K163103 · Clinical Laserthermia Systems, AB · GEX · Mar 23, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163103
Device Name
Tranberg CLS Diffusor Laser Fiber
Applicant
Clinical Laserthermia Systems, AB
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 23, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TRANBERGCLS Diffusor Laser Fiber is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology, and urology, for wavelengths 980nm through 1064nm.
Device Story
The TRANBERG CLS Diffusor Laser Fiber is a sterile, single-use accessory for powered surgical laser instruments. It transfers laser energy from a laser unit to a target tissue site for interstitial thermal therapy. The device features a 550 µm core, 0.22 numerical aperture, and a 15 mm diffusing tip designed to create cylindrical lesions. It connects to laser units via a standard SMA 905 connector. Used by clinicians in various surgical specialties, the fiber enables precise tissue coagulation or necrosis. The device is operated within a clinical setting; the provider monitors the laser application to achieve the desired therapeutic effect, benefiting the patient through minimally invasive thermal tissue treatment.
Clinical Evidence
No clinical data submitted. Evidence consists of non-clinical bench testing, including engineering comparative temperature testing in ex vivo bovine cardiac tissue (37°C) to evaluate lesion size, carbonization, and fiber integrity. Biocompatibility (ISO 10993-1, -5, -10) and sterilization/shelf-life (ISO 11135-1, 10993-7, 11607-1/2) testing were performed.
Technological Characteristics
Accessory to powered surgical laser; 550 µm core diameter; 0.22 numerical aperture; 3-12m length; SMA 905 connector; 15 mm diffusing tip; 1.55 mm tip assembly diameter. Supports continuous wave laser operation at 980-1064 nm. Sterilized via Ethylene Oxide (ISO 11135-1).
Indications for Use
Indicated for soft tissue necrosis or coagulation via interstitial irradiation or thermal therapy in patients undergoing cardiovascular thoracic (excluding heart/vessels in pericardial sac), dermatological, ENT, gastrointestinal, general, gynecological, head/neck, neurosurgical, plastic, pulmonary, radiological, or urological procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Clinical LaserThermia Systems TranbergCLS Laser Fiber (K151569)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other. The profiles are stylized and appear to be connected, forming a single, flowing shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 23, 2017
Clinical Laserthermia Systems AB % Mr. David Makanani OMEDtech, LLC 1725 Signal Ridge Drive Suite 150 Edmond, Oklahoma 73013
Re: K163103
Trade/Device Name: Tranberg CLS Diffusor Laser Fiber Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: February 6, 2017 Received: February 21, 2017
Dear Mr. Makanani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163103
Device Name The TRANBERG CLS/Diffusor Laser Fiber
#### Indications for Use (Describe)
The TRANBERG CLS/Diffusor Laser fiber is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery. neurosurgery, pulmonology, radiology, and urology, for wavelengths 980mm through 1064nm.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Date | March 22, 2017 |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER | Lars-Erik Eriksson, CEO<br>Clinical Laserthermia Systems, AB<br>Scheelevagen 2<br>Lund, Sweden 22381 |
| CONTACT PERSON | Lars-Erik Eriksson, CEO<br>Clinical Laserthermia Systems, AB<br>Scheelevagen 2<br>Lund, Sweden 22381<br>Tel: +4646152100<br>Email: lee@clinicallaser.se |
| DEVICE NAME | |
| Classification | Class II |
| Trade Name | TRANBERGCLS Diffusor Laser fiber |
| Common Name | TRANBERGCLS Diffusor Laser fiber |
| Classification | 21 CFR 878.4810 |
| Product Code | GEX - Powered Laser Surgical Instrument |
| Review Panel | General and Plastic Surgery |
| PREDICATE DEVICE: | K151569, Clinical LaserThermia Systems TranbergCLS<br>Laser Fiber |
| INTENDED USE: | The TRANBERGCLS Diffusor Laser Fiber is indicated for<br>use to necrotize or coagulate soft tissue through<br>interstitial irradiation or thermal therapy in medicine<br>and surgery in cardiovascular thoracic surgery<br>(excluding the heart and the vessels in the pericardial<br>sac), dermatology, ear-nose-throat surgery,<br>gastroenterology, general surgery, gynecology, head<br>and neck surgery, neurosurgery, plastic surgery,<br>pulmonology, radiology, and urology, for wavelengths<br>980nm through 1064nm. |
# DEVICE DESCRIPTION:
The TRANBERG45| Diffusor Laser fiber is used to transfer laser energy from the laser unit to the location for the treatment.
Clinical Laserthermia Systems, AB ● 223 81 Lund, Sweden
{4}------------------------------------------------
The diffusor laser fiber is designed with a core of 550 µm. The fiber length is 3 to 12m and it has a standard connector SMA 905 to fit the laser unit. The numerical aperture is at 0.22.
The TRANBERGCS|Diffusor Laser fiber is delivered sterile and for single use only.
# TECHNOLOGICAL CHARACTERISTIC AND SUBSTANTIAL EQUIVALENCE:
The following table provides more detailed information regarding the basis for the determination of substantial equivalence:
| Parameter | | |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Laser Fiber | Image: Laser Fiber | Image: Laser Fiber |
| Product name | Tranberg CLS Laser fiber | Tranberg CLS Diffusor Laser Fiber |
| Manufacturer | Clinical LaserThermia Systems CLS,<br>Sweden | Clinical LaserThermia Systems CLS,<br>Sweden |
| Indications for use | The Tranberg CLS Laser fiber is indicated<br>for use to necrotize or coagulate soft<br>tissue through interstitial irradiation or<br>thermal therapy in medicine and surgery<br>in cardiovascular thoracic surgery<br>(excluding the heart and the vessels in<br>the pericardial sac), dermatology, ear-<br>nose-throat surgery, gastroenterology,<br>general surgery, gynecology, head and<br>neck surgery, neurosurgery, plastic<br>surgery, pulmonology, radiology, and<br>urology, for wavelengths 800nm through<br>1064nm. | The Tranberg CLS Laser fiber is indicated<br>for use to necrotize or coagulate soft<br>tissue through interstitial irradiation or<br>thermal therapy in medicine and surgery<br>in cardiovascular thoracic surgery<br>(excluding the heart and the vessels in<br>the pericardial sac), dermatology, ear-<br>nose-throat surgery, gastroenterology,<br>general surgery, gynecology, head and<br>neck surgery, neurosurgery, plastic<br>surgery, pulmonology, radiology, and<br>urology, for wavelengths 980nm through<br>1064nm. |
| Device Regulatory | Accessory to powered surgical laser<br>instrument | Accessory to powered surgical laser<br>instrument |
| Classification | FDA 878.4810 | FDA 878.4810 |
| Product Code | GEX | GEX |
{5}------------------------------------------------
| Parameter | | |
|-------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|
| Device Class | Accessory to powered surgical laser<br>instrument<br>Class 2 | Accessory to powered surgical laser<br>instrument<br>Class 2 |
| 510(k) number | K151569 | To be obtained |
| Fiber core diameter: | 550 μm | 550 μm |
| Numerical aperture: | 0.22 | 0.22 |
| Fiber length: | 3 m | 3 - 12m |
| Proximal connector: | SMA 905 | SMA 905 |
| Wavelength: | 1064 nm | 1064 nm |
| Laser operation mode: | Continuous Wave | Continuous Wave |
| Diffusing region length: | 1 mm | 15 mm |
| Diffusing tip assembly<br>diameter: | 1.55 mm | 1.55 mm |
| Lesion Shape: | Ring shape | Cylindrical shape |
| Max power: | 8 W for 550 μm | 20 W for 550 μm |
| Lesion volume | 0.8 cm³ at 1min/8W;<br>2.7 cm³ at 2min/8W | 4.6 cm³ at 2min/15W |
# PERFORMANCE TESTING - (NON-CLINICAL) BENCH
The Tranberg 65 Diffusor Laser fiber has been determined through engineering bench testing to support substantial equivalence with this device and the predicates. This testing showed the Tranberg 65 Diffusor Laser fiber to meet applicable ISO, IEC and FDA safety and performance standards.
Non-clinical bench testing showed the device performed to specification and included:
- Engineering comparative temperature testing
Performance testing of the fiber was performed in an ex vivo model using bovine cardiac tissue kept at 37 ℃. At a wavelength of 1064 nm using different power settings and exposure times the following parameters where studied:
- o Lesions size
- o Occurrence of carbonization
- o Integrity of the laser fiber
- Biocompatibility Testing
- o ISO 10993-1, Biological Evaluation of Medical Devices
- ISO 10993-5, Tests for in vitro cytotoxicity o
- ISO 10993-10, Tests for irritation and skin sensitization o
- Sterilization/Shelf Life Testing
- o ISO 11135-1, Sterilization of Health-Care Products: Ethylene Oxide
{6}------------------------------------------------
- ISO 10993-7, Biological Evaluation of Medical Devices: Ethylene Oxide O
- O ISO 11607-1, Packaging for Terminally Sterilized Medical Devices: Requirements
- ISO 11607-2, Packaging for Terminally Sterilized Medical Devices: Validation O
### PERFORMANCE TESTING - CLINICAL
There are no clinical data submitted with this Notification.
#### CONCLUSION:
Based on the results of non-clinical testing, the TRANBERGCSS| Diffusor Laser Fiber performs according to specifications, and as intended, and the comparative discussion of intended use, principle of operation, and technological characteristics, has determined that the Tranberg 65 Diffusor Laser fiber is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.