DUET TRS RELOADS, DUET TRS RELOADS WITH TRI-STAPLE TECHNOLOGY, ENDO GIA UNIVERSAL AND ULTRA UNIVERSAL STAPLERS
K120258 · Covidien, Formerly US Surgical A Divison of Tyco H · GDW · Feb 29, 2012 · General, Plastic Surgery
Device Facts
Record ID
K120258
Device Name
DUET TRS RELOADS, DUET TRS RELOADS WITH TRI-STAPLE TECHNOLOGY, ENDO GIA UNIVERSAL AND ULTRA UNIVERSAL STAPLERS
Applicant
Covidien, Formerly US Surgical A Divison of Tyco H
Product Code
GDW · General, Plastic Surgery
Decision Date
Feb 29, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Endo GIA™ Universal Staplers with Duet TRS™ reloads have applications in abdominal, gynecologic and pediatric surgery for resection, transection and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas. The Duet TRS™ Reloads with Tri-Staple™ Technology have applications in abdominal, gynecologic and pediatric surgery for resection, transection and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
Device Story
Duet TRS™ Reloads and Duet TRS™ Reloads with Tri-Staple™ Technology are surgical stapler cartridges used with Endo GIA™ Universal and Ultra Universal Staplers. The device places two, triple-staggered rows of titanium staples while simultaneously dividing tissue and an integrated synthetic absorbable reinforcement film. The reinforcement material is composed of glycolide, dioxanone, and trimethylene carbonate, secured with BIOSYN™ synthetic absorbable suture. Used in open or endoscopic surgical procedures (via trocar access) by surgeons to perform resections, transections, and anastomoses in abdominal, gynecologic, and pediatric applications. The device provides mechanical tissue approximation and division, aiding in hemostasis and tissue management. No software or electronic components are involved.
Clinical Evidence
No clinical data. Performance evaluations were not required to support this labeling modification as the devices are identical to the predicate devices.
Technological Characteristics
Disposable surgical stapler reloads. Materials: Titanium staples, synthetic absorbable film (glycolide, dioxanone, trimethylene carbonate), BIOSYN™ synthetic absorbable suture. Materials comply with ISO 10993-1. Mechanical actuation via Endo GIA™ stapler handle. No software, electronics, or energy sources.
Indications for Use
Indicated for patients undergoing abdominal, gynecologic, or pediatric surgery requiring resection, transection, or creation of anastomoses, including liver substance, hepatic vasculature, biliary structures, and pancreas.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Duet TRS™ Reloads with Tri-Staple™ Technology (K103263, K111825)
Endo GIA™ Universal and Ultra Universal Staplers (K111825)
Submission Summary (Full Text)
{1}------------------------------------------------
KI20258
## 510(k) Summary of Safety and Effectiveness
FEB 2 9 2012
| SUBMITTER: | Surgical Devices, a global business unit<br>of Tyco Healthcare Group LP (d/b/a Covidien)<br>60 Middletown Avenue<br>North Haven, CT 06473<br>Tel. No.: (203) 492-5352 |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Jennifer Brennan Manager, Regulatory Affairs
DATE PREPARED:
CONTACT PERSON:
January 20, 2012
Duet TRS™ Reloads. TRADE/PROPRIETARY NAME:
Duet TRS™ Reloads with Tri-Staple™ Technology
Endo GIA™ Universal and Ultra Universal Staplers
Surgical Stapler with Implantable Staples
COMMON/USUAL NAME:
PREDICATE DEVICE(S):
Staples, Implantable CLASSIFICATION NAME:
Duet TRS Reloads (K080898, K111825)
Duet TRS™ Reloads with Tri-Staple™ Technology (K103263, K111825)
Endo GIA ™ Universal and Ultra Universal Staplers (K111825)
DEVICE DESCRIPTION:
The Duet TRS™ Reloads when used with Endo GIA™ Staplers place two, triple-staggered rows of titanium staples along with two layers of absorbable tissue reinforcement material (one layer on cartridge side and one laver on anvil side) and simultaneously divides the tissue and reinforcement material between the two, triple-staggered staple rows.
The Duet TRS™ Reloads with Tri-Staple™ Technology when used with Endo GIA™ Staplers places two, triple-staggered rows of titanium staples along with two layers of absorbable tissue reinforcement material (one layer on cartridge side and one layer on anvil side) and simultaneously divides the tissue and reinforcement material between the two, triple-staggered staple rows.
The staple line reinforcement material is a synthetic absorbable film prepared from synthetic polyester composed of glycolide, dioxanone, and trimethylene carbonate. The staple line reinforcement material supplied on each Reload is undyed (natural) and secured to the anvil and cartridge with BIOSYN™ synthetic absorbable suture.
{2}------------------------------------------------
The Endo GIA™ Universal and Ultra Universal Staplers (K111825) with associated staple cartridge reloads are articulating, disposable surgical staplers that simultaneously transect and staple various types of internal tissues. Each can be used in both endoscopic and open surgical procedures, is available in multiple sizes, is for endoscopic procedures and can be introduced and used through appropriately sized trocar endoscopic access cannulae.
The Endo GIA™ Universal Staplers with Duet TRS™ reloads have applications in abdominal. gynecologic and pediatric surgery for resection, transection and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas
The Duet TRS™ Reloads with Tri-Staple™ Technology have applications in abdominal, gynecologic and pediatric surgery for resection, transection and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas
Duet TRS™ Reloads and the Duet TRS™ Reloads with Tri-Staple™ Technology and the Endo GIA™ Staplers Universal and Ultra Universal are identical to the predicate devices.
All components of the Duet TRS™ Reloads and the Duet TRS™ Reload with Tri-Staple™ Technology and the Endo GIA™ Staplers (Universal and Ultra Universal) are comprised of materials that are in accordance with ISO Standard 10993-1. ·
There have been no changes to the design of the Duet TRS™ Reloads, the Duet TRS™ Reloads with Tri-Staple™ Technology, or the Endo GIA™ Universal and Ultra Universal Staplers Performance evaluations were not required to support this labeling modification.
## INTENDED USE:
TECHNOLOGICAL CHARACTERISTICS:
MATERIALS:
## PERFORMANCE DATA:
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Covidien % Ms. Jennifer Brennan Manager, Regulatory Affairs 60 Middletown Avenue · North Haven, Connecticut 06473
FEB 2 9 2012
Re: K120258
Trade/Device Name: Duet TRS™ Reloads Duet TRS™ Reloads with Tri-Staple™ Technology Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: January 25, 2012 Received: January 27, 2012
Dear Ms. Brennan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Ms. Jennifer Brennan
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vou
ely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Duet TRS™ Reloads Device Name: Duet TRS™ Reloads with Tri-Staple™ Technology
Indications For Use:
The Endo GIA™ Universal Staplers with Duet TRS™ reloads have applications in abdominal, The Lindo Of . . . Onliverour Station, transection and creation of anastomoses. They gyncologic and poulatio ourgery for version of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas
The Duet TRS™ Reloads with Tri-Staple™ Technology have applications in abdominal, gynecologic and pediatric surgery for resection and creation of anastomoses. They gynecologic and poularle ourger resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas
Prescription Use (Part 21 CFR 801 Subpart D) AND/ÓR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bail Kmefer, WXAU
(Division Sign-Off)
Division of Surgical, Orthopedic. and Restorative Devices
510(k) Number K120258
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.