AMEDICA DRUG SCREEN THC, COC, OPI, AMP, MET, PCP, MDMA, BAR, BZO, M, MODELS C1210, C1220

K063379 · Amedica Biotech, Inc. · DKZ · Dec 11, 2006 · Clinical Toxicology

Device Facts

Record IDK063379
Device NameAMEDICA DRUG SCREEN THC, COC, OPI, AMP, MET, PCP, MDMA, BAR, BZO, M, MODELS C1210, C1220
ApplicantAmedica Biotech, Inc.
Product CodeDKZ · Clinical Toxicology
Decision DateDec 11, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.3100
Device ClassClass 2

Indications for Use

The Amedica Drug Screen Test Cup II THC/COC/OPI/AMP/MET/PCP/MDMA/BAR/BZO/MTD/TCA/PPX/OXY test is intended for the qualitative detection of drug metabolites in human urine. The test is intended for professional use only.

Device Story

Drug screen test cup; collects human urine; detects up to 12 drug metabolites; utilizes permanent test strips; features tamper-proof design preventing access to strips; two cup configurations (A and B) with varying drug panels; professional use only; provides qualitative results; aids in clinical drug screening.

Clinical Evidence

No clinical data provided. Device performance is established through analytical bench testing, including sensitivity, specificity, and cutoff verification for each drug analyte.

Technological Characteristics

Lateral flow immunoassay; qualitative visual readout; urine specimen; multi-panel test cup format. Detects drugs at specific ng/mL cutoffs (e.g., THC 50, COC 300, OPI 300/2000, AMP 1000, MET 1000, PCP 25, MDMA 500, BAR 300, BZO 300, MTD 300, TCA 1000, PPX 300, OXY 300).

Indications for Use

Indicated for professional use in the qualitative detection of drug metabolites (THC, COC, OPI, AMP, MET, PCP, MDMA, BAR, BZO, MTD, TCA, PPX, OXY) in human urine samples.

Regulatory Classification

Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification ODE Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER K063379 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for the Amedica Drug Screen Test Cup II THC/COC/OPI/AMP/MET/PCP/MDMA/BAR/BZO/MTD/TCA/PPX/OXY test incorporates a modified cup and permanent test strips. This product includes a collection cup that can now test up to 12 drugs and has a tamper proof feature to prevent access to the test strips. The test strips are permanently placed in two versions of the cup (A and B). Cups A and B contain different drug test configurations. The previous device could accommodate up to 10 drugs and the strips were interchangeable. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, immunoassay technology, test protocol, performance, assay principle, cutoff level and reagents. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...