K012159 · American Bio Medica Corp. · DKZ · Jul 25, 2001 · Clinical Toxicology
Device Facts
Record ID
K012159
Device Name
RAPID DRUG SCREEN
Applicant
American Bio Medica Corp.
Product Code
DKZ · Clinical Toxicology
Decision Date
Jul 25, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.3100
Device Class
Class 2
Indications for Use
The ABM Rapid Drug Screen™ is for the determination of Cocaine, Tricyclics, Methamphetamine and Opiates, Marijuana, PCP, Barbiturates, Benzodiazepines, Qualitative results and is intended for professional use only.
Device Story
Rapid Drug Screen™ is an in vitro diagnostic device for qualitative detection of drugs of abuse in human urine. Device utilizes lateral flow immunoassay technology; urine sample applied to test device; specific drug analytes bind to labeled antibodies; visual color bands indicate presence or absence of drugs at specific cutoff concentrations. Intended for professional use in clinical settings. Healthcare providers interpret visual results to assist in clinical decision-making regarding substance use screening. Benefits include rapid, point-of-care identification of drug presence.
Clinical Evidence
No clinical data provided; device relies on bench testing for analytical performance validation.
Indicated for the qualitative determination of Cocaine, Tricyclics, Methamphetamine, Opiates, Marijuana, PCP, Barbiturates, and Benzodiazepines in human urine. Intended for professional use only.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circle's perimeter.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
American Bio Medica Corporation c/o Ms. Fran White Regulatory Consultant MDC Associates 163 Cabot Street Beverly, MA 01915
510(k) Number: K012159 Re: Trade/Device Name: Rapid Drug Screen™ Trade/Device Name: Rapid Drug Screen
Regulation Number: 862.3100, 862.3250, 862.3650, 862.3170, 862.3170,
Regulation Number: 862.3100, 862.3210 862.3610, 862.3910
JUL 2 5 2001
Regulatory Class: II Regulatory Class: 11
Product Code: DKZ, DIO, LDJ, DJG, DIS, JXM, DJC, LCM, LFI Dated: July 13, 2001 Received: July 17, 2001
Dear Ms. White:
We have reviewed your Section 510(k) notification of intent to market the devices merketed in We have reviewed your Section 510(x) notificantially equivalent to devices marketed in
above and we have determined the device is substantially of the Modicel Davice above and we nave determined the device is the enactment date of the Medical Device interstate commerce pror to May 20, 1978, the encordance with the provisions of the dayine, subject to Amendments, of to devices man have been receifice, market the device, subject to
Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, of the Ast include Federal Food, Drug, and Cosmone Free (110).
the general controls provisions of the Act. The general controls provisions proctine proctine argentiae, lobaling the general controls provisions of the Act. "The general vood manufacturing practice, labeling, requirembitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III) If your device is classified (sec above) and cuch additional controls. Existing major regulations
(Premarket Approval), it may be subject to such additions, Title 21, Parts (Premarket Approval), it may be subject to suen actively Regulations, Title 21, Parts 800 to 895.
affecting your device can be found in the Code of Federal Regulations, Title affecting your device can be found in the Ossumes compliance with the Cood Manufacturing A substantially equivalent actemination assummation (21 CFR Part 820) and that, through Practice Iol Medical Devices: "Sentral (Children) (FDA) will verify such periodic GMP Inspections, the I ood and Drog Namestion may result in regulatory action. In
assumptions. Failure to comply with the GMP regulation may result wise in the Fed assumptions. Fallure to comply with the GMT regulation your device in the Federal addition, FDA may publish further announcemarket notification submission does not affect
Register. Please note: this response to your premarket notification and the Register. Please note: this response to your printered 542 of the Act for devices under the any obligation you might have under becauses or other Federal laws or regulations.
{2}------------------------------------------------
## Page 2
This letter will allow you to begin marketing your device as described in your 510(k) prematket This letter will allow you to begin marketing your device to a legally marketed
notification. The FDA finding of substantial equivalee of your device to notification. The FDA finding of substantial equivalence of your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your devices), please contact the Office of Compliance at
additionally 809.10 for in vitto diagnostic devices), please contact the Office of additionally 809.10 for in vitto dagnostic nevestions on the promotion and advice, and device, and (301) 594-4588. Additionally, for questions on une promote the regulation
please contact the Office of Compliance at (311) 594-4639. Also, please note the regulation please contact the Office of Compliance at (301) - 1037 - 1201 (FR 807.97). Other general
entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Oth entitled, "Misbranding by reference of prematics (professor the Division of Small
information on your responsibilities under the Act may be obtained from the Division of Smal information on your responsibilities under (800) 638-2041 or (301) 443-6597 or at its
Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at Manufacturers Assistance at its ton tres collections and mamain.html".
internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Office of Dovices and Radiological Health
Enclosure
{3}------------------------------------------------
## Special 510k Rapid Drug Screen Original 510 (k) number: K002447 American BioMedica Corporation
Page 9 of 14
## 510(k) Number: Kora 159 Device Name: Rapid Drug Screen™
Indication for Use: The ABM Rapid Drug Screen™ is for the determination of Cocaine,
Triavelier, The ABM Rapid Drug Screen Trievelies, Mothemphetamine and Indication for Use: The ADM Rapid Driven Tricyclics, Methampletamine and Opiates, Marijuana, PCF, Barontales, Donzoulasspance, Barry, Qualitative results and is intended for professional use only.
Fred Lacy
(Division Sign-Off) (Division Sigil-On)
Division of Clinical Laboratory Deviews K012159 510(k) Number .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
..............................................................................................................................................................................
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over The Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.