K172373 · Medtech Products, Inc. · GEH · Nov 2, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172373
Device Name
Compound W Nitro-Freeze
Applicant
Medtech Products, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Nov 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Compound W® Nitro-Freeze is intended for over-the-counter treatment of common warts and plantar warts in adults and children four years of age or older.
Device Story
Compound W Nitro-Freeze is a pen-like, over-the-counter cryogenic device for wart removal. It uses pressurized nitrous oxide (N2O) as a cryogen. The user activates the device by pressing the pen into the cap for 2 seconds, which charges a disposable foam applicator tip. The device is applied topically to the wart; the cryogen causes evaporative cooling, destroying infected tissue through freezing. It is intended for home use by lay people. The device includes a pressure release feature for safety. Clinical output is the physical destruction of the wart tissue, which the user observes as the wart clears. The device provides a faster, more effective alternative to existing cryogenic wart removers, potentially reducing the number of treatments required for the patient.
Clinical Evidence
Single-center, randomized, multi-arm study (n=138) compared Compound W Nitro-Freeze to Wartner and Wartie. Primary endpoint: wart clearance after 1-3 treatments. Results: Nitro-Freeze showed 82.0% overall clearance vs 47.4% (Wartner) and 52.8% (Wartie) (p=0.001). First-treatment clearance was 34.0% vs 10.5% and 13.9% (p=0.013). Adverse events were minor (blistering, pain, burning) and considered offset by increased effectiveness.
Technological Characteristics
Pen-like cryogenic device; pressurized nitrous oxide (N2O) cryogen; disposable foam applicator tips. Thermal energy removal principle. Home-use, non-electric, non-sterile. Features a pressure release mechanism validated for safety at elevated temperatures (38°C-50°C).
Indications for Use
Indicated for over-the-counter treatment of common warts and plantar warts in adults and children 4 years of age or older.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 2, 2017
Medtech Products Inc. Mr. Vincent Argiro Director, Regulatory Affairs 660 White Plains Rd. Suite 250 Tarrytown, New York 10591
Re: K172373
Trade/Device Name: Compound W Nitro-Freeze Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: October 27, 2017 Received: October 30, 2017
Dear Mr. Argiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@Ida.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K172373
Device Name Compound W® Nitro-Freeze
Indications for Use (Describe)
Compound W® Nitro-Freeze is intended for over-the-counter treatment of common warts in adults and children four years of age or older.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for Medtech, a Prestige Brands company. The word "MEDTECH" is in large, bold, blue letters. Below the word is a blue line, and below that is the text "a PrestigeBrands company", with the word "PrestigeBrands" in red.
# Section 5 – 510(k) Summary
## Compound W® Nitro-Freeze
K #
### 1. Submitter
| Name & Address: | Medtech Products Inc.<br>660 White Plains Road<br>Tarrytown, NY, 10591 |
|-----------------|------------------------------------------------------------------------|
| Contact: | Vincent Argiro, RAC |
| Title: | Director, Regulatory Affairs |
| Phone number: | (914) 524-8721 |
| Email: | vargiro@prestigebrands.com |
**2. Date Prepared** August 4, 2017
**3. Device Identification**
| Trade/Proprietary Name: | Compound W® Nitro-Freeze |
|-------------------------|-----------------------------------------|
| Common/Usual Name: | Over-the-counter cryogenic wart remover |
| Classification Name: | Cryosurgical unit and accessories |
| Regulation Number: | 21 C.F.R. § 878.4350 |
| Product Code: | GEH [Unit, Cryosurgical, Accessories] |
| Device Class: | Class II |
| Classification Panel: | General & Plastic Surgery Devices |
#### 4. Legally Marketed Predicate Device(s)
| Predicate No. 1: | Wartner® Wart Removal System (K032271), Product Code GEH |
|------------------|----------------------------------------------------------|
| Predicate No. 2: | Wartie® Wart Remover (K140314), Product Code GEH |
{4}------------------------------------------------
#### 5. Device Description
Compound W® Nitro-Freeze is a pen-like cryogenic device that utilizes pressurized nitrous oxide (N,O, liquefied) to destroy wart tissue through evaporative cooling when the cryogen is delivered via a disposable foam applicator tip. Each Compound W® Nitro-Freeze device is supplied in an outer carton with an Instructions-for-Use leaflet and a bag containing replacement tips.
#### 6. Intended Use
The Compound W® Nitro-Freeze is an over-the-counter (OTC) device that is intended to be used by lay people to treat common warts and plantar warts.
The Compound W® Nitro-Freeze bears the following indications for use statement:
Compound W® Nitro-Freeze is intended for over-the-counter treatment of common warts and plantar warts in adults and children four years of age or older.
Compound W® Nitro-Freeze has the same intended use as the predicate devices and very minor differences between Compound W® Nitro-Freeze and the predicates with respect to the wording of the indications for use statements do not alter the devices' intended therapeutic effect.
#### 7. Substantial Equivalence Discussion
The following table compares Compound W® Nitro-Freeze to the predicate devices with respect to intended use and technological characteristics. This comparison of the devices provides detailed information demonstrating the basis for the determination of substantial equivalence.
| | Proposed Device | Predicate 1 | Predicate 2 | Differences |
|--------------------------------|-----------------------------------------|---------------------------------|---------------------------|-------------|
| Trade Name<br>/Device Name | Compound W®<br>Nitro-Freeze | Wartner® Wart<br>Removal System | Wartie® Wart<br>Remover | N/A |
| 510(k) Number | TBD | K032271 | K140314 | N/A |
| Date Cleared | 90 days from date<br>of receipt by FDA1 | 09/10/2003 | 06/12/2014 | N/A |
| Original applicant | Medtech Products<br>Inc. | Wartner USA B.V. | YonMedical B.V. | N/A |
| REGULATORY CLASSIFICATION | | | | |
| Regulatory Class | Class II | Class II | Class II | None |
| Name of Generic<br>Device Type | Cryogenic wart<br>remover | Cryogenic wart<br>remover | Cryogenic wart<br>remover | None |
| | | Table 5-1: Comparison of Characteristics |
|--|--|------------------------------------------|
|--|--|------------------------------------------|
<sup>1</sup> Projected based on Section 510(k) of the FDCA
{5}------------------------------------------------
| Regulation | 21 CFR § 878.4350 | 21 CFR § 878.4350 | 21 CFR § 878.4350 | None | | apply tip to wart | sec.; use handle to<br>apply tip to wart | use device to apply<br>tip to wart | |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Product Code | GEH | GEH | GEH | None | Duration of<br>treatments | Dependent upon<br>size and location of<br>wart (max 20 sec.<br>for common warts<br>and max 40 sec.<br>for plantar warts | Dependent upon<br>size and location of<br>wart (max 20 sec.<br>for common warts<br>and max 40 sec.<br>for plantar warts | Dependent upon<br>size and location ot<br>wart (max 20 sec.<br>for common warts<br>and max 40 sec.<br>for plantar warts | None |
| Applicable<br>Performance<br>Standards or<br>Special Controls | N/A | N/A | N/A | None | Cryogen and<br>Applicator | Nitrous oxide<br>delivered via foam<br>applicator | Dimethyl ether<br>and propane<br>delivered via foam<br>applicator | Dimethyl ether<br>used to cool a<br>metal applicator<br>bead | Yes (see discussion<br>in Sections 8-10 of<br>this summary) |
| DEVICE DESCRIPTION - SUBSTANTIAL EQUIVALENCE COMPARATORS | | | | | Temperature at tip<br>(after 40 seconds, in<br>bench testing) | -82.83°C | -33.06°C | -24.01°C | Yes (see discussion<br>in Sections 8-10 of<br>this summary) |
| Intended Use | Allow consumers<br>direct access to a<br>safe and effective<br>means to remove<br>common warts and<br>plantar warts. | Allow consumers<br>direct access to a<br>safe and effective<br>means to remove<br>common warts and<br>plantar warts. | Allow consumers<br>direct access to a<br>safe and effective<br>means to remove<br>common warts and<br>plantar warts. | None | Gas flammability | No | Yes | Yes | Improved safety |
| OTC or Rx | OTC | OTC | OTC | None | Pressure release<br>feature | Yes | No | No | Yes (see discussion<br>in Section 8 of this<br>summary) |
| Indications for Use | For over-the-<br>counter treatment<br>of common warts<br>and plantar warts<br>in adults and<br>children four years<br>of age or older. | For the over-the-<br>counter treatment<br>of common warts<br>and plantar warts. | For the over-the-<br>counter treatment<br>of common warts<br>and plantar warts<br>for patients aged 4<br>years and older. | None | Multiple Use Device | Yes | Yes | Yes | None |
| Target Population | Adults and children<br>4 years and older | Adults and<br>children 4 years<br>and older | Adults and<br>children 4 years<br>and older | None | Sterile Device | No | No | No | None |
| Technological<br>Characteristics | Pressurized gas to<br>generate extreme<br>cold at an<br>applicator tip | Pressurized gas to<br>generate extreme<br>cold at an<br>applicator tip | Pressurized gas to<br>generate extreme<br>cold at an<br>applicator tip | None | Leaflet included | Yes | Yes | Yes | None |
| DEVICE DESCRIPTION - DESIGN FEATURES | | | | | Use environment | Home | Home | Home | None |
| Energy | Thermal (removal<br>of heat from<br>treated skin) | Thermal (removal<br>of heat from<br>treated skin) | Thermal (removal<br>of heat from<br>treated skin) | None | Anatomical site of<br>use | Topical - wart | Topical - wart | Topical - wart | None |
| Design /<br>Presentation | Pen-like device<br>with disposable<br>foam applicators | Table-top device<br>with disposable<br>foam applicators | Pen-like device<br>with reusable<br>applicator bead | Minor (similar to<br>both predicates) | Electricity or<br>radiation used | No | No | No | None |
| Mode of Action | Destruction of<br>infected cells by<br>freezing | Destruction of<br>infected cells by<br>freezing | Destruction of<br>infected cells by<br>freezing | None | | | | | |
| Directions for use | Place device on flat<br>surface; push pen<br>into cap for 2 sec.;<br>remove from cap<br>and use device to | Place device on flat<br>surface; assemble<br>tip and handle;<br>push tip/handle<br>into device for 2-3 | Place on flat<br>surface; re-align<br>cap; press cap<br>down for 2 sec.;<br>remove cap and | Minor (similar to<br>both predicates) | | | | | |
{6}------------------------------------------------
In accordance with section 513(i)(1)(A) of the FDCA, a device is substantially equivalent (SE) when it has the same intended use and technological characteristics as a legally marketed predicate device. As demonstrated in this traditional 510(k), any differences between the subject device and the cited predicates do not raise different questions of safety or effectiveness and data (including clinical data) establishes that the device is as safe and effective as the predicates. It is on this basis that Compound W® Nitro-Freeze is SE to the cited predicate devices.
#### Non-Clinical Performance Data 8.
While the design differences between Compound W® Nitro-Freeze and the predicates are minor and non-clinical testing is therefore unnecessary, the cryogenic performance of the device has been properly validated. An in vitro experiment was conducted to determine the temperature generated by Compound W® Nitro-Freeze in comparison to three other over-the-counter cryogenic wart removers currently on the market: the predicates, Wartner and Wartie Wart Remover, and Urgo
{7}------------------------------------------------
#### K172373
(marketed in the U.S. as "Wart Freeze" under K130599). Product samples were tested at room temperature after being activated and introduced to an application substrate using a calibrated measuring instrument with a thermocouple sensor. This study clearly demonstrated the freezing performance of Compound W® Nitro-Freeze throughout the maximum treatment time of 40 seconds; the product generated temperatures ranging from -86.08°C (T=0) to -82.83°C (T=40) versus -28.82°C to -33.06°C (Wartner), -20.38°C to -24.01°C (Wartie Wart Remover), and -24.36°C to -1.93°C (Urgo).
The performance of the device's pressure release feature was validated in a study involving placement of 20 device samples in a laboratory oven at progressive temperatures of 35°C, 40°C, and 50°C or 38°C and 50°C for 24-hour periods. [10 samples started at 35°C and 10 samples started at 38°C; devices that were not disarmed moved on to the next stage in their respective test groups.] After each exposure period, the devices were cooled at room temperature and then tested to determine which had been successfully disarmed. To determine whether a device was disarmed, the analyst attempted to load (charge) the applicator tip with cryogen; if no cryogen remained in the canister, the device was considered disarmed. As intended, no devices were disarmed at 35°C and all were disarmed at either 38°C, 40°C or 50°C.
The non-clinical testing of Compound W® Nitro-Freeze demonstrates the device's freezing performance and the performance of its pressure release feature and therefore supports the substantial equivalence of Compound W® Nitro-Freeze to the cited predicates.
#### 9. Clinical Performance Data
While there are no differences in intended use and technological characteristics between the Compound W® Nitro-Freeze and the predicates that necessitate conducting a clinical trial, clinical data was provided in the 510(k) submission.
A single-center, randomized, multi-arm study in 138 patients (124 per-protocol) was conducted to compare the safety and effectiveness of Compound W® Nitro-Freeze to two over-the-counter cryogenic wart treatments available in the U.S. and abroad (the predicate devices, Wartner and Wartie Wart Remover). That study involved 1-3 treatments with the selected product according to its labeled instructions (second and third treatments only if needed), each followed by evaluator assessments. The results are shown in the table below.
| Treatment | Cured after X treatment(s) | | | Not cured after 3 treatments |
|--------------|----------------------------|---------------|---------------|------------------------------|
| | X=1 | X=2 | X=3 | |
| Nitro-Freeze | 34.0% (17/50) | 28.0% (14/50) | 20.0% (10/50) | 18.0% (9/50) |
| Wartner | 10.5% (4/38) | 18.4% (7/38) | 18.4% (7/38) | 52.6% (20/38) |
| Wartie | 13.9% (5/36) | 8.3% (3/36) | 30.6% (11/36) | 47.2% (17/36) |
#### Table 5-2: Primary effectiveness criterion in Study no. 14E2344
{8}------------------------------------------------
### K172373
This study demonstrated statistically significant superior effectiveness for Compound W® Nitro-Freeze ("Pixie") versus Wartner and Wartie in both overall wart clearance (82.0% vs. 47.4% and 52.8%, respectively after final treatment, p = 0.001) and first-treatment wart clearance (34.0% vs. 10.5% and 13.9%, p = 0.013) with no serious product-related adverse events recorded and an increase in only minor treatment-related symptoms (i.e., wart-site blistering, pain, and burning sensation). With more people experiencing wart removal after one treatment with Compound W® Nitro-Freeze than after two treatments with either of the comparators, any modest increase in nonserious adverse events can be considered offset by the significant increase in product effectiveness.
Study no. 14E2344 confirms and demonstrates that Compound W® Nitro-Freeze is substantially equivalent to the legally marketed predicate devices.
#### 10. Statement of Substantial Equivalence
Compound W® Nitro-Freeze has the same intended use as the predicate devices, Wartner (K032271) and Wartie Wart Remover (K140314), and utilizes the same basic principle of operation (introduction of extreme cold to destroy infected tissue). Like the predicates, Compound W® Nitro-Freeze uses pressurized gas as a cryogen, which it delivers using a disposable foam applicator – the same delivery method employed by the predicate, Wartner. Compound W® Nitro-Freeze and the predicate devices are of roughly the same size and shape and require depression of a device component for 2/3 seconds to release the cryogen and charge the applicator tip, and all three employ the same sizeand location-dependent application times.
Although the design of Compound W® Nitro-Freeze relies on a different cryogen than the predicates (nitrous oxide vs. dimethyl ether and propane or dimethyl ether alone), that difference does not raise any new questions of safety and effectiveness that are not associated with the predicate devices and the non-clinical data provided in this submission demonstrate that Compound W® Nitro-Freeze is as safe and effective as the predicates. Compound W® Nitro-Freeze is therefore substantially equivalent to the cited predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.