K050317 · Wavelight Laser Technologie AG · GEX · Feb 24, 2005 · General, Plastic Surgery
Device Facts
Record ID
K050317
Device Name
BURANE XL
Applicant
Wavelight Laser Technologie AG
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 24, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The BURANE XL is intended to be used in small and large joint arthroscopy, laparoscopic procedures, general surgical procedures for incision/excision, vaporization, ablation, and coagulation of soft tissue and cartilage. All soft tissues encountered in an open surgical procedure are included in this indication such as, skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage, meniscus, mucous membrane, lymph vessels and nodes, organs and glands. The BURANE XL is indicated for use in skin resurfacing. The BURANE XL is indicated for use in medicine and surgery, in the following medical specialities: Dermatology, Plastic Surgery, Gastroenterology, ENT, Thoracic Surgery, Oral & Maxillofacial Surgery, Ophtalmology & Podiatry.
Device Story
The BURANE XL is an Er:YAG laser system (2.94 µm wavelength) designed for dermatological and aesthetic interventions. It utilizes triple-mode technology, allowing the physician to select between Ablation Mode, Coagulation Mode, or a combination of both. The device operates by delivering laser energy to target soft tissue; the 2.94 µm wavelength is highly absorbed by water, enabling precise ablation with minimal thermal damage to surrounding tissue, thereby promoting optimal wound healing and reducing scar formation. The treating physician selects parameters including spot size, laser energy, pulse sequence, and coagulation function to adjust for individual patient requirements. It is intended for use by trained medical professionals in clinical settings across multiple surgical specialties.
Clinical Evidence
No clinical data presented; bench testing only.
Technological Characteristics
Er:YAG laser system; wavelength 2.94 µm. Features triple-mode technology (Ablation, Coagulation, and combination). Adjustable parameters include spot size, laser energy, and pulse sequence. Designed for dermatological and surgical soft tissue applications.
Indications for Use
Indicated for surgical incision, excision, vaporization, ablation, and coagulation of soft tissue and cartilage across various specialties including dermatology, plastic surgery, gastroenterology, ENT, thoracic surgery, oral/maxillofacial surgery, ophthalmology, and podiatry. Specific applications include skin resurfacing, wrinkle treatment, epidermal nevi, telangiectasia, spider veins, actinic keratoses, scar revision, debulking benign tumors, decubitus ulcers, genitourinary lesions (condyloma, polyps), gynecological conditions (CIN, adhesions, cysts), ENT lesions, oral/glossal lesions, gingivectomy, and podiatric procedures (warts, matrixectomy).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Aesculap-Medited MCL 29 Dermablate Laser System (K992707 and K964128)
{0}------------------------------------------------
Wavelight Burane XL: 510(k) Review January 20, 2005
K 050317
510K Summary
FEB 2 4 2005
WaveLight®
510(k) Summary of Safety and Effectiveness
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance rills 810(t) Samments of the SMDA 1990 and 21 CFR 807.92.
### 1. General Information
WaveLight Laser Technologie, AG Submitter: Am Wolfsmantel 5 91058 Erlangen Germany Alexander Popp Contact Person: Am Wolfsmantel 5 91058 Erlangen Germany Telephone: +49 (0)9131-6186-121 Fax: +49 (0)9131-6186-202 July 30, 2004 Summary Preparation Date:
- 2. Names
Device Name:
BURANE XL
Classification Name:
Laser Instrument, Surgical Powered Product Code: GEX Panel: Dermatology and Plastic Surgery
## 3. Predicate Devices
The BURANE XL laser system is substantially equivalent to the Aesculap-Medited MCL 29 Dermablate Laser System (K992707 and K964128), the Dornier Medilas E Laser (K981438), the Cynosure CO3 Er:YAG Laser (K983034), the Laserscope Laser VELA (K971843) and the Sciton Er:YAG Laser PROFILE 3000 (K040005).
## 4. Device Description
BURANE XL is a Er:YAG laser system for dermatological interventions and for aesthetic laser applications. The triple mode technology makes BURANE for aestheric ideor upphodibensial controllable Coagualtion Mode for different AL an export dother the Ablation Mode for classical ablation procedures, and the combination of both techniques.
0029
{1}------------------------------------------------
K050J17
2/3
Wavelight Burane XL: 510(k) Review January 20, 2005
510K Summary (con't)
S WaveLight®
510(k) Summary of Safety and Effectiveness
Furthermore, the laser system enables a quick, precise ablation with minimal Furthermore, the laser System enables a quion process is achieved.
thermal tissue damage. Thus an optimal wound health PULPANE YL's thermal tissue damage. Thus an optimal wear in the BURANE XL's
-- the risk of scar formation is minimized. This is due to the BURANE XL's -- the risk of scar lomation is minimized. This is and is water and is wavelength of 2.94 pm, which has an absorplane in thermal damage zone is thus minimized.
While the parameters wavelength and pulse form are specified by the device, While the parameters wavelengin and patool coagulation function can be
spot size, laser energy, pulse sequence, and coaggitation function can be spot size, laser encry), pullos soquenest ...
selected by the treating physician for optimal adjustment to individual requirements.
## 5. Indications for Use
The BURANE XL is indicated:
The BURANE XL is intended to be used in small and large joint arthroscopy, The BURANE XL is intendou to be ad all surgical procedures for incision/excision, vaporization, ablation, and coagulation of soft tissue and including the may be many ablem, and surgical procedures are included in this indication such as, skin, cutaneous tissue, subcutaneous tissue, in this indication such as, Skiff, Gutanoous neartilage, meniscus, mucous membrane, stratted and smooth tissue, music, organs and glands. The BURANE XL is indicated lymph vessels and nodes, organs and grands. Indicated for use in medicine
for use in skin resurfacing. The BURANE XL is indicated for use in medicine for use in skirl resultacing. The Born specialities: Dermatology, Plastic and surgery, in the following medical Spotlallabor Spotland of July 2017 - 11:40
Gastroenterology, ENT, Thoracic Surgery, Oral & Maxillofacial Surgery, Ophtalmology & Podiatry.
Aesthetic Surgery
Skin resurfacing and treatment of wrinkles.
## Dermatology/Plastic Surgery
Indications include, epidermal nevi, telangiectasia, spider veins, actinio Indications Include, epidennal nevi, colangiocacial, spiratoses, scar revision, debulking benign tumors, decubitis ulcers.
#### Gastroenterology
#### General Surgery
The Er:YAG laser is intended for the surgical incision/excision, vaporization The E.TAG laser is intended for the ourgery application where skin and coaguiation of Solt lissue duning general tumors and lesions, complete Incision, itssue dissection, oxolori of oxecions, tissue ablation and/or vessel coagulation.
0030
{2}------------------------------------------------
K050317
3/3
Wavelight Burane XL: 510(k) Review January 20, 2005
510K Summary (con't)
Image /page/2/Picture/3 description: The image shows the word "WaveLight" with a symbol to the left of the word. The symbol is three horizontal lines stacked on top of each other, with each line slightly curved. The word "WaveLight" is written in a simple, sans-serif font. There is a registered trademark symbol to the upper right of the word.
# 510(k) Summary of Safety and Effectiveness
#### Genitourinary
Genitourinary
Indications include lesions of the external genitalia, urethra and anus, penis, Indications Include lesions of the external geniture, giant perineal scrotum and urethra (includes condyloma deaminene) ,
condyloma and verrucous carcinoma), vulvar lesions, polyps and familial polyps of the colon.
#### Gynecology
്യന്മാവിമയ്ക്കും.
Indications include cervical intraepithelial neoplasia (CIN), herpes simplex, molometrial adhesions, cysts and condyloma.
#### ENT
Indications include ear, nose and throat lesions, polyps, cysts, indications include car, nooo and the tic tissue, oral leukoplakia.
Oral/Maxillofacial Oral/Maxilloracial
Indications include benign oral tumors, oral and glossal lesions and gingivectomy.
Ophtalmology Indications include soft tissue surrounding the eye and orbit and anterior capsulotomy.
#### Podiatry
Podiatry
Indications include warts, plantar verrucae, large mosaic verrucae and matrixectomy.
- 6. Performance Data None presented.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars representing the agency's mission to promote health, well-being, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.
Public Health Service
FEB 2 4 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Wavelight Laser Technologie AG c/o Mr. William J. Sammons Underwriters Laboratories, Inc. 12 Laboratory Drive P.O. Box 13995 Research Triangle Park, North Carolina 27709
Re: K050317 Trade/Device Name: Burane XL Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 20, 2005 Received: February 9, 2005
Dear Mr. Sammons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. William J. Sammons
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Stipt Rhodes
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Wavelight Burane XL: 510(k) Review January 20, 2005 Indication for use
Image /page/5/Picture/1 description: The image contains the word "WaveLight" with a registered trademark symbol. To the left of the word is a symbol of three horizontal lines that are stacked on top of each other. The lines are curved to resemble a wave.
11 >
#### 510 (k) Indications for Use
Indications for Use 510(k) Number (if known): N# K 05D 317 BURANE XL Device Name:
#### Indications for Use:
The BURANE XL is intended to be used in small and large joint arthroscopy, The BURANE XE is intendou to be ad all surgical procedures for laparoscopic procedures, gonoral ation, and coagulation of soft tissue and incision/excision, vaporization, abliation, and surgical procedures are included in cartilage. All soll lissues encountered in an subcutaneous tissue, straited and this indication such as, skin, outainous, meniscus, membrane, lymph vessels
smooth tissue, muscle, cartilage, meniscus, meniscus, men in clin smooth tissue, muscle, cartilage, membodo, macered for use in skin and nodes, organis and grands. The Beard for use in medicine and surgery, in the resultacing. The DONANE AE is included in Plastic Gastroenterology, ENT, following medical specialities: Domiatology, Ophtalmology & Podiatry.
Aesthetic Surgery
Skin resurfacing and treatment of wrinkles.
### Dermatology/Plastic Surgery
Dermatology/Flastic ാവ്യല്
Indications include, epidermal nevi, telangiectasia, spider veins, adabulking Indications include, epidential nove, totanges, scar revision, debulking benign tumors, decubitis ulcers.
Gastroenterology
#### General Surgery
General Surgery
The Er:YAG laser is intended for the surgical incision/excision, vaporization and The E.Y.AG laser is intended for the ourgery application where skin indision, coagulation of solt trisde duning general cargony desions, complete or partial tissue dissection, excision of extornal tamble and lesions, tissue ablation and/or vessel coaqulation.
#### Genitourinary
Indications include lesions of the external genitalia, urethra and anus, penis, Indications Include lesions of the oxtomal genminate, giant perineal condyloma
scrotum and urethra (includes condyloma acuminate, giant perineal condyloma scrolum and their a (inoludos contactions, polyps and familial polyps of the colon.
#### Gynecology
Gynecology
Indications include cervical intraepithelial neoplasia (CIN), herpes simplex, muloations include ons, cysts and condyloma.
0027
{6}------------------------------------------------
## 510 (k) Indications for Use
## ENT
Indications include ear, nose and throat lesions, polyps, cysts, hyperkeratosis, excision of carcinogenic tissue, oral leukoplakia.
## Oral/Maxillofacial
Indications include benign oral tumors, oral and glossal lesions and qinqivectomy.
### Ophtalmology
Indications include soft tissue surrounding the eye and orbit and anterior capsulotomy.
### Podiatry
Indications include warts, plantar verrucae, large mosaic verrucae and matrixectomy.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|---|
|-------------------------------------------------|---|
AND/OR
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | N/A |
|------------------------------------------------|-----|
|------------------------------------------------|-----|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEEDED)## Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
Page 1 ol (indication for use only
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number**: K050317
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.