K063074 · Candela Corp. · GEX · Dec 27, 2006 · General, Plastic Surgery
Device Facts
Record ID
K063074
Device Name
CANDELA 3630 LASER SYSTEM
Applicant
Candela Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 27, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
755nm The Candela 3630 Laser System is indicated for the temporary hair reduction. Stable long-term or permanent hair reduction through selective targeting of melanin in hair follieles. Premanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. On all skin types (Fitzpatrick I-IV) including tanned skin. Treatment of benign pigmented lessons. Treatment of wrinkles. 1064nm The Candela 3630 Laser System is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The lasers are indicted on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemoangioma, warts, telandievas sia, rosseea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft thesue. Bening pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tatoos (significant reduction in the intensity of black and/or blue-black tartoos) and plaques a The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that harvenot responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Device Story
Dual-wavelength laser system (755nm Alexandrite, 1064nm Nd:YAG) for dermatological procedures; utilizes solid-state laser rods and xenon flashlamps; water-cooled with air-cooled heat exchanger; IGBT-switched power supply. Delivers laser energy via lens-coupled optical fiber and handpiece; pulse durations 0.25-300ms; repetition rates up to 10Hz; spot sizes 1.5-18mm. Operated by clinicians in medical settings; controlled via microprocessor-based system with touch screen interface for parameter selection (energy density, repetition rate) and system monitoring. Provides photocoagulation, hemostasis, and selective melanin targeting; enables hair reduction, lesion removal, and wrinkle treatment; improves patient outcomes through targeted tissue interaction.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, functional features, and intended use comparisons to predicate devices.
Indicated for hair reduction (temporary, long-term, or permanent), treatment of benign pigmented/vascular lesions, wrinkles, PFB, and scar reduction. Applicable to Fitzpatrick skin types I-IV (755nm) and I-VI (1064nm), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Candela Gentle Y AG Family of Laser Systems (K033172)
Candela GentleLASE Family of Laser Systems (K024371)
Candela GentleLASE Family of Laser Systems (K024335)
Candela GentleLASE Family of Laser Systems (K024260)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Candela. The logo consists of a stylized flame-like shape on the left, with horizontal lines suggesting heat or energy emanating from it. To the right of the flame is the word "CANDELA" in a simple, sans-serif font.
K 063074
## 510(k) Summary
DEC 2 7 2006
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92.
### Submitter Information
| Company Name: | Candela Corporation |
|------------------------|----------------------------------------------------|
| Company Address: | 530 Boston Post Road<br>Wayland, MA 01778 |
| Company Phone: | 508-358-7400 |
| Company Facsimile: | 508-358-5602 |
| Contact Person: | Mr. Jeffrey Roberts<br>Manager, Regulatory Affairs |
| Date summary Prepared: | 10/05/06 |
### Device Identification
| | Device Trade/Proprietary Name:Candela 3630 Laser System |
|----------------------------|-----------------------------------------------------------|
| Common Name: | Medical Laser System |
| Classification Name: | Laser Surgical Instrument, for use in General and Plastic |
| | Surgery and Dermatology |
| Classification Regulation: | 21 CFR § 878.4810 |
| Device Classification: | II |
## Identification of Predicate Device
Predicate Device(s): Candela Gentle Y AG Family of Laser Systems, K033172 Candela GentleLASE Family of Laser Systems, K024371 Candela GentleLASE Family of Laser Systems. K024335 Candela GentleLASE Family of Laser Systems, K024260 Cynosure Apogee Elite Laser, K034030
## Device Description
The Candela 3630 Laser System contains two separate laser heads (Alexandrite and Nd Y AG) which produce laser light outputs of 755 mm and 1064 nm. The outputs of each laser head are optically combined on the laser rail so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system which can output either 755 mm, 1064 nm wavelengths.
Each laser head contains the appropriate solid state laser rod and high intensity xenon flashlamps to excite the laser medium. The laser heads are water cooled with a self contained circulating water system that includes a water to air heat exchanger so the system can be fully air-cooled. The temperature of the laser heads are regulated by the circulation of distilled water at a
{1}------------------------------------------------
controlled temperature. A high voltage power supply is used to charge a storage capacitor which provides energy to the flashlamps. An IGBT (high voltage switch) is used to discharge the capacitor through the flashlamp. Each laser head has its own dedicated IGBT switch which is how the system controls which wavelength is produced. The resulting flash of the flashlamp excites the laser rod which causes emission of a pulse of laser energy.
The Candela 3630 Laser System delivers laser energy with various pulse durations from 0.25 milliseconds to 300 milliseconds (See attached specification 9914-92-0400 for further details). The output of this laser is delivered to the area of treatment by means of a lens coupled user replaceable optical fiber with a treatment handpiece attached to its distal end. A trigger switch (fingerswitch or footswitch) is used to control the delivery of laser pulses. The user may choose to deliver a single pulse each time the trigger switch is depressed, or pulses may be delivered repetitively as long as the switch is depressed, at repetition rates up to 10 pulses per second (depending on the chosen pulse duration).
A microprocessor based system controller is used to monitor and direct all system functions. Users of the laser select parameters such as desired energy density (fillerice) level and reportion rate and monitor operation via a touch screen and display panel. The touch screen panel can also be used to enable or disable the triggering of the laser, to initiate the calibration fantal and to obtain feedback from the system, such as the number of pulses delivered or spot size selected. The Candela 3630 Laser System supports spot sizes from 1.5 to 18 mm.
## Description of Intended Use
## 755nm
The Candela 3630 Laser System is indicated for the temporary hair reduction. Stable long-term or permanent hair reduction through selective targeting of melanin in hair follieles. Premanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. On all skin types (Fitzpatrick I-IV) including tanned skin.
Treatment of benign pigmented lessons. Treatment of wrinkles.
## 1064nm
The Candela 3630 Laser System is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The lasers are indicted on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemoangioma, warts, telandievas sia, rosseea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft thesue. Bening pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tatoos (significant reduction in the intensity of black and/or blue-black tartoos) and plaques a
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that harvenot responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
{2}------------------------------------------------
## Rationale for Substantial Equivalence
The Candela 3630 Laser System has the same intended use, utilizes similar functional features (including power output, spot size, repetition rate, energy, and fluence) and matches key design aspects (including wavelength, light generation medium, power supply, cooling and controls system) . as the predicate devices.
The Candela 3630 Laser System shares similar methods of assembly, method of operation, and intended uses, and therefore is substantially equivalent to the currently legally marketed Candela GentleY AG Family of Laser Systems, K033172, Candela GentleLASE Family of Laser Systems, K024371. K024335. K024260, and Cynosure Apoyee Elite Laser, K034030 predicate devices.
## Safety and Effectiveness Information
The Candela 3630 Laser System is substantial equivalent to the currently legally marketed Candela Gentle YAG Family of Laser Systems, K033172, Candela GentleLASE Family of Laser Systems, K024371, K024335, K024260, and Cynosure Apogee Elite Laser, K034030 predicate devices in intended use and technological features and therefore the risks and benefits are comparable to the predicate devices.
We therefore believe that there are no new questions of safety or effectiveness raised by the introduction of the Candela 3630 Laser System.
#### Conclusion
Base on the similarities in indications for use, design features, and functional features the Candela 3630 Laser System has been shown to be substantially equivalent to the current legally marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Candela Corporation % Mr. Jeffrey Roberts Manager, Regulatory Affairs 530 Boston Post Road Wayland, Massachusetts 01778
DEC 2 7 2006
Re: K063074 Trade/Device Name: Candela 3630 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery
and in dermatology Regulatory Class: II Product Code: GEX Dated: October 5, 2006 Received: October 10, 2006
Dear Mr. Roberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Jeffrey Roberts
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Ваввие Вискио
for
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K063074
## Device Name: Candela 3630 Laser System
Indications for Use:
### 755mm
The Candela 3630 Laser System is indicated for the temporary hair reduction. Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. On all skin types (Fitzpatrick I-IV) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
### 1064nm
The Candela 3630 Laser System is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The lasers are indicted on all skin types Fitzpatrick I-VI including tarned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemoangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Bening pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheio keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tuttoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lessons that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
Jarbare Bucklund
(Division Sign-Off) Division of General, Restorative,
and Ivouronoging and leveres
**510(k) Number** K063074
{6}------------------------------------------------
#### Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Barbara Buchner
(Division Sign-Off) (Division Signeral, Restorative, Division of Neurological Devices
**510(k) Number** K063074
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.