K013760 · Nidek, Inc. · HQF · Feb 11, 2002 · Ophthalmic
Device Facts
Record ID
K013760
Device Name
DC-3300 LASER DIODE PHOTOCOAGULATOR
Applicant
Nidek, Inc.
Product Code
HQF · Ophthalmic
Decision Date
Feb 11, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nidek DC-3300 Laser Diode Photocoagulator is indicated for all retinal photocoagulation procedures, such as limited and pan-retinal, transpupillary laser photocoagulation, endophotocoagulation and transscleral photocoagulation, and glaucoma procedures, such as laser trabeculoplasty and iridotomy. The DC-3300 is used in combination with various delivery systems, such as slit lamps, binocular indirect ophthalmoscopes, endoprobes, and transscleral probes.
Device Story
The DC-3300 is a portable, continuous-wave 808 nm laser diode photocoagulator used for ophthalmic procedures. It features a 633 nm red diode aiming laser. The system includes a main console, footswitch, and various delivery attachments: slit lamps (Nidek SL-1600), binocular indirect ophthalmoscopes (Keeler All-Pupil, Heine Omega 180), endoprobes, and transscleral probes. Operated by clinicians in clinical settings, the device delivers laser energy to ocular tissue to perform photocoagulation or glaucoma treatments. The clinician visualizes the target tissue through the delivery system and controls laser emission via the footswitch. The device provides therapeutic thermal effects to treat retinal conditions and glaucoma, aiming to preserve or improve patient vision.
Clinical Evidence
Bench testing only. The device underwent system and component performance testing based on product specifications and risk analysis. It demonstrated conformance to international safety standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-22, IEC 60825-1, and UL 2601-1. No clinical data was required or presented.
Iriderm Iris OcuLight Diode Laser With Slit Lamp Adapter (K912918)
Submission Summary (Full Text)
{0}------------------------------------------------
### FEB 1 1 2002
# K013760 1/3
### SUMMARY OF SAFETY AND EFFECTIVENESS SECTION 14:
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92.
#### SUBMITTER INFORMATION 14.1
| a. | Company Name: | Nidek, Inc. |
|----|--------------------------------------|-----------------------------------------------|
| b. | Company Address: | 47651 Westinghouse Drive<br>Fremont, CA 94539 |
| c. | Company Phone:<br>Company Facsimile: | (510) 226-5700<br>(510) 226-5750 |
| d. | Contact Person: | Hiro Matsuzaki<br>Quality Assurance Manager |
| e. | Date Summary Prepared: | February 4, 2002 |
#### DEVICE IDENTIFICATION 14.2.
| a. Trade/Proprietary Name: | DC-3300 Laser Diode Photocoagulation |
|----------------------------|-------------------------------------------|
| b. Classification Name: | Ophthalmic Laser<br>21 CFR 886.4390 HQF |
#### IDENTIFICATION OF PREDICATE DEVICES 14.3
| Company | Device | 510(k) No. | Date Cleared |
|-------------|-----------------------------------------------------|------------|--------------|
| Nidek, Inc. | DC-3000 Laser Diode<br>Photocoagulator | K903639 | 02/04/1991 |
| Iriderm | Iris OcuLight SLx | K913430 | 11/15/1991 |
| Iriderm | Iris OcuLight Diode Laser<br>With Slit Lamp Adapter | K912918 | 08/23/1991 |
{1}------------------------------------------------
K613760 2/3
#### DEVICE DESCRIPTION 14.4
The DC-3300 Laser Diode Photocoagulator is an 808 nm continuous wave device. The DC-3300 consists of a compact, lightweight main console, footswitch and carrying case. The aiming laser of the DC-3300 is a red diode with a wavelength of 633 nm. A number of delivery systems are available with the DC-3300. A Slit Lamp Delivery Unit (Nidek SL-1600) Multipurpose Delivery Unit for the attachment of slit lamps (Nidek SL-1600), Endophotocoagulation Set, Binocular Indirect Ophthalmoscope (Models A and B), and Transscleral Photocoagulation probes.
#### 14.5 SUBSTANTIAL EQUIVALENCE
The DC-3300 Laser Diode Photocoagulator is substantially equivalent to the Nidek DC-3000 Laser Diode Photocoagulator and the Iriderm Iris OcuLight SLx.
The fundamental technical characteristics and device specifications of the DC-3300 Laser Diode Photocoagulator are the same as those of the predicate devices. The DC-3300 and the predicate devices are diode photocoagulation lasers with a wavelength of 810 nm. The DC-3300 and the OcuLight SLx use a variety of delivery systems, including slit lamps, indirect ophthalmoscopes, endoprobes, and transscleral probes. The DC-3300 and the OcuLight SLx are indicated for all retinal photocoagulation and glaucoma procedures.
#### INDICATIONS FOR USE 14.6
The Nidek DC-3300 Laser Diode Photocoagulator is indicated for all retinal photocoagulation procedures, such as limited and pan-retinal, transpupillary laser photocoagulation, endophotocoagulation and transscleral photocoagulation, and glaucoma procedures, such as laser trabeculoplasty and iridotomy. The DC-3300 is used in combination with various delivery systems, such as slit lamps, binocular indirect ophthalmoscopes, endoprobes and transscleral probes.
{2}------------------------------------------------
Nidek DC3300 Laser Diode Photocoagulator Original Premarket 510(k) Notification
#### 14.7 TECHNOLOGICAL CHARACTERISTICS
The DC-3300 Laser Diode Photocoagulator is a continuous wave 808 nm laser for ophthalmic photocoagulation procedures. The compact design of the system makes it portable and easily relocated. The aiming laser of the DC-3300 is a red diode with a wavelength of 633 nm. The delivery systems include the Slit Lamp for the attachment of the Nidek SL-1600, the Multipurpose for the attachment of the Nidek SL-1600 slit lamp, Endoprobes (two angled and two straight configurations), Transscleral probes (contact, non-contact and retinal), and two models of the Binocular Indirect Ophthalomoscope (Keeler All-Pupil and Heine Omega 180). The technological characteristics of the DC-3300 are equivalent to those of the predicate devices.
KO137b0 3/3
#### 14.8 PERFORMANCE DATA
Performance testing was conducted on the DC-3300 Laser Diode Photocoagulator and delivery systems. System and component testing was completed based on product specifications and hazard effects determined from the risk analysis. The device was found to be in conformance with the following international safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-22, IEC 60825-1, and UL 2601-1. The DC-3300 was found to perform as intended.
#### 14.9 CONCLUSION
This notification contains all information required by 21 CFR 807.87. The DC-3300 Laser Diode Photocoagulator was found to perform as intended during validation testing. The DC-3300 is substantially equivalent to the Nidek DC-3000 and Iriderm Iris OcuLight SLx devices. Product specifications and technical characteristics are within the same range of specifications for the predicate devices. The evaluation of the product specifications, performance testing and risk analysis does not raise any new issues of safety or effectiveness. The DC-3300 is intended for all retinal photocoagulation and glaucoma procedures and is combined with various delivery systems to perform its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the border of the circle. Inside the circle is a stylized image of an eagle with three heads.
FEB 1 1 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nidek. Inc. Ms. Carol Patterson c/o Patterson Consulting Group, Inc. 21911 Erie Lane Lake Forest, California 92630
Re: K013760
Trade Name: Nidek C-3300 Laser Diode Photocoagulation Regulation Number: 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: II Product Code: HQF; GEX Dated: November 9, 2001 Received: November 13, 2001
Dear Ms. Patterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Carol Patterson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
- Nidek DC-3300 Laser Diode Photocoagulator Nidex DC-5500 Lasel Dk) Notification: #K013760
## INDICATION FOR USE
| 510(k) Number: | K013760 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Nidek DC-3300 Laser Diode Photocoagulator |
| Indications for Use: | The Nidek DC-3300 Laser Diode Photocoagulator is indicated<br>for all retinal photocoagulation procedures, such as limited and<br>pan-retinal, transpupillary laser photocoagulation,<br>endophotocoagulation and transscleral photocoagulation, and<br>glaucoma procedures, such as laser trabeculoplasty and<br>iridotomy. The DC-3300 is used in combination with various<br>delivery systems, such as slit lamps, binocular indirect<br>ophthalmoscopes, endoprobes, and transscleral probes. |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
______________________________________________________________________________________________________________________________________________________________________________
______________________________________________________________________________________________________________________________________________________________________________
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div style="text-align:center">Miriam Provost</div> <div style="text-align:center">(Division Sign-Off)</div> <div style="text-align:center">Division of General, Restorative</div> <div style="text-align:center">and Neurological Devices</div> |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription Use | <div style="display:inline-block"><span style="text-decoration: overline;">X</span></div> 510(k) Number K013760 |
| | OR |
| | Over-The-Counter Use |
| (Per 21 CFR 801.109) | |
: : … "
-
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.