K202175 · Raja Trading Company, Inc. · GEX · Dec 2, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202175
Device Name
OxyLight 2.0
Applicant
Raja Trading Company, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Oxylight 2.0 is intended for dermatological use by physicians and healthcare professionals for the following: LED Technology is intended for: -Blue LED 465nm - to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris. -Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions. - Yellow LED 590nm - treatment of periorbital wrinkles and rhytides. Microdermabrasion is intended for exfoliation of the skin. Oxygen spray is intended to refresh the skin.
Device Story
OxyLight 2.0 is a multi-modal skin therapy system; includes LED light panel, microdermabrasion handpiece, oxygen spray, and two new handheld LED 'MyoLight' applicators (face and body). Device inputs: user-selected treatment parameters via display screen. Principle of operation: LED light emission at specific wavelengths (465nm, 625nm, 590nm) for phototherapy; mechanical exfoliation via microdermabrasion; oxygen delivery for skin refreshing. Used in clinical settings by physicians and healthcare professionals. Output: light energy delivered to skin surface; mechanical skin exfoliation; oxygen spray. Clinical benefit: treatment of acne, vascular/pigmented lesions, and wrinkles; skin rejuvenation. System is an evolution of the predicate OxyLight (K200104) with added handheld LED functionality.
Clinical Evidence
No clinical data included. Substantial equivalence is supported by bench testing, software verification/validation, and compliance with international safety standards (IEC 60601 series) and biocompatibility requirements (ISO 10993).
Technological Characteristics
System includes LED panel (840 LEDs) and handheld LED applicators (8 LEDs for face, 18 for body). Wavelengths: 465nm (Blue), 625nm (Red), 590nm (Yellow). Energy/power outputs identical to predicate. Connectivity: display screen interface. Safety standards: IEC 60601-1, IEC 60601-1-2. Biocompatibility: ISO 10993.
Indications for Use
Indicated for patients requiring dermatological treatment for moderate inflammatory acne vulgaris, superficial benign vascular and pigmented lesions, and periorbital wrinkles and rhytides. Also indicated for skin exfoliation and refreshing.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 2, 2021
Raja Trading Company, Inc. % Yolanda Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114
Re: K202175
Trade/Device Name: OxyLight 2.0 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, GFE Dated: January 6, 2021 Received: January 7, 2021
Dear Yolanda Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202175
Device Name OxyLight 2.0
Indications for Use (Describe)
The Oxylight 2.0 is intended for dermatological use by physicians and healthcare professionals for the following: LED Technology is intended for:
-Blue LED 465nm - to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
-Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions.
- Yellow LED 590nm - treatment of periorbital wrinkles and rhytides.
Microdermabrasion is intended for exfoliation of the skin.
Oxygen spray is intended to refresh the skin.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of 21CFR Section 807.92.
510(k) Number: K202175
- Date of Preparation 1. 12/02/2021
- 2. Applicant
| Name: | RAJA Trading Company Inc. |
|-----------------|------------------------------------------------|
| Address: | 2801 Juniper Street, Suite 2, Fairfax, VA 2203 |
| Contact Person: | Robert J. Adipietro |
| Title: | Vice President |
| Telephone: | 561-868-4600 |
| Fax: | 561-258-0207 |
| Email: | rja@rajamedical.com |
- 3. Identification of the Proposed Device Trade Name: OxyLight 2.0 Common Name: LED Phototherapy and Microdermabrasion Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Classification: Class II Product Code: GEX, GFE Regulation Number: 21 CFR 878.4810 Review Panel: General& Plastic Surgery
| Attachment | Regulation Name | Regulation No. | Product<br>code | Class |
|--------------------------------|----------------------------------------|----------------|-------------------|-------------------------|
| LED Panel | Laser Instrument, Surgical<br>Powered, | 878.4810 | GEX | 2 |
| LED Handpieces | Laser Instrument, Surgical<br>Powered, | 878.4810 | GEX | 2 |
| Microdermabrasion<br>Handpiece | Brush, Dermabrasion<br>Powered | 878.4820 | GFE | Class I, 510k<br>exempt |
| Oxygen Spray<br>Handpieces | Not Applicable | Not Applicable | Not<br>Applicable | General<br>Wellness |
- 4. Identification of Predicate Device
| 510(k) Number: | K200104 |
|----------------|---------------------------|
| Product Name: | OxyLight |
| Manufacturer: | RAJA Trading Company Inc. |
- 5. Device Description
The subject device Oxylight 2.0 is a skin therapy system that was modified from the predicate device OxyLight cleared previously under K200104 for the LED Light Panel Therapy (Class 2), Microdermabrasion (Class 1, 510k exempt) and Oxygen Spray (General Wellness). The OxyLight 2.0 system includes two new LED handpieces called MyoLight. The MyoLight handpieces connect to the main unit. The small handpiece (diameter: 3.9 cm, Length: 14 cm) is generally used on the face and the large handpiece (diameter: 6.2 cm, Length: 15 cm) is generally used on the body.
- 6. Indications for Use
The Oxylight 2.0 is intended for dermatological use by physicians and healthcare professionals for the following: LED Technology is intended for:
-Blue LED - 465nm - to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
-Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions.
{4}------------------------------------------------
- Yellow LED 590nm - treatment of periorbital wrinkles and rhytides.
Microdermabrasion is intended for exfoliation of the skin.
Oxygen spray is intended to refresh the skin.
- Substantially Equivalent (SE) Comparison 7.
| Specifications | Subject Device | Predicate Device | Comment |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Device Name | OxyLight 2.0 | OxyLight | N/A |
| 510(k) Number | K202175 | K200104 | N/A |
| Applicant | RAJA Trading Company,<br>Inc. | RAJA Trading Company,<br>Inc. | N/A |
| Intended use/ Product<br>Code | Powered laser surgical<br>instrument,<br>Power dermabrasion<br>Class II/GEX/878.4810<br>Class I/GFE/878.4820 | Powered laser surgical<br>instrument,<br>Power dermabrasion<br>Class II/GEX/878.4810<br>Class I/GFE/878.4820 | |
| Indications LED | -Blue LED Light - to treat<br>dermatological conditions<br>and specifically indicated<br>to treat mild to moderate<br>inflammatory acne<br>vulgaris.<br>-Red LED Light - for<br>treatment of superficial,<br>benign vascular and<br>pigmented lesions.<br>- Yellow LED Light -<br>treatment of periorbital<br>wrinkles and rhytides. | -Blue LED Light - to treat<br>dermatological conditions<br>and specifically indicated<br>to treat mild to moderate<br>inflammatory acne<br>vulgaris.<br>-Red LED Light - for<br>treatment of superficial,<br>benign vascular and<br>pigmented lesions.<br>- Yellow LED Light -<br>treatment of periorbital<br>wrinkles and rhytides. | Identical |
| Design and Mode of<br>Action | Panel with an array of<br>LED Lights. The LEDs<br>emit blue (465 nm), red<br>(625 nm) or yellow (590<br>nm) light. In addition, the<br>OxyLight 2.0 includes 2<br>hand-held handpieces that<br>house LED lights. The<br>hand-held handpieces<br>deliver the light to the<br>skin as they are moved<br>over the skin surface. | Panel with an array of LED<br>Lights. The LEDs emit<br>blue (465 nm), red (625<br>nm) or yellow (590 nm)<br>light. | Similar |
| No. of LEDs | Panel - 840<br>Face Handpiece - 8<br>Body Handpiece - 18 | Panel - 840 | Similar power<br>density and<br>energy flux |
| Treatment Time | 20 minutes | 20 minutes | Identical |
| Light Source | LED | LED | Identical |
| Operation Interface | Display Screen | Display Screen | Identical |
| Wavelength (Blue) | 465 nm +/- 5nm | 465 nm +/- 5nm | Identical |
| Energy Output (Blue) | 54 J/cm² | 54 J/cm² | Identical |
| Power Output (Blue) | 45 mW/cm² | 45 mW/cm² | Identical |
| Wavelength (Red) | 625 nm +/- 5nm | 625 nm +/- 5nm | Identical |
| Energy Output (Red) | 120 J/cm² | 120 J/cm² | Identical |
| Power Output (Red) | 100 mW/cm² | 100 mW/cm² | Identical |
| Wavelength (Yellow) | 590 nm +/- 5nm | 590 nm +/- 5nm | Identical |
| Energy Output (Yellow) | 42 J/cm² | 42 J/cm² | Identical |
| Power Output (Yellow) | 35 mW/cm² | 35 mW/cm² | Identical |
{5}------------------------------------------------
- 8. Non-Clinical Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The following tests were conducted:
- > IEC 60601-1:2005/A1:2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance*;
- > IEC 60601-1-2:2014/EN 60601-1-2:2015, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests**.
- > Biocompatibility Tests per ISO 10993 and FDA Guidance
- ハ Software Validation & Verification Test
- > Bench Testing to verify the performance
*The OxyLight 2.0 was tested with the LED Handpieces under the name OxyLight was cleared by the FDA K200104 with the LED Panel, Microdermabrasion and Oxygen Spray Modalities. The OxyLight 2.0 is the OxyLight with the addition of the LED handpieces.
** Sapphire 3 OxyLight is the same test article. Sapphire 3 is a trademark brand name that RAJA Trading Company, Inc. uses for its class of skin care devices. In addition, the OxyLight 2.0 was tested with the LED Handpieces under the name OxyLight. The OxyLight was cleared by the FDA K200104 with the LED Panel, Microdermabrasion and Oxygen Spray Modalities. The OxyLight 2.0 is the OxyLight with the addition of the LED handpiece.
#### 9. Clinical Testing
No clinical study is included in this submission.
- 10. Conclusion
Based on the comparison and analysis above, the proposed subject device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.