The At Home Drug Test is a rapid, qualitative immunoassay for the detection of the target drugs/drug metabolites in urine. The cut-off concentration for this test is as follows: THC (or marijuana): 50 ng/ml. This assay is intended for use in the home to assist in the prevention of drug abuse. This kit is designed to incorporate a mechanism for anonymous confirmation testing to be performed at a certified laboratory.
Device Story
At Home Drug Test (Model 9078) is a lateral flow immunoassay for qualitative detection of THC metabolites in urine; cut-off 50 ng/ml. Device utilizes visual color sandwich immunochemical assay principle; specific antibody-drug complexes form on test strip. Intended for home use by lay individuals for preliminary screening; kit includes mechanism for anonymous laboratory confirmation. Output is visual color change on test strip indicating presence or absence of drug. Results assist users in drug abuse prevention. No electronic processing or software involved.
Clinical Evidence
Clinical sample correlation study and blind labeled spiked study performed. Correlation >98% compared to Behring EMIT II and GC/MS. Professional user accuracy >97%. Consumer study (n=4715) showed 96.65% overall accuracy; THC-specific consumer interpretation accuracy 97.0% (915/943).
Technological Characteristics
Lateral flow immunoassay; visual color sandwich immunochemical principle. Standalone, non-electronic, single-use test kit. No software or energy source required.
Indications for Use
Indicated for home use by lay users for the preliminary qualitative detection of THC (marijuana) metabolites in urine at a 50 ng/ml cut-off concentration to assist in drug abuse prevention.
Regulatory Classification
Identification
A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.
Special Controls
*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
QuickScreen At Home Drug Test
Applied Biotech SureStep Test
Reference Devices
Behring EMIT II
Related Devices
K013408 — AT HOME DRUG TEST, MODEL 9079 · Phamatech, Inc. · Dec 6, 2001
K993663 — FIRST CHECK HOME DRUG TEST (THC) · Worldwide Medical Corp. · Jun 23, 2000
K991641 — AT HOME COCAINE & THC SCREENING TEST, MODEL 9170 · Phamatech · Jun 1, 1999
K992217 — AT HOME DRUG TEST, MODEL 9175 · Phamatech · Jul 13, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
K994220
## 510 (k) SUMMARY AS REQUIRED BY SECTION 807.92(C)
At Home Drug Test (Model 9078) Identification:
Description: Immunoassay for the qualitative detection of THC in urine
Name Of Manufacturer:
Phamatech 9265 Activity Road #112 San Diego, California 92126, USA
Intended Use: The At Home Drug Test is a rapid, qualitative immunoassay for the detection of the target drugs/drug metabolites in urine. The cut-off concentration for this test is as follows: THC (or marijuana): 50 ng/ml. This assay is intended for use in the home to assist in the prevention of drug abuse. This kit is designed to incorporate a mechanism for anonymous confirmation testing to be performed at a certified laboratory.
The At Home Drug Test, like many commercially available drug screening test Technology: kits, qualitatively measures the presence of target drugs or their metabolites by visual color sandwich one step immunoassay technology. Examples of such predicate devices include the QuickScreen At Home Drug Test and the Applied Biotech SureStep Test (San Diego, CA 92121). All of the above devices rely on the basic immunochemical sandwich assay principle of recognition and formation of specific antibody / target drug / antibody / complexes.
Performance: The product performance characteristics of the At Home Drug Test were evaluated in a clinical sample correlation study and a blind labeled spiked study. The results of these studies demonstrate the Phamatech At Home Drug Test to be substantially equivalent to the reported performance characteristics of other commercially available tests for the qualitative detection of the stated target drugs in urine. Correlations studies, using clinical specimens, produced a >98% correlation when compared to the Behring EMIT II (Cupertino, CA 95014) and GC/MS methodology. Clinical studies, performed at two independent laboratories, were also performed. In them the Phamatech QuickScreen™ exhibited excellent overall accuracy (>97%) in the hands of professional users. A comsumer study was also performed. The At Home Drug Test exhibited excellent overall accuracy (4557/4715 correct or 96.65%) in the hands of lay users. Consumer interpretation of the THC test in this study showed accuracy to be 915/943 or 97.0%.
Conclusion: For the reasons mentioned above, it may be concluded that the Phamatech At Home Drug Test Model 9078 is substantially equivalent to a variety of detection tests currently in commercial distribution and is safe in the hands of the lay user.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 3 0 1999
Mr. Carl A. Mongiovi Vice President of Operations Phamatech 9265 Activity Road Suite #112 San Diego, California 92126
Re: K994220
> Trade Name: At Home Drug Test (Model 9078) Regulatory Class: II Product Code: LDJ Dated: August 25, 1999 Received: December 15, 1999
Dear Mr. Mongiovi:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
## Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS FOR USE
Applicant: Phamatech
16994220 510 (k) Number (if known):
Device Name: At Home Drug Test (Model 9078)
Indications for Use:
A home drug screening, and if needed, confirmation service. This kit provides a preliminary result for the detection/presence of the following drugs of abuse in urine: THC
Sein Coogh
(Division Sign-Off)
Division of Clinical Laboratory Levices 510(k) Number .
## PLEASE DO NOT WRITE BELOW THIS LINE
Concurrence of the CDRH Office of Device Evaluation (ODE)
Division Sign-off Division of Clinical Laboratory Devices 510 (k) Number:
Over the Counter: OR Prescription Use: Per 21 CFR 801.109
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.