ENDO GIA REINFORCED RELOAD WITH TRI-STAPLE TECHNOLOGY
K133938 · Covidien · GDW · Jan 22, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133938
Device Name
ENDO GIA REINFORCED RELOAD WITH TRI-STAPLE TECHNOLOGY
Applicant
Covidien
Product Code
GDW · General, Plastic Surgery
Decision Date
Jan 22, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Endo GIA™ reinforced reload with Tri-Staple™ Technology preloaded with polyglycolic acid staple line reinforcement has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas.
Device Story
Surgical stapler reload; utilizes titanium staples and synthetic absorbable polyglycolic acid reinforcement material. Device places two, triple-staggered rows of staples while simultaneously dividing tissue and reinforcement material. Used in abdominal, gynecologic, pediatric, and thoracic surgeries; operated by surgeons. Provides mechanical tissue closure and transection; reinforcement material secured to anvil and cartridge via absorbable sutures. Benefits include staple line reinforcement during resection and anastomosis procedures.
Clinical Evidence
Bench and animal performance evaluations only. Bench testing included insertion/removal force, firing force, retraction force, staple formation, iDrive Ultra testing, tissue pull-apart, sequential/perpendicular/L-shaped firings, and biocompatibility. Animal testing evaluated atraumatic reload, free bleed time, air leaks, burst pressure, and additional firings. No human clinical data provided.
Technological Characteristics
Titanium staples; synthetic absorbable polyglycolic acid reinforcement material; synthetic absorbable sutures. Articulating 45mm and 60mm cartridges. Staple sizes: 3.0-4.0mm or 4.0-5.0mm. Biocompatibility per ISO 10993-1. Mechanical actuation.
Indications for Use
Indicated for patients undergoing abdominal, gynecologic, pediatric, or thoracic surgery requiring resection, transection, or anastomosis of tissue, including liver substance, hepatic vasculature, biliary structures, and pancreas.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
NEOVEIL™ Reinforcement Staple Line Material (K130997)
Submission Summary (Full Text)
{0}------------------------------------------------
K133938 Page 1 of 4
# 510(k) Summary
JAN 2 2 2014
This 510(k) summary information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.98.
| | Submitter Information |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name: | Covidien IIc |
| Address: | 15 Hampshire Street |
| | Mansfield, MA 02048 |
| Establishment Registration: | 1282497 |
| Name of contact person: | Jose Marquez |
| | Regulatory Affairs Manager |
| | Covidien |
| | 15 Hampshire Street |
| | Mansfield, MA 02048 USA |
| | Phone: (508) 452-4160 |
| Date prepared: | January 22, 2014 |
| Trade or proprietary name: | Endo GIA™ Reinforced Reload with Tri-Staple™ Technology |
| Common or usual name: | Surgical Stapler with Implantable Staple |
| Classification name: | Staple, Implantable |
| Classification panel: | General and Plastic Surgery (79) |
| Regulation: | 21 CFR 878.4750 |
| Product Code(s): | GDW, ORQ |
| Legally marketed devices to<br>which equivalence is claimed: | <ul><li>Endo GIA™ Stapler (K111825)</li><li>NEOVEIL™ Reinforcement Staple Line Material (K130997)</li></ul> |
| Reason for 510(k) submission: | To obtain market clearance for the Endo GIA™ Reinforced<br>Reload with Tri-Staple™ Technology. |
| Device description: | The Endo GIA™ Reinforced Reload with Tri-Staple™<br>Technology places two, triple-staggered rows of titanium staples<br>with a reinforcement material and simultaneously divides the |
{1}------------------------------------------------
tissue and the reinforcement material between the two, triple-The absorbable staple line reinforcement staggered rows. material, comprised of a synthetic polyester composed of pure alvcolide, will be secured to the anvil and cartridge sides of the reload with four synthetic absorbable sutures. The size of the staples is determined by the selection of the 3.0-4.0 mm or 4.0-5.0 mm Reload. The Endo GIA™ Reinforced Reload with Tri-Staple™ Technology is only available in the articulating 45 mm and 60 mm length cartridges with the 3.0-4.0 mm or 4.0-5.0 mm staples, respectively.
Endo GIA™ reinforced reload with Tri-Staple™ Technology Intended use of the device: preloaded with polyglycolic acid staple line reinforcement has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas.
Summary comparing the technological characteristics of the subject and predicate devices:
Materials:
Performance Data:
GIA™ Reinforced Reload with Tri-Staple ™ The Endo Technology is substantially equivalent to the predicate devices staple line reinforcement regard to stapling and with technologies.
All components of the Endo GIA™ Reinforced Reload with Tri-Staple™ Technology are comprised of materials which are in
> Bench and animal performance evaluations were completed to verify that the Endo GIA™ Reinforced Reload with Tri-Staple™ Technology are safe and effective and perform as intended.
> The tests performed to show substantial equivalence of the Endo GIA™ Reinforced Reload with Tri-Staple™ Technology to the predicate device are as follows:
- . In Vitro
- Insertion/removal force o
accordance with ISO Standard 10993-1.
- Firing force 0
- Retraction force 0
- Staple formation o
- iDrive Ultra testing O
- Twek Pull apart 0
- Sequential firings 0
- Perpendicular firings O
- L-shaped firings O
- In Vivo · .
- Atraumatic reload O
{2}------------------------------------------------
.
Free bleed time ্ O Air leaks
- -
:
. .
.
. . . . . . .
.
. .
.
{3}------------------------------------------------
- Burst o
- Additional firings o
- Biocompatibility .
Conclusion:
The results of the tests performed demonstrate that the subject device, Endo GIA™ Reinforced Reload with Tri-Staple™ Technology, is substantially equivalent to the predicate device(s). .
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font and is black. The background is white.
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with three stylized wing segments.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 22, 2014
Covidien Ilc Mr. Jose Marquez Regulatory Affairs Manager 15 Hampshire Street Mansfield, Massachusetts 02048
Re: K133938
Trade/Device Name: Endo GIA™ Reinforced Reload with Tri-Staple™ Technology Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, ORO Dated: December 20, 2013 Received: December 23, 2013
Dear Mr. Marquez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Jose Marquez
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
For
Sincerely yours,
Joshua FDMipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510{k} Number (if known) K133938
Device Name
Endo GLA™ Reinforced Reload with Tri-Staple™ Technology
#### Indications for Use (Describe)
Endo GIA™ reinforced reload with Tri-Staple™ Technology preloaded with polyglycolic acid staple line reinforcement has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas.
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Slatement on last page.
Type of Use (Select one or both, as applicable)
[Z] Prescription Use (Part 21 CFR 801 Subpart D)
O Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY STATE AND STATE Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (9/13)
Page Lof 2
PSC Publishion Screlocs (3011 443-674)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.