (1000 ng/ml), benzoyl ecgonine (300 ng/ml), cannabinoids (50 ng/ml), phencyclidine (25 ng/ml), and opiates (2000 ng/ml). Manual dip-test format. No electronic components or software.
Indications for Use
Indicated for professional use for the simultaneous qualitative detection of methamphetamines, benzoyl ecgonine, cannabinoids, phencyclidine, and opiates in human urine. Provides preliminary results requiring confirmatory testing (e.g., GC/MS). Not for over-the-counter sale.
Regulatory Classification
Identification
A methadone test system is a device intended to measure methadone, an addictive narcotic pain-relieving drug, in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of methadone use or overdose and to determine compliance with regulations in methadone maintenance treatment.
Special Controls
*Classification.* Class II (special controls). A methadone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
K021114 — 'RAPIDTEC' 5A MULTIPLE DIP TEST · American Bio Medica Corp. · Jul 23, 2002
K153192 — Chemtrue Multi-Panel Drug Screen Dip Card/Cup Tests, Chemtrue Multi-Panel Drug Screen Dip Card/Cup with OPI 2000 Tests · Chemtron Biotech, Inc. · Apr 18, 2016
K111322 — CHEMTRUE SINGLE/MULTI-PANEL DRUG SCREEN CASSETTE/DIPCARD · Chemtron Biotech, Inc. · Sep 14, 2011
K993151 — RADID DRUG SCREEN 5-PANEL TEST · American Bio Medica Corp. · Jan 10, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
# APR 3 0 2003
## 510 (k) Summary
#### Submitter's Name/Address: American Bio Medica Corportion 122 Smith Road
Kinderhook, NY 12106
### Contact Person:
Henry Wells VP Product Development Phone: 410 992-4734 Fax: 410 992-0328
Date of Preparation of this Summary:
Device Trade or Proprietary Name:
Device Common/Usual Name or Classification Name:
November 14, 2002
'RapidTec'-5M-Multiple Dip Test
Multi Drug Test System
Classification Number/Class:
[no classification regulation]/Class II
This 510(k) Summary is being submitted in accordance with the requirement of 21 CFR 807.92.
The assigned 510(k) number is:
Predicate Device: American Bio Medica Corp. 'Rapid Drug Screen'-9-Panel. (510(k) No. K012159.)
#### Test Description:
The assays employed in the 'RapidTec'-5M-Multiple Dip Test is based on the same principle of highly specific reactions between antigens and antibodies.
This assay is a one-step, competitive, immunoassay for the detection of marijuana, opiates, phencyclidine, cocaine and methamphetamine in human urine. The test device consists of a membrane strip onto which drug conjugates have been immobilized and a colloidal gold-multi-antibody complex dried at one end of the membrane. In the absence of any drug in the urine sample, the colloidal gold-antibody complex moves with the urine by capillary action to contact the immobilized drug conjugates. Antibody-antigen reactions occur forming visible lines in the 'test' area.
When drug is present in the urine sample, the drug or metabolite will compete with its corresponding drug conjugate in the test area for the limited antibody sites on the colloidal gold-labeled antibody complex. If sufficient amount of drug is present, it will fill all of the available antibody binding sites, thus preventing attachment of the labeled antibody to the drug conjugate. An absence of a color band (line) in the 'test' area is indicative of a positive result.
{1}------------------------------------------------
A control band (line), comprised of a different antibody/antigen reaction, is present on the membrane strip. The control line is not influenced by the presence or absence of drug in the urine, and therefore, should be present on all reactions.
A negative urine will produce six colored bands, and a positive sample will produce only one band.
### Intended use:
'RapidTec'-5M-Multiple Drug Test is used for the qualitative detection of the following abused substances in human urine: methamphetamines, benzoyl ecgonine, phencyclidine, opiates and cannabinoids. This immunoassay is a simplified screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e. gas chromatography/mass spectrometry (GC/MS.)
#### Performance Characteristics:
'RapidTec' - 5M-Multiple Dip Test will detect drugs of abuse in human urine at the following levels:
| Methamphetamine | 1000 ng/ml |
|------------------|------------|
| Benzoyl ecgonine | 300 ng/ml |
| Cannabinoids | 50 ng/ml |
| Phencyclidine | 25 ng/ml |
| Opiates | 2000 ng/ml |
| | 300 ng/ml |
Reproducibility was evaluated using control urines containing concentrations above and below the stated cut-off. Negative controls were also used. All concentrations were verified by GC/MS. Each sample was tested four times, twice daily, for five days. The results confirmed the reproducibility of the 'RapidTec' -5M-Multiple Dip Test performance.
#### Conclusion:
'RapidTec'-5M-Multiple Dip Test is substantially equivalent to the previously cleared 'Rapid Drug screen' - 9-Panel (510(k) No. K012159). Both systems utilize the same antibodies. The only difference is that in 'Rapid Drug Screen'-9-Panel, each assay occupies a separate channel, while in 'RapidTec'-5M-Multiple Dip Test each independent assay is 'stacked' in separate lines, one above the other.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight. The symbol is made up of three curved lines that converge at the bottom.
APR 3 0 2003
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Henry Wells, Ph.D. V.P. Product Development American Bio Medica Corporation 9110 Red Branch Road Columbia, MD 21045
k023869 Re:
Trade/Device Name: 'RapidTec'-5M-Multiple Dip Test Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine Test System Regulatory Class: Class II Product Code: DJR, JXM, LDJ, DIG, LCM Dated: February 24, 2003 Received: February 25, 2003
Dear Dr. Wells:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 –
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page ನ
K023869 SIO(K) NUMBER (IF KNOWN) :
DEVICE NAME : 'RapidTec'-5M-Multiple Dip Test
INDICATIONS FOR USE:
'RapidTec'-5M-Multiple Dip Test is a onc-step lateral flow immunoassay for the simultaneous qualitative detection of methamphetamines, benzoyl ecgonine, cannabinoids, phencyclidine, and opiates in urine.
'RapidTec'-5Mmultiple Dip Test will dctect these analytes at the following concentrations:
| Methamphetamines: | 1000 ng/ml |
|-------------------|------------|
| Benzoyl ecgonine | 300 ng/ml |
| Cannabinoids | 50 ng/ml |
| Phencylidine | 25 ng/ml |
| Opiates | 2000 ng/ml |
'Rapid Tec'-5M-Multiple Dip Test is intended for professional use. It is not intended for over-the-counter sale to nonprofessionals. The ussay is casy to perform, but should not be used without proper supervision. This immunoassay is a simplified, qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e. gas chromatography/mass spectrometry (GC/MS).
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
. シ
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--------------------------------------------------|
| | Han Cooper |
| | (Division Sign-Off) |
| | Division of Clinical Laboratory Devices |
| 510(k) Number | K023869 |
| Prescription Use<br>(Per 21 CFR 801.109) | |
| | OR |
| | Over-The-Counter-Use<br>(Optional Format 1-2-96) |
{5}------------------------------------------------
'RapidTec'-5M-Multiple Dip Test provides only a preliminary analytical result. A more specific alternate chemical method must be used in order to obtain. A more confirmed result. GC/MS is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.