Covidien, Formerly US Surgical A Divison of Tyco H
Published clinical literature
A clinical literature review was conducted to demonstrate and support the clinical application of the Endo GIA, DST Series GIA, and DST Series TA staplers for the transection and resection of the pancreas.
Literature review; Pancreatic surgery; Clinical application
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical literature review; Literature review
Patients undergoing pancreatic surgery
Not applicable for this study
Clinical application for transection and resection of pancreas
Indications for Use
Endo GIA™ Ultra Universal Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas. The Endo GIA™ Ultra Universal Staplers when used with the Endo GIA™ Curved-Tip Reloads can be used to blunt dissect or separate target tissue from other certain tissue. The Endo GIA™ Ultra Universal Staplers with Duet TRS™ Reloads with Tri-Staple™ Technology have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas. The Endo GIA™ Ultra Universal Staplers when used with the Duet TRS™ Curved-Tip Reloads can be used to blunt dissect or separate target tissue from other certain tissue. The Endo GIA™ Radial Reload with Tri-Staple™ Technology has application in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection. It may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas. The Endo GIA™ Universal Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection and creation and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas. The Endo GIA™ Ultra Universal Staplers with Duet TRS™ Reloads have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomosis. They may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas. The Auto Suture™ GIA™ Reloadable Staplers and the SGIA™ Knifeless Stapler have applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis. The SGIA™ Knifeless stapler may be used for occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas. The Auto Suture™ TA™ Reloadable staplers have applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection of tissue and creation of anastomosis, including occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas.
Device Story
Endo GIA, DST Series GIA, and DST Series TA staplers are surgical instruments used for tissue resection, transection, and anastomosis. Devices utilize single-use reloads containing titanium staples; staples are placed in double or triple-staggered rows while simultaneously dividing tissue. Some models feature Tri-Staple technology or Duet TRS staple line reinforcement. Operated by surgeons in OR settings for abdominal, gynecologic, pediatric, and thoracic procedures. Devices facilitate tissue management in complex anatomical areas, including liver, pancreas, and pelvis. Clinical benefit derived from standardized, reliable tissue closure and division during surgery. No changes to device design or function; submission supports expanded indication for pancreatic tissue.
Clinical Evidence
Clinical literature review performed to support the use of Endo GIA, DST Series GIA, and DST Series TA staplers for transection and resection of the pancreas. No new clinical trials were conducted for this submission.
Technological Characteristics
Mechanical surgical staplers using titanium staples. Materials comply with ISO 10993-1. Features include single-use reloads with varying staple sizes (2.0mm to 5.0mm) and lengths (30mm to 100mm). Some models incorporate Tri-Staple technology or Duet TRS reinforcement. Manual operation; no software or electronic components.
Indications for Use
Indicated for patients undergoing abdominal, gynecologic, pediatric, or thoracic surgery requiring tissue resection, transection, or anastomosis. Includes procedures involving liver substance, hepatic vasculature, biliary structures, and pancreas. Specific applications include low anterior resection (pelvic) and left atrial appendage occlusion (SGIA/TA models).
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness Surgical Devices, a global business unit SUBMITTER: of Tyco Healthcare Group LP (d/b/a Covidien) 60 Middletown Avenue North Haven, CT 06473 Tel. No .: (203) 492-5352 CONTACT PERSON: Frank Gianelli Senior Associate, Regulatory Affairs October 19, 2011 DATE PREPARED: Endo GIA™ Staplers TRADE/PROPRIETARY NAME: DST Series™ GIA™ Staplers DST Series™ TA™ Staplers Staple, Implantable COMMON/USUAL NAME: Staple, Implantable CLASSIFICATION NAME: a. Panel no and product code: 79-GDW b. Regulation no: 21 CFR 878.4750 Endo GIA™ Staplers (K061095, K083519, K093410, K101444, PREDICATE DEVICE(S): K103263, K102291, K080898) DST Series™ GIA™ Staplers (K032696) DST Series™ TA™ Staplers (K032696) ENDOPATH Linear Cutters and Staplers (K070887) (K070887 Echelon Linear Cutters and Staplers The Endo GIA™ Stapler places two, triple-staggered rows of DEVICE DESCRIPTION: titanium staples and simultaneously divides the tissue between the two, triple-staggered rows. The size of the staples is determined by the selection of the appropriate single use reload that is available in 30, 45 and 60 mm lengths as well as being available with optional Duet TRS™ staple line reinforcement material and optional curved-tip anvil for reloads with Tri-Staple™ Technology: · 2.0 mm staple size single use reload (gray cartridge) · 2.5 mm staple size single use reload (white cartridge · 3.5 mm staple size single use reload (blue cartridge) · 4.8 mm staple size single use reload (green cartridge) · 2.0, 2.5, 3.0 mm staple size single use reload with Tri-Staple™ Technology (tan cartridge) · 3.0, 3.5, 4.0 mm staple size single use reload with Tri-Staple™ Technology (purple cartridge) · 4.0, 4.5, 5.0 mm staple size single use reload with Tri-Staple™ Technology (black cartridge) The Endo GIA™ Single Use Radial Reload with Tri-Staple™ Technology places two, triple-staggered rows of titanium staples and simultaneously divides the tissue between the two, triple-staggered rows. The size of the staples is determined by the selection of the appropriate single use reload that is
- available in a 60 mm length only: · 3.0, 3.5, 4.0 mm staple size single use radial reload with Tri-Staple™ Technology (purple cartridge)
OCT 2 5 2011
{1}------------------------------------------------
The DST Series™ GIA™ Stapler place two double staggered rows of titanium staples and simultaneously divides the tissue between the two double rows. The size of the staples is determined by the selection of the appropriate single use reload that is available in 60, 80 and 100 mm lengths:
KIII825 0. 2 of
- · 2.5 mm staple size single use reload (white cartridge) (60, 80, and 100 mm lengths)
- · 3.5 mm staple size single use reload (blue cartridge) (80 and 100 mm lengths only
- · 4.8 mm staple size single use reload (green cartridge) (80 and 100 mm lengths only)
The DST Series™ SGIA™ Knifeless Stapler places two double staggered rows of titanium staples. The size of the staples is determined by the selection of a 2.5 mm staple size single use reload available in a 60 mm length only.
The DST Series™ TA™ Stapler places a double staggered row of titanium staples. The size of the staples is determined by the selection of the appropriate single use reload that is available in 30, 45, 60 and 90 mm lengths:
- · 2.5 mm staple size single use reload (white cartridge) (30, 45, 60 and 90 mm lengths)
- · 3.5 mm staple size single use reload (blue cartridge) (45, 60 and 90 mm lengths only
- · 4.8 mm staple size single use reload (green cartridge) (45, 60 and 90 mm lengths only)
## INTENDED USE:
## Endo GIA™ Staplers:
The Endo GIA™ Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
## Endo GIA™ Radial Reload with Tri-Staple™ Technology:
The Endo GIA™ Radial Reload with Tri-Staple™ Technology has application in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
## DST Series™ GIA™ Staplers:
The DST Series™ GIA™ Staplers and the DST Series™ SGIA™ Knifeless Stapler have applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis. The SGIA™ Knifeless stapler may be used for occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas.
{2}------------------------------------------------
## DST Series TM TATM Staplers:
The Auto Suture™ TA™ Reloadable staplers have applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue and creation of anastomosis, including occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas.
The Endo GIA™ Staplers, DST Series™ GIA™ Staplers and TECHNOLOGICAL DST Series™ TA™ Staplers are each identical to their CHARACTERISTICS: predicate device. The only change is the inclusion of a specific indication concerning each device's use for the transection and resection of pancreas as a subset of the general indication for each of these devices.
- All components of the Endo GIA™ Staplers, DST Series™ MATERIALS: GIA™ Staplers and DST Series™ TA™ Staplers are comprised of materials, which are in accordance with ISO Standard 10993-1.
#### A clinical literature review was performed to demonstrate and PERFORMANCE DATA. support the clinical application of the Endo GIA™ Staplers, DST Series™ GIA™ Staplers and DST Series™ TA™ Staplers for transection and resection of pancreas.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circle's perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 2 5 2011
Covidien % Mr. Frank Gianelli Senior Associate, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473
Re: K111825
Trade/Device Name: Endo GIA"" Stapler, DST Series™ GIA™ Stapler and
DST Series™ Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: September 29, 2011 Received: September 30, 2011
Dear Mr. Gianelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Frank Gianelli
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Eric Keith
ForMark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
111825
# Indications For Use
# 510(k) Number (if known): 长川贸25
# Device Name: Endo GIA™ Stapler, DST Series™ GIA™ Stapler and DST Series™ TA™ Stapler
### Indications For Use:
#### Endo GIA™ Staplers:
The following indications are applicable to the Endo GIA™ Ultra Universal Stapler with Endo GIA™ Reloads with Tri-Staple™ Technology including Curved Tip Reloads:
The Endo GIA™ Ultra Universal Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
The Endo GIA™ Ultra Universal Staplers when used with the Endo GIA™ Curved-Tip Reloads can be used to blunt dissect or separate target tissue from other certain tissue.
The following indications are applicable to the Endo GIA™ Ultra Universal Stapler with Duet TRS™ Reloads with Tri-Staple™ Technology including Duet TRS™ Curved Tip Reloads:
The Endo GIA™ Ultra Universal Staplers with Duet TRS™ Reloads with Tri-Staple™ Technology have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
The Endo GIA™ Ultra Universal Staplers when used with the Duet TRS™ Curved-Tip Reloads can be used to blunt dissect or separate target tissue from other certain tissue.
The following indications are applicable to the Endo GIA™ Ultra Universal Stapler with Radial Reload with Tri-Staple™ Technology:
The Endo GIA™ Radial Reload with Tri-Staple™ Technology has application in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection.
It may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
The following indications are applicable to the Endo GIA™ Universal Stapler with Endo GIA™ Straight and Roticulator ™ SULUs:
The Endo GIA™ Universal Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection and creation and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
The following indications are applicable to the Endo GIA™ Universal Stapler with Duet TRS™ Reloads:
The Endo GIA™ Ultra Universal Staplers with Duet TRS™ Reloads have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of
{6}------------------------------------------------
anastomosis. They may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
DST Series ™ GIA™:
The Auto Suture™ GIA™ Reloadable Staplers and the SGIA™ Knifeless Stapler have applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis. The SGIA™ Knifeless stapler may be used for occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas.
## DST Series™ TA™:
The Auto Suture™ TA™ Reloadable staplers have applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection of tissue and creation of anastomosis, including occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Eriel Keith
Sign Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111825
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.