The sponsor utilized a literature review of clinical investigations to determine safe and effective default treatment parameters for the Frax 1550 laser, ensuring equivalence to consensus recommendations and predicate device settings.
Literature review; Clinical consensus; Treatment parameters
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Ellipse literature study (RA18704); Literature review
—
Not applicable for this study
Safe and effective treatment parameters for non-ablative skin resurfacing
Indications for Use
The Frax 1550 nm laser is indicated for use in dermatological procedures requiring the coagulation of soft tissue, as well as for skin resurfacing procedures.
Device Story
The Ellipse Nordlys is a multi-application system utilizing Intense Pulsed Light (IPL), Nd:YAG laser (1064 nm), and the Frax 1550 nm laser. The system consists of a console with power unit, control electronics, and a display panel. The Frax 1550 handpiece delivers fractional laser energy to the skin for non-ablative resurfacing and soft tissue coagulation. The device is operated by healthcare professionals in a clinical setting. The provider selects treatment parameters via the console interface; these settings are based on established clinical consensus and predicate device standards. The energy delivery induces controlled thermal effects in the skin, intended to improve skin texture and appearance. The system is designed to be safe and effective for dermatological applications, with performance verified through bench testing and comparison to existing cleared fractional laser technologies.
Clinical Evidence
Bench testing only. Compliance with performance standards (IEC 60601-1, IEC 60601-2-57, IEC 60825-1, IEC 60601-2-22) and FDA 21 CFR 1040.10/1040.11 was demonstrated. Default treatment settings were validated against clinical literature and standard parameters used by predicate devices.
Technological Characteristics
System includes console with power unit and control electronics. Emits 1550 nm fractional laser energy. Complies with IEC 60601-1 (3rd ed), IEC 60601-2-57, IEC 60825-1, and IEC 60601-2-22. Manufactured under ISO 13485 and 21 CFR 820 quality systems.
Indications for Use
Indicated for dermatological procedures requiring soft tissue coagulation and skin resurfacing.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of a stylized eagle with three human profiles incorporated into its design.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 16, 2016
Ellipse A/S Ole Kofod QA/RA Manager Agern Alle 11 Hoersholm, DK-2970 DK
Re: K161162
Trade/Device Name: Frax 1550 For Ellipse Nordlys/ Ellipse Nordlys+, Ellipse Sirius, Ellipse Infinity, Ellipse Mjolner, Ellipse Nordlys Ultra Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX Dated: August 15, 2016 Received: August 18, 2016
Dear Ole Kofod:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161162
Device Name
Ellipse Nordlys
Indications for Use (Describe)
Ellipse Nordlys system is intended to be used in dermatology, as tabled below:
• Permanent Hair Reduction (defined as the long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime) (overall 600-950nm).
- · Treatment of Telangiectasias (530-750nm or 555-950mm).
- · Treatment of Port Wine Stains (530-750nm or 555-950nm).
· Treatment of Benign Pigmented Lesions (eg Mottled Pigmentation, Ephilides) and Benign Vascular Lesions (ex Diffuse Redness) (530-750nm or 555-950nm).
- · Treatment of Rosacea (530-750nm or 555-950nm).
- · Treatment of Poikiloderma of Civatte (530-750nm or 555-950mm).
- · Treatment of Benign Epidermal Pigmented Lesions (eg Lentigo Solaris) (400-720nm).
- · Treatment of Inflammatory Acne Vulgaris (530-750mm).
Using Nd:YAG Laser, (1064 nm):
- · Treatment of Leg vessels (0.1 -3.0 mm diameter).
- · Treatment of Benign Vascular Lesions.
- Treatment of Venous Lakes.
- · Treatment of Port Wine Stains.
- · Treatment for Clear Nail defined as: Temporary increase of clear nail in patients with onychomycosis (e.g.,
dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
- · Treatment of benign cutaneous lesions, such as warts.
- · Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue), including:
- Matrixectomy.
- Periungal and subungal warts
- Plantar warts.
Using Frax 1550 Laser, (1550 nm):
· The Frax 1550 nm laser is indicated for use in dermatological procedures requiring the coagulation of soft tissue, as well as for skin resurfacing procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary - K161162
#### Ellipse Frax 1550 for Nordlys
This 510(k) summary is submitted in accordance with the requirements of 21 CFR Part 807.87(h) and Part 807.92.
#### Submitter Information:
Date of the summary: 13 September 2016
Submitted by/manufacturer:
Ellipse A/S (Establishment Registration no. 3005112495) Agern Alle 11 2970 Hoersholm, Denmark Tel: + 45 4576 8808 Fax: + 45 4517 6851
Contact person:
Ole Kofod
#### Device Identification:
Device Trade Name 1:
Device Model number 1:
Alternative Device Trade Name and Model:
Common Name: Classification name:
Device classification: Product code:
#### Predicate Devices:
Predicate devices (for other than Frax 1550) legally marketed to which Ellipse A/S claims substantial equivalence:
Predicate devices (for Frax 1550) legally marketed to which Ellipse A/S claims substantial equivalence:
Frax 1550 for Ellipse Nordlys (with IPL and Nd: YAG hand pieces/applicators). 9APP7829 for 9SYS7751.
Ellipse Sirius system 9SYS7773 Ellipse Infinity system 9SYS7774 Ellipse Mjølner system 9S YS7775 Ellipse Nordlys+ system 9SYS7776 Ellipse Nordlys ultra system 9SYS7777
Intense Pulsed Light (IPL) & Laser. Laser surgical instrument for use in general and plastic surgery and in dermatology (per 21 CFR Part 878.4810). Class II (per 21 CFR 870.1250). GEX
Ellipse Nordlys (K150907) Ellipse A/S Agern Alle 11 DK-2970-Hørsholm, Denmark
Palomar Icon Aesthetic System (K142376) Cynosure Inc. 5 Carlisle Road Westford, MA 01886, USA
Fraxel Dual (K130193) Solta medical, Inc. 25881 Industrial Blvd. Hayward, CA 94545, USA
{5}------------------------------------------------
Lutronic Mosaic (K080932) Lutronic Coorporation #403-2,3,4, Ilsan Technotown 1141-1Baeksok-Dong, Ilsan-Gu Goyang-Si, Gyeonggi-Do, 410-722 Republic of Korea
Lumenis ResurFX (K130028) Lumenis Ltd. 31 Haavoda St. Binyamina, IL 30500, USA
#### Description of Ellipse Nordlys with associated applicators (hand pieces):
Ellipse Nordlys is an Intense Pulsed Light (IPL) and Laser system (Nd:YAG 1064 nm and Fractional 1550 nm) used for long-term removal of unwanted hair; for treatment of sun-damaged skin, including uneven pigmentation, age spots, large pores, diffuse redness, and for the treatment of telangiectasias, port wine stains, rosacea, poikiloderma of civatte and inflammatory acne in the area of dermatology.
The systems consist of a console containing power unit and control electronics with control and display panel including software.
IPL Applicators/hand-pieces are connected to the systems in order to generate light energy for treatment in the waveband 400 nm - 950 nm.
Additionally a Nd:Y AG hand piece (1064 nm) and a Fractional (1550 mm) hand piece can be connected to the Ellipse Nordlys for treatment of: Vascular lesions as stated below under Intended Use/Indications for Use for the Nd: YAG and fractional treatments for non-ablative skin resurfacing for Frax 1550 as stated below.
#### Intended Use/Indications for Use:
Ellipse Nordlys system is intended to be used in dermatology, as tabled below:
- Permanent Hair Reduction (defined as the long-term, stable reduction in the number of hairs re-growing when ● measured at 6, 9, and 12 months after the completion of a treatment regime) (overall 600-950nm).
- . Treatment of Telangiectasias (530-750nm or 555-950nm).
- 0 Treatment of Port Wine Stains (530-750nm or 555-950mm).
- Treatment of Benign Pigmented Lesions (eg Mottled Pigmentation, Ephilides) and Benign Vascular Lesions 0 (eg Diffuse Redness) (530-750nm or 555-950nm).
- 0 Treatment of Rosacea (530-750nm or 555-950nm).
- . Treatment of Poikiloderma of Civatte (530-750nm or 555-950nm).
- . Treatment of Benign Epidermal Pigmented Lesions (eg Lentigo Solaris) (400-720nm).
- Treatment of Inflammatory Acne Vulgaris (530-750nm). �
#### Using Nd: YAG Laser, (1064 nm):
- Treatment of Leg vessels (0.1 -3.0mm diameter).
- . Treatment of Benign Vascular Lesions.
- @ Treatment of Venous Lakes.
- 0 Treatment of Port Wine Stains.
- Treatment for Clear Nail defined as: Temporary increase of clear nail in patients with onychomycosis (e.g., 0 dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
- 0 Treatment of benign cutaneous lesions, such as warts.
- Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue), including:
- Matrixectomy.
- Periungal and subungal warts
- Plantar warts.
#### Using Frax 1550 Laser, (1550 nm):
- The Frax 1550 nm Laser is indicated for use in dermatological procedures requiring the coagulation of soft 0 tissue, as well as for skin resurfacing procedures.
{6}------------------------------------------------
#### Performance Standards utilized
The Ellipse Nordlys Intense Pulsed Light (IPL) and Laser system has been tested according to and comply with:
- US FDA 21 CFR 1040.10 and 1040.11 for class IV Laser Products. .
- IEC 60601-1 3rd edition, UL 60601-1 and CSA C22.2 No. 601.1. .
- IEC 60601-2-57. .
- IEC 60825-1 and IEC 60601-2-22. .
- IEC 60601-1-2. .
- 0 Complies with the European Medical Device Directive 93/42/EEC (Annex II).
- 0 Manufactured under ISO13485 Quality Management System certified by Presafe/DGM and QMI and also complies with the US FDA 21CFR Part 820.
#### Performance Data
Testing certificates by UL International and Delta according to above listed Performance Standards were provided which demonstrates that Ellipse Nordlys with the listed applicators/handpieces utilize technical parameters which are as safe and effective as the predicate devices.
{7}------------------------------------------------
#### Substantial Equivalence conclusion:
The Ellipse Nordlys system as regards its IPL and Nd: Y AG applications is identical to the previously cleared Ellinse Nordlys system (K150907).
An Ellipse Frax 1550 hand piece has now been added to the system (included in this application K16162). The Ellipse Frax 1550 hand piece has been evaluated and compared to the 1550mm hand piece of the predicate systems Cynosure/Palomar Icon® 1540 mm (K142376), Fraxel Dual® 1550 nm hand piece (K130193), Lumenis ResurFX (K130028) and Lutronic Mosaic (K080932).
Ellipse Frox 1550 is found to be substantial equivalent to these with respect to clinical and biological issues of these systems.
The default start settings for the Frax 1550 are based on the most relevant clinical investigations found to be safe and effective in the Ellipse literature study (RA18704, appendix 2) and equivalent to what are used as standard settings for the same treatments for Frax SR1550 (K130193) and Icon Lux1540 (K142376).
It is hereby proven that the chosen default parameters for Ellipse Frax1550 both are equivalent to settings already shown to be safe an effective by FDA cleared equipment (predicate devices) as well as being inside the mentioned consensus parameter recommendations. Furthermore, it is proven that the possible maximum energy settings for the individual treatments are inside what have already been considered to be safe and effective.
The new, added fractional application, Non-ablative Skin Resurfacing, is found to be substantially equivalent to the same application of the mentioned predicate devices.
Based on this analysis of the overall performance characteristics of the mentioned predicate devices Ellipse A/S believes that no significant differences exist. The Ellinse Nordlys system, now with the Ellipse Frax 1550 Laser applicator/hand piece added, should not raise new issues of safety and effectiveness and are judged to be substantially equivalent to the mentioned predicate devices.
Dee Roford
(Signature)
Ole Kofod (Typed Name)
13-September-2016 (Date)
K161162 (Premarket Notification 510(K) number).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.