K021006 · Starplex Scientific, Inc. · DKZ · Jun 10, 2002 · Clinical Toxicology
Device Facts
Record ID
K021006
Device Name
STARTOX DRUG OF ABUSE SCREENING TEST
Applicant
Starplex Scientific, Inc.
Product Code
DKZ · Clinical Toxicology
Decision Date
Jun 10, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3100
Device Class
Class 2
Indications for Use
STARTOX™ Drugs of Abuse Screening Test is a one-step lateral flow immunoassay intended for the simultaneous detection of multiple drug analytes in urine. STARTOX™ is intended for use in the qualitative detection of drugs of abuse at the following Substance Abuse Mental Health Services Administration (SAMHSA) recommended levels: Compound Abbreviation Level Amphetamine (d-amphetamine sulfate) AMP 1000 ng/ml Opiates (morphine-3-P-D glucuronide) OPIATES 2000 ng/ml Phencyclidine (phencyclidine HCl) PCP 25 ng/ml Cocaine (benzoylecgonine) COCAINE 300 ng/ml Cannabinoids (11-nor-Δ9-THC-9-carboxylic-acid) THC 50 ng/ml STARTOX™ Drugs of Abuse Screening Test provide only a preliminary qualitative test result. Use a more specific alternate quantitative analytical method to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Apply clinical and professional judgment to any drug of abuse test result, particularly when preliminary positive results are obtained. For professional use only.
Device Story
Lateral flow immunoassay for qualitative detection of drugs of abuse in human urine; simultaneous multi-analyte screening. Device utilizes immunochemical reaction; visual readout of test results. Intended for professional use in clinical settings. Provides preliminary qualitative results; requires confirmatory testing via GC/MS for clinical decision-making. Benefits include rapid screening for drug presence to guide further diagnostic or clinical management.
Clinical Evidence
Bench testing only. Performance evaluated using 90 urine samples per drug (50 negative, 40 positive). Positive samples confirmed via Emit and GC/MS. Accuracy reported as >99% compared to Medtox Profile II. Reproducibility confirmed using controls above and below cutoff levels.
Technological Characteristics
Lateral flow immunoassay; qualitative visual readout. Detects AMP, OPIATES, PCP, COCAINE, and THC in urine at specified ng/ml concentrations. Standalone test device.
Indications for Use
Indicated for professional use in the qualitative detection of Amphetamine, Opiates, Phencyclidine, Cocaine, and Cannabinoids in human urine at SAMHSA-recommended cutoff levels. Provides preliminary results requiring confirmation by GC/MS.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
ABMC Rapid Drug Screen multi-drug products (K012159)
Reference Devices
Medtox Profile II
Related Devices
K980872 — RAPID DRUG SCREEN' 8 PANEL · American Bio Medica Corp. · Mar 20, 1998
K032575 — FIRST SIGN DRUG OF ABUSE SCREENING TEST · W.H.P.M., Inc. · May 19, 2004
K012159 — RAPID DRUG SCREEN · American Bio Medica Corp. · Jul 25, 2001
K022388 — STARTOX DRUG OF ABUSE SCREENING TEST (4) · Starplex Scientific, Inc. · Feb 27, 2003
K080872 — AMEDICHECK DRUG SCREEN THC, COC, OPI, AMP, MET, PCP, MDMA,BAR, BZ · Amedica Biotech, Inc. · Aug 19, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
JUN 1 0 2002
# Starplex Scientific Inc. 510k Submission K021006 Additional Information 6/6/02
# 510 (K) SUMMARY
Page 3 of 19
#### Date of Summary
June 4, 2002
Product Name:
STARTOX™
#### Sponsor and Manufacturer:
Starplex Scientific, Inc. 50 Steinway Blvd. Etobicoke, Ontario, Canada M9W 6Y3
#### Correspondent:
Fran White MDC Associates 163 Cabot Street Beverly, MA 01915
#### Substantially Equivalent Device:
Product: Rapid Drug Screen - 9 panel Manufactured by: American BioMedica Corp 510k Number: K002447
#### Product Description:
A lateral flow immunoassay for the detection of drugs of abuse.
{1}------------------------------------------------
### Starplex Scientific Inc. 510k Submission K021006 Additional Information 6/6/02
### Page 4 of 19
#### INTENDED USE:
STARTOX™ Drugs of Abuse Screening Test is a one-step lateral flow immunoassay intended for the simultaneous detection of multiple drug analytes in urine. STARTOX™ is intended for use in the qualitative detection of drugs of abuse at the following Substance Abuse Mental Health Services Administration (SAMHSA) recommended levels:
| Compound | Abbreviation | Level |
|---------------------------------------------------|--------------|------------|
| Amphetamine (d-amphetamine sulfate) | AMP | 1000 ng/ml |
| Opiates (morphine-3-P-D glucuronide) | OPIATES | 2000 ng/ml |
| Phencyclidine (phencyclidine HCl) | PCP | 25 ng/ml |
| Cocaine (benzoylecgonine) | COCAINE | 300 ng/ml |
| Cannabinoids<br>(11-nor-Δ9-THC-9-carboxylic-acid) | THC | 50 ng/ml |
STARTOX™ Drugs of Abuse Screening Test provide only a preliminary qualitative test result. Use a more specific alternate quantitative analytical method to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Apply clinical and professional judgment to any drug of abuse test result, particularly when preliminary positive results are obtained.
### PERFORMANCE CHARACTERISTICS:
STARTOX™ drugs of abuse screening test detects 5 drugs in human urine at the levels indicated.
STARTOX™ is substantial equivalent to Rapid Drug Screen 9 panel manufactured by American BioMedica Corporation 510k number K002447.
Product performance was compared Medtox Profile II manufactured by Medtox Diagnostics. Ninety (90) samples were tested against each drug, 50 negative which were confirmed to be drug free and 40 positive specimens for each drug were screened as positive by Emit, confirmed and quantified by GC/MS.
The STARTOX™ Drugs of Abuse Screening Test was compared to a FDA substantially equivalent approved device, Medtox Profile II. STARTOX™ demonstrated greater than 99% accuracy with when compared to a legally marketed device.
{2}------------------------------------------------
# Starplex Scientific Inc. 510k Submission K021006 Additional Information 6/6/02
Page 5 of 19
Reproducibility was evaluated using control urines containing drug concentrations above and below the stated cut-off. Negative controls were also tested. The results confirmed the reproducibility of the STARTOXTM Drugs of Abuse Screening Test.
### CONCLUSION:
STARTOX™ Drug of Abuse Screening Test is substantially equivalent to ABMC Rapid Drug Screen multi-drug products (K012159, K002447) previously cleared for market as demonstrated by the results obtained in the studies.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three horizontal lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# JUN 1 0 2002
Starplex Scientific, Inc. c/o Ms. Fran White Regulatory Consultant MDC Associates 163 Cabot Street Beverly, MA 01915
Re: k021006
Trade/Device Name: Startox™ Drug of Abuse Screening Test Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: DKZ; DJG; LCM; DIO; LDJ Dated: March 27, 2002 Received: March 28, 2002
Dear Ms. White:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Starplex Scientific Inc. 510k Submission K021006 Additional Information 6/6/02
Page 6 of 19
510(k) Number: Device Name: STARTOXTM Drug of Abuse Screening Test
#### Indication for Use:
STARTOX™ Drugs of Abuse Screening Test is a one-step lateral flow immunoassay intended for the simultaneous detection of multiple drug analytes in urine. STARTOX™ is intended for use in the qualitative detection of drugs of abuse at the following Substance Abuse Mental Health Services Administration (SAMHSA) recommended levels:
| Compound | Abbreviation | Level |
|---------------------------------------------------|--------------|------------|
| Amphetamine (d-amphetamine sulfate) | AMP | 1000 ng/ml |
| Opiates (morphine-3-P-D glucuronide) | OPIATES | 2000 ng/ml |
| Phencyclidine (phencyclidine HCl) | PCP | 25 ng/ml |
| Cocaine (benzoylecgonine) | COCAINE | 300 ng/ml |
| Cannabinoids<br>(11-nor-Δ9-THC-9-carboxylic-acid) | THC | 50 ng/ml |
STARTOX™ Drugs of Abuse Screening Test provide only a preliminary qualitative test result. Use a more specific alternate quantitative analytical method to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Apply clinical and professional judgment to any drug of abuse test result, particularly when preliminary positive results are obtained.
For professional use only.
Cooger
(Division -Off)
Division o Clinical Laboratory Devices
510(k) Numes: K221006
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over the Counter Use (Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.