K200104 · Raja Trading Company, Inc. · GEX · Jun 4, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200104
Device Name
Oxylight
Applicant
Raja Trading Company, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 4, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Oxylight is intended for dermatological use by physicians and healthcare professionals for the following: LED Technology is intended for: -Blue LED = 465nm - to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris. -Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions. - Yellow LED 590nm - treatment of periorbital wrinkles and rhytides. Microdermabrasion is intended for exfoliation of the skin. Oxygen spray is intended to refresh the skin.
Device Story
OxyLight is a multi-modality skin therapy system for use by physicians and healthcare professionals in clinical settings. It integrates three modalities: LED light therapy (Blue 465nm, Red 625nm, Yellow 590nm), microdermabrasion, and oxygen spray. The system utilizes a universal control unit with a display screen to power various attachments, including an LED light panel and separate handpieces for microdermabrasion and oxygen delivery. LED therapy treats acne, vascular/pigmented lesions, and wrinkles via light-based phototherapy. Microdermabrasion provides mechanical skin exfoliation, while oxygen spray refreshes the skin. The device is operated by clinicians to provide non-invasive dermatological treatments. Clinical decision-making is guided by the practitioner's assessment of the patient's skin condition, with the device providing targeted therapeutic light or mechanical/topical skin care. Benefits include non-invasive treatment of specific dermatological conditions and skin rejuvenation.
Clinical Evidence
No clinical data was performed. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1) and electromagnetic compatibility (EN 60601-1-2) testing.
Technological Characteristics
System utilizes LED arrays for phototherapy (465nm, 625nm, 590nm), microdermabrasion handpieces, and oxygen spray. Energy source is a universal control unit. Dimensions: 41.4cm x 23cm x 19.2cm. Operation via display screen interface. No patient contact during LED therapy. Sterilization/biocompatibility not specified. Software/algorithm details not provided.
Indications for Use
Indicated for dermatological use by physicians/healthcare professionals to treat moderate inflammatory acne vulgaris (Blue LED), superficial benign vascular and pigmented lesions (Red LED), and periorbital wrinkles/rhytides (Yellow LED). Also indicated for skin exfoliation (microdermabrasion) and refreshing skin (oxygen spray).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
June, 4, 2020
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
RAJA Trading Company, Inc. % Yolanda Smith Consultant Smith Associates 1468 Harwell Ave Crofton, Maryland 21114
Re: K200104
Trade/Device Name: Oxylight Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, GFE Dated: January 14, 2020 Received: March 10, 2020
Dear Yolanda Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200104
Device Name OxyLight
Indications for Use (Describe)
The Oxylight is intended for dermatological use by physicians and healthcare professionals for the following:
LED Technology is intended for:
-Blue LED = 465nm - to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
-Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions.
- Yellow LED 590nm - treatment of periorbital wrinkles and rhytides.
Microdermabrasion is intended for exfoliation of the skin.
Oxygen spray is intended to refresh the skin.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K200104 510(k)Summary
## SPONSOR
| Company Name: | RAJA Trading Company Inc. |
|------------------|--------------------------------|
| Company Address: | 801 South Olive Avenue |
| | Suite 124 |
| | West Palm Beach, Florida 33401 |
| Telephone: | 561-868-4600 |
| Fax: | 561-258-0207 |
| Contact Person: | Robert J. Adipietro |
| Title: | VP |
Summary Preparation Date: June 4, 2020
## DEVICE NAME
| Trade Name: | OxyLight |
|----------------------|----------------------------------------------------------------------------------------|
| Common/Usual Name: | LED Phototherapy, Microdermabrasion, Oxygen Spray |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology |
| Regulation Number: | 21 CFR 878.4810; 878.4820 |
| Product Code: | GEX, GFE |
| Device Class: | Class II |
| Attachment | Regulation Name | Regulation No. | Product code | Class |
|--------------------------------|----------------------------------------|----------------|----------------|----------------------|
| Panel | Laser Instrument, Surgical<br>Powered, | 878.4810 | GEX | 2 |
| Microdermabrasion<br>Handpiece | Brush, Dermabrasion<br>Powered | 878.4820 | GFE | Class I, 510k exempt |
| Oxygen Spray<br>Handpieces | Not Applicable | Not Applicable | Not Applicable | General Wellness |
#### Legally Marketed Equivalent Devices
| LED BLUE Light<br>- Primary | BioPhotas, Inc. | BioPhotas LifeLight | K122237 |
|-------------------------------|-----------------------------------------|------------------------------|---------|
| LED BLUE Light<br>- Reference | Photo Therapeutics Ltd | Omnilux Blue | K030883 |
| LED RED Light<br>- Primary | Photo Therapeutics Ltd | Omnilux Revive | K030426 |
| LED YELLOW -<br>- Primary | Quantel Derma GMBH AM<br>Wolfsmantel 46 | LEDA — Applicator<br>SCR 585 | K090762 |
{4}------------------------------------------------
The Oxylight system is a skin therapy system that uses three major modalities delivered by different attachments powered from a single energy source and operated from a universal control unit. The treatments include, LED Light Panel Therapy (Class 2), Microdermabrasion (Class 1, 510k exempt) and Oxygen Spray. The Oxygen Spray is delivered through the same panel used for LED light therapy and separate handpieces.
#### DEVICE INDICATIONS FOR USE
The Oxylight is intended for dermatological use by physicians and healthcare professionals for the following:
LED Technology is intended for:
- -Blue LED 465nm to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
- -Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions.
- -Yellow LED 590nm treatment of periorbital wrinkles and rhytides.
Microdermabrasion is intended for exfoliation of the skin.
Oxygen spray is intended to refresh the skin.
| COMPARISON OF TECHNICAL CHARACTERISTICS | | | | |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| LED PDT Blue<br>Light Panel<br>Component | Subject Device | Primary Predicate Device | Reference<br>Predicate Device | Substantial<br>Equivalent |
| Device Name | OxyLight | BioPhotas LifeLight | Omnilux Blue | |
| Applicant | RAJA Trading<br>Company, Inc. | BioPhotas, Inc. | Photo<br>Therapeutics<br>Limited | |
| Listing | K200104 | K122237 | K030883 | |
| Classification<br>Name<br>Class/Product<br>Code/Regulation | Laser Instrument,<br>Surgical Powered<br>Class II/GEX/878.4810 | Laser Instrument,<br>Surgical Powered<br>Class<br>II/GEX/878.4810 | Laser Instrument,<br>Surgical Powered<br>Class<br>II/GEX/878.4810 | Yes |
| Indications for<br>Use | To treat dermatological<br>conditions, specifically<br>to treat mild to<br>moderate acne<br>vulgaris. | The blue light<br>spectrum is intended<br>to reduce mild to<br>moderate acne<br>vulgaris. | Dermatological<br>conditions and<br>specifically to<br>treat moderate<br>inflammatory<br>acne vulgaris. | Yes |
| Wavelength of<br>Blue Light | 465 nm +/- 5nm | 464 nm +/- 5nm | 415 +/-5 nm | Yes |
| Energy Output<br>in<br>Joules/cm2 | 54 joules/cm2 | 19.6 joules/cm2 | 48 joules/cm² | Different from<br>the Primary<br>Predicate Device<br>but similar in<br>technological<br>characteristics to<br>the reference<br>predicate device. |
| Power Output<br>Level mW/cm² | 45 milliwatts/cm² | Unknown | 40 milliwatts/cm² | Different from<br>the Primary<br>Predicate Device<br>but similar in<br>technological |
| | | | | characteristics to<br>the reference<br>predicate device. |
| Design and<br>Mode of Action | Panel with an array of<br>LED Lights | Panel with an array<br>of LED Lights | Panel with an<br>array of LED Lights | Yes |
| Panel<br>Dimensions | 41.4cm x 23cm x<br>19.2cm | Unknown | 31.8cm x 35.6cm x<br>8.9cm | Yes |
| Recommended<br>Treatment Time | 20 minutes | 30 minutes | 20 minutes | Yes |
| Patient Contact | None | None | None | Yes |
| Laser Type | LED | LED | LED | Yes |
| Operation<br>Interface | Display Screen | Unknown | Display Screen | Yes |
#### PARISON OF TECHNICAL CHARA
{5}------------------------------------------------
| Red LED Light<br>Panel<br>Component | Subject Device | Predicate Device | Substantial<br>Equivalent |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|---------------------------|
| Device Name<br>Applicant | OxyLight<br>RAJA Trading<br>Company, Inc. | Omnilux Revive<br>Photo Therapeutics<br>Limited | |
| Listing | K200104 | K030426 | |
| Classification<br>Name<br>Class/Product<br>Code/Regulation | Laser Instrument,<br>Surgical Powered<br>Class II/GEX/878.4810 | Laser Instrument,<br>Surgical Powered<br>Class II/GEX/878.4810 | Yes |
| Indications for<br>Use | Indicated for use in<br>dermatology for<br>treatment of<br>superficial, benign<br>vascular, and<br>pigmented lesions. | Indicated for use in<br>dermatology for<br>treatment of<br>superficial, benign<br>vascular, and<br>pigmented lesions. | Yes |
| Wavelength of<br>Red Light | 625 nm +/- 5nm | 633 +/- 5 nm | Yes |
| Energy Fluency<br>Joules/cm2 | 120 joules/cm2 | 126 joules/cm² | Yes |
| Power Output at<br>Skin Level<br>mW/cm² | 100 milliwatts/cm² | 105 milliwatts/cm² | Yes |
| Design and<br>Mode of Action | Panel with an array of<br>LED Lights | Panel with an array<br>of LED Lights | Yes |
| Panel<br>Dimensions | 41.4cm x 23cm x<br>19.2cm | 31.8cm x 35.6cm x<br>8.9cm | Yes |
| Recommended<br>Treatment Time | 20 minutes | 20 minutes | Yes |
| Patient Contact | None | None | Yes |
| Laser Type | LED | LED | Yes |
| Operation<br>Interface | Display Screen | Display Screen | Yes |
{6}------------------------------------------------
| Yellow LED PDT<br>Light Panel<br>Component | Subject Device | Reference Predicate<br>Device | Substantial<br>Equivalent |
|------------------------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|---------------------------|
| Device Name | OxyLight | LEDA - Applicator SCR<br>585 | |
| Applicant | RAJA Trading<br>Company, Inc. | Quantel Derma<br>GMBH<br>AM Wolfsmantel 46 | |
| Listing | K200104 | K090762 | |
| Classification<br>Name<br>Class/Product<br>Code/Regulation | Laser Instrument,<br>Surgical Powered<br>Class II/GEX/878.4810 | Laser Instrument,<br>Surgical Powered<br>Class<br>II/GEX/878.4810 | Yes |
| Intended Use | Generally indicated for<br>treatment of<br>periorbital wrinkles<br>and rhytides. | Generally indicated<br>for treatment of<br>periorbital wrinkles<br>and rhytides. | Yes |
| Wavelength of<br>Yellow Light | 590 nm +/- 5nm | 585 nm +/- 5nm | Yes |
| Energy Fluency<br>Joules/cm2 | 42 joules/cm2 | 0.1 - up to 100<br>joules/cm2 | Yes |
| Power Output at<br>Skin Level<br>mW/cm² | 35 milliwatts/cm² | 4mW/cm² –<br>120mW/cm² | Yes |
| Panel<br>Dimensions | 41.4cm x 23cm x<br>19.2cm | 45cm x 41cm x 24 cm | Yes |
| Design and<br>Mode of Action | Panel with an array of<br>LED Lights | Panel with an array<br>of LED Lights | Yes |
| Recommended<br>Treatment Time | 20 minutes | Unknown | Yes |
| Patient Contact<br>Materials | None | None | Yes |
| Laser Type | LED | LED | Yes |
| Operation<br>Interface | Display Screen | Display Screen | Yes |
# PERFORMANCE DATA
# Safety Testing
| Standard | Test Title | Test Article | Test Results |
|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--------------|
| IEC 60601-1:2005 +<br>CORR. 1:2006 +<br>CORR. 2:2007 +<br>AM1:2012 | Medical electrical equipment Part 1: General<br>requirements for basic safety and essential<br>performance | OxyLight | Pass |
| EN 60601-1-2:2015<br>EN 61000-3-2:2014<br>EN 61000-2-3:2013 | Medical electrical equipment -Part 1-2:<br>General requirements for basic safety and<br>essential performance-<br>Collateral standard: Electromagnetic<br>compatibility- Requirements and tests | Sapphire 3<br>OxyLight* | Pass |
{7}------------------------------------------------
ISO 14971:2012 – Medical devices – Application of risk management to medical devices. Clinical testing was not performed with this device.
*Sapphire 3 OxyLight is the same test article. Sapphire 3 is a trademark brand name that RAJA Trading Company, Inc. uses for its class of skin care devices.
#### CONCLUSION
The OxyLight is similar to the predicate devices in indications for use, principle of operation and technological characteristics. Differences do not introduce new issues of safety and effectiveness and are thus substantially equivalent to the predicate. Non-clinical tests (IEC 60601-1:2005 + CORR. 2:2007 + AM1:2012, and EN 60601-1-2:2015) demonstrate that the proposed device is as safe and perform as well as the predicate devices. The proposed device is Substantially Equivalent (SE) to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.