Ellipse Nordlys system is intended to be used in dermatology, as tabled below: •Permanent Hair Reduction (defined as the long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime) (overall 600-950nm). ·Treatment of Telangiectasias (530-750nm or 555-950mm). · Treatment of Port Wine Stains (530-750nm or 555-950nm). · Treatment of Benign Pigmented Lesions (eg Mottled Pigmentation, Ephilides) and Benign Vascular Lesions (ex Diffuse Redness) (530-750nm or 555-950nm). · Treatment of Rosacea (530-750nm or 555-950mm). · Treatment of Poikiloderma of Civatte (530-750nm or 555-950nm). - · Treatment of Benign Epidermal Pigmented Lesions (eg Lentigo Solaris) (400-720nm). · Treatment of Inflammatory Acne Vulgaris (530-750mm). · Using Nd: Y AG laser, (1064nm) (Ellipse MultiFlex+ system, only): - · Treatment of Leg vessels (0.1 -3.0mm diameter). - · Treatment of Benign Vascular Lesions - •Treatment of Venous Lakes •Treatment of Port Wine Stains • Treatment for Clear Nail defined as: Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). · Treatment of benign cutaneous lesions, such as warts. - •Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue), including - Matrixectomy. - Periungal and subungal warts - Plantar warts.
Device Story
System comprises console with power unit, control electronics, and display panel; connects to IPL applicators (400-950nm) or Nd:YAG laser handpiece (1064nm). Operated by clinicians in dermatology/podiatry settings. Device delivers light energy to target tissues; thermal effects achieve hair reduction, lesion treatment, or soft tissue ablation. Clinician selects parameters via control panel; output affects tissue via photothermolysis or vaporization. Benefits include non-invasive or minimally invasive treatment of dermatological and podiatric conditions.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-2-57, IEC 60825-1, IEC 60601-2-22, and 21 CFR 1040.10/1040.11 standards.
Technological Characteristics
Intense Pulsed Light (IPL) and Nd:YAG laser system. IPL waveband 400-950nm; Nd:YAG 1064nm. Console-based with control electronics and software interface. Complies with IEC 60601-1 (3rd ed), IEC 60601-2-57, IEC 60825-1, and IEC 60601-2-22. Manufactured under ISO 13485 and 21 CFR 820.
Indications for Use
Indicated for dermatology patients requiring hair reduction, treatment of vascular/pigmented lesions, rosacea, poikiloderma of Civatte, acne vulgaris, onychomycosis, and podiatric soft tissue procedures (ablation, incision, excision, coagulation).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle are three stylized human profiles facing to the right, with flowing lines representing hair or ribbons.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 7, 2015
Ellipse A/S Ole Kofod Quality Assurance and Regulatory Affairs Manager Agern Alle 11 DK-2970 Hoersholm Denmark
Re: K150907
Trade/Device Name: Ellipse Nordlys Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 8, 2015 Received: June 17, 2015
Dear Mr. Kofod:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150907
Device Name Ellipse Nordlys
Indications for Use (Describe)
Ellipse Nordlys system is intended to be used in dermatology, as tabled below:
•Permanent Hair Reduction (defined as the long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime) (overall 600-950nm).
·Treatment of Telangiectasias (530-750nm or 555-950mm).
· Treatment of Port Wine Stains (530-750nm or 555-950nm).
· Treatment of Benign Pigmented Lesions (eg Mottled Pigmentation, Ephilides) and Benign Vascular Lesions (ex Diffuse Redness) (530-750nm or 555-950nm).
· Treatment of Rosacea (530-750nm or 555-950mm).
· Treatment of Poikiloderma of Civatte (530-750nm or 555-950nm).
- · Treatment of Benign Epidermal Pigmented Lesions (eg Lentigo Solaris) (400-720nm).
· Treatment of Inflammatory Acne Vulgaris (530-750mm).
· Using Nd: Y AG laser, (1064nm) (Ellipse MultiFlex+ system, only):
- · Treatment of Leg vessels (0.1 -3.0mm diameter).
- · Treatment of Benign Vascular Lesions
- •Treatment of Venous Lakes
·Treatment of Port Wine Stains
• Treatment for Clear Nail defined as: Temporary increase of clear nail in patients with onychomycosis (e.g.,
dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
· Treatment of benign cutaneous lesions, such as warts.
- •Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue), including
- Matrixectomy.
- Periungal and subungal warts
- Plantar warts.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary – Section 3
## Ellipse Nordlys
This 510(k) summary is submitted in accordance with the requirements of 21 CFR Part 807.87(h) and Part 807.92.
## Submitter Information:
| Date of the summary: | 30 March 2015 |
|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by/manufacturer: | Ellipse A/S (Establishment Registration no. 3005112495)<br>Agern Alle 11<br>2970 Hoersholm, Denmark<br>Tel: +45 4576 8808<br>Fax: + 45 4517 6851 |
| Contact person: | Ole Kofod |
| <b>Device Identification:</b> | |
| Device Trade Name 1: | Ellipse Nordlys (with IPL and Nd:YAG hand pieces/applicators). |
| Device Model number 1: | 9SYS7751. |
| Alternative Device Trade Name and Model: | Ellipse Sirius 9SYS7773<br>Ellipse Infinity 9SYS7774<br>Ellipse Mjølner 9SYS7775<br>Ellipse Nordlys+ 9SYS7776<br>Ellipse Nordlys ultra 9SYS7777 |
| Common Name: | Intense Pulsed Light (IPL) & Laser. |
| Classification name: | Laser surgical instrument for use in general and plastic surgery and in dermatology (per 21 CFR Part 878.4810). |
| Device classification: | Class II |
| Product code: | GEX |
| <b>Predicate Devices:</b> | |
| Predicate devices legally marketed to which Ellipse A/S claims substantial equivalence: | Ellipse MultiFlex+ <sup>TM</sup> (K140670).<br>Ellipse A/S<br>Agern Alle 11, DK-2970 Hoersholm, Denmark.<br>(Laser surgical instrument for use in general and plastic surgery and in dermatology (per 21 CFR Part 878.4810)). |
| | Genesisplus Laser System <sup>TM</sup> (K122493)<br>Cutera, Inc., 3240 Bayshore Blvd., Brisbane, CA94005, USA.<br>(Laser surgical instrument for use in general and plastic surgery and in dermatology (per 21 CFR Part 878.4810)). |
| | Candela V-beam <sup>TM</sup> (K033461)<br>Candela Corporation, 530 Boston Post Rd., Wayland, MA01778, USA.<br>(Laser surgical instrument for use in general and plastic surgery and in dermatology (per 21 CFR Part 878.4810)). |
{5}------------------------------------------------
## Description of Ellinse Nordlys:
Ellipse Nordlys is an Intense Pulsed Light (IPL) and Nd:YAG Laser system used for long-term removal of unwanted hair; for treatment of sun-damaged skin, including uneven pigmentation, age spots, large pores, diffuse redness, and for the treatment of telangiectasias, port wine stains, rosacea, poikiloderma of civatte and inflammatory acne in the area of dermatology.
The systems consist of a console containing power unit and control electronics with control and display panel including software.
IPL Applicators/hand-pieces are connected to the systems in order to generate light energy for treatment in the waveband 400 nm - 950 nm.
Additionally a Nd: YAG handpiece (1064 nm) can be connected to the Ellipse Nordlys for treatment of vascular lesions as stated below under Intended Use/Indications for Use.
## Intended Use/Indications for Use:
Ellipse Nordlys system is intended to be used in dermatology, as tabled below:
- Permanent Hair Reduction (defined as the long-term, stable reduction in the number of hairs re-growing when measured ● at 6, 9, and 12 months after the completion of a treatment regime) (overall 600-950nm).
- . Treatment of Telangiectasias (530-750nm or 555-950nm).
- Treatment of Port Wine Stains (530-750nm or 555-950nm). ●
- Treatment of Benign Pigmented Lesions (eg Mottled Pigmentation, Ephilides) and Benign Vascular Lesions (eg Diffise ● Redness) (530-750nm or 555-950nm).
- Treatment of Rosacea (530-750nm or 555-950nm).
- Treatment of Poikiloderma of Civatte (530-750nm or 555-950nm).
- . Treatment of Benign Epidermal Pigmented Lesions (eg Lentigo Solaris) (400-720nm).
- . Treatment of Inflammatory Acne Vulgaris (530-750nm).
- . Using Nd:YAG laser, (1064nm) (Ellipse MultiFlex+ system, only):
- Treatment of Leg vessels (0.1 -3.0mm diameter).
- . Treatment of Benign Vascular Lesions
- Treatment of Venous Lakes
- Treatment of Port Wine Stains
- Treatment for Clear Nail defined as: Temporary increase of clear nail in patients with onychomycosis (e.g., ● dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
- . Treatment of benign cutaneous lesions, such as warts.
- Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue), including ●
- Matrixectomy.
- Periungal and subungal warts
- Plantar warts.
#### Performance Standards
The Ellipse Nordlys Intense Pulsed Light (IPL) and Laser system has been tested according to and comply with:
- US FDA 21 CFR 1040.10 and 1040.11 for class IV Laser Products. ●
- IEC 60601-1 3td edition, UL 60601-1 and CSA C22.2 No. 601.1. ●
- IEC 60601-2-57. ●
- IEC 60825-1 and IEC 60601-2-22. ●
- IEC 60601-1-2. ●
- Complies with the European Medical Device Directive 93/42/EEC (Annex II). .
- Manufactured under ISO13485 Quality Management System certified by Presafe/DGM and QMI and also complies with ● the US FDA 21CFR Part 820.
## Substantial Equivalence conclusion:
The Ellipse Nordlys system are substantially equivalent in terms of technological characteristics, performance, intended use/indications for use to the predicate devices listed on page 1 of this document.
The Ellipse Nordlys has been evaluated and compared to the above mentioned predicate systems and to their application modules. The Ellipse Nordlys as far as the identical modules, applications, parameters, and intended uses/indications are concerned, have been judged to be substantially equivalent to the mentioned predicate devices. Based on this analysis of the overall performance of the mentioned predicate devices Ellipse A/S believes that no significant differences exist. The Ellipse Nordlys system should not raise new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.