K070388 · Sciton, Inc · GEX · May 2, 2007 · General, Plastic Surgery
Device Facts
Record ID
K070388
Device Name
PROFILE MULTI-PLATFORM SYSTEM
Applicant
Sciton, Inc
Product Code
GEX · General, Plastic Surgery
Decision Date
May 2, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1064 nm Indications for Use: Coagulation and hemostasis of benign vascular lesions such as, but not limited to-telangiectasia and rosacea. Removal of unwanted hair, for the stable long term, or permanent hair reduction through selective targeting of melanin in hair follicles, and for the treatment for pseudofolliculitis barbae (PFB). The Profile 1064 Laser Systems and Accessories are indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin. The Profile 1064 Laser Systems and Accessories are indicated for the treatment of facial wrinkles. Incision/excision and cutting, vaporization, ablation, coagulation/hemostasis of soft tissue in the performance of surgical applications. All soft tissue is included, such as skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs and glands. It is further indicated for laser assisted lipolysis. 1320 nm Indications for Use: It is indicated for the treatment of fine lines and wrinkles. It is also indicated for the treatment of back acne, atrophic acne scars and mild to moderate inflammatory acne vulgaris. Profile Multi-Platform System with fiber delivery is indicated for the treatment of reflux of great and small saphenous veins associated with varicose veins and varicosities, and for treatment of incompetence and reflux of superficial veins in the lower extremity. 2940 nm Indications for Use; Skin resurfacing and treatment of wrinkles. Pulsed Light (300-1400 nm) Indications for Use: The Profile Multi-Platform Systems and Accessories are indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery and dermatology. It is intended for use for: The treatment of benign pigmented lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles); (515nm LP filter, 560nm LP filter) The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiornas, poikiloderma of Civatte, leg veins and venous malformations; (560nm LP filter, 590nm LP filter) The removal of unwanted hair from all skin types, and to effect stable long-term, or permanent, hair reduction: (590nm LP filter, 640nm LP filter, 695nm LP filter) Treatment of facial wrinkles, treatment of fine lines and wrinkles; and (590nm LP filter, 640nm LP filter, 695nm LP filter, 800nm LP filter) Topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, it may also help muscle spasms, minor sprains and strains, and minor muscular back pain. (800nm LP filter)
Device Story
Modular multi-wavelength laser/light system; delivers thermal energy via scanning/focusing optics. Components: control console, power supply, cooling system, fiber optic/articulated arm delivery system, handpiece/scanner. Used in clinical settings by physicians (dermatology, plastic/general surgery). Operates via selective photothermolysis or direct tissue ablation/coagulation. Physician visualizes treatment site; adjusts parameters (fluence, spot size, repetition rate) to achieve clinical endpoints (hemostasis, tissue removal, hair reduction). Benefits: minimally invasive surgical/aesthetic outcomes; versatile multi-modality platform.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance similarities to legally marketed predicate devices.
Technological Characteristics
Modular laser/light system; wavelengths include 1064 nm, 1320 nm, 2940 nm, and pulsed light (300-1400 nm). Features fiber optic or articulated arm delivery, integrated cooling, and control console. Energy delivery via scanning/focusing optics. Class II device.
Indications for Use
Indicated for patients requiring soft tissue surgery, dermatological procedures, or aesthetic treatments. Includes treatment of vascular/pigmented lesions, hair reduction (all skin types I-VI), acne, wrinkles, scars, venous reflux, and topical heating for pain relief. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Profile Multi-Platform Laser System and Accessories (K060033)
{0}------------------------------------------------
K070388
## Attachment IV
# 510(k) Summary
MAY - 2 2007
:
.
| Submitter | Sciton, Inc. |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 925 Commercial Street, Palo Alto, CA 94303 |
| Phone: | (650) 493-9155 |
| Fax : | (650) 493-9146 |
| Contact Person: | Jay M. Patel, VP of Regulatory Affairs |
| Date Prepared: | April 10, 2007 |
| Device Trade Name: | Profile Multi-Platform System |
| Common Name: | Laser/Light Powered Surgical Device (and Accessories) |
| Classification Name: | Laser Surgical Instrument, 21 CFR 878.4810. |
| Legally Marketed<br>Predicate Device: | K060033: Profile Multi-Platform Laser System and Accessories<br>K062321: Cynosure SmartLipo Nd:YAG Laser System<br>K062822, K051287: Vascular Solutions Vari-Lase Lasers<br>K062210, K061618: CoolTouch CTEV Nd:YAG Laser Systems<br>K023954: Altus Medical CoolGlide Aesthetic Lasers<br>K003715: CoolTouch Nd:YAG Laser System<br>K050673, K043251, K033461, K033331: Candela Pulsed Dye Laser Systems<br>K053616, K031671: Polaris WR, ST Applicator<br>K0041242: Candela Smoothbeam Laser<br>K051255: Opusmed LumiPhase-R<br>K031425: Light BioScience GentleWaves LED Photomodulation Device<br>K042165: Cutera Titan Tabletop Product<br>K042630: RevLight Skin Care System |
| Description of<br>Profile Multi-Platform<br>System | The Profile Multi-Platform System is a modular, multi-wavelength laser/light system.<br>The system uses scanning and focusing optics to deliver a pattern of thermal<br>energy to the treatment site. The system consists of control console which houses the<br>power supply, cooling system, fiber optic delivery system and/or articulated arm<br>delivery system with handpiece and/or scanner. |
| Intended Use: | 1064 nm Indications for Use: |
| | Coagulation and hemostasis of benign vascular lesions such as, but not limited<br>to-telangiectasia and rosacea.<br>Removal of unwanted hair, for the stable long term, or permanent hair<br>reduction through selective targeting of melanin in hair follicles, and for<br>the treatment for pseudofolliculitis barbae (PFB). The Profile 1064 Laser<br>Systems and Accessories are indicated for use on all skin types<br>(Fitzpatrick I-VI), including tanned skin.<br>The Profile 1064 Laser Systems and Accessories are indicated for the<br>treatment of facial wrinkles.<br>Incision/excision and cutting, vaporization, ablation,<br>coagulation/hemostasis of soft tissue in the performance of surgical<br>applications. All soft tissue is included, such as skin, cutaneous tissue,<br>subcutaneous tissue, striated and smooth tissue, muscle, cartilage<br>meniscus, mucous membrane, lymph vessels and nodes, organs and<br>glands. It is further indicated for laser assisted lipolysis. |
:
{1}------------------------------------------------
### 1320 nm Indications for Use:
It is indicated for the treatment of fine lines and wrinkles. It is also indicated for the treatment of back acne, atrophic acne scars and mild to moderate inflammatory acne vulgaris.
Profile Multi-Platform System with fiber delivery is indicated for the treatment of reflux of great and small saphenous veins associated with varicose veins and varicosities, and for treatment of incompetence and reflux of superficial veins in the lower extremity.
### 2940 nm Indications for Use;
Skin resurfacing and treatment of wrinkles.
### Pulsed Light (300-1400 nm) Indications for Use:
The Profile Multi-Platform Systems and Accessories are indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery and dermatology.
It is intended for use for.
- · The treatment of benign pigmented lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles); (515nm LP filter, 560nm LP filter)
- · The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiornas, poikiloderma of Civatte, leg veins and venous malformations; (560nm LP filter, 590nm LP filter)
- · The removal of unwanted hair from all skin types, and to effect stable long-term, or permanent, hair reduction: (590nm LP filter, 640nm LP filter, 695nm LP filter)
- · Treatment of facial wrinkles, treatment of fine lines and wrinkles; and (590nm LP filter, 640nm LP filter, 695nm LP filter, 800nm LP filter)
- · Topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, it may also help muscle spasms, minor sprains and strains, and minor muscular back pain. (800nm LP filter)
| Technological<br>Characteristics | The Profile Multi-Platform System shares the same indications for use, similar design<br>features (including wavelength, laser/light medium and delivery systems, power<br>supply, cooling and control system), functional features (including power output,<br>repetition rate, energy, spot size and fluence), and is therefore substantially<br>equivalent to the above legally marketed predicate devices. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety and<br>Effectiveness | The indications for use are based upon the indications for use for predicate systems.<br>Technologically, the Profile Multi-Platform System is substantially equivalent to the<br>listed predicate devices. Therefore, the risks and benefits for the Profile Multi-Platform<br>System are comparable to the predicate devices. |
| Conclusion | The Profile Multi-Platform System shares similar indications for use, design features,<br>and similar functional features as, and therefore is substantially equivalent to, the<br>currently marketed predicate devices. |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 6 2008
Sciton, Inc % Mr. Jav M. Patel 925 Commercial Street Palo Alto, California 94303
Re: K070388
Trade/Device Name: Profile Multi-Platform System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery in dermatology Regulatory Class: II Product Code: GEX Dated: February 8, 2007 Received: February 13, 2007
Dear Mr. Patel:
This letter corrects our substantially equivalent letter of May 2, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Jay M. Patel
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Neil R.E. Ogden
for
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
### Attachment III
### Statement of Indications for Use
| 510(k) Number (if known): | K070388 |
|---------------------------|---------|
|---------------------------|---------|
Profile Multi-Platform System Device Name:
Indications for Use:
#### 1064 nm Indications for Use:
Coagulation and hemostasis of benign vascular lesions such as, but not limited to telangiectasia and rosacea.
Removal of unwanted hair, for the stable long term, or permanent hair reduction through selective targeting of melanin in hair follicles, and for the treatment for pseudofolliculitis barbae (PFB). The Profile 1064 Laser Systems and Accessories are indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin. The Profile 1064 Laser Systems and Accessories are indicated for the treatment of facial winkles.
Incision/excision and cutting, vaporization, ablation, coagulation/hemostasis of soft tissue in the performance of surgical applications. All soft tissue is included, such as skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs and glands. It is further indicated for laser assisted lipolysis.
#### 1320 nm Indications for Use:
It is indicated for the treatment of fine lines and wrinkles. It is also indicated for the treatment of back acne, atrophic acne scars and mild to moderate inflammatory acne vulgaris.
Profile Multi-Platform System with fiber delivery is indicated for the treatment of reflux of great and small saphenous veins associated with varicose veins and varicosities, and for treatment of incompetence and reflux of superficial veins in the lower extremity.
### 2940 nm Indications for Use:
Skin resurfacing and treatment of wrinkles.
| Prescription Use<br>(Per 21CFR801) | <b>X</b> |
|------------------------------------|----------|
| OR Over-The-Counter Use | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
Page 1 of 2
1 of 2
**510(k) Number** K070388
{5}------------------------------------------------
### Pulsed Light (300-1400 nm) Indications for Use:
The Profile Multi-Platform Systems and Accessories are indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery and dermatology. It is intended for use for:
- · The treatment of benign pigmented lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles); (515nm LP filter, 560nm LP filter)
- · The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations;
- (560nm LP filter, 590nm LP filter)
- · The removal of unwanted hair from all skin types, and to effect stable long-term, or permanent, hair reduction; and
- (590nm LP filter, 640nm LP filter, 695nm LP filter)
- · Topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, it may also help muscle spasms, minor sprains and strains, and minor muscular back pain. (800nm LP filter)
Nil R.P. Ogle
(Division Sign-Off) for mxm
Division of General, Restorative, and Neurological Devices
**510(k) Number** K070388
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.