FIRST CHECK HOME DRUG TEST (THC)

K993663 · Worldwide Medical Corp. · LDJ · Jun 23, 2000 · Clinical Toxicology

Device Facts

Record IDK993663
Device NameFIRST CHECK HOME DRUG TEST (THC)
ApplicantWorldwide Medical Corp.
Product CodeLDJ · Clinical Toxicology
Decision DateJun 23, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3870
Device ClassClass 2

Intended Use

The First Check Home Drug Test (THC) is an educational screening test for the rapid detection of THC (and its metabolites) in human urine at a cut off for the rapid dececeion of the (antended for consumer use as the first step in a two step process to provide consumers, including but not limited to concerned a two seep process to provious concerning the presence of THC (and its parents, with information obliozing is beneficial to consumer metabolices) in a arine baspeable laws and/or societal expectations. Information regarding second step confirmatory testing is provided.

Device Story

First Check Home Drug Test (THC) is a rapid, one-step chromatographic immunoassay for qualitative detection of THC and metabolites in human urine; cut-off concentration 50ng/mL. Device utilizes antibody/target drug/antibody complex formation principle. Intended for lay-user/consumer self-testing; no special training or licensure required. User applies urine sample to test card well; visual color change indicates presence or absence of target. Results serve as preliminary screening; information provided for mandatory second-step confirmatory testing. Benefits include providing consumers/parents with information regarding drug presence for personal or societal awareness.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and performance equivalence to the predicate device. The device relies on established one-step chromatographic immunoassay principles.

Technological Characteristics

One-step rapid chromatographic immunoassay. Principle: recognition and formation of specific antibody/target drug/antibody complexes. Cut-off concentration: 50ng/mL. Form factor: test card. No electronic components or software.

Indications for Use

Indicated for consumer use as a preliminary, rapid screening test for the detection of THC and its metabolites in human urine. Intended as the first step in a two-step process to provide information regarding the presence or absence of THC in a urine sample.

Regulatory Classification

Identification

A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.

Special Controls

*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### JUN 2 3 2000 Image /page/0/Picture/3 description: The image shows a black parallelogram. The parallelogram is oriented diagonally, with the top left corner being higher than the bottom right corner. The shape is solid black, with no visible texture or details within the shape. Section D 510(k) Summary 21 C.F.R. § 807.92 - I. Submitter: - Name: Worldwide Medical Corporation A. - Address:199 Technology Drive, Suite 150, Irvine, California B. 92618 - Phone and Fax Numbers: Phone: 949/727-0711 C. Fax: 949/727-0602 - Contact Person: H. Thad Morris D. - Date of Preparation of this Summary: September 29, 1999 II. - III. Trade Name: First Check® Home Drug Test (THC) - Common Name: At home drugs of abuse rapid screening test for THC in IV. urine - Classification Name: Immunoassay for the qualitative detection of THC in V. urine - The Marketed Products to Which Equivalence is Claimed: The First VI. Check® Home Drug Test (THC) that is the subject of this submission is identical to the Applied Biotech SureStep™ Drug Screen THC Test in terms of intended use, product design, performance characteristics, materials of construction, and manufacturing process. It is also substantially equivalent to the Phamatech QuickScreen™ At Home Drug Test and other commercially available drug screening tests which qualitatively measure the presence of target drugs or metabolites by visual color one step immunoassay technology. - Statement of Intended Use Compared to Other Products: The intended use VII. of the First Check® Home Drug Test (THC) is substantially equivalent to the listed products; it is a preliminary, rapid screening test for the detection of THC and its metabolites in urine. This product is intended to be the first step in a two step process to provide consumers, including concerned parents. with information regarding the presence or absence of THC (and its metabolites) in a urine sample. Information regarding the second step, confirmatory testing, is provided. 199 Technology Drive Suite 150 Irvine, CA 92618 ·49-727-0711 800-788-5716 Fax: 949-727-0602 www.wwmed.com Stock symbo : WMED {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows a black parallelogram. The parallelogram is tilted slightly to the left. The background is white. - Discussion of Technological Characteristics: The First Check® Home VIII. Drug Test (THC), like many commercially available drug screening tests, qualitatively measures the presence or absence of THC and its metabolites in urine using a one step rapid chromatographic immunoassay which operates under the principle of recognition and formation of specific antibody/target drug/antibody complexes. The cut-off concentration for THC is 50ng/mL. - IX. Safety and Effectiveness: Because the First Check® Home Drug Test (THC) is identical to the Applied Biotech SureStep™ Drug Screen THC test that is legally marketed under K960744, and because no special skills, training, education, or licensure are required to transfer a dropper level of urine sample into the test card well, there is no issue regarding the safety or effectiveness of the product to perform its intended function, i.e., to screen urine for the presence or absence of THC and its metabolites. Because the labeling of the First Check® Home Drug Test (THC) is substantially equivalent to a variety of rapid screening tests currently in commercial distribution, including the Phamatech QuickScreen™ At Home Drug Test, and there have been no reports of consumer inability to follow instructions or interpret results during the twenty-four months the product has been purchased by the general public and used in quantities exceeding 87,000 tests, it is concluded that the product can be used effectively by the lay user. 199 Technology D. ... : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : - 2 ( 2 0 2 ) 3 Fer 949 721 Coll www.wwmed.com Stock symbol WMED {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines extending from its head, representing the department's mission to protect the health of all Americans and provide essential human services. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 ## JUN 2 3 2000 Worldwide Medical Corporation c/o Mr. Larry R. Pilot McKenna & Cueno, L.L.P. 1900 K Street, N.W. Washington, D.C. 20006-1108 Re: K993663 > Trade Name: First Check® Home Drug Test (THC) Regulatory Class: II Product Code: LDJ Dated: May 23, 2000 Received: May 23, 2000 Dear Mr. Pilot: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours. Steven Gutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page I of 1 # 510(k) Number (if known):_K993663 First Check Home Drug Test (THC) Device Name: Indications For Use: The First Check Home Drug Test (THC) is an educational screening test for the rapid detection of THC (and its metabolites) in human urine at a cut off for the rapid dececeion of the (antended for consumer use as the first step in a two step process to provide consumers, including but not limited to concerned a two seep process to provious concerning the presence of THC (and its parents, with information obliozing is beneficial to consumer metabolices) in a arine baspeable laws and/or societal expectations. Information regarding second step confirmatory testing is provided. Jean Cooper (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K993663 (Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) OVER-The -COUNTA USE (Optional Format 3-10-98) 4
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%