K052806 · Wavelight Laser Technologie AG · GEX · Oct 17, 2005 · General, Plastic Surgery
Device Facts
Record ID
K052806
Device Name
BURANE
Applicant
Wavelight Laser Technologie AG
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 17, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The BURANE is intended to be used in small and large joint arthroscopy, laparoscopic procedures, general and all surgical procedures for incision/excision, vaporization, ablation, and coagulation of soft tissue and cartilage. All soft tissues encountered in all surgical procedures are included in this indication such as, skin, cutaneous tissue, subcutaneous tissue, straited and smooth tissue, muscle, cartilage, meniscus, mucous membrane, lymph vessels and nodes, organs and glands. The BURANE is indicated for use in skin resurfacing. The BURANE is indicated for use in medicine and surgery, in the following medical specialities: Dermatology, Plastic Gastroenterology, ENT, Thoracic Surgery, Oral & Maxillofacial Surgery, Ophtalmology & Podiatry.
Device Story
BURANE is an Er:YAG laser system (2.94 µm wavelength) for dermatological and surgical soft tissue interventions. Device utilizes high water absorption of 2.94 µm wavelength to achieve precise tissue ablation with minimal thermal damage. Physician selects parameters including spot size, laser energy, pulse sequence, and coagulation mode to tailor treatment to individual patient requirements. Used by surgeons and clinicians in clinical settings for procedures ranging from skin resurfacing to internal organ resection. Output is laser energy delivered to target tissue; enables controlled incision, excision, vaporization, and coagulation. Benefits include optimized wound healing and reduced thermal damage to surrounding tissue.
Clinical Evidence
No clinical data presented. Bench testing only.
Technological Characteristics
Er:YAG laser system; 2.94 µm wavelength; Ablation and Coagulation modes. Adjustable parameters: spot size, laser energy, pulse sequence. Class II medical device.
Indications for Use
Indicated for incision, excision, vaporization, ablation, and coagulation of soft tissue and cartilage across various specialties including dermatology, plastic surgery, gastroenterology, ENT, thoracic, oral/maxillofacial, ophthalmology, and podiatry. Specific applications include skin resurfacing, wrinkle treatment, lesion removal (nevi, keloids, verrucae, tags, keratoses, tumors, ulcers), scar revision, vessel coagulation, and various procedures in gynecology (CIN, adhesions, cysts), genitourinary, and podiatry (matrixectomy).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Aesculap-Meditec MCL 29 Dermablate Laser System (K992707 and K964128)
{0}------------------------------------------------
052806 1 of 3
OCT 17 2005
Image /page/0/Picture/1 description: The image shows the word "WaveLight" with a trademark symbol. To the left of the word is a symbol of three horizontal lines that are curved to resemble waves. The word "WaveLight" is written in a sans-serif font. The trademark symbol is located to the upper right of the word.
# 510(k) Summary of Safety and Effectiveness
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
# 1. General Information
Submitter:
WaveLight Laser Technologie, AG Am Wolfsmantel 5 91058 Erlangen Germany
Contact Person:
Alexander Popp Am Wolfsmantel 5 91058 Erlangen Germany Telephone: +49 (0)9131-6186-121 Fax: +49 (0)9131-6186-202
Summary Preparation Date:
- 2. Names
Device Name:
Classification Name:
BURANE
July 30, 2004
Laser Instrument, Surgical Powered Product Code: GEX Panel: Dermatology and Plastic Surgery
# 3. Predicate Devices
The BURANE laser system is substantially equivalent to the Aesculap-Meditec MCL 29 Dermablate Laser System (K992707 and K964128), the Domier Medilas E Laser (K981438), the Cynosure CO3 Er:YAG Laser (K983034), the Laserscope Laser VELA (K971843), the WaveLight BURANE XL (K050317) and the Sciton Er:YAG Laser PROFILE 3000 (K040005).
# 4. Device Description
BURANE is a Er:YAG laser system for dermatological interventions and for aesthetic laser applications. The BURANE has an Ablation Mode and a Coagulation Mode.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains a sequence of handwritten digits and a letter. The sequence starts with the letter 'K', followed by the digits '052806'. The handwriting style is simple and clear, with each character distinct and easily readable. The characters are arranged horizontally, forming a single line of text.
S WaveLight®
# 510(k) Summary of Safety and Effectiveness
2 of 3
Furthermore, the laser system enables a quick, precise ablation with minimal r arthernore, the lamage. Thus an optimal wound healing process is achieved thornial toous admission is minimized. This is due to the BURANE's wavelength of 2.94 µm, which has an absorption maximum in water and is wavelong. I of Elect pery well by tissue water. The thermal damage zone is thus minimized.
While the parameter wavelength is specified by the device, spot size, laser energy, pulse sequence, and coagulation function can be selected by the treating physician for optimal adjustment to individual requirements.
# 5. Indications for Use
The BURANE is indicated:
The BURANE is intended to be used in small and large joint arthroscopy, laparoscopic procedures, general and all surgical procedures for incision/excision, vaporization, ablation, and coagulation of soft tissue and cartilage. All soft tissues encountered in all surgical procedures are included in this indication such as, skin, cutaneous tissue, subcutaneous tissue, straited and smooth tissue, muscle, cartilage, meniscus, mucous membrane, Ismoh vessels and nodes, organs and glands. The BURANE is indicated for use in skin resurfacing. The BURANE is indicated for use in medicine and surgery, in the following medical specialities: Dermatology, Plastic Gastroenterology, ENT, Thoracic Surgery, Oral & Maxillofacial Surgery, Ophtalmology & Podiatry.
# Aesthetic Surgery
Skin resurfacing and treatment of wrinkles.
# Dermatology/Plastic Surgery
Indications include, epidermal nevi, telangiectasia, spider veins, actinic incheditis, keloids, verrucae, skin tags, anal tags, keratoses, scar revision, debulking benign tumors, decubitis ulcers.
# Gastroenterology
# General Surgery
The Er:YAG laser is intended for the surgical incision/excision, vaporization and coagulation of soft tissue during general surgery application where skin incision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors and lesions, tissue ablation and/or vessel coagulation.
{2}------------------------------------------------
K052806 3 of 3
![]()WaveLight®
# 510(k) Summary of Safety and Effectiveness
# Genitourinary
Genitourinaly
Indications include lesions of the external genitalia, urethra and anus, penis, Indications include tesions of the extransmate, giant perineal scrotum and urethra (Includes contylema actinoma), vulvar lesions, polyps and familial polyps of the colon.
#### Gynecology
Gynecology
Indications include cervical intraepithelial neoplasia (CIN), herpes simplex, endometrial adhesions, cysts and condyloma.
#### ENT
Indications include ear, nose and throat lesions, polyps, cysts, mulcations include our, noos ania and the sue, oral leukoplakia.
# Oral/Maxillofacial
Indications include benign oral tumors, oral and glossal lesions and gingivectomy.
#### Ophtalmology
Indications include soft tissue surrounding the eye and orbit and anterior capsulotomy.
#### Podiatry
Podially
Indications include warts, plantar verrucae, large mosaic verrucae and matrixectomy.
- 6. Performance Data
None presented.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Public Health Service
OCT 17 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Wavelight Laser Technologie AG c/o Jeffrey D. Rongero Senior Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Drive, P.O. Box 13995 Research Triangle Park, North Carolina 27709-3995
Re: K052806
Trade/Device Name: BURANE Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: August 29, 2005 Received: October 4, 2005
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2- Jeffrey D. Rongero
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Mark N. Mullen
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K052806
WaveLight®
# 510 {k} Indications for Use
of 2
Indications for Use
510(k) Number (if known): N/A BURANE Device Name:
Indications for Use:
The BURANE is intended to be used in small and large joint arthroscopy, taparoscopic procedures, general and all surgical procedures for incision/excision, vaporization, ablation, and coagulation of soft tissue and inololokolori, Tapes encountered in all surgical procedures are included in cantiage. An Soft tissue, oncentraneous tissue, subcutaneous tissue, straited and this indication odor as, chari, satamoniscus, mucous membrane, lymph vessels smooth tissue, massio, caraids. The BURANE is indicated for use in skin and frodoo, organo and glansend for use in medicine and surgery, in the following medical specialities: Dermatology, Plastic Gastroenterology, ENT, Thoracic Surgery, Oral & Maxillofacial Surgery, Ophtalmology & Podiatry.
#### Aesthetic Surgery
Skin resurfacing and treatment of wrinkles.
# Dermatology/Plastic Surgery
Indications include, epidermal nevi, telangiectasia, spider veins, actinic cheilitis, mulcatoris mondo, oplaamaanal tags, keratoses, scar revision, debulking benign tumors, decubitis ulcers.
#### Gastroenterology
# General Surgery
The Er:YAG laser is intended for the surgical incision/excision, vaporization and The L. . FO laser to intonebal to the reserv application where skin incision, tissue dissection, excision of external tumors and lesions, complete or partial tresection of internal organs, tumors and lesions, tissue ablation and/or vessel coagulation.
#### Genitourinary
Indications include lesions of the external genitalia, urethra and anus, penis, Indications frolude foolone of the additional acuminate, giant perineal condyloma servicing and arothing (ona), vulvar lesions, polyps and familial polyps of the colon.
#### Gynecology
510(k) N
Indications include cervical intraepithelial neoplasia (CIN), herpes simplex, endometrial adjhesions, cysts and condyloma.
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page of 2 (indication for use only)
0031
Jumber_ 405 2806
{6}------------------------------------------------
K052806
Image /page/6/Picture/1 description: The image shows the handwritten text "2 of 2". The numbers are written in a cursive style. The text is in black ink on a white background.
# S WaveLight®
# 510 (k) Indications for Use
# ENT
Indications include ear, nose and throat lesions, polyps, cysts, hyperkeratosis, excision of carcinogenic tissue, oral leukoplakia.
# Oral/Maxillofacial
Indications include benign oral tumors, oral and glossal lesions and gingivectomy.
# Ophtalmology
Indications include soft tissue surrounding the eye and orbit and anterior capsulotomy.
# Podiatrv
Indications include warts, plantar verrucae, large mosaic verrucae and matrixectomy.
Over-The-Counter Use N/A Prescription Use V AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
2
(Division Sign-Off)
Page (indication for use only
Division of General, Restorative, and Neurological Devices
510(k) Num! rr_k os Z &O(z
0032
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.