K093087 · Litecure, LLC · GEX · May 12, 2010 · General, Plastic Surgery
Device Facts
Record ID
K093087
Device Name
BWF-5 MEDICAL LASER SERIES
Applicant
Litecure, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
May 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The BWF-5 Medical Laser Series (810nm, 930nm, 980nm, 1080nm and 1320nm) are intended for delivery of laser light to soft tissue in the contact or non-contact mode during surgical procedures, including via endoscopes, introducers, or catheters. The BWF-5 Medical Laser Series are generally indicated for incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery and thoracic surgery. The device is specifically indicated for use as follows: LASER 980nm: Indicated for use with the VARI-LASE Procedure Kit in the Endovascular coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux. LASER 1320nm: Indicated for use with the VARI-LASE Procedure Kit in the treatment of reflux of great saphenous veins associated with varicose veins and varicosities.
Device Story
Compact medical laser system; delivers laser light via flexible optical fiber to soft tissue in contact or non-contact modes. Operator enables laser via foot switch; touch-screen panel adjusts output levels. Operates in continuous wave or controlled pulse modes. Used in clinical/surgical settings by physicians for incision, excision, vaporization, ablation, hemostasis, or coagulation. Specific endovascular applications require VARI-LASE Procedure Kit. Output allows precise tissue interaction; benefits include controlled surgical outcomes and minimally invasive vein treatment.
Clinical Evidence
No clinical data provided. Substantial equivalence based on technological characteristics and performance standards compliance.
Indicated for soft tissue incision, excision, vaporization, ablation, hemostasis, or coagulation in various surgical specialties including ENT, dental, arthroscopy, GI, general, derm, plastic, podiatry, urology, GYN, neuro, pulmonary, and thoracic surgery. Specifically indicated for endovascular coagulation of the Greater Saphenous Vein in patients with superficial vein reflux (980nm) and treatment of reflux of great saphenous veins associated with varicose veins (1320nm).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Biolitec, Inc., Ceralas E 980nm Diode Laser (K082263)
QUANTA SYSTEM POLYSURGE DIODE LASER FAMILY (K083613)
BioTex, Inc., PhoTex 15 Diode Laser Series (K060304)
Laserscope, 800 Series Surgical Laser System (K972575)
B&W Tek, Inc., BWF-5 Medical Laser Series (K062363)
Sciton, Inc., Profile Multi-Platform System (K060033)
Biolitec, Inc., 100W CERALAS D 980NM DIODE LASER (K071295)
Sciton, Inc., Profile Multi-Platform System (K070388)
Submission Summary (Full Text)
{0}------------------------------------------------
# MAY 1 2 2010
1693087
----
斗
# 510(k) SUMMARY
The Summary of Safety and Effectiveness on the LiteCure, LLC BWF-5 Medical Laser Series reflects data available and represented at the time the submission was prepared, but caution should be exercised in interpreting the data. The results of future studies and or tests may require alterations of the conclusions or recommendations set forth.
| Applicant | LiteCure, LLC<br>250 Corporate Blvd., Suite B<br>Newark, Delaware 19702<br>Tel: 302-709-0408 |
|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date | April 02, 2010 |
| Correspondent | Liang Lu, Regulatory Affairs Manager |
| Device Name | BWF-5 Medical Laser Series |
| Classification | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology, 21 CFR 878.4810 |
| Predicate Device | K053540 INTERmedic Arfran, S.A., INTERmedic Diode Laser<br>Family 810 nm and 980 nm and the delivery accessories that are used<br>with them to deliver, ultrasound and RF Surgical 15TM, 30TM, 50TM,<br>100TM, 120TM and SR 15 OFTTM, MULTIDIODE PL3DTM,<br>ContrAgeTM, RF ContrAgeTM,MULTIDIODE VARIUSTM, VR1000TM,<br>MULTIDIODE ODONTTM; K082263 Biolitec, Inc., Ceralas E 980nm Diode Laser (Models<br>E15/980, E30/980); K083613 QUANTA SYSTEM POLYSURGE DIODE LASER<br>FAMILY; K060304 BioTex, Inc., PhoTex 15 Diode Laser Series: 810, 940, 980,<br>market clearance date March 21, 2006; and K972575 Laserscope, 800 Series Surgical Laser System Orion<br>Surgical Laser System Angled Delivery Devices (ADD Family<br>Product Line), market clearance date July 17, 1998. K062363 B&W Tek, Inc., BWF-5 Medical Laser Series. K060033 Sciton, Inc., Profile Multi-Platform System. K071295 Biolitec, Inc., 100W CERALAS D 980NM DIODE LASER K070388 Sciton, Inc., Profile Multi-Platform System |
| Intended Use | The BWF-5 Medical Laser Series (810nm, 930nm, 980nm, 1080nm<br>and 1320nm) are intended for delivery of laser light to soft tissue in the<br>contact or non-contact mode during surgical procedures, including via<br>endoscopes, introducers, or catheters. The BWF-5 Medical Laser Series are<br>generally indicated for incision, excision, vaporization, ablation,<br>hemostasis or coagulation of soft tissue in ear, nose and throat and oral<br>surgery (otolaryngology), dental procedures, arthroscopy,<br>gastroenterology, general surgery, dermatology, plastic surgery, podiatry,<br>urology, gynecology, neurosurgery (peripheral nervous system),<br>pulmonary surgery and thoracic surgery. The device is specifically<br>indicated for use as follows: |
| Intended Use<br>(continued) | LASER 980nm:<br>Indicated for use with the VARI-LASE Procedure Kit in the Endovascular<br>coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients<br>with Superficial Vein Reflux.<br><br>LASER 1320nm:<br>Indicated for use with the VARI-LASE Procedure Kit in the treatment of<br>reflux of great saphenous veins associated with varicose veins and<br>varicosities. |
| Description | BWF-5 Medical Laser Series are compact medical laser systems. The laser<br>light delivery system consists of a flexible optical fiber. Activation occurs<br>when the operator enables the laser and presses the foot switch. Release<br>the foot switch to deactivate the laser. Depending on laser system<br>configuration, the foot switch can function as on/off switch. A touch-<br>screen display panel allows the operator to adjust or set laser output level.<br>The laser can operate in continuous wave mode or controlled pulse mode. |
| Technological<br>Characteristics<br>and Substantial<br>Equivalence | The BWF-5 Medical Laser Series is as safe and effective as the predicate<br>devices. The BWF-5 Medical Laser Series has the same intended uses and<br>similar indications, technological characteristics (such as wavelength, laser<br>safety class, etc), and principles of operation as its predicate device. The<br>minor technological differences between the BWF-5 Medical Laser Series<br>and its predicate devices raise no new issues of safety or effectiveness.<br>Thus, the BWF-5 Medical Laser Series is substantially equivalent to its<br>predicate devices. |
| Compliance | 21 C.F.R. § 1040.10 & 1040.11 Performance Standards for Light Emitting Products;<br>IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety,<br>1988; Amendment 1, 1991-11, Amendment 2, 1995 (2nd Edition);<br>IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements for Safety<br>- Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition<br>2:2001 with Amendment 1:2004);<br>IEC 60601-2-22 1995, 2nd Edition, "Medical electrical equipment - Part 2: Particular<br>requirements for the safety of diagnostic and therapeutic laser equipment"<br>IEC 60825-1 Ed. 2.0 (2007), Safety of laser products - Part 1: Equipment classification,<br>and requirements. |
{1}------------------------------------------------
### 510(k) SUMMARY continued
The Summary of Safety and Effectiveness on the LiteCure, LLC BWF-5 Medical Laser Series reflects data available and represented at the time the submission was prepared, but caution should be exercised in interpreting the data. The results of future studies and or tests may require alterations of the conclusions or recommendations set forth.
:
:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The eagle is positioned to the right of a circular seal. The seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
LiteCure, LLC % Intertek Testing Services Mr. Jay Y. Kogoma 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087
MAY 1 2 2010
Re: K093087
Trade/Device Name: BWF-5 Medical Laser Series Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and Plastic surgery and in dermatology Regulatory Class: Class II
Product Code: GEX Dated: May 05, 2010 Received: May 07, 2010
Dear Mr. Kogoma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Mr. Jay Y. Kogoma
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely vours.
Sincerely yours,
Mark A. Milliken
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name
Indications For Use:
The BWF-5 Medical Laser Series (810nm, 930nm, 1080mm and 1320nm) are intended for delivery of laser light to soft tissue in the contact or non-contact mode during surgical procedures, including via endoscopes, introducers, or catheters. The BWF-5 Medical Laser Series are generally indicated for incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery and thoracic surgery.
The device is specifically indicated for use as follows:
#### LASER 980nm:
Indicated for use with the VARI-LASE Procedure Kit in the Endovascular coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.
#### LASER 1320nm:
Indicated for use with the VARI-LASE Procedure Kit in the treatment of reflux of great saphenous veins associated with varicose veins and varicosities.
Over-The-Counter Use Prescription Use AND/OR X (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Oyler for mxm
(Digital Sign Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K093087
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.