K201466 · Clinical Laserthermia Systems, AB · GEX · Mar 26, 2021 · General, Plastic Surgery
Device Facts
Record ID
K201466
Device Name
Tranberg CLS Laser Applicator
Applicant
Clinical Laserthermia Systems, AB
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TRANBERG® Laser Applicator is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology, and urology, for wavelengths 980nm through 1064nm.
Device Story
The TRANBERG® Laser Applicator is a sterile, single-use fiber optic device designed to transfer laser energy from a compatible laser unit to target soft tissue. It consists of a 550 μm core fiber (3m or 12m length) with an SMA 905 connector, a 0.22 numerical aperture, and an introducer system (stylet and catheter with fiber lock). Used in various surgical specialties, the device facilitates interstitial laser therapy to necrotize or coagulate tissue. The clinician operates the device by inserting the applicator into the target site; laser energy is delivered in continuous wave mode to induce thermal effects. The device is identical in construction and design to the predicate, ensuring equivalent performance in clinical settings.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by non-clinical bench testing, including engineering evaluations confirming compliance with applicable ISO, IEC, and FDA safety and performance standards.
Technological Characteristics
Fiber optic laser applicator; 550 μm core diameter; 0.22 numerical aperture; 3m and 12m lengths; SMA 905 proximal connector; 1.55 mm tip diameter; 1-25 mm diffusing region. Operates in continuous wave mode for 980nm-1064nm wavelengths. Delivered sterile for single use; includes introducer stylet and catheter with fiber lock.
Indications for Use
Indicated for soft tissue necrosis or coagulation via interstitial irradiation or thermal therapy. Applicable to cardiovascular thoracic (excluding heart/pericardial vessels), dermatology, ENT, gastroenterology, general, gynecology, head and neck, neurosurgery, plastic surgery, pulmonology, radiology, and urology. For use with 980nm-1064nm laser wavelengths.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Clinical LaserThermia Systems Tranberg® Laser Diffuser Fiber (K163103)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
March 26, 2021
Clinical Laserthermia Systems, AB % David Makanani CEO OMEDtech, LLC 1725 Signal Ridge Drive, Suite 150 Edmond, Oklahoma 73013
Re: K201466
Trade/Device Name: Tranberg CLS Laser Applicator Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: February 22, 2021 Received: February 26, 2021
Dear David Makanani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indication for Use
510(k) Number: K201466
Device Name: TRANBERG에 Laser Applicator
Indications for Use:
The TRANBERGOS| Laser Applicator is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, earnose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology, for wavelengths 980nm through 1064nm.
Prescription Use: X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use: NO (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
510(k)
{3}------------------------------------------------
## 510(K) SUMMARY
| Date | May 23, 2020 |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER | Lars-Erik Eriksson, CEO<br>Clinical Laserthermia Systems, AB<br>Scheelevagen 2<br>Lund, Sweden 22381 |
| CONTACT PERSON | Lars-Erik Eriksson, CEO<br>Clinical Laserthermia Systems, AB<br>Scheelevagen 2<br>Lund, Sweden 22381<br>Tel: +4646152100<br>Email: lee@clinicallaser.se |
| DEVICE NAME<br>Classification<br>Trade Name<br>Common Name<br>Classification<br>Product Code<br>Review Panel | Class II<br>TRANBERG® Laser Applicator<br>TRANBERG® Laser Applicator<br>21 CFR 878.4810<br>GEX - Powered Laser Surgical Instrument<br>General and Plastic Surgery |
| PREDICATE DEVICE: | K163103, Clinical LaserThermia Systems Tranberg®<br>Laser Diffuser Fiber |
| INTENDED USE: | The TRANBERG® Laser Applicator is indicated for use<br>to necrotize or coagulate soft tissue through<br>interstitial irradiation or thermal therapy in medicine<br>and surgery in cardiovascular thoracic surgery<br>(excluding the heart and the vessels in the pericardial<br>sac), dermatology, ear-nose-throat surgery,<br>gastroenterology, general surgery, gynecology, head<br>and neck surgery, neurosurgery, plastic surgery,<br>pulmonology, radiology, and urology, for<br>wavelengths 980nm through 1064nm. |
#### DEVICE DESCRIPTION:
The TRANBERG® Laser Applicator is used to transfer laser energy from the laser unit to the location for the treatment.
Clinical Laserthermia Systems, AB ● 223 81 Lund, Sweden
{4}------------------------------------------------
The laser Applicator is designed with a core of 550 μm. The fiber length is 3 and 12m and it has a standard connector SMA 905 to fit the laser unit. The numerical aperture is at 0.22.
The Laser Applicator is used with an introducer and both (fiber and introducer), are delivered sterile and for single use only. The introducer consists of an introducer stylet and introducer catheter with a fiber lock
#### TECHNOLOGICAL CHARACTERISTIC AND SUBSTANTIAL EQUIVALENCE:
This device is identical in construction and design as the predicate device.
The following table provides more detailed information regarding the basis for the determination of substantial equivalence:
| Parameter | Predicate | Device |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Laser Fiber | | |
| Product name | Tranberg® Laser Diffuser Fiber | Tranberg® Laser Applicator |
| Manufacturer | Clinical LaserThermia Systems CLS,<br>Sweden | Clinical LaserThermia Systems CLS,<br>Sweden |
| Indications for use | The Tranberg® Laser fiber is indicated for<br>use to necrotize or coagulate soft tissue<br>through interstitial irradiation or thermal<br>therapy in medicine and surgery in<br>cardiovascular thoracic surgery (excluding<br>the heart and the vessels in the<br>pericardial sac), dermatology, ear-nose-<br>throat surgery, gastroenterology, general<br>surgery, gynecology, head and neck<br>surgery, neurosurgery, plastic surgery,<br>pulmonology, radiology, and urology, for<br>wavelengths 980nm through 1064nm | The Tranberg® Laser Applicator is<br>indicated for use to necrotize or<br>coagulate soft tissue through interstitial<br>irradiation or thermal therapy in<br>medicine and surgery in cardiovascular<br>thoracic surgery (excluding the heart and<br>the vessels in the pericardial sac),<br>dermatology, ear-nose-throat surgery,<br>gastroenterology, general surgery,<br>gynecology, head and neck surgery,<br>neurosurgery, plastic surgery,<br>pulmonology, radiology, and urology, for<br>wavelengths 980nm through 1064nm |
| Device Regulatory<br>Classification | Accessory to powered surgical laser<br>instrument<br>FDA 878.4810 | Accessory to powered surgical laser<br>instrument<br>FDA 878.4810 |
| Product Code | GEX | GEX |
Clinical Laserthermia Systems, AB ● 223 81 Lund, Sweden
{5}------------------------------------------------
| Parameter | Predicate | Device |
|----------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|
| Device Class | Accessory to powered surgical laser instrument<br>Class 2 | Accessory to powered surgical laser instrument<br>Class 2 |
| 510(k) number | K163013 | To be obtained |
| Fiber core diameter: | 550 μm | 550 μm |
| Numerical aperture: | 0.22 | 0.22 |
| Fiber length: | 3m, 12m standard | 3m, 12m standard |
| Proximal connector: | SMA 905 | SMA 905 |
| Wavelength: | 980nm-1064 nm | 980nm-1064 nm |
| Laser operation mode: | Continuous Wave | Continuous Wave |
| Diffusing region length: | 10-15 mm | 1-25 mm |
| Diffusing tip assembly diameter: | 1.55 mm | 1.55 mm |
| Lesion Shape: | Elliptical shape | Elliptical or Round shape |
## PERFORMANCE TESTING - (NON-CLINICAL) BENCH
The Tranberg® Laser Applicator has been determined through engineering testing to support substantial equivalence with this device and the predicate. This testing showed the Tranberg® Laser Applicator to meet applicable ISO, IEC and FDA safety and performance standards.
### PERFORMANCE TESTING - CLINICAL
There are no clinical data submitted with this Notification.
#### CONCLUSION:
Based on the results of non-clinical testing, the TRANBERG® Laser Applicator performs according to specifications, and as intended, and the comparative discussion of intended use, principle of operation, and technological characteristics, has determined that the Tranberg® Laser Applicator is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.