K182365 · Cynosure, Inc. · GEI · Oct 24, 2018 · General, Plastic Surgery
Device Facts
Record ID
K182365
Device Name
TempSure System
Applicant
Cynosure, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 24, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids. The 25mm, and 30mm handpieces provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The massage device is intended to provide a temporary reduction in the appearance of cellulite. The following surgical modes are applicable to the generator: Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed. Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty. Blended Cutting and Coagulation: submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma. nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and development of skin flaps. Fulguration: basal cell carcinoma. papilloma. cyst destruction, tumors. verrucae, hemostasis. Bipolar: pinpoint precise coagulation. pinpoint hemostasis in any, field (wet or dry). snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment and turbinate shrinkage.
Device Story
TempSure System is a radiofrequency (RF) generator for aesthetic and surgical applications; delivers RF energy to patient tissue to induce heating and cellular alteration. System includes generator, temperature-sensing handpieces, footswitch, neutral pad, and electrosurgical accessories. Used in clinical settings by physicians; energy output controlled via GUI and switches. Surgical modes include Cut, Coagulation, Blend, Fulgurate, and Bipolar. Temperature-sensing handpieces provide feedback for aesthetic treatments (wrinkles, pain relief, cellulite reduction). Output affects tissue via thermal effects; provider monitors device settings to achieve desired clinical outcome. Benefits include precise tissue heating and surgical hemostasis/cutting.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by non-clinical bench testing, including ex-vivo tissue samples (liver, kidney, muscle) to compare thermally affected zones (TAZ) against predicate devices, and electrical safety/EMC testing per IEC 60601-1, 60601-1-2, and 60601-2-2.
Technological Characteristics
RF generator; Monopolar and Bipolar modalities. Output: 4.0 MHz Sin-wave CW, Fully/Partially Rectified; 1.7 MHz for Bipolar. Max power 300W. Materials: Gold-Plated Brass, Polytheramide, Loctite, Delrin. Features temperature-sensing handpieces with <1s response time. Connectivity: Standalone. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2.
Indications for Use
Indicated for patients requiring non-ablative treatment of mild to moderate facial wrinkles/rhytids, temporary relief of pain, muscle spasms, increased local circulation, temporary reduction in cellulite appearance, and general surgical procedures requiring electrosurgical coagulation, cutting, or hemostasis (urologic, thoracic, plastic, reconstructive, gynecological).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 24, 2018
Cynosure, Inc Amy Tannenbaum Regulatory Affairs Specialist III 5 Carlisle Road Westford, Massachusetts 01886
Re: K182365
Trade/Device Name: TempSure System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 29, 2018 Received: August 30, 2018
Dear Amy Tannenbaum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K182365
Device Name TempSure System
#### Indications for Use (Describe)
The 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.
The 25mm, and 30mm handpieces provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.
The massage device is intended to provide a temporary reduction in the appearance of cellulite
The following surgical modes are applicable to the generator:
Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed
Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty.
Blended Cutting and Coagulation: submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma. nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and development of skin flaps.
Fulguration: basal cell carcinoma. papilloma. cyst destruction, tumors. verrucae, hemostasis.
Bipolar: pinpoint precise coagulation. pinpoint hemostasis in any, field (wet or dry). snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment and turbinate shrinkage
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary for Cynosure TempSure System
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a)(1) Submitter Information | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Cynosure, Inc |
| Address | 5 Carlisle Road, Westford MA, 01886 |
| Phone Number | 781-993-2454 |
| Fax Number | 978-256-6556 |
| Establishment Registration Number | 1222993 |
| Contact Person | Amy Tannenbaum |
| Preparation Date | August 29, 2018 |
| 807.92(a)(2) Name of Device | |
| Trade or Proprietary Name | TempSure System |
| Common or Usual Name | Surgical RF Generator |
| Classification Name | Electrosurgical, Cutting / Coagulation / Accessories |
| Classification Panel | General & Plastic Surgery |
| Regulation | 21 CFR 878.4400 |
| Regulatory Class | II |
| Product Code(s) | GEI, PBX |
| 807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed | |
| Predicate Devices | Cynosure TempSure (K171262)<br>Cynosure dba Ellman Surgitron 4.0 Dual RF S5 - Pelleve (K123366)<br>ValleyLab ForceTriad (K051644) |
| 807.92(a)(4) Device Description | |
| | The Cynosure TempSure™ System is a radiofrequency generator with<br>a variety of applications both aesthetic and surgical procedures. The<br>intended action is achieved through application of radiofrequency<br>energy to the patient which results in minimization of heat dissipation<br>and cellular alteration. Output of energy is controlled via the guided<br>user interface (GUI) and the foot and/or hand-switch.<br><br>The TempSure generator has been upgraded to include additional<br>surgical modes – Cut, Bipolar, Fulgurate, and Blend, along with the<br>existing Coag. The power of the device for Surgical applications is a<br>maximum of 300W. There have been no changes to the Smart<br>Handpiece treatment with temperature sensitive handpieces. |
| | TempSure system is used with the same existing Ellman |
| | electrosurgical accessories. |
| | The TempSure™ System includes: |
| | TempSure™ Generator<br>• |
| | Temperature Sensing Handpieces<br>• |
| | IEC Power Cord<br>• |
| | Footswitch<br>• |
| | Disposable/Reusable Neutral Pad<br>• |
| | Surgical Fingerswitch/Foot Controlled Handpieces<br>• |
| | Monopolar Cables<br>• |
| | Disposable/Reusable Electrodes, Forceps<br>• |
| 807.92(a)(5) Intended Use of the Device | |
| | The TempSure System has the following indications for use: |
| | The 10mm, 15mm, and 20mm handpieces are indicated for non-<br>ablative treatment of mild to moderate facial wrinkles and rhytids. |
| | The 25mm, and 30mm handpieces provide heating for the purpose of<br>elevating tissue temperature for selected medical conditions such as<br>temporary relief of pain, muscle spasms, and increase in local<br>circulation. |
| | The massage device is intended to provide a temporary reduction in<br>the appearance of cellulite |
| | The following surgical modes are applicable to the generator: |
| | <b>Coagulation/Hemostasis:</b> General surgical procedures including<br>urologic, thoracic, plastic, reconstructive, and gynecological<br>procedures where electrosurgical coagulation of tissue is performed |
| | <b>Cutting:</b> snoring. submucosal palatal shrinkage, traditional<br>uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage<br>control. epistaxis treatment, turbinate shrinkage. skin incisions,<br>biopsy, cysts, abscesses. tumors. cosmetic repairs, development of'<br>skin flaps. skin tags and blepharoplasty. |
| | <b>Blended Cutting and Coagulation:</b> snoring. submucosal palatal<br>shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with<br>effective hemorrhage control. epistaxis treatment, turbinate shrinkage,<br>skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma. |
| nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and<br>development of skin flaps. | |
| <b>Fulguration:</b> basal cell carcinoma. papilloma. cyst destruction,<br>tumors. verrucae, hemostasis. | |
| <b>Bipolar:</b> pinpoint precise coagulation. pinpoint hemostasis in any,<br>field (wet or dry). snoring. submucosal palatal shrinkage, traditional<br>uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage<br>control. epistaxis treatment and turbinate shrinkage | |
{5}------------------------------------------------
{6}------------------------------------------------
### 807.92(b)(1) Non-clinical tests submitted
The following non-clinical tests have been included in this 510k submission in determination of substantial equivalence between the test device and the referenced predicates. These tests have been conducted in accordance with the FDA Guidance "Premarket Notification 510k Submissions for electrosurgical Devices for General Surgery" - Section XI. Performance Data. E - System Testing, Thermal Effects on Tissue. See Section 18 - Performance Testing, Bench. A brief summary is provided below.
#### Thermal Effects on Tissue
Testing was performed on ex-vivo tissue samples to compare the thermally affected zone (TAZ) of the TempSure and predicate devices. Measurements were taken in triplicate on three tissue samples (liver, kidney, and muscle), and three power settings. For each tissue and power setting combination, the TempSure and the predicate device(s) were compared. This test shows that the TempSure is substantially equivalent to the ForceTriad and Surgitron devices.
### Electromagnetic Compatibility and Electrical Safety
Electrical safety testing for the Cynosure TempSure System was also completed to prove the safe use of the device. These test reports are provided in accordance with FDA Guidance "Premarket Notification 510k Submissions for Electrosurgical Devices for General Surgery" - Section XII -Electrical Safety and Electromagnetic Compatibility". The following test reports are available in Section 17 - Electromagnetic Compatibility and Electrical Safety.
- IEC 60601-1, Medical Electrical Equipment Part 1: General requirements for basic safety . and essential performances
- IEC 60601-1-2, Medical Electrical Equipment Part 1 -2: General requirements for basic ● safety and essential performance - Collateral standard: Electromagnetic disturbances requirements and tests
- IEC 60601-2-2, Medical electrical equipment Part 2 -2: Particular requirements for the basic ● safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
### 807.92(b)(2) Clinical tests submitted – N/A - No clinical tests submitted
### 807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted
{7}------------------------------------------------
The nonclinical tests demonstrate that the TempSure™ system is safe and effective and performs as well as the legally marketed predicate devices. The Thermal Effects on Tissue test showed that the thermally affected zone created by the TempSure™ device was similar to that created by the predicate devices, Surgitron and ForceTriad, in ex-vivo tissue testing. Despite their difference in maximum output power, the devices created a similar TAZ and therefore the devices can be considered substantially equivalent. In addition, the Electromagnetic Compatibility and Electrical Safety testing shows that the device is safe to use and meets required standards. The non-clinical testing shows that the device meets design specifications as well as performance requirements.
{8}------------------------------------------------
### Cynosure TempSure K182365
| Characteristic | Cynosure TempSure™<br>(KPending) | Cynosure TempSure<br>(K171262) | Ellman Surgitron 4.0<br>Dual RF S5 (Pelleve<br>S5) (K123366) | Valley Labs Force<br>Triad (K051644) |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The 10mm, 15mm, and<br>20mm Smart handpieces<br>are indicated for non-<br>ablative treatment of mild<br>to moderate facial<br>wrinkles and rhytids | The 10mm, 15mm, and<br>20mm handpieces are<br>indicated for non-ablative<br>treatment of mild to<br>moderate facial wrinkles<br>and rhytids | <b>Cutting:</b> Snoring,<br>submucosal palatal<br>shrinkage, traditional<br>uvulopalatoplasty<br>(RAUP), myringotomy<br>with effect hemorrhage<br>control, epistaxis<br>treatment, turbinate<br>shrinkage, skin<br>incisions biopsy, cysts,<br>abscesses, tumors,<br>cosmetic repairs,<br>development of skin<br>flaps, skin tags, and<br>blepharoplasty | General (including<br>urologic, thoracic,<br>plastic and<br>reconstructive,<br>arthroscopic),<br>laparoscopic, and<br>gynecological<br>procedures where<br>electrosurgical cutting<br>and coagulation of<br>tissue, and sealing<br>(fusion) of vessels and<br>tissue bundles is<br>performed, including<br>such procedures as<br>bowel resections,<br>hysterectomies (both<br>vaginal and<br>abdominal),<br>laparoscopic,<br>cholecystectomies,<br>laparoscopically<br>assisted vaginal<br>hysterectomies, gall<br>bladder procedures,<br>Nissen fundoplication,<br>adhesiolysis,<br>oophorectomy, etc. |
| | The 25mm, and 30mm<br>Smart handpieces provide<br>heating for the purpose of<br>elevating tissue<br>temperature for selected<br>medical conditions such<br>as temporary relief of<br>pain, muscle spasms, and<br>increase in local<br>circulation. | The 18mm, 25mm, and<br>30mm handpieces provide<br>heating for the purpose of<br>elevating tissue<br>temperature for selected<br>medical conditions such<br>as temporary relief of<br>pain, muscle spasms, and<br>increase in local<br>circulation. | <b>Blended Cutting and<br/>Coagulation:</b> Snoring,<br>submucosal palatal<br>shrinkage, traditional<br>uvulopalatoplasty<br>(RAUP), myringotomy<br>with effective<br>hemorrhage control,<br>epistaxis treatment,<br>turbinate shrinkage,<br>skin tags, papilloma<br>keloids, keratosis,<br>verrucae, basal cell<br>carcinoma, nevi,<br>fistulas, epithelidma,<br>cosmetic repairs, cysts,<br>abscesses, and | Vessels (arteries,<br>veins, lymph) 7mm<br>and smaller in<br>diameter, and bundles<br>as large as will fit in |
| | The massage device is<br>intended to provide a<br>temporary reduction in the<br>appearance of cellulite | The massage device is<br>intended to provide a<br>temporary reduction in the<br>appearance of cellulite | | |
| | The following surgical<br>modes are applicable to<br>the generator: | <b>Coagulation/Hemostasis:</b><br>Using the surgical<br>handpieces and<br>accessories, general<br>surgical procedures<br>including urologic,<br>thoracic, plastic,<br>reconstructive, and<br>gynecological procedures<br>where electrosurgical | | |
| | <b>Coagulation/Hemostasis:</b><br>General surgical<br>procedures including<br>urologic, thoracic, plastic,<br>reconstructive, and<br>gynecological procedures | | | |
| Characteristic | Cynosure TempSure™<br>(KPending) | Cynosure TempSure<br>(K171262) | Ellman Surgitron 4.0<br>Dual RF S5 (Pelleve S5) (K123366) | Valley Labs Force<br>Triad (K051644) |
| | where electrosurgical<br>coagulation of tissue is<br>performed | coagulation of tissue is<br>performed | development of skin<br>tags. | the jaws of the devices<br>can be sealed with<br>Ligasure vessel sealing<br>(tissue fusion) output. |
| | <b>Cutting:</b> snoring.<br>submucosal palatal<br>shrinkage, traditional<br>uvulopalatoplasty<br>(RAUP), myringotomy<br>with effective hemorrhage<br>control. epistaxis<br>treatment, turbinate<br>shrinkage. skin incisions,<br>biopsy, cysts, abscesses.<br>tumors. cosmetic repairs,<br>development of skin<br>flaps. skin tags and<br>blepharoplasty. | | <b>Hemostasis:</b> Control of<br>bleeding, epilation,<br>telangiectasia. | |
| | <b>Blended Cutting and<br/>Coagulation:</b> snoring.<br>submucosal palatal<br>shrinkage, traditional<br>uvulopalatoplasty<br>(RAUP). myringotomy<br>with effective hemorrhage<br>control. epistaxis<br>treatment, turbinate<br>shrinkage, skin tags,<br>papilloma keloids.<br>keratosis, verrucae, basal<br>cell carcinoma. nevi.<br>fistulas. epithelidma | | <b>Fulguration:</b> Basal<br>cell carcinoma,<br>papilloma, cyst<br>destruction, tumors,<br>verrucae, hemostasis. | |
| | | | <b>Bipolar:</b> Pinpoint<br>precise coagulation,<br>pinpoint hemostasis in<br>any field (wet or dry),<br>snoring, submucosal<br>palatal shrinkage,<br>traditional<br>uvulopalatoplasty<br>(RAUP), myringotomy<br>with effective<br>hemorrhaging control,<br>epistaxis treatment and<br>turbinate shrinkage. | |
| | | | Wrinkles: Non-ablative<br>treatment of mild to<br>moderate facial<br>wrinkles and rhytids. | |
| Characteristic | Cynosure TempSure™<br>(KPending) | Cynosure TempSure<br>(K171262) | Ellman Surgitron 4.0<br>Dual RF S5 (Pelleve<br>S5) (K123366) | Valley Labs Force<br>Triad (K051644) |
| | cosmetic repairs. cysts.<br>abscesses. and<br>development of skin flaps. |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.