WART FREEZE

K130599 · Koninklijke Utermohlen NV · GEH · Jun 5, 2013 · General, Plastic Surgery

Device Facts

Record IDK130599
Device NameWART FREEZE
ApplicantKoninklijke Utermohlen NV
Product CodeGEH · General, Plastic Surgery
Decision DateJun 5, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4350
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

Wart Freeze (private label trade mark: Wart Freezer, Cryogenic Wart Remover) is indicated for the removal of common and plantar warts. The wart remover product is intended to be used in adults and children 4 years of age and older.

Device Story

Wart Freeze is an OTC cryosurgical device for treating common and plantar warts. It consists of a pressurized canister containing 38ml dimethylether (DME) and a permanently attached polypropylene applicator. The user operates the device by holding the canister upside down and pressing the bottom three times; this activates a safety valve that releases the cryogen into the applicator. The applicator tip reaches -50°C and is applied directly to the wart. The device includes disposable 70% alcohol swabs for cleaning the applicator between uses. The mechanism provides a controlled freeze duration to ensure effective treatment while maintaining safety via the valve. The device is intended for patient self-use at home.

Clinical Evidence

Bench testing only. Comparative laboratory testing using a phantom skin model demonstrated effectiveness similar to predicate devices. Applicator temperature testing via infrared camera confirmed a tip temperature of -50°C. Dose technique testing evaluated 3-short presses (1.31s freeze) versus 3-long presses (1.99s freeze), confirming safety valve performance. Biocompatibility testing (ISO 10993-5) confirmed the polypropylene applicator is non-cytotoxic. Chemical residual testing confirmed no leachables between the applicator and DME. Inhalation studies (NTP) on DME in animal models showed no acute or developmental toxicity at consumer exposure levels.

Technological Characteristics

Cryosurgical system; pressurized canister containing 100% dimethylether (DME). Applicator: polypropylene, permanently attached. Energy source: pressurized gas (DME). Operating principle: cryotherapy via direct application of -50°C cryogen. Safety features: integrated safety valve requiring 3-press actuation. Sterilization: not specified. Connectivity: none (standalone).

Indications for Use

Indicated for removal of common and plantar warts in adults and children 4 years of age and older.

Regulatory Classification

Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K130599 # 510K SUMMARY | Submitter's name: | Koninklijke Utermöhlen N.V.<br>De Overweg 1, 8471 ZA Wolvega, the Netherlands | | | |------------------------------|-------------------------------------------------------------------------------|--|--| | Address: | | | | | Telephone number: | 431 26 1 693 366 | | | | Title: | CEO | | | | Contact Person: | D van der Vat | | | | Date of 510k special: | 1 March 2013 | | | | Device Proprietary Name: | Wart Freeze | | | | Device common or usual name: | OTC wart removal system | | | | Classification name. | Cryosurgical unit, accessories | | | | Classification code: | GEI-I | | | | Regulation Number: | 21 CFR 878.4350 | | | # Substantial equivalence / predicate devices (Confidential) The following device listed in the table is predicate device based on technology, method of application and materials for this 510k Special. | Name of device | Manufacturer | Predicate<br>comparison | 510(k)<br>Number | | |----------------|---------------------------|-------------------------------------------|------------------|----------------------------------| | Wart Freeze | Koninklijke<br>Utermöhlen | Intended use,<br>technology,<br>materials | K101049 | Primary predicate<br>for changes | ### Description of the device: Wart Freeze is a private label (private label trade mark: Wart Freezer, Cryogenic Wart Remover) over the counter cryosurgery product for the treatment of common and plantar warts. The device consists of the following: - A pressurised dispenser (canister) containing as ingredient 38ml dimethylether (DME) with a . polypropylene applicator that is permanently attached (fixed) to the dispenser. The gas does not harm the ozone layer. The applicator is used to administer the cryogen directly to the wart. The applicator is cleaned after each use. - Disposable 70% alcohol cleansing swabs for cleaning the applicator after each use . - Instruction for use . #### Indications for use statement Wart Freeze (private label trade mark: Wart Freezer, Cryogenic Wart Remover) is indicated for the removal of common and plantar warts. #### Intended use The wart remover product is intended to be used in adults and children 4 years of age and older. 5 2013 JUN {1}------------------------------------------------ #### Statement of technological characteristics of the device - The Wart Freeze (private label trade mark: Wart Freezer, Cryogenic Wart Remover) is a 21. cryosurgical system comprised of a dispenser containing as ingredient, dimethylether (DME), and a permanently attached (fixed) applicator to apply the cryogen directly to the wart. #### ﺷ Laboratory (functionality) testing (CONFIDENTIAL) The following testing was conducted: - 1. Comparative laboratory testing using predicate devices using a phantom skin model. The predicates were used in accordance with their instructions. The functionality testing indicated that the various competitor products are similar in effectiveness. - Applicator temperature laboratory test using an infrared camera technology was conducted to 2. determine the degree of cold generated by the cryogen on the applicator. The applicator surface at the tip of the applicator (part placed over the wart after dispensing the cryogen) is -50 ° C. - 3. Dose technique testing using infrared (IR) camera technology to determine the dispensing freeze time between the two potential dose techniques, i.e., 3 short presses of the dispenser versus 3 long presses of the dispenser. The test also evaluated effectiveness of freeze time between the dose techniques. The effectiveness of 3-short presses resulted in a freeze time of 1.31 seconds versus 1.99 seconds for 3-long presses. The maximum acceptable concentration of DME administered was 0,37 mg/ml during both dose techniques as result of safety (dose) valve used. #### Biocompatibility testing c. Biocompatibility testing was performed on the applicator in line with ISO 10993:5 Tests for in vitro cyloloxicity. The test was conducted to determine the safety of the cryogen and applicator material, polypropylene. The applicator material when in contact with the cryogen was found to be non-lysic. (CONFIDENTIAL) #### d. Chemical residual testing Laboratory testing was conducted to assess and identify any potential chemical residuals including leachable or degradable components. The test concluded that there were no chemical residuals in the dimethylether/polypropyelene extracts whereby there is no leachable between the applicator and cryogen, dimethylether. (CONFIDENTIAL) #### e. Inhalation studies on the dimethylether Dimethylether inhalation studies were conducted by the suppliers. The US National Toxicological Programme (NTP) conducted acute toxicity studies (LC:o)) on mouse and rabbit. The study concluded that DME did not produce acute toxicity in animal models or bacteria (genus: Ames Salmonella Typhimiurium). Developmental toxicity studies in animal models also showed that DME did not have any significant toxic effect on the mice or rabbit pups. There is however a limitation in the availability of human data when it comes to DME developmental and reproductive toxicity in human beings. DME received a ranking of not acutely toxic by the NTT. Current exposure limits for DME has been set at 1000 ppm or 0.1% DME for an 8-hour daily life-time exposures of the exposures for consumer products where DME is used have been shown to be in the range of 100 ppm for short {2}------------------------------------------------ periods of time (<15 minutes/day). This is far below the levels to induce central nervous system effects. (CONFIDENTIAL) #### ﻧﮯ Comparison to predicate devices ### Application The product administers cryogen directly onto the wart by way of an applicator permanently fixed to the dispenser. ## Applicator effectiveness duration The applicator maintains a temperature of -50 ° C. The effectiveness of predicates is identical. ## Cryogen The product uses 100% dimethylether, same as predicate device. #### Safe!Ease of use The design of the product incorporates the use of a safety valve. The dispenser is held upside down and actuated by pressing the bottom of the dispenser 3 times before the safety valve allows cryogen to be administered into the applicator and onto the wart. The predicate device has identical safety valve mechanism. ### Indications for use The product is indicated for over-the-counter treatment of common and plantar warts. The predicate device is indicated for identical intended use. #### Conclusions Based on the information presented above, it is concluded that the from the data supports the safety and effectiveness of the Wart Freezer. It is indicated that based on the safety data related to the ingredient, dimethylether, and similar labelling to the predicate device, this product should be qualified for the special 510k submission. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. # DEPARTMENT OF HEALTH & HUMAN SERVICES Letter dated: June 5, 2013 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Koninklijke Utermohlen NV % KRM Associates Kenneth R. Michael, Pharm.D., RAC, ASC 11956 Bernardo Plaza Drive, #417 San Diego, California 92128-1317 Re: K130599 ﺭ Trade/Device Name: Wart Freeze Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: April 29, 2013 Received: May 06, 2013 Dear Dr. Michael: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval.of.a.premarket.approval.application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Kenneth R. Michael, Pharm.D., RAC, ASC device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N.J.Mielkerson -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K130599 Device Name, Wart Freeze Indications For Use: The product is indicated for the removal of common and plantar warts. Prescription Use (Part-21-CFR-801-Subpart-D)- AND/OR Over-The-Counter Use × -(21-CFR-801-Subpart-C)- (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) Neil R Ogdens 2013.06.06 11:2:10:11:9 -04'00' (Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number__K130599 Page 1 of _1_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...