K023464 · Quantel Medical · GEX · Jan 13, 2003 · General, Plastic Surgery
Device Facts
Record ID
K023464
Device Name
VIRIDIS TWIN OPHTHALMIC PHOTOCOAGULATOR
Applicant
Quantel Medical
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 13, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Viridis Twin laser is indicated for the photocoagulation or ablation of pigmented tissue within the eye. The 532 nm wavelength is indicated for use for: retinal photocoagulation proliferative diabetic retinopathy, macular degeneration, and retinal detachment. peripheral photocoagulation proliferative diabetic retinopathy, macular degeneration, and retinal detachment. endophotocoagulation (EPCP) intraocular photocoagulation as an adjunct in vitrectomy surgery, complicated rhegmatogeous, tractional retinal detachments, proliferative vitreoretinopathy, proliferative diabetic retinopathy, retinopathy, and retinal vascular tumors. The 810 nm wavelength is indicated for use for: photocoagulation or ablation of pigmented tissue within the eye, transscleral ciliary body ablation (treatment is reserved for patients with chronic glaucoma and those not responding to conventional treatments), limited and pan-retinal photocoagulation, transpupillary photocoagulation, endophotocoagulation, treatment of complicated rhegmatogeous, tractional retinal detachments, proliferative vitreoretinopathy, proliferative diabetic retinopathy, macular degeneration, peripheral photocoagulation (recumbent patients), transpupillary photocoagulation of choriodal neovasculature, and age-related macular degeneration (AMD) treatments.
Device Story
Viridis Twin Ophthalmic Photocoagulator is a dual-wavelength laser system combining a 532 nm frequency-doubled Nd:YAG laser and an 810 nm diode laser. Device delivers coherent light energy to ocular tissues to perform photocoagulation or ablation. User selects desired wavelength and attaches compatible delivery systems—such as slit lamp adapters, biomicroscopes, endoprobes, or indirect ophthalmoscopes—to the laser aperture. Operated by ophthalmologists in clinical settings. Output energy is directed to treatment sites to manage conditions like diabetic retinopathy, macular degeneration, and glaucoma. Device provides therapeutic benefit by enabling precise tissue coagulation or ablation, aiding in the management of retinal and choroidal pathologies.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Dual-wavelength ophthalmic laser system: 532 nm frequency-doubled Nd:YAG and 810 nm diode laser. Class II device (21 CFR 886.4390). Compatible with standard ophthalmic delivery systems including slit lamp adapters, biomicroscopes, endoprobes, and indirect ophthalmoscopes. Standalone system.
Indications for Use
Indicated for patients requiring photocoagulation or ablation of pigmented ocular tissue, including those with proliferative diabetic retinopathy, macular degeneration, retinal detachment, proliferative vitreoretinopathy, retinal vascular tumors, and chronic glaucoma unresponsive to conventional treatments.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
## SECTION 10
## 510(k) SUMMARY
K023464
This 510(k) summary of safety and effectiveness for the modification in the Indications for Use for the Viridis Twin laser is submitted in accordance with the requirements of SMDA 1990 and follows Office of Device Evaluation guidance concerning the organization and content of a 510(k) summary.
Quantel Medical Applicant: QUANTEL MEDICAL Address: 21 rue Newton Z.I. du Br zet 63039 Clermont-Ferrand Cedex 2 FRANCE +33 (0)473 745 745 +33 (0)473 745 700 (Fax) Contact Person: Mr. Jean Abascal Regulatory Affairs Manager (+33) 169 29 17 25 : (+33) 169 29 17 29 Preparation Date: September 2002 (of the Summary) Viridis Twin Ophthalmic Photocoagulator Device Name: Ophthalmic Laser, frequency doubled Nd: Y AG (532 nm) laser Common Name: Laser Surgical Instrument, frequency doubled Nd: Y AG (532 nm) laser Ophthalmic Laser, 810 nm diode laser Laser Surgical Instrument; 810 nm diode laser Classification Ophthalmic laser (see 21 CFR 886.4390) Name: Laser surgical instrument (see: 21 CFR 878.4810). Product Code: HQF; Panel: 86 Product Code: GEX; Panel: 79 Predicate devices: Viridis Laser (K960867); Novus Verdi Delivery Systems (K991258) -and the Viridis laser
{1}------------------------------------------------
IRIS Medical OcuLight SL/SLx (K020374); Nidek DC-3300 (K013760) The predicate lasers utilize several delivery systems, including slit lamp adapters, biomicroscopes, endoprobes, and indirect ophthalmoscopes to deliver the laser energy to the treatment sites.
The delivery systems used with the Viridis Twin Ophthalmic Photocoagulator are the same as or similar to delivery systems supplied with or used with other ophthalmic photocoagulators operating at 532 nm and/or 810 nm. These devices are independently marketed by their respective manufacturers which are responsible for complying with regulatory and marketing requirements. The cited models are compatible for use with the Viridis Twin Ophthalmic Photocoagulator.
- Device description: The Viridis Twin Ophthalmic Photocoagulator is a combination system consisting of a Nd:YAG frequency doubled laser which emits a beam of coherent light at 532 nm and a diode laser which emits a beam of coherent light 810 microns. The treatment beam is selected by the user, the delivery system is attached to the appropriate laser aperture, and the energy is delivered to the treatment site(s) via delivery systems/devices.
- Quantel Viridis Twin Ophthalmic Photocoagulator is indicated Indications: The for the photocoagulation or ablation of pigmented tissue within the eye.
The 532 nm wavelength is indicated for use for:
retinal photocoagulation proliferative diabetic retinopathy, macular degeneration, and retinal detachment. peripheral photocoagulation proliferative diabetic retinopathy, macular degeneration, and retinal detachment. endophotocoagulation (EPCP) intraocular photocoagulation as an adjunct in vitrectomy surgery, complicated rhegmatogeous, tractional retinal detachments, proliferative vitreoretinopathy, proliferative diabetic retinopathy, retinopathy, and retinal vascular tumors.
The 810 nm wavelength is indicated for use for:
photocoagulation or ablation of pigmented tissue within the eye, transscleral ciliary body ablation (treatment is reserved for patients with chronic glaucoma and those not responding to conventional treatments), limited and pan-retinal photocoagulation, transpupillary photocoagulation,
{2}------------------------------------------------
endophotocoagulation,
treatment of complicated rhegmatogeous, tractional retinal detachments, proliferative vitreoretinopathy, proliferative diabetic retinopathy, macular degeneration, peripheral photocoagulation (recumbent patients),
transpupillary photocoagulation of choriodal neovasculature, and
age-related macular degeneration (AMD) treatments.
Performance Data: None required.
- CONCLUSION: Based on the information in this notification Quantel Medical concludes that the Viridis Twin Ophthalmic Photocoagulator is substantially equivalent to the cited legally marketed predicates. Information regarding delivery systems which may be used with the laser is supplied by Quantel Medical; the individual devices or systems are marketed by their respective manufacturers who have the responsibility for complying with applicable regulations and marketing requirements.
Revised: January 2003
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three stripes forming its body and wing. The eagle faces left and is enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the border of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 3 2003
Quantel Medical c/o Roger W. Barnes 342 Sunset Bay Road Hot Springs, Arkansas 71913
Re: K023464
Trade/Device Name: Viridis Twin Ophthalmic Photocoagulator Regulation Number: 886.4390 Regulation Name: Laser surgical instrument Regulatory Class: Class II Product Code: GEX Dated: October 9, 2002 Received: October 15, 2002
Dear Mr. Barnes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in
{4}------------------------------------------------
Page 2 – Mr. Roger W. Barnes
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97): Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N. Milkinson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## SECTION 7
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Viridis Twin laser
Indications for Use Statement:
The Viridis Twin laser is indicated for the photocoagulation or ablation of pigmented tissue within the eye.
The 532 nm wavelength is indicated for use for:
retinal photocoagulation proliferative diabetic retinopathy, macular degeneration, and retinal detachment. peripheral photocoagulation proliferative diabetic retinopathy, macular degeneration, and retinal detachment. endophotocoagulation (EPCP) intraocular photocoagulation as an adjunct in vitrectomy surgery, complicated rhegmatogeous, tractional retinal detachments, proliferative vitreoretinopathy, proliferative diabetic retinopathy, retinopathy, and retinal vascular tumors.
The 810 nm wavelength is indicated for use for:
photocoagulation or ablation of pigmented tissue within the eye, transscleral ciliary body ablation (treatment is reserved for patients with chronic glaucoma and those not responding to conventional treatments),
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation | |
|--------------------------------------------------|----|
| Prescription Use | OR |
| (Per 21 CFR 801.109) | |
Over-The Counter Use
{6}------------------------------------------------
limited and pan-retinal photocoagulation, transpupillary photocoagulation,
endophotocoagulation,
treatment of complicated rhegmatogeous, tractional retinal detachments, proliferative vitreoretinopathy, proliferative diabetic retinopathy, macular degeneration, peripheral photocoagulation (recumbent patients), transpupillary photocoagulation of choriodal neovasculature, and
age-related macular degeneration (AMD) treatments.
revised: January 2003
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.